INSTANT-VIEW MULTI-DRUG SCREEN URINE TEST

K022564 · Alfa Scientific Designs, Inc. · DKZ · Aug 30, 2002 · Clinical Toxicology

Device Facts

Record IDK022564
Device NameINSTANT-VIEW MULTI-DRUG SCREEN URINE TEST
ApplicantAlfa Scientific Designs, Inc.
Product CodeDKZ · Clinical Toxicology
Decision DateAug 30, 2002
DecisionSESE
Submission TypeSpecial
Regulation21 CFR 862.3100
Device ClassClass 2

Indications for Use

The Instant-View™ Multi-Drug Screen Urine Test is a qualitative immunoassav intended to be used to detect any combination (variants from 2 to 12) of the following drugs and/or drug metabolites in human urine at the specified cutoff levels. It is intended for professional use only. Instant-View™ Multi-Drug Screen Urine Test provides only a preliminary analytical test result. A more specific alternate chemical must be used to obtain a confirmed analytical result. Gas Chromatography/Mass Spectrometry (GC/MS) is the preferred confirmatory method. Clinical consideration and professional judgment should be applied to any drug of abuse test result, particularly when preliminary positive results are used.

Device Story

Device is an assemblage of multiple single-drug urine test strips; enables simultaneous detection of 2 to 12 drugs/metabolites in one urine sample. Operates via immunochemical recognition principle; each strip functions independently. Used in professional clinical settings; provides preliminary qualitative results. Results require confirmation by alternate chemical method, preferably GC/MS. Clinical judgment required for interpretation. Benefits include rapid, multi-analyte screening in a single test frame.

Clinical Evidence

Bench testing only. In-house evaluation confirmed no interference between individual test strips within the multi-drug assembly. Accelerated degradation studies demonstrated a minimum device lifetime of two years.

Technological Characteristics

Assemblage of multiple immunochemical test strips. Qualitative immunoassay. Detects drugs/metabolites in human urine. Cutoff levels range from 25 ng/ml to 2000 ng/ml depending on analyte. No electronic components or software.

Indications for Use

Indicated for professional use to detect drugs or drug metabolites (AMP, BAR, BZD, COC, THC, MTD, METH, MOR, PCP) in human urine at specified cutoff levels.

Regulatory Classification

Identification

An amphetamine test system is a device intended to measure amphetamine, a central nervous system stimulating drug, in plasma and urine. Measurements obtained by this device are used in the diagnosis and treatment of amphetamine use or overdose and in monitoring levels of amphetamine to ensure appropriate therapy.

Special Controls

*Classification.* Class II (special controls). An amphetamine test system is not exempt if it is intended for any use other than employment or insurance testing or is intended for Federal drug testing programs. The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 862.9, provided the test system is intended for employment and insurance testing and includes a statement in the labeling that the device is intended solely for use in employment and insurance testing, and does not include devices intended for Federal drug testing programs (*e.g.,* programs run by the Substance Abuse and Mental Health Services Administration (SAMHSA), the Department of Transportation (DOT), and the U.S. military).

