TRANSEND 300 ES GUIDEWIRE, MODEL 46-814 & TRANSEND 300 FLOPPY GUIDEWIRE, MODEL 46-815
K022357 · Boston Scientific, Target · DQX · Oct 16, 2002 · Cardiovascular
Device Facts
Record ID
K022357
Device Name
TRANSEND 300 ES GUIDEWIRE, MODEL 46-814 & TRANSEND 300 FLOPPY GUIDEWIRE, MODEL 46-815
Applicant
Boston Scientific, Target
Product Code
DQX · Cardiovascular
Decision Date
Oct 16, 2002
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 870.1330
Device Class
Class 2
Indications for Use
The Transend 300 ES Guidewire and Transend 300 Floppy Guidewire are intended for general intravascular use, including the neuro and peripheral vasculature. The guidewires can be torqued to facilitate the selective placement of diagnostic or therapeutic catheters. These devices are not intended for use in coronary arteries. A torque device (pin vise) is included with the guidewires to facilitate directional manipulation of the guidewires.
Device Story
Transend 300 ES and 300 Floppy are 0.014" diameter, 300cm length intravascular guidewires; designed for exchange of therapeutic devices without extension wires. Operated by physicians in clinical settings; used to navigate neuro and peripheral vasculature. Manual torque device (pin vise) included for directional manipulation. Facilitates selective placement of diagnostic or therapeutic catheters. Benefits include procedural efficiency by enabling device exchange without additional hardware.
Clinical Evidence
Bench testing only. In vitro performance testing included particle analysis, friction testing, angular rotation, tip buckling, and tip stiffness. Biocompatibility testing performed per ISO-10993.
Indicated for general intravascular use in neuro and peripheral vasculature to facilitate selective placement of diagnostic or therapeutic catheters. Contraindicated for use in coronary arteries.
Regulatory Classification
Identification
A catheter guide wire is a coiled wire that is designed to fit inside a percutaneous catheter for the purpose of directing the catheter through a blood vessel.
Special Controls
*Classification.* Class II (special controls). The device, when it is a torque device that is manually operated, non-patient contacting, and intended to manipulate non-cerebral vascular guide wires, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 870.9.
{0}------------------------------------------------
# SECTION 2 – 510(k) SUMMARY OF SAFETY AND EFFECTIVENESS
#### July 17, 2002 a. Date Prepared:
#### b. Contact Person
George J. Prendergast Regulatory Affairs Specialist II Boston Scientific Target 47900 Bayside Parkway Fremont, CA 94538
#### c. Trade Name
Transend™ 300 ES Guidewire Transend™ 300 Floppy Guidewire
#### d. Common Name
Catheter Guidewire
#### e. Classification Name
Catheter Guidewire (21 CFR Section 870.1330)
#### f. Identification of Predicate Devices
| Number | Description | Clearance Date |
|---------|------------------------------------------------------------------|----------------|
| K950069 | DasherTM - 14 Guidewire | 12 April 1995 |
| K950113 | ChoICETM Extra Support<br>Guidewire<br>ChoICETM Floppy Guidewire | 31 March 1995 |
Transend™ 300 ES Guidewire, Transend™ 300 Floppy Guidewire 510(k) Submission
OCT 1 6 2002
{1}------------------------------------------------
#### g. Device Description
The Transend™300 ES Guidewire and Transend™ 300 Floppy Guidewire are intended for general intravascular use, including the neuro and peripheral vasculature. The 300cm length allows exchange of therapeutic devices without the use of other exchange devices, i.e. extension wires. The Transend™300 ES Guidewire and Transend™ 300 Floppy Guidewire have a nominal diameter of 0.014".
All device materials, manufacturing processes and product dimensions are identical to the Boston Scientific Scimed ChoICE™ ES 300cm exchange guidewire and the ChoICE™ 300cm Floppy exchange guidewire.
#### h. Intended Use
The Transend™300 ES Guidewire and Transend™ 300 Floppy Guidewire Indications For Use are as follows:
The Transend 300 ES Guidewire and Transend 300 Floppy Guidewire are intended for general intravascular use, including the neuro and peripheral vasculature. The guidewires can be torqued to facilitate the selective placement of diagnostic or therapeutic catheters. These devices are not intended for use in coronary arteries. A torque device (pin vise) is included with the guidewires to facilitate directional manipulation of the guidewires.
