# CARECOMPANION PATIENT STATION & NURSE STATION (K022274)

_Neptec Design Group, Ltd. · DRG · May 29, 2003 · Cardiovascular · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K022274

## Device Facts

- **Applicant:** Neptec Design Group, Ltd.
- **Product Code:** [DRG](/productcode/DRG.md)
- **Decision Date:** May 29, 2003
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 870.2910
- **Device Class:** Class 2
- **Review Panel:** Cardiovascular

## Indications for Use

The Patient Station is intended to be used upon prescription of an authorized healthcare provider by patients as a means to collect and transmit patient vital signs information over standard telephone lines between the patient, typically at home, and a health care professional at the health care provider's site. The information includes: blood pressure, pulse rate, blood glucose level, weight, blood oxygen saturation level and heart, lung and bowel sounds. The information is collected upon request and direction of the healthcare provider. The CareCompanion Patient Station is intended to be used in conjunction with the CareCompanion Nurse Station to provide two-way video, audio and data communications between the patient and the health care professional. The device does not send any real-time alarms. The device is a diagnostic aid. Clinical judgement and experience are required to check and interpret the information transmitted. The device is not intended as a substitute for medical care. The device is contraindicated for patients requiring direct medical supervision or emergency intervention.

## Device Story

Tele-homecare system comprising Patient Station (home) and Nurse Station (clinic). Patient Station integrates vital signs measurement devices (blood pressure, pulse, glucose, weight, pulse oximeter, stethoscope) to collect data; transmits real-time video, audio, and physiological data over standard telephone lines via modems. Nurse Station (PC with peripherals and videophone) receives data for display and database recording; allows caregiver to view/speak with patient and listen to auscultatory sounds via headphones. System acts as diagnostic aid; no real-time alarms. Caregiver interprets data to inform clinical decisions. Benefits include remote monitoring capability, reducing need for in-person visits.

## Clinical Evidence

Bench testing only. Testing validated functional performance of integrated vital signs measurement devices compared to independent operation. Pulse oximeter verified against manufacturer specifications. System subjected to mechanical, electrical, and environmental standards testing.

## Technological Characteristics

System consists of Patient Station and Nurse Station PC with videophone. Connectivity via modems over standard telephone lines. Integrates various vital signs sensors (blood pressure, pulse, glucose, weight, pulse oximeter, stethoscope). Data management via PC database. No specific materials or software architecture details provided.

## Regulatory Identification

A radiofrequency physiological signal transmitter and receiver is a device used to condition a physiological signal so that it can be transmitted via radiofrequency from one location to another, e.g., a central monitoring station. The received signal is reconditioned by the device into its original format so that it can be displayed.

## Predicate Devices

- Neptec CareCompanion ([K020584](/device/K020584.md))

## Reference Devices

- Nonin Palmsat Pulse Oximeter Model 2500 ([K002690](/device/K002690.md))
- 9303 Neonatal/Adult Vital Signs Monitor ([K982776](/device/K982776.md))
- Escort II Modular Telemetry System ([K970763](/device/K970763.md))
- DR180-R/Oxy Holter and Pulse Oximetry Recorder ([K983576](/device/K983576.md))
- Digital EEG and Sleep Acquisition System ([K990522](/device/K990522.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the word "neptec" in a bold, sans-serif font. To the right of the word is a geometric design composed of several cubes. The cubes are arranged in a way that creates a sense of depth and dimension. The overall design is simple and modern.

MAY 2 9 2003

Image /page/0/Picture/2 description: The image shows a document with the text "K022274 page 1 of 3" written at the top. Below the text is a logo with the text "ISO 9001 Registered" inside of it. The logo is a diamond shape with a square inside of it. The text is centered inside of the square.

# 510(k) Summary

#### Submitter

Neptec Design Group Ltd. 302 Legget Drive Kanata, Ontario Canada K2K 1Y5

#### Contacts

John Schneider Neptec Design Group Ltd. Tel: 613-599-7602 Fax: 613-599-7604

Date Prepared

July 5, 2002

# Device Information

| Trade Name:          | CareCompanion Nurse Station / CareCompanion Patient Station  |
|----------------------|--------------------------------------------------------------|
| Common Name:         | Tele Homecare System                                         |
| Classification Name: | Radiofrequency Physiological Signal Transmitter and Receiver |

# Device Description

The CareCompanion System consists of two components: a transportable Patient Station installed typically in a patient's home; and the Nurse Station, installed in a healthcare provider or professional caregiver's office. The two components communicate with each other through modems over standard telephone lines and transmit real-time video, audio and data between them.

1

Tel .: (613) 599-7602 Fax: (613) 599-7604

{1}------------------------------------------------

K022274
page 2 of 3

The real-time video and audio communications allow the patient and the caregiver to view and speak with each other.

