K022191 · Medtronic Sofamor Danek, Inc. · KWP · Aug 28, 2002 · Orthopedic

Device Facts

Record IDK022191
Device NameMODIFICATION TO TENOR SPINAL SYSTEM
ApplicantMedtronic Sofamor Danek, Inc.
Product CodeKWP · Orthopedic
Decision DateAug 28, 2002
DecisionSESE
Submission TypeSpecial
Regulation21 CFR 888.3050
Device ClassClass 2
AttributesTherapeutic

Indications for Use

The TENOR™ Spinal System, when used for pedicle screw fixation, is intended only for patients: (a) having severe spondylolisthesis (Grades 3 and 4) of the fifth lumbar -- first sacral (L5-S1) vertebral joint; (b) who are receiving fusions using autogenous bone graft only: (c) who are having the device fixed or attached to the lumbar and sacral spine (levels of pedicle screw fixation may be from L3 to sacrum); and/or (d) who are having the device removed after the development of a solid fusion mass. TENOR™ Plates are intended for the L5-S1 pedicle screw indication described above only. The TENORTM Spinal System, when used as a non-pedicle screw fixation system, is intended for the following indications: 1) degenerative disc disease (as defined by back pain of discogenic origin with degeneration of the disc confirmed by patient history and radiographic studies), 2) pseudarthrosis, 3) stenosis, 4) spondy|olisthesis, 5) spinal deformities: scoliosis, kyphosis, 6) fracture, 7) unsuccessful previous attempts at spinal fusion, and/or 8) tumor resection. When used for posterior non-pedicle screw fixation, the TENORTM System is intended for thoracic, lumbar, and sacral (T1-Sacrum) fixation only.

Device Story

Spinal stabilization system providing temporary, bilateral fixation to augment spinal fusion. Components include titanium alloy or stainless steel clamps, cross connectors, nuts, and screws. Used in conjunction with GDLH™ 5.5mm rods, TSRH® hooks/connectors, and various CROSSLINK® plates. Surgeon-implanted in OR; components assembled to fit patient-specific anatomy. Provides mechanical support to spine during fusion process; removed after solid fusion mass development. Benefits patient by stabilizing spinal segments, facilitating bone healing, and correcting deformities.

Clinical Evidence

No clinical data provided; bench testing only.

Technological Characteristics

Materials: medical grade titanium alloy or stainless steel. Components: clamps, cross connectors, nuts, screws. System is modular, designed for posterior thoracic, lumbar, and sacral fixation. Mechanical stabilization principle.

Indications for Use

Indicated for patients requiring spinal stabilization/fusion. Pedicle screw use: severe spondylolisthesis (Grades 3-4) at L5-S1; requires autogenous bone graft; fixation levels L3-sacrum. Non-pedicle screw use: degenerative disc disease, pseudarthrosis, stenosis, spondylolisthesis, spinal deformities (scoliosis, kyphosis), fracture, failed fusion, or tumor resection; fixation levels T1-sacrum.

Regulatory Classification

Identification

A spinal interlaminal fixation orthosis is a device intended to be implanted made of an alloy, such as stainless steel, that consists of various hooks and a posteriorly placed compression or distraction rod. The device is implanted, usually across three adjacent vertebrae, to straighten and immobilize the spine to allow bone grafts to unite and fuse the vertebrae together. The device is used primarily in the treatment of scoliosis (a lateral curvature of the spine), but it also may be used in the treatment of fracture or dislocation of the spine, grades 3 and 4 of spondylolisthesis (a dislocation of the spinal column), and lower back syndrome.

Reference Devices

Submission Summary (Full Text)

