TOTAL BILIRUBIN

K022339 · Abbott Laboratories · CIG · Sep 13, 2002 · Clinical Chemistry

Device Facts

Record IDK022339
Device NameTOTAL BILIRUBIN
ApplicantAbbott Laboratories
Product CodeCIG · Clinical Chemistry
Decision DateSep 13, 2002
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 862.1110
Device ClassClass 2

Indications for Use

The Total Bilirubin assay is used for the quantitation of total bilirubin in human serum and plasma. Measurement of total bilirubin, an organic compound formed during the normal and abnormal destruction of red blood cells, is used in the diagnosis and treatment of liver, hemolytic hematological, and metabolic disorders, including hepatitis and gall bladder block.

Device Story

In vitro diagnostic assay for quantitative determination of total bilirubin in human serum and plasma; utilizes diazo reagent and surfactant to form azobilirubin; absorbance measured at 548 nm; concentration proportional to absorbance increase. Used in clinical laboratory settings on automated systems (AEROSET, ARCHITECT c8000); operated by laboratory personnel. Output provided as concentration value (mg/dL) to healthcare providers for clinical assessment of liver, hemolytic, and metabolic disorders.

Clinical Evidence

Bench testing only. Method comparison studies on AEROSET and ARCHITECT c8000 systems vs. Roche Total Bilirubin assay (Hitachi 717) showed high correlation (r=0.999, slope=0.93). Precision studies (within-run, between-run, between-day) demonstrated total %CVs ranging from 0.6% to 4.6% across four control levels. Assay range 0.1-35.7 mg/dL; limit of quantitation 0.03-0.08 mg/dL.

Technological Characteristics

In vitro colorimetric assay; diazo reagent/surfactant chemistry; 548 nm absorbance measurement; automated system integration; quantitative measurement range 0.1 to 35.7 mg/dL.

Indications for Use

Indicated for the quantitative determination of total bilirubin in human serum and plasma to aid in the diagnosis and treatment of liver, hemolytic hematological, and metabolic disorders, including hepatitis and gall bladder block.