Using vital signs measurement devices integrated with the Patient Station, the Patient Station is designed to monitor the patient's blood pressure, pulse rate, blood glucose level, weight, blood oxygen saturation level and/or heart, lung and bowel sounds, and transmit this data to the Nurse Station. The data is displayed to the caregiver operating the Nurse Station and also automatically recorded in a patient information database. The heart, lung and bowel sounds may be listened to by the caregiver using a set of headphones supplied with the system.

The Nurse Station consists of two sub-components, the Nurse Station PC, which is a standard PC with supporting peripherals connected to a videophone, which provides the video conferencing functions for the Nurse Station. The Nurse Station PC may also operate as a standalone device for patient data management and record keeping functions.

#### Substantial Equivalence

The Neptec CareCompanion with the addition of the pulse oximeter is substantially equivalent to the predicate Neptec CareCompanion (#K020584) that has previously been approved by the FDA.

The CareCompanion and its predicate have the same general use to provide the capability for health care professionals to monitor the vitals signs of their patients from remote locations.

The main difference between the systems is the addition of a pulse oximeter for the collection of blood oxygen saturation data.

The pulse oximeter is substantially equivalent to the following FDA approved devices:

- Nonin Palmsat Pulse Oximeter Model 2500 (#K002690) by Nonin Medical, Inc. .
- 9303 Neonatal/Adult Vital Signs Monitor (#K982776) by CAS Medical Systems, Inc. .
- Escort II Modular Telemetry System (#K970763) by Medical Data Electronics, Inc. .
- DR180-R/Oxy Holter and Pulse Oximetry Recorder (#K983576) by Northeast . Monitoring, Inc.
- Digital EEG and Sleep Acquisition System (#K990522) by La Mont Medical. Inc. .

# Intended Use

The Patient Station is intended to be used upon prescription of an authorized healthcare provider by patients as a means to collect and transmit patient vital signs information over standard telephone lines between the patient, typically at home, and a health care professional at the health care provider's site. The information includes: blood pressure, pulse rate, blood glucose level, weight, blood oxygen saturation level and

{2}------------------------------------------------

K022274
page 3 of 3

heart. Jung and bowel sounds. The information is collected upon request and direction of the healthcare provider.

The CareCompanion Patient Station is intended to be used in conjunction with the CareCompanion Nurse Station to provide two-way video, audio and data communications between the patient and the health care professional.

The device does not send any real-time alarms. The device is a diagnostic aid. Clinical iudgement and experience are required to check and interpret the information transmitted. The device is not intended as a substitute for medical care. The device is contraindicated for patients requiring direct medical supervision or emergency intervention.

#### Performance Testing

Testing was performed to validate the functional performance of the CareCompanion. In particular, testing was performed with each vital signs measurement device to show that they operate equivalently when integrated with CareCompanion as when operated as independent devices. The pulse oximeter was tested to verify comformance with manufacturer's specifications.

In addition, the CareCompanion Patient Station and the CareCompanion Nurse Station have been subjected to performance testing to applicable mechanical, electrical and environmental standards.

#### Conclusion

The results of the test indicate that the device is substantially equivalent to its predicate device and does not raise any new questions of safety or effectiveness.

{3}------------------------------------------------

Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized image of three human profiles facing right, stacked on top of each other. The profiles are rendered in black and have a flowing, ribbon-like appearance. Encircling the profile image is the text "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" in a circular arrangement.

Public Health Service

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

MAY 2 9 2003

Neptec Design Group, Ltd. c/o Mr. John Schneider HINS Project Manager 302 Legget Drive Kanata, Ontario Canada K2K 1Y5

Re: K022274

Trade Name: Carecompanion patient station & nurse station Regulation Number: 21 CFR 870.2910 Regulation Name: Radiofrequency physiological signal transmitter and receiver Regulatory Class: Class II (two) Product Code: DRG Dated: March 6, 2003 Received: March 10, 2003

Dear Mr. Schneider:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{4}------------------------------------------------

Page 2 - Mr. John Schneider

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (301) 594-4646. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html

Sincerely yours,

R. Zeller M.D.

Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{5}------------------------------------------------

K02274

Page 1 of 1

510(k) Number (if known):

Device Name: CareCompanion Nurse Station / CareCompanion Patient Station

Indications For Use:

The CareCompanion Patient Station is intended to be used in conjunction with the CareCompanion Nurse Station to provide two-way video, audio and data communications between the patient and the health care professional.

The CareCompanion Patient Station is used upon prescription of an authorized healthcare provider by patients where regular monitoring of vital signs information is indicated. The information is collected from the CareCompanion Patient Station and transmitted over standard telephone lines to a health care professional.

The device does not send any real-time alarms. The device a aid. Clinical judgement and experience are required to check and interpret the information transmitted. The device is not intended as a substitute for medical care. The device is contraindicated for patients requiring direct medical supervision or emergency intervention.

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

# Prescription Use Only

10-98)

(Optional Format 3-

V. Oellette

---

**Source:** [https://fda-staging.innolitics.com/device/K022274](https://fda-staging.innolitics.com/device/K022274)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).