← Product Code [OWB](/productcode/OWB) · K022322

# INNOVA AND INNOVA 2000 S (K022322)

_Ge Medical Systems, Inc. · OWB · Oct 8, 2002 · Radiology · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K022322

## Device Facts

- **Applicant:** Ge Medical Systems, Inc.
- **Product Code:** [OWB](/productcode/OWB.md)
- **Decision Date:** Oct 8, 2002
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 892.1650
- **Device Class:** Class 2
- **Review Panel:** Radiology

## Indications for Use

The Digital Fluoroscopic Imaging Systems are indicated for use in diagnostic and interventional cardiac angiography and optionally, rotational cardiac angiography procedures of human anatomy. It is intended to replace image intensifier fluoroscopic systems in all diagnostic or interventional cardiac angiography procedures.

## Device Story

Digital Fluoroscopic Imaging System (Innova 2000/2000 S) performs fluoroscopic X-ray examinations. System components: angiographic monoplane positioner, table, X-ray system, and amorphous silicon/cesium iodide flat-panel detector. Input: X-ray radiation; Output: digital images. Data transmitted via Fiber Channel to acquisition equipment and network (DICOM) for postprocessing, viewing, printing, and archiving. Used in clinical settings by healthcare professionals. Features include optional rotational cardiac angiography (200° range) and SuperFast Gantry (SFG) with capacitive sensors and collision avoidance software. Hardware/software redundancies prevent single-point failures during motion. Benefits include replacement of traditional image intensifier technology with digital imaging, enabling diagnostic and interventional cardiac procedures.

## Clinical Evidence

No new clinical data provided. The digital technology (flat-panel detectors) was previously clinically compared to image-intensified technology in the predicate submission (K993037) and found substantially equivalent.

## Technological Characteristics

Solid-state X-ray imaging system using amorphous silicon detector with cesium iodide scintillator. Energy source: 480 VAC 50/60Hz. Connectivity: Fiber Channel link, DICOM network. Construction compliant with UL 2601. Features: monoplane positioner, rotational C-arm (200°), capacitive sensor-based collision avoidance. Software includes redundancy for motion control.

## Regulatory Identification

An image-intensified fluoroscopic x-ray system is a device intended to visualize anatomical structures by converting a pattern of x-radiation into a visible image through electronic amplification. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.

## Special Controls

*Classification.* Class II (special controls). An anthrogram tray or radiology dental tray intended for use with an image-intensified fluoroscopic x-ray system only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9. In addition, when intended as an accessory to the device described in paragraph (a) of this section, the fluoroscopic compression device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.

## Predicate Devices

- LCV+ Version 2 ([K993037](/device/K993037.md))
- LCA system ([K945375](/device/K945375.md))

## Reference Devices

- LC system ([K890348](/device/K890348.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows a date, "OCT 08 2002". The month is October, the day is the 8th, and the year is 2002. The text is in a bold, sans-serif font. The image is clear and easy to read.

K022322

GE Medical Systems

P.O. Box 414, W-709 Milwaukee. Wi 53201 USA

## 510(K) SUMMARY OF SAFETY AND EFFECTIVENESS

This 510(k) summary of safety and effectiveness information is submitted in accordance with the requirements of 21 CFR Part 807.87(h),

| Identification of Submitter: | Larry A. Kroger, Ph.D.             |
|------------------------------|------------------------------------|
|                              | Senior Regulatory Programs Manager |
|                              | GE Medical Systems                 |
|                              | Tel. (414) 544-3894                |
|                              | Summary prepared: 5 June, 2002     |

| Identification of Product(s): | Innova 2000 and Innova 2000 S                                                  |
|-------------------------------|--------------------------------------------------------------------------------|
| Classification Name:          | Solid State X-ray Imaging System                                               |
| Manufacturer:                 | GE Medical Systems Europe<br>283, rue de la Minière<br>78530 Buc Cedex, France |
| Distributed by:               | GE Medical Systems, Milwaukee, WI                                              |

Marketed Devices: The Digital Fluoroscopic Imaging Systems are substantially equivalent to the currently marketed cardiographic system LCV+ Version 2 (K993037) that complies with the same or equivalent standards. -The optional Fast Spin rotational angiography feature in the Innova 2000 system is substantially equivalent to the LCA system (K945375) for this feature.

Device Description: The Digital Fluoroscopic Imaging System is designed to perform fluoroscopic x-ray examinations. The detector is comprised of amorphous silicon with a cesium iodide scintillator. The resulting digital image can be sent through a Fiber Channel link to an acquisition equipment then to network (in using DICOM) for applications such as postprocessing, printing, viewing and archiving. Digital Fluoroscopic Imaging System consists of an angiographic monoplane positioner, a table, an X-RAY system and a digital detector.

