The Patient Station is intended to be used upon prescription of an authorized healthcare provider by patients as a means to collect and transmit patient vital signs information over standard telephone lines between the patient, typically at home, and a health care professional at the health care provider's site. The information includes: blood pressure, pulse rate, blood glucose level, weight, blood oxygen saturation level and heart, lung and bowel sounds. The information is collected upon request and direction of the healthcare provider. The CareCompanion Patient Station is intended to be used in conjunction with the CareCompanion Nurse Station to provide two-way video, audio and data communications between the patient and the health care professional. The device does not send any real-time alarms. The device is a diagnostic aid. Clinical judgement and experience are required to check and interpret the information transmitted. The device is not intended as a substitute for medical care. The device is contraindicated for patients requiring direct medical supervision or emergency intervention.
Device Story
Tele-homecare system comprising Patient Station (home) and Nurse Station (clinic). Patient Station integrates vital signs measurement devices (blood pressure, pulse, glucose, weight, pulse oximeter, stethoscope) to collect data; transmits real-time video, audio, and physiological data over standard telephone lines via modems. Nurse Station (PC with peripherals and videophone) receives data for display and database recording; allows caregiver to view/speak with patient and listen to auscultatory sounds via headphones. System acts as diagnostic aid; no real-time alarms. Caregiver interprets data to inform clinical decisions. Benefits include remote monitoring capability, reducing need for in-person visits.
Clinical Evidence
Bench testing only. Testing validated functional performance of integrated vital signs measurement devices compared to independent operation. Pulse oximeter verified against manufacturer specifications. System subjected to mechanical, electrical, and environmental standards testing.
Technological Characteristics
System consists of Patient Station and Nurse Station PC with videophone. Connectivity via modems over standard telephone lines. Integrates various vital signs sensors (blood pressure, pulse, glucose, weight, pulse oximeter, stethoscope). Data management via PC database. No specific materials or software architecture details provided.
Indications for Use
Indicated for patients requiring regular monitoring of vital signs (blood pressure, pulse, blood glucose, weight, blood oxygen saturation, heart/lung/bowel sounds) via prescription. Contraindicated for patients requiring direct medical supervision or emergency intervention.
Regulatory Classification
Identification
A radiofrequency physiological signal transmitter and receiver is a device used to condition a physiological signal so that it can be transmitted via radiofrequency from one location to another, e.g., a central monitoring station. The received signal is reconditioned by the device into its original format so that it can be displayed.
DR180-R/Oxy Holter and Pulse Oximetry Recorder (K983576)
Digital EEG and Sleep Acquisition System (K990522)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the word "neptec" in a bold, sans-serif font. To the right of the word is a geometric design composed of several cubes. The cubes are arranged in a way that creates a sense of depth and dimension. The overall design is simple and modern.
MAY 2 9 2003
Image /page/0/Picture/2 description: The image shows a document with the text "K022274 page 1 of 3" written at the top. Below the text is a logo with the text "ISO 9001 Registered" inside of it. The logo is a diamond shape with a square inside of it. The text is centered inside of the square.
# 510(k) Summary
#### Submitter
Neptec Design Group Ltd. 302 Legget Drive Kanata, Ontario Canada K2K 1Y5
#### Contacts
John Schneider Neptec Design Group Ltd. Tel: 613-599-7602 Fax: 613-599-7604
Date Prepared
July 5, 2002
# Device Information
| Trade Name: | CareCompanion Nurse Station / CareCompanion Patient Station |
|----------------------|--------------------------------------------------------------|
| Common Name: | Tele Homecare System |
| Classification Name: | Radiofrequency Physiological Signal Transmitter and Receiver |
# Device Description
The CareCompanion System consists of two components: a transportable Patient Station installed typically in a patient's home; and the Nurse Station, installed in a healthcare provider or professional caregiver's office. The two components communicate with each other through modems over standard telephone lines and transmit real-time video, audio and data between them.
1
Tel .: (613) 599-7602 Fax: (613) 599-7604
{1}------------------------------------------------
K022274
page 2 of 3
The real-time video and audio communications allow the patient and the caregiver to view and speak with each other.
Using vital signs measurement devices integrated with the Patient Station, the Patient Station is designed to monitor the patient's blood pressure, pulse rate, blood glucose level, weight, blood oxygen saturation level and/or heart, lung and bowel sounds, and transmit this data to the Nurse Station. The data is displayed to the caregiver operating the Nurse Station and also automatically recorded in a patient information database. The heart, lung and bowel sounds may be listened to by the caregiver using a set of headphones supplied with the system.
