← Product Code [LLZ](/productcode/LLZ) · K022310

# ADVANTAGE WINDOWS X-RAY/MR FUSION (K022310)

_Ge Medical Systems, Inc. · LLZ · Aug 1, 2002 · Radiology · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K022310

## Device Facts

- **Applicant:** Ge Medical Systems, Inc.
- **Product Code:** [LLZ](/productcode/LLZ.md)
- **Decision Date:** Aug 1, 2002
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 892.2050
- **Device Class:** Class 2
- **Review Panel:** Radiology
- **Attributes:** Software as a Medical Device, 3rd-Party Reviewed

## Indications for Use

Advantage Windows /MR Fusion provides an easy means for comparison of three dimensional (3D) images from X-ray angiography (XR: providing anatomical imaging) and Magnetic Resonance (MR: providing anatomical imaging). It allows registration between two volumetric acquisitions, which may come from different acquisition modalities (X-Ray and MR), to help physicians in diagnostic radiology or therapy planning.

## Device Story

Advantage Windows X-Ray/MR Fusion is a software package for the Advantage Windows workstation. It takes 3D volumetric image data from X-ray angiography and Magnetic Resonance (MR) as input. The device performs 3D registration between these two volumetric acquisitions, enabling the fusion of angiographic and anatomical images. It is used by physicians in diagnostic radiology or therapy planning to facilitate image comparison. The output is a fused 3D visualization displayed on the workstation, which assists clinicians in diagnostic assessment and planning of therapeutic interventions.

## Clinical Evidence

No clinical data provided. Safety and effectiveness are supported by software development, validation, and verification processes, and adherence to industry and international standards.

## Technological Characteristics

Software-based image processing package for Advantage Windows workstation. Performs 3D registration and fusion of volumetric datasets from X-ray and MR modalities. Operates as an option on existing GE Advantage Windows platforms. Software development follows established verification and validation protocols.

## Regulatory Identification

A medical image management and processing system is a device that provides one or more capabilities relating to the review and digital processing of medical images for the purposes of interpretation by a trained practitioner of disease detection, diagnosis, or patient management. The software components may provide advanced or complex image processing functions for image manipulation, enhancement, or quantification that are intended for use in the interpretation and analysis of medical images. Advanced image manipulation functions may include image segmentation, multimodality image registration, or 3D visualization. Complex quantitative functions may include semi-automated measurements or time-series measurements.

## Special Controls

*Classification.* Class II (special controls; voluntary standards—Digital Imaging and Communications in Medicine (DICOM) Std., Joint Photographic Experts Group (JPEG) Std., Society of Motion Picture and Television Engineers (SMPTE) Test Pattern).

## Predicate Devices

- Advantage Windows Fusion ([K983256](/device/K983256.md))
- Advantage Windows CT/PET Fusion ([K010336](/device/K010336.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the General Electric (GE) logo. The logo consists of the letters "GE" intertwined and enclosed within a circle. There are decorative swirls or flourishes around the outer edge of the circle, giving it a dynamic and stylized appearance. The logo is in black and white.

## AUG 01 2002

GE Medical Systems

K022310

# 510(k) SUMMARY OF SAFETY AND EFFECTIVENESS

This 510(k) summary of safety and effectiveness information is submitted in accordance with the requirements of 21 CFR Part 807.87(h).

- Submitter Larry A. Kroger, Ph.D. Senior Regulatory Program Manager Telephone: (414) 544-3894 Fax: (262) 544-3863 GE Medical Systems 3000 North Grandview Blvd. Waukesha, WI 53188 USA Date Prepared: May 15, 2002.
### PRODUCT IDENTIFICATION

| Name:                | Advantage Windows X-Ray/MR Fusion                                                     |  |
|----------------------|---------------------------------------------------------------------------------------|--|
| Classification Name: | Accessory to X-Ray                                                                    |  |
| Manufacturer :       | General Electric Medical Systems<br>283, rue de la Minière<br>78533 Buc Cedex, FRANCE |  |
| Distributor:         | General Electric Medical Systems, Buc, France.                                        |  |
| Marketed Devices     | The X-Ray/MR Fusion is substantially equivalent to the devices listed below:          |  |
| Model:               | Advantage Windows Fusion (CT/MR Fusion)                                               |  |
| Manufacturer:        | General Electric Medical Systems, Milwaukee, WI                                       |  |
| 510(k) #:            | K983256                                                                               |  |
| Model:               | Advantage Windows Fusion (CT/PET Fusion)                                              |  |
| Manufacturer:        | General Electric Medical Systems, Milwaukee, WI                                       |  |
| 510(k) #:            | K010336                                                                               |  |

#### Device Description:

The GEMS Advantage Windows X-Ray/MR Fusion software package is an option on Advantage Windows that provides easy comparison of three dimensional (3D) images from X-ray angiography and Magnetic Resonance (MR). It allows 3D registration between two volumer. : acquisitions, which may come from different acquisition modalities, producing fusion of angiographic and other anatomical images.