{0}------------------------------------------------ AUG 2 8 2002 K022/91 page 1 of 1 TENOR™ Spinal System 510(k) Summary August 2002 I. Medtronic Sofamor Danek USA Company: 1800 Pyramid Place Memphis, TN 38132 (901) 396-3133 ### II. Proposed Proprietary Trade Name: TENOR™ Spinal System - III. Classification Name: Orthosis, Spondylolisthesis Spinal Fixation (per 21 CFR Section 888.3070) ### IV. Product Description The TENOR™ Spinal System is a spinal device intended to provide temporary, bilateral stabilization and augment the development of a solid spinal fusion. The system comprises a variety of shapes and sizes of clamps, cross connectors, nuts, and screws made of medical grade titanium alloy or stainless steel. The TENOR™ Spinal System is used in conjunction with GDLH™ 5.5mm rods, TSRH® hooks and connectors, TSRH® Low Profile CROSSLINK® plates, CD HORIZON( Low Profile MULTI-SPAN™ CROSSLINK® plates and/or MULTI AXIAL Low Profile MULTI-SPAN™ CROSSLINK® plates for attachment to the posterior thoracic and lumbar spine. These components are assembled to fit the patient's specific anatomic needs. The purpose of this submission is to include additional screws to the system. ### V. Indications The TENOR™ Spinal System, when used for pedicle screw fixation, is intended only for patients: (a) having severe spondylolisthesis (Grades 3 and 4) of the fifth lumbar -- first sacral (L5-S1) vertebral joint; (b) who are receiving fusions using autogenous bone graft only: (c) who are having the device fixed or attached to the lumbar and sacral spine (levels of pedicle screw fixation may be from L3 to sacrum); and/or (d) who are having the device removed after the development of a solid fusion mass. TENOR™ Plates are intended for the L5-S1 pedicle screw indication described above only. The TENORTM Spinal System, when used as a non-pedicle screw fixation system, is intended for the following indications: 1) degenerative disc disease (as defined by back pain of discogenic origin with degeneration of the disc confirmed by patient history and radiographic studies), 2) pseudarthrosis, 3) stenosis, 4) spondy|olisthesis, 5) spinal deformities: scoliosis, kyphosis, 6) fracture, 7) unsuccessful previous attempts at spinal fusion, and/or 8) tumor resection. When used for posterior non-pedicle screw fixation, the TENORTM System is intended for thoracic, lumbar, and sacral (T1-Sacrum) fixation only. ### VI. Substantial Equivalence Documentation was provided that demonstrated the TENOR™ Spinal System to be substantially equivalent to itself. {1}------------------------------------------------ ## DEPARTMENT OF HEALTH & HUMAN SERVICES Image /page/1/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is an abstract symbol that resembles an eagle or bird in flight, composed of three stylized human profiles facing to the right. Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 AUG 2 8 2002 Mr. Richard W. Treharne, Ph.D. Senior Vice President, Regulatory Affairs Medtronic Sofamor Danek 1800 Pyramid Place Memphis, Tennessee 38132 Re: K022191 Trade/Device Name: TENOR® Spinal System Regulation Number: 21 CFR §888.3050 and §888.3070 Regulation Name: Spinal interlaminal fixation orthosis and Pedicle screw spinal system Regulatory Class: Class II Product Code: KWP and MNH Dated: July 3, 2002 Received: July 5, 2002 Dear Dr. Treharne: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. {2}------------------------------------------------ # Page 2 – Mr. Richard W. Treharne, Ph.D. This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 21 CFR Part 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4659. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html Sincerely yours, R. Mark N. Milliken Celia M. Witten, Ph.D., M.D. Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health {3}------------------------------------------------ ## 510/k) Number (if known): KO22 (9) #### TENOR™ Spinal System Device Name: ## Indications for Use: The TENOR™ Spinal System, when used for pedicle screw fixation, is intended only for patients: (a) having severe spondylolisthesis (Grades 3 and 4) of the fifth lumbar -- first sacral (LS-S1) vertebral joint; (b) who are receiving fusions using autogenous bone graft only; (c) who are having the device fixed or attached to the lumbar and sacral spine (levels of pedicle screw fixation may be from L3 to sacrum); and/or (d) who are having the device removed after the development of a solid fusion mass. TENOR™ Plates are intended for the L5-S1 pedicle screw indication described above only. The TENOR™ Spinal System, when used as a non-pedicle screw fixation system, is intended for the following indications: 1) degenerative disc disease (as defined by back pain of discogenic origin with degeneration of the disc confirmed by patient history and radiographic studies), 2) pseudatthrosis, 3) stenosis, 4) spondylolisthesis, 5) spinal deformities: scoliosis, lordosis, 6) fracture, 7) unsuccessful previous attempts at spinal fusion, and/or 8) tumor resection. When used for posterior non-pedicle screw fixation, the TENOR™ Spinal System is intended for thoracic, lumbar, and sacral (T1-Sacrum) fixation only. (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) ## Concurrence of CDRH, Office of Evaluation (ODE) Prescription Use (Per 21 CFR 801.109) (Optional 1-2-96) OR Over-the-counter Use Mark N Millman ivision Sign-Off) Division of General, Restorative and Neurological Devices 510(k) Number K0221960047
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...