MODIFICATION TO CARDEON AEGIS CATHETER

K022174 · Cardeon Corp. · DWF · Jul 29, 2002 · Cardiovascular

Device Facts

Record IDK022174
Device NameMODIFICATION TO CARDEON AEGIS CATHETER
ApplicantCardeon Corp.
Product CodeDWF · Cardiovascular
Decision DateJul 29, 2002
DecisionSESE
Submission TypeSpecial
Regulation21 CFR 870.4210
Device ClassClass 2
AttributesTherapeutic

Indications for Use

The Cardeon® Aegis™ Catheter is intended to perfuse the aorta during open chest (sternotomy) procedures on cardiopulmonary bypass up to 6 hours.

Device Story

Cardeon® Aegis™ Catheter is a vascular cannula used during open chest (sternotomy) cardiac surgery. Device directly cannulates aorta to return perfusion from extracorporeal circuit to patient. Operated by surgical team in OR. Design facilitates directed flow pattern for arterial return. Device supports cardiopulmonary bypass for up to 6 hours. Clinical benefit includes safe, effective arterial perfusion during cardiac procedures.

Clinical Evidence

Bench testing only. Assessment included biocompatibility, in vitro, and in vivo performance testing comparing modified Aegis device to original Aegis and predicate devices. Results demonstrated safety and effectiveness equivalent to predicates.

Technological Characteristics

Vascular cannula for cardiopulmonary bypass. Features directed flow pattern for arterial return. Designed for standard connection to extracorporeal circuits. Class II device (21 CFR 870.4210).

Indications for Use

Indicated for patients undergoing open chest (sternotomy) procedures requiring cardiopulmonary bypass for up to 6 hours.

Regulatory Classification

Identification

A cardiopulmonary bypass vascular catheter, cannula, or tubing is a device used in cardiopulmonary surgery to cannulate the vessels, perfuse the coronary arteries, and to interconnect the catheters and cannulas with an oxygenator. The device includes accessory bypass equipment.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ ## 510(k) SUMMARY ## 510(k) number: K022174 This summary of information is being provided for the modified Aegis Catheter in accordance with 21 CFR 807.92(a). | Applicant Information: | Cardeon Corporation<br>10161 Bubb Road<br>Cupertino, CA 95014 | |------------------------|---------------------------------------------------------------| | Contact Person: | Jane Beggs<br>Regulatory Affairs<br>Cardeon Corporation | | Date: | 15 July 2002 | | Device Trade Name: | Cardeon® Aegis™ Catheter | | Device Common Name | Catheter, cannula and tubing, vascular, cardiopulmonary | | Regulation No.: | bypass 870.4210 | | Classification / Code: | Class II / DWF | Indications for Use: The Cardeon® Aegis™ Catheter is intended to perfuse the aorta during open chest (sternotomy) procedures on cardiopulmonary bypass up to 6 hours. Summary of Substantial Equivalence: The modified Aegis Catheter is substantially equivalent to the original Aegis Catheter and currently marketed devices with regard to intended use/indications, device performance and technological characteristics. These devices are used to directly cannulate the aorta and return perfusion to patients undergoing general cardiac surgery through standard connections to the extracorporeal circuit. Substantial equivalence is supported through comparison with several marketed devices with the same indications for use, including arterial return cannulae with directed flow (specified flow pattern). Differences between the Aegis Catheter and other devices do not alter or diminish the rate of arterial return required during cardiopulmonary bypass. Based on comparisons to premarket devices, the modified Aegis Catheter is substantially equivalent to the original Aegis Catheter and currently marketed devices. Product Testing: The determination of substantial equivalence was also based on an assessment of original and modified Aegis devices. Tests for biocompatibility, in vitro and/or in vivo performance were conducted. Results of product testing demonstrated that the Aegis Catheter functions as safely and effectively as predicate devices. {1}------------------------------------------------ Image /page/1/Picture/2 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle or bird symbol with three human profiles incorporated into its design. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular fashion around the bird symbol. Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 JUL 2 9 2002 Cardeon Corporation c/o Ms. Jane Beggs Vice President, Regulatory Affairs 10161 Bubb Road Cupertino, CA 95014 Re: K022174 Trade Name: Cardeon® Aegis™ Catheter Regulation Number: 21 CFR 870.4210 Regulation Name: Cardiopulmonary bypass vascular Catheter, Cannula, or Tubing. Regulatory Class: Class II (two) Product Code: DWF Dated: July 2, 2002 Received: July 3, 2002 Dear Ms. Beggs: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. {2}------------------------------------------------ Page 2 - Ms. Jane Beggs Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 21 CFR Part 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4646. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html Sincerely vours. Aepant Mayn Donna-Bea Tillman, Ph.D. Acting Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {3}------------------------------------------------ 1022174 510(k) Number (if known): _ K422474 Page 1 of 1 Device Name: Cardeon® Aegis™ Catheter Indications for Use: The Cardeon Aegis™ Catheter is intended to perfuse the aorta during open chest procedures on cardiopulmonary bypass up to 6 hours. Concurrence of CDRH, Office of Device Evaluation (ODE) Prescription Use OR Over-the Counter Use _________ M.G. Van Meter for D.B. Tillman Richard L. Moore ion Sibn-Off Division of Cardiovascular and Respiratory Devices 510(k) Number K022174
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...