The Nurse Station consists of two sub-components, the Nurse Station PC, which is a standard PC with supporting peripherals connected to a videophone, which provides the video conferencing functions for the Nurse Station. The Nurse Station PC may also operate as a standalone device for patient data management and record keeping functions.
#### Substantial Equivalence
The Neptec CareCompanion with the addition of the pulse oximeter is substantially equivalent to the predicate Neptec CareCompanion (#K020584) that has previously been approved by the FDA.
The CareCompanion and its predicate have the same general use to provide the capability for health care professionals to monitor the vitals signs of their patients from remote locations.
The main difference between the systems is the addition of a pulse oximeter for the collection of blood oxygen saturation data.
The pulse oximeter is substantially equivalent to the following FDA approved devices:
- Nonin Palmsat Pulse Oximeter Model 2500 (#K002690) by Nonin Medical, Inc. .
- 9303 Neonatal/Adult Vital Signs Monitor (#K982776) by CAS Medical Systems, Inc. .
- Escort II Modular Telemetry System (#K970763) by Medical Data Electronics, Inc. .
- DR180-R/Oxy Holter and Pulse Oximetry Recorder (#K983576) by Northeast . Monitoring, Inc.
- Digital EEG and Sleep Acquisition System (#K990522) by La Mont Medical. Inc. .
# Intended Use
The Patient Station is intended to be used upon prescription of an authorized healthcare provider by patients as a means to collect and transmit patient vital signs information over standard telephone lines between the patient, typically at home, and a health care professional at the health care provider's site. The information includes: blood pressure, pulse rate, blood glucose level, weight, blood oxygen saturation level and
{2}------------------------------------------------
K022274
page 3 of 3
heart. Jung and bowel sounds. The information is collected upon request and direction of the healthcare provider.
The CareCompanion Patient Station is intended to be used in conjunction with the CareCompanion Nurse Station to provide two-way video, audio and data communications between the patient and the health care professional.
The device does not send any real-time alarms. The device is a diagnostic aid. Clinical iudgement and experience are required to check and interpret the information transmitted. The device is not intended as a substitute for medical care. The device is contraindicated for patients requiring direct medical supervision or emergency intervention.
#### Performance Testing
Testing was performed to validate the functional performance of the CareCompanion. In particular, testing was performed with each vital signs measurement device to show that they operate equivalently when integrated with CareCompanion as when operated as independent devices. The pulse oximeter was tested to verify comformance with manufacturer's specifications.
In addition, the CareCompanion Patient Station and the CareCompanion Nurse Station have been subjected to performance testing to applicable mechanical, electrical and environmental standards.
#### Conclusion
The results of the test indicate that the device is substantially equivalent to its predicate device and does not raise any new questions of safety or effectiveness.
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized image of three human profiles facing right, stacked on top of each other. The profiles are rendered in black and have a flowing, ribbon-like appearance. Encircling the profile image is the text "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" in a circular arrangement.
Public Health Service
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
MAY 2 9 2003
Neptec Design Group, Ltd. c/o Mr. John Schneider HINS Project Manager 302 Legget Drive Kanata, Ontario Canada K2K 1Y5
Re: K022274
Trade Name: Carecompanion patient station & nurse station Regulation Number: 21 CFR 870.2910 Regulation Name: Radiofrequency physiological signal transmitter and receiver Regulatory Class: Class II (two) Product Code: DRG Dated: March 6, 2003 Received: March 10, 2003
Dear Mr. Schneider:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{4}------------------------------------------------
Page 2 - Mr. John Schneider
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (301) 594-4646. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html
Sincerely yours,
R. Zeller M.D.
Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
K02274
Page 1 of 1
510(k) Number (if known):
Device Name: CareCompanion Nurse Station / CareCompanion Patient Station
Indications For Use:
The CareCompanion Patient Station is intended to be used in conjunction with the CareCompanion Nurse Station to provide two-way video, audio and data communications between the patient and the health care professional.
The CareCompanion Patient Station is used upon prescription of an authorized healthcare provider by patients where regular monitoring of vital signs information is indicated. The information is collected from the CareCompanion Patient Station and transmitted over standard telephone lines to a health care professional.
The device does not send any real-time alarms. The device a aid. Clinical judgement and experience are required to check and interpret the information transmitted. The device is not intended as a substitute for medical care. The device is contraindicated for patients requiring direct medical supervision or emergency intervention.
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
# Prescription Use Only
10-98)
(Optional Format 3-
V. Oellette
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.