{1}------------------------------------------------

#### Indications for Use:

Advantage Windows /MR Fusion provides an easy means for comparison of three dimensional (3) 1 images from X-ray angiography (XR: providing anatomical imaging) and Magnetic Resonance (Mi providing anatomical imaging). It allows registration between two volumetric acquisitions, which may come from different acquisition modalities (XRay and MR), to help physicians in diagnostic radiology or therapy planning.

#### Comparison with Predicate:

The Advantage Windows X-Ray/MR Fusion option allows merged 3D registration of anatomica. images from X-Rav with anatomical images from MR. The functional features of this package are substantially equivalent to that of the following devices:

| Device Name                     | FDA Clearance Number |
|---------------------------------|----------------------|
| Advantage Windows Fusion        | K983256              |
| Advantage Windows CT/PET Fusion | K010336              |

#### Adverse Effects on Health:

The potential hazards are identified in a risk management summary (hazard analysis) and are controlled by:

- · Software Development, Validation and Verification Process to ensure performance to specifications, Federal Regulations and user requirements.
- · Adherence to industry and international standards.

#### Conclusions:

The Advantage Windows X-Ray/MR Fusion does not result in any new potential safety risks and performs as well as devices currently on the market. GE considers features of the Advantage Windows X-Ray/MR Fusion to be equivalen (K983256) and Advantage Windows (CT/PET) 1 usion (K010336).

{2}------------------------------------------------

Image /page/2/Picture/1 description: The image shows the seal of the Department of Health & Human Services (HHS). The seal features a stylized eagle with three stripes extending from its back, symbolizing health, services, and people. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" are arranged in a circular pattern around the eagle.

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

AUG 01 2002

General Electric Medical Systems % Mr. Wolfram Gmelin Technical Manager TÜV Rheinland of North America 12 Commerce Road NEWTON CT 06470

Re: K022310 Trade/Device Name: Advantage Windows X-ray/MR Fusion Regulation Number: 21 CFR 892.2050 Regulation Name: Picture archiving and communications system Regulatory Class: II Product Code: 90 LLZ Dated: July 16, 2002 Received: July 17, 2002

Dear Mr. Gmelin:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or 10 devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. Iisting of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (sections 531-542 of the Act); 21 CFR 1000-1050.

{3}------------------------------------------------

This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at one of the following numbers, based on the regulation number at the top of this letter:

| 8xx.1xxx                         | (301) 594-4591 |
|----------------------------------|----------------|
| 876.2xxx, 3xxx, 4xxx, 5xxx       | (301) 594-4616 |
| 884.2xxx, 3xxx, 4xxx, 5xxx, 6xxx | (301) 594-4616 |
| 892.2xxx, 3xxx, 4xxx, 5xxx       | (301) 594-4654 |
| Other                            | (301) 594-4692 |

Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers. International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html.

Sincerely yours,

Nancy C. Brogdon

Nancy C. Brogdon Director, Division of Reproductive, Abdominal, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

General Electric Medical Systems

Image /page/4/Picture/1 description: The image shows the General Electric (GE) logo. The logo consists of the letters 'G' and 'E' intertwined within a circular border. The logo is black and white.

## STATEMENT OF INTENDED USE

Device name: XR/MR Fusion

022310

Intended Use:

Advantage Windows /MR Fusion provides an easy means for comparison of three dimensional (3D) images from X-ray angiography (XR: providing anatomical imaging) and Magnetic Resonance (MR: providing anatomical imaging). It allows registration between two volumetric acquisitions, which may come from different acquisition modalities (X-Ray and MR), to help physicians in diagnostic radiology or therapy planning.

(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Prescription Use (Per 21 CFR 801.109)

510(k) Number

-OR-

Over-The-Counter Use _

Nancy C Brigdon

(Division Sign-Off)
Division of Reproductive, Abdominal,
and Radiological Devices 4022310

---

**Source:** [https://fda-staging.innolitics.com/device/K022310](https://fda-staging.innolitics.com/device/K022310)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).