{2}------------------------------------------------
## i. Product Feature Comparison
| Feature | FasDasher™ – 14<br>Guidewire<br>Predicate device | Transend™ 300 ES<br>Guidewire<br>Transend™ 300 Floppy<br>Guidewire<br>Subject devices | ChoICETM PTCA Exchange<br>Guidewire<br>ChoICETM PTCA Exchange<br>Guidewire<br>Predicate devices |
|-------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Intended Use | FasDasher-14 Guidewire is<br>intended to assist delivery<br>of catheters to selected<br>vascular sites (neuro<br>peripheral and coronary<br>vasculature). | The Transend 300 ES<br>Guidewire and<br>Transend 300 Floppy<br>Guidewire are intended<br>for general intravascular<br>use, including the neuro<br>and peripheral<br>vasculature. These devices<br>are not intended for use in<br>coronary arteries. | The ChoICE ES 300cm<br>Exchange Guidewire<br>ChoICE Floppy 300cm<br>Exchange Guidewire are<br>intended to facilitate the<br>placement of balloon<br>dilatation catheters or<br>other therapeutic devices<br>during PTCA. The ChoICE<br>guidewires are not<br>intended for use in the<br>cerebral vasculature. |
| Design: | | | |
| Length | 195cm | 300cm | 300cm |
| Distal<br>Outside<br>Diameter | 0.014" minimum | 0.014" minimum | 0.014" minimum |
| Materials | Predominately stainless<br>steel, platinum,<br>polytetrafluoroethylene,<br>polyurethane, lubricious<br>coating | Predominately stainless<br>steel, platinum,<br>polytetrafluoroethylene,<br>polyurethane, lubricious<br>coating | Predominately stainless<br>steel, platinum,<br>polytetrafluoroethylene,<br>polyurethane, lubricious<br>coating |
Transend™ 300 ES Guidewire, Transend™ 300 Floppy Guidewire 510(k) Submission
{3}------------------------------------------------
#### i. Comparison of Technological Characteristics
The Transend™300 ES Guidewire and Transend™ 300 Floppy Guidewire are substantially equivalent to the predicate devices in terms of functionality, intended use, design, materials, method of operation, size, length, sterilization methods, packaging, biocompatibility, and labeling.
#### k. Testing
In vitro performance testing of the Transend 300 ES Guidewire and the Transend 300 Floppy Guidewire included particle analysis, friction testing, angular rotation, tip buckling, and tip stiffness to gain clearance for a neurovascular and peripheral indications.
Biocompatibility testing was verified according to ISO-10993, Biological Evaluation of Medical Devices. Test results confirm biocompatibility of the Transend 300 ES Guidewire and the Transend 300 Floppy Guidewire.
All testing indicates that the devices are safe and perform according to their intended use.
### 1. Summary of Substantial Equivalence
Boston Scientific Target's determination of substantial equivalence to Boston Scientific Scimed's ChoICE™ Extra Support Guidewire and ChoICE™ Floppy Guidewire and Boston Scientific Target's FasDasher™- 14 Guidewire as predicate devices is based on the following.
The subject guidewires are substantially equivalent to the FasDasher-14 Guidewire with respect to the following:
- size
- intended use ●
- labeling ●
- . biocompatibility
- packaging
- functionality
- method of operation
Transend™ 300 ES Guidewire, Transend™ 300 Floppy Guidewire 510(k) Submission
Boston Scientific Target
July 2002
{4}------------------------------------------------
The subject guidewires are substantially equivalent to the Scimed's ChoICE™ Extra Support Guidewire and ChoICE™ Floppy Guidewire with respect to the following:
- materials .
- . size
- . length
- . labeling
- packaging ●
- sterilization methods .
- . functionality
- design
- method of operation ●
- biocompatibility ●
Based on the above information provided in this submission, Boston Scientific Target's Transend™ 300 ES Guidewire and Transend™ 300 Floppy Guidewire are substantially equivalent to Boston Scientific Target's FasDasher ™-14 Guidewire and Boston Scientific Scimed's ChoICE Extra Support Guidewire and ChoICE Floppy Guidewire.
Transend™ 300 ES Guidewire, Transend™ 300 Floppy Guidewire 510(k) Submission
{5}------------------------------------------------
Image /page/5/Picture/2 description: The image is a black and white logo for the U.S. Department of Health & Human Services. The logo features a stylized image of an eagle's head and neck, with three distinct lines forming the head and neck. The eagle is facing to the left. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular pattern around the eagle image.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
OCT 1 6 2002
Boston Scientific Target c/o Mr. George J. Prendergast 47900 Bayside Parkway Fremont, CA 94538
K022357 Re:
> Transend™ 300 ES Guidewire and Transend™ 300 Floppy Guidewire Regulation Number: 870.1330 Regulation Name: Catheter Guidewire Regulatory Class: Class II (two) Product Code: DQX Dated: July 17, 2002 Received: July 19, 2002
Dear Mr. Prendergast:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{6}------------------------------------------------
Page 2 - Mr. George J. Prendergast ·
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 21 CFR Part 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4646. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html
Sincerely yours,
Dalatium
Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{7}------------------------------------------------
Image /page/7/Picture/0 description: The image shows the logo for Boston Scientific Target. The words "Boston Scientific" are stacked on top of the word "TARGET". The word "TARGET" is underlined. The font is a serif font.
INDICATIONS FOR USE STATEMENT
510(k) Number: _ K022.357
Device Name: Transend™ 300 ES Guidewire and Transend™ 300 Floppy Guidewire
Indications for Use:
The Transend™ 300 ES Guidewire and Transend™ 300 Floppy Guidewire are intended for general intravascular use, including the neuro and peripheral vasculature. The guidewires can be torqued to facilitate the selective placement of diagnostic or therapeutic catheters. These devices are not intended for use in coronary arteries. A torque device (pin vise) is included to facilitate directional manipulation of the guidewires.
#### (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
OR
Prescription Use
(Per 21 CFR 801.109)
Over The Counter Use
Division of Cardiovascular & Respiratory Devices
510(k) Number K022357
Transend™ 300 ES Guidewire, Transend™ 300 Floppy Guidewire 510(k) Submission
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.