K022542 · Philips Consumer Electronics · FLL · Aug 20, 2002 · General Hospital
Device Facts
Record ID
K022542
Device Name
PHILIPS BABY THERMOMETER SET (SBC SC530)
Applicant
Philips Consumer Electronics
Product Code
FLL · General Hospital
Decision Date
Aug 20, 2002
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 880.2910
Device Class
Class 2
Attributes
Pediatric
Indications for Use
The Philips Baby Thermometer Set is intended for determination of a baby's oral, rectal, or axillary temperature. The set contains three items: - A pacifier thermometer to detect baby's oral temperature. - An ordinary pacifier can be used as a daily use pacifier to help accustom the baby to the shape of the pacifier-thermometer. - A digital flexible-tip thermometer to detect baby's rectal or underarm (axillary) temperature.
Device Story
Philips Baby Thermometer Set (SBC SC530) comprises a pacifier-style thermometer for oral temperature measurement and a digital flexible-tip thermometer for rectal or axillary measurement. Intended for home use by parents to monitor infant health. The device provides temperature readings to assist in clinical decision-making regarding infant care. The set includes an ordinary pacifier to acclimate the infant to the pacifier-thermometer shape.
Clinical Evidence
No new clinical data provided; relies on clinical testing previously performed for the predicate Paci-Temp and Pro-Flex devices.
Technological Characteristics
Electronic thermometers; flexible-tip probe; pacifier-integrated sensor. Conforms to ASTM1112 standard for electronic thermometers. Meets CPSC 1511 requirements for baby pacifiers regarding choking hazards and EPA requirements for nitro-amines.
Indications for Use
Indicated for determination of oral, rectal, or axillary temperature in infants/babies.
Regulatory Classification
Identification
A clinical electronic thermometer is a device used to measure the body temperature of a patient by means of a transducer coupled with an electronic signal amplification, conditioning, and display unit. The transducer may be in a detachable probe with or without a disposable cover.
Special Controls
(1) Device is not a clinical thermometer with telethermographic functions;
(2) Device is not a clinical thermometer with continuous temperature measurement functions; and
(3) Appropriate analysis and testing (such as that outlined in the currently FDA-recognized editions, as appropriate, of ISO 80601-2-56, “Medical electrical equipment—Part 2-56: Particular requirements for basic safety and essential performance of clinical thermometers for body temperature measurement,” or ASTM E1965, “Standard Specification for Infrared Thermometers for Intermittent Determination of Patient Temperature,” or ASTM E1112, “Standard Specification for Electronic Thermometer for Intermittent Determination of Patient Temperature,” or ASTM E1104, “Standard Specification for Clinical Thermometer Probe Covers and Sheaths”) must validate specifications and performance of the device.
Pro-Flex Digital Thermometer with Flexible Probe (K961357)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows a handwritten document with the text "K022542" at the top. Below this, the text "Page 1 of 2" is written in a smaller font. The handwriting appears to be somewhat rough, but the text is still legible.
Image /page/0/Picture/2 description: The image is a logo for Philips. The logo is a shield shape with the word "PHILIPS" at the top. Inside the shield is a circle with two wavy lines across the middle and two four-pointed stars, one in the upper left and one in the lower right.
## Philips Consumer Electronics Company
## 510(k) Summary
(As required by 21 CFR 807.92)
Premarket Notification Number:
- 1. Submitter's Identification:
Philips Consumer Electronics One Philips Drive, Knoxville, TN 37914 Phone: (865) 521-4701 Fax: (865) 521-3402 Contact: Mr. Bob Pooler
Date Summary Prepared: July 18, 2002
- 2. Name of the Device:
Philips Baby Thermometer Set (SBC SC530)
## 3. Predicate Device Information:
- 1) The Paci-Temp Pacifier Thermometer, K952073, Intelligent Products L.C., Orem, Utah (presently owned by Acute Ideas Co., Ltd.)
- 2) The Pro-Flex Digital Thermometer with Flexible Probe, K961357, Intelligent Products L.C., Orem, Utah (presently owned by Acute Ideas Co., Ltd.)