The Innova 2000 S system is identical to the Innova 2000 system but without the optional features.

{1}------------------------------------------------

New optional features : The offset C-arm permits rotational cardiac angiography over a total 200° at variable speed from 20° to 40°/sec with Cranio/Caudal anqulation.

The SuperFast Gantry (SFG) includes capacitive sensor technology and optimized collision avoidance software that permits an increase of pivot and C-arm speed of up to 20°/sec.

Materials: All construction and materials are compliant with UL 2601.

Design: There are hardware and software redundancies to prevent from single point failures that could cause unintended motion.

Energy Source: 480 VAC 50/60Hz.

Indications for Use: The Digital Fluoroscopic Imaging Systems are indicated for use in generating fluoroscopic images of human anatomy for diagnostic and interventional cardiac angiography. and optionally, rotational cardiac angiography procedures. It is intended to replace fluoroscopic images obtained through the image intensifier technology.

Comparison with

predicate

These systems are substantially equivalent to the Fluoroscopic system LCV+ Version 2 (K993037). The optional Fast Spin rotational angiography feature in the Innova 2000 system is substantially equivalent to the LCA system (K945375) for this feature.

- Summary of Studies: Not applicable as the digital technology (SSXD, i.e., flat panel detectors) has already been clinically compared in the submission of LCV+ Version 2 (K993037) with the image intensified technology system LC introduced in 1989 (cleared under K890348) and found substantially equivalent.
- Conclusions: GE considers the systems to be equivalent with the predicate devices. The systems provide fluoroscopic images that are equivalent to the diagnostic capabilities of the predicate The potential hazards, e.g., wrong device images. measurements, misdiagnosis and increased gantry speeds are controlled by a risk management plan including:
	- · A hazard identification
	- · A risk evaluation
	- · A Software Development and Validation Process

SumSE Rev D.doc

Ch. 1 - 2/2

{2}------------------------------------------------

Image /page/2/Picture/0 description: The image shows the logo for the Department of Health & Human Services. The logo consists of a stylized blue graphic on the left, resembling a human figure, followed by the text "DEPARTMENT OF HEALTH & HUMAN SERVICES" in blue font. The text is aligned horizontally with the graphic.

Public Health Service

Food and Drug Administration 10903 New Hampshire Avenue Document Control Room - WO66-G609 Silver Spring, MD 20993-0002

Larry A. Kroger, Ph.D. Senior Regulatory Programs Manager GE Medical Systems P.O. Box 414 MILWAUKEE WI 53201

JULL 30 2012

Re: K022322

Trade/Device Name: Innova 2000 and Innova 2000S Regulation Number: 21 CFR 892.1650 Regulation Name: Image-intensified fluoroscopic x-ray system Regulatory Class: II Product Code: OWB and JAA Dated: July 16, 2002 Received: July 17, 2002

Dear Dr. Kroger:

This letter corrects our substantially equivalent letter of October 8, 2002.

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into class II (Special Controls), it may be subject to such additional controls. Existing major regulations affecting your device can be found in Title 21, Code of Federal Regulations (CFR), Parts 800 to 895. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); medical device reporting (reporting of

{3}------------------------------------------------

medical device-related adverse events) (21 CFR 803); and good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820). This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Parts 801 and 809), please contact the Office of In Vitro Diagnostic Device Evaluation and Safety at (301) 796-5450. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.

Sincerely Yours,

Janine M. Morris

Acting Director Division of Radiological Devices Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

## STATEMENT OF INTENDED USE

· · 510(k) Number (if known): ________________________________________________________________________________________________________________________________________________

Device Name: Digital Fluoroscopic Imaging System - Innova 2000 and Innova 2000 S

## Indications for Use

. .

. •

The Digital Fluoroscopic Imaging Systems are indicated for use in diagnostic and interventional cardiac angiography and optionally, rotational cardiac angiography procedures of human anatomy. It is intended to replace image intensifier fluoroscopic systems in all diagnostic or interventional cardiac angiography procedures.

## (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Prescription Use (Per 21 CFR 801-109)

OR Over-The-Counter Use_

David A. Elynn

(Division Simm-Off) Proseers of eproductive, Abdominal, . . cal Devices 11223 . : (Yk) N. .: .. .er .

---

**Source:** [https://fda-staging.innolitics.com/device/K022322](https://fda-staging.innolitics.com/device/K022322)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
