← Product Code [EJT](/productcode/EJT) · K022467

# PORTA IP SOLDER V-2 (K022467)

_Wieland Dental + Technik GmbH & Co. KG · EJT · Sep 17, 2002 · Dental · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K022467

## Device Facts

- **Applicant:** Wieland Dental + Technik GmbH & Co. KG
- **Product Code:** [EJT](/productcode/EJT.md)
- **Decision Date:** Sep 17, 2002
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 872.3060
- **Device Class:** Class 2
- **Review Panel:** Dental

## Indications for Use

Porta IP Solder V-2 is intended for use as a solder before firing in operations in which dental ceramic alloy(s) parts are joined to form a dental restoration. It should be used only together with recommended dental alloys with suitable melting ranges and compositions.

## Device Story

Porta IP Solder V-2 is a yellow dental brazing alloy composed of precious metals (gold, platinum, iridium, silver). Used by dental technicians in laboratory settings to join dental alloy components into restorations like bridges. Applied as filler material before firing process. High corrosion resistance. Benefits patient by enabling fabrication of durable, biocompatible dental restorations. Device is a material, not an active system; no software or automated processing involved.

## Clinical Evidence

Bench testing only. Device complies with international standard ISO 9333 for dental brazing alloys.

## Technological Characteristics

Yellow dental brazing alloy; 99.4% precious metal content (Gold, Platinum, Iridium, Silver). Complies with ISO 9333. Solid form factor for manual application by dental technicians.

## Regulatory Identification

A noble metal alloy is a device composed primarily of noble metals, such as gold, palladium, platinum, or silver, that is intended for use in the fabrication of cast or porcelain-fused-to-metal crown and bridge restorations.

## Special Controls

*Classification.* Class II (special controls). The special control for these devices is FDA's “Class II Special Controls Guidance Document: Dental Noble Metal Alloys.” The devices are exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 872.9. See § 872.1(e) for availability of guidance information.

## Predicate Devices

- Argesol 600 ([K954360](/device/K954360.md))

## Reference Devices

- Porta Geo Ti
- Porta SMK 82
- Portalloy 54
- Simidur S1S
- Simidur KF plus

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

SEP 17 2002 (022467)

## Premarket Notification 510(k)

# Porta IP Solder V-2

### 5. 510 (k) Summary

| Submitter of 510(k): | Wieland Dental + Technik GmbH & Co. KG |
|----------------------|----------------------------------------|
|                      | Schwenninger Str. 13                   |
|                      | D-75179 Pforzheim                      |
|                      | Germany                                |
|                      | Phone: +49-7231-3705-0                 |

| Contact person: | Dr. Gerhard Polzer               |
|-----------------|----------------------------------|
| Phone:          | +49-7231-3705-219                |
| Fax:            | +49-7231-357959                  |
| e-mail:         | gerhard.polzer@wieland-dental.de |

Date of Summary: 2002-06-18

Trade name: PORTA IP SOLDER V-2

Classification name: Product code: C.D.R section: Classification:

Alloy, gold based, for clinical use EJT 872.3060 Class II

Legally marketed equivalent device: Argesol 600 510(k) number: K 954360

#### Device description

Porta IP Solder V-2 is a dental brazing alloy, that can be used by dental technicians to fabricate dental appliances for patients. It is a yellow alloy with very high contents of precious metals (Gold, Platinum, Iridium and Silver: 99,4%).

Porta IP Solder V-2 is suitable for use as a filler material in operations in which dental alloy(s) parts are joined to form a dental restoration, e.g. bridges.

It is intended as solder before firing for dental ceramic alloys with appropriate melting ranges and compositions, especially Porta Geo Ti, Porta SMK 82, Portalloy 54, and Simidur S1S and Simidur KF plus, which are manufactured by Wieland Dental + Technik GmbH & Co. KG, Germany.

Porta IP Solder V-2 is highly corrosion resistant. It fully complies to the international standard ISO 9333 and fulfills the essential requirements of the European directive 93/42/ECC concerning medical devices.

{1}------------------------------------------------

#### DEPARTMENT OF HEALTH & HUMAN SERVICES

Image /page/1/Picture/1 description: The image shows the logo for the Department of Health & Human Services - USA. The logo is a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized image of an eagle with three lines representing its wings or feathers.

#### Public Health Service

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

SFP 1 7 2002

Dr. Gerhard Polzer Director, Regulatory Affairs Wieland Dental + Technik GmbH & Company KG Schwenninger Straße 13 D-75179 Pforzheim GERMANY

Re: K022467

Trade/Device Name: Porta IP Solder V-2 Regulation Number: 21 CFR 872.3060 Regulation Name: Gold-Based Alloys and Precious Metal Alloys for Clinical Use Regulatory Class: II Product Code: EJT Dated: July 22, 2002 Received: July 26, 2002

Dear Dr. Polzer:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting vour device can be found in the Code of Federal Regulations. Title 21. Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies.

{2}------------------------------------------------

Page 2 – Dr. Polzer

You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus. permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 21 CFR Part 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4613. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html

Sincerely yours

Timothy A. Ulatowski

Timoth A. Ulatowski Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{3}------------------------------------------------

Page 1 of 1

510(k) Number orta IP Solder V-2

Device Name:_

Indications For Use:

Porta IP Solder V-2 is intended for use as a solder before firing in operations in which dental ceramic alloy(s) parts are joined to form a dental restoration.

It should be used only together with recommended dental alloys with suitable melting ranges and compositions.

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

RShetz DDS for Dr. Susan Kummer

on of Anesthesiology, General Hospital, Infection Control, Dental

510(k) Number: K022467

ાર

Prescription Use_ (Per 21 CFR 801.109) Over-The-Counter Use__

(Optional Formal 1-2-96)

---

**Source:** [https://fda-staging.innolitics.com/device/K022467](https://fda-staging.innolitics.com/device/K022467)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).