- 4. Device Description:
The Philips Baby Thermometer set allows parents to accurately and conveniently measure the baby's temperature. The baby thermometer set is designed for he baby's safety, comfort, well-being and, therefore, parents own peace of mind.
The set contains three items:
- A soother thermometer to reliably measure baby's temperature. It is very convenient to ー use both for parents and baby.
Philips Consumer Electronics Company A Division of North American Philips Corporation Product Safety and Compliance One Philips Drive, P.O. Box 14810 Knoxville, TN 37914-1810
{1}------------------------------------------------
K0 22542
page 2 of 2
- A digital flexible-tip thermometer that measures baby's body temperature rectally if parents want added reassurance.
- 5. Intended Use:
The Philips Baby Thermometer Set is intended for determination of baby's temperature. The set contains three items:
- A pacifier thermometer to detect baby's oral temperature. -
- An ordinary pacifier can be used as a daily used regular pacifier -
- A digital flexible-tip thermometer to detect baby's rectal or underarm (axillary) temperature.
- 6. Comparison to Predicate Devices:
The Philips Baby Thermometer Set contains a pacifier thermometer and a digital flexible-tip thermometer which are substantially equivalent to the Pacifier thermometer and Pro-Flex Digital Thermometer with Flexible Probe.
- 7. Discussion of Non-Clinical Tests Performed for Determination of Substantial Equivalence are as follows:
The Philips Baby Thermometer Set conforms to physical requirements and operating parameters outlined in ASTM1112, "Standard Specification for Electronic Thermometer for Intermittent Determination of Patient Temperature". It also meets EPA requirements for nitro-amines and Consumer Product Safety Commission (CPSC 1511) requirements for Baby Pacifiers (choking hazards).
- 8. Discussion of Clinical Tests Performed:
Clinical testing previously performed for the Paci-Temp and Pro-Flex also applies to the Philips Baby Thermometer Set.
- 9. Conclusions:
The Philips Baby Thermometer Set is substantially equivalent in intended use, design, material and technology to the Paci-Temp and Pro-Flex, which contains both of these items in one set. Thus, when compared to the predicate devices, the Philips Baby Thermometer Set does not incorporate any significant changes in intended use, method of operation, material, or design that could affect safety or effectiveness.
{2}------------------------------------------------
Image /page/2/Picture/1 description: The image is a black and white seal for the Department of Health & Human Services USA. The seal is circular, with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" arranged around the perimeter of the circle. Inside the circle is a stylized image of three human profiles facing to the right, with flowing lines above them.
Public Health Service
AUG 2 0 2002
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
Mr. Robert Pooler Manager, Product Safety & Compliance Division Phillips Consumer Electronics One Phillips Drive Knoxville, Tennessee 37914
Re: K022542
Trade/Device Name: Philips Baby Thermometer Set (SBC SC530) Regulation Number: 880.2910 Regulation Name: Regulatory Class: II Product Code: FLL Dated: July 18, 2002 Received: August 1, 2002
Dear Mr. Pooler:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies.
{3}------------------------------------------------
Page 2 - Mr. Pooler
You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act): 21 CFR 1000-1050.
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 21 CFR Part 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4618. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html
Sincerely yours.
Susan Runner
Timothy A. Ulatowski Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
510(k) Number (if known): . . . . . . . . . . . .
______________________________________________________________________________________________________________________________________________________________________________ Philips Baby Thermometer Set (SBC SC530) Device Name :
Indications For Use :
The Philips Baby Thermometer Set is intended for determination of a baby's oral, rectal, or axillary temperature.
The set contains three items:
- A pacifier thermometer to detect baby's oral temperature. .
- An ordinary pacifier can be used as a daily use pacifier to help accustom the baby to . the shape of the pacifier-thermometer.
- A digital flexible-tip thermometer to detect baby's rectal or underarm (axillary) . temperature.
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE)
Prescription Use ( Per 21 CFR 801.109 OR
Over-The-Counter Use ﻣﺴﺴﺴﻪ
Patricio Cucumite
510(k) Number: K022542
(Optional Format 1-2-96)
। ସ
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.