Regulatory Classification

Identification

A bilirubin (total or direct) test system is a device intended to measure the levels of bilirubin (total or direct) in plasma or serum. Measurements of the levels of bilirubin, an organic compound formed during the normal and abnormal distruction of red blood cells, if used in the diagnosis and treatment of liver, hemolytic hematological, and metabolic disorders, including hepatitis and gall bladder block.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ # SEP 1 3 2002 # 510(k) Summary Submitter's Name/Address Abbott Laboratories 1920 Hurd Drive Irving, TX 75038 Contact Person Michele Smith-Waheed Senior Regulatory Specialist MS 1-8 Regulatory Affairs (972) 518-7466 Fax (972) 753-3367 | Date of Preparation of this Summary: | July 15, 2002 | |--------------------------------------------------|-----------------| | Device Trade or Proprietary Name: | Total Bilirubin | | Device Common/Usual Name or Classification Name: | Total Bilirubin | | Classification Number/Class: | CIG, Class II | This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR 807.92. The assigned 510(k) number is: ko223339 #### Test Description: Total Bilirubin is an in vitro diagnostic assay for the quantitative determination of total bilirubin in human serum and plasma. Total (conjugated and unconjugated) bilirubin couples with the diazo reagent in the presence of a surfactant to form azobilirubin. The increase in absorbance at 548 nm due to azobilirubin formation is directly proportional to the total bilirubin concentration. {1}------------------------------------------------ #### Substantial Equivalence: The Total Bilirubin assay is substantially equivalent to the Roche Total Bilirubin assay (K910591) on the Hitachi® 717 Analyzer. Both assays yield similar Performance Characteristics. Similarities: - Both assays are in vitro colorimetric assays. . - Both assays can be used for the quantitative determination of total bilirubin. . - Both assays yield similar clinical results. - Both assays are based on a similar assay principle. . - Human serum and plasma are suitable specimens for both assays. . - . Both assays have similar assay ranges. - . Both assays require a calibration with calibrators. Differences: - Suitable specimens for the Roche Total Bilirubin assay include neonatal specimens. . #### Intended Use: The Total Bilirubin assay is used for the quantitation of total bilirubin in human serum and plasma. {2}------------------------------------------------ #### Performance Characteristics: Comparative performance studies were conducted using the AEROSET® System and ARCHITECT® c8000™ System. The Total Bilirubin assay method comparison yielded acceptable correlation with the Roche Total Bilirubin assay on the Hitachi 717 Analyzer. On the AEROSET System, the correlation coefficient = 0.999, slope = 0.93, and the Y-intercept = 0.16 mg/dL. On the ARCHITECT c8000 System, the correlation coefficient = 0.999, slope = 0.93, and the Y-intercept = 0.15 mg/dL. Precision studies were conducted using the Total Bilirubin assay. Within-run, between-run, and between-day studies were performed using four levels of control material. On the AEROSET System, the total %CV for Level 1 ranged from 2.5 to 4.6%, Level 2 ranged from 1.0 to 1.9%, Level 3 ranged from 0.9 to 1.3%, and Level 4 ranged from 0.9 to 1.2 %. On the ARCHITECT c8000 System, the total %CV for Level 1 ranged from 2.2 to 3.4%, Level 2 ranged from 1.1 to 1.2%, Level 3 ranged from 0.7 to 0.9%, and Level 4 ranged from 0.6 to 0.9%. The Total Bilirubin assay range is 0.1 to 35.7 mg/dL. The limit of quantitation (sensitivity) of the Total Bilirubin assay is 0.03 mg/dL on the AEROSET System and 0.08 mg/dL on the ARCHITECT c8000 System. These data demonstrate that the performance of the Total Bilirubin assay is substantially equivalent to the performance of the Roche Total Bilirubin assay on the Hitachi 717 Analyzer. #### Conclusion: The Total Bilirubin assay is substantially equivalent to the Roche Total Bilirubin assay on the Hitachi 717 Analyzer as demonstrated by results obtained in the studies. {3}------------------------------------------------ #### DEPARTMENT OF HEALTH & HUMAN SERVICES Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a stylized caduceus symbol, which is a staff with two snakes coiled around it, and the words "DEPARTMENT OF HEALTH & HUMAN SERVICES" arranged in a circular pattern around the symbol. The logo is black and white. Food and Drug Administration 2098 Gaither Road Rockville MD 20850 Ms. Michele Smith-Waheed Senior Regulatory Specialist Abbott Laboratories 1920 Hurd Drive Irving, TX 75038 ### SFP 1 3 2002 k022339 Re: > Trade/Device Name: Total Bulirubin Regulation Number: 21 CFR 862.1110 Regulation Name: Bilirubin test system Regulatory Class: Class II Product Code: CIG; JIT Dated: July 15, 2002 Received: July 18, 2002 Dear Ms. Smith Waheed: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirement; as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. {4}------------------------------------------------ Page 2 - This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and 1 additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4588. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.gov/cdrb/dsma/dsmamain.html". Sincerely yours, Steven Butman Steven I. Gutman, M.D., M.B.A. Director Division of Clinical Laboratory-Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {5}------------------------------------------------ 510(k) Number (if known): KO22339 Device Name: _________________________________________________________________________________________________________________________________________________________________ Indications For Use: The Total Bilirubin assay is used for the quantitation of total bilirubin in human serum and plasma. Measurement of total bilirubin, an organic compound formed during the normal and abnormal destruction of red blood cells, is used in the diagnosis and treatment of liver, hemolytic hematological, and metabolic disorders, including hepatitis and gall bladder block. ## (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED) | Concurrence of CDRH, Office of Device Evaluation (ODE) | | | |--------------------------------------------------------|----|----------------------| | Prescription Use | OR | Over-The-Counter Use | | (Per 21 CFR 801.109) | | | (Optional Format 1-2-96) tean (Division Sign-Off) Division of Clinical Labor \$10(k) Number K022329. Total Bilirubin 510(k) 7/15/2002 Bil T_5_RI do 000007
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...