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Special 510(k) ## AUG 3 0 2002 KO22564 Alfa Scientific Designs Inc. # 510 (k) SUMMARY AS REQUIRED BY SECTION 807.92(c) Product Name: Instant-View™ Multi-Drug Screen Urine Test L ## II. Manufacturer: Alfa Scientific Designs, Inc. 12330 Stowe Drive Poway, CA 92064 (858) 513-3888 Telephone: Fax: (858) 513-8388 info@alfascientific.com Email: ### III. Common Name of the Device: Multiple Drugs of Abuse Urine Test IV. Trade Name of the Device: Instant-View ™Multi-Drug Screen Urine Test ### V. Establishment Registration Number: 2060833 ### VI. Classification of the Device: The Test Device is classified as Class II ## VII. Intended Use: #### The intended use of the modified device, Instant-View™ Multi-Drug Screen Urine Test, has not changed as a result of the modification(s). The Instant-View™ Multi-Drug Screen Urine Test is a qualitative immunoassav intended to be used to detect any combination (variants from 2 to 12) of the following drugs and/or drug metabolites in human urine at the specified cutoff levels. It is intended for professional use only. | Drug | Cutoff Level (ng/ml) | |------------------------|----------------------| | Amphetamine (AMP) | 1000 | | Barbiturate (BAR) | 200 | | Benzodiazepine (BZD) | 300 | | Cocaine (COC) | 300 | | Marijuana (THC) | 50 | | Methadone (MTD) | 300 | | Methamphetamine (METH) | 1000 | | Methamphetamine (METH) | 500 | | Morphine (MOR) | 300 | | Morphine (MOR) | 2000 | | Phencyclidine (PCP) | 25 | {1}------------------------------------------------ Instant-View™ Multi-Drug Screen Urine Test provides only a preliminary analytical test result. A more specific alternate chemical must be used to obtain a confirmed analytical result. Gas Chromatography/Mass Spectrometry (GC/MS) is the preferred confirmatory method. Clinical consideration and professional judgment should be applied to any drug of abuse test result, particularly when preliminary positive results are used. ## VIII. Fundamental Scientific Technology: The Instant-View™ Multi-Drug Screen Urine Test device is an assemblage of single test devices. It works similarly as the single devices because each device functions independently, but conveniently provides the detection of multiple drugs instead of only one in the same time frame. The 510(k)-approved Instant View™ single test devices are used for substantial equivalence predicate kit, for there is no change for each device in the intended use, in the way the test actually runs, and in the expected result. All the devices are based on the same immunochemical principle of recognition and formation. #### IX. Performance Summary: The performance characteristics of Instant-View™ Multi-Drug Screen Urine Test are significantly based on the single test devices, which are independently 510(k) cleared. During the in-house evaluation, no interference was observed between different single test devices while each test device performed normally. Moreover, the study on accelerated degradation of the device demonstrates that the device have a lifetime of at least two years, further indicating that the assembled devices perform independently as single devices and do not affect each other. #### X. Conclusion: For the reasons mentioned above, it is possible to conclude that the Instant-View™ Multi-Drug Screen Urine Test is substantially equivalent to the Instant-View™ single drug of abuse urine tests presently distributed commercially. {2}------------------------------------------------ Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized depiction of an eagle or bird-like figure with three curved lines forming its body and wings. The logo is surrounded by the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged in a circular fashion. Food and Drug Administration 2098 Gaither Road Rockville MD 20850 Naishu Wang, M.D., Ph.D. President Alfa Scientific Designs, Inc. 12330 Stowe Drive Poway, California 92064 AUG 3 0 2002 Re: k022564 > Trade/Device Name: Instant-View™ Multi-Drug Screen Urine Test Regulation Number: 21 CFR § 862.3100 Regulation Name: Amphetamine Test System Regulatory Class: II Product Code: DKZ, DIO, DIS, DJC, DJR, DOE, DPK, JXM, LAF, LCM, LDJ Dated: August 15, 2002 Received: August 19, 2002 Dear Dr. Wang: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices. good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807): labeling (21 CFR Part 801): good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. {3}------------------------------------------------ Page 2 This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4588. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsma/dsmamain.html". Sincerely yours, Steven Sutman Steven I. Gutman, M.D., M.B.A. Director Division of Clinical Laboratory Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {4}------------------------------------------------ # 510(K) NUMBER (IF KNOWN): DEVICE NAME : Instant-View™ Multi-Drug Screen Urine Test #### INDICATIONS FOR USE: The Instant-View™ Multi-Drug Screen Urine Test is a qualitative immunoassay intended to be used to detect any combination (variants from 2 to 12) of the following drugs and/or drug metabolites in human urine at the specified cutoff levels. It is intended for professional use only. | Drug | Cutoff Level (ng/ml) | |------------------------|----------------------| | Amphetamine (AMP) | 1000 | | Barbiturate (BAR) | 200 | | Benzodiazepine (BZD) | 300 | | Cocaine (COC) | 300 | | Marijuana (THC) | 50 | | Methadone (MTD) | 300 | | Methamphetamine (METH) | 1000 | | Methamphetamine (METH) | 500 | | Morphine (MOR) | 300 | | Morphine (MOR) | 2000 | | Phencyclidine (PCP) | 25 | Instant-View™ Multi-Drug Screen Urine Test provides only a preliminary andlytical test result. A more specific alternate chemical method must be used to obtain a confirmed Gas Chromatography/Mass Spectrometry (GC/MS) is the preferred analytical result. confirmatory method. Clinical consideration and professional judgment should be applied to any drug of abuse test result, particularly when preliminary positive results are used. Stan Cooper (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED . ) Concurrence of CDRH, Office of Device Evaluation (ODE) OR Prescription Use (Per 21 CFR 801.109) Over-The-Counter Use (Optional Format 1-2-96)
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...