The Accu-Chek Compact system is designed to quantitatively measure the concentration of glucose in capillary and venous whole blood. The device is indicated for professional use and over-the-counter sale. The Accu-Chek Compact system is indicated for lay person use with capillary whole blood samples drawn from the fingertips, forearm, upper arm, thigh, calf, and palm.
Device Story
Accu-Chek Compact System is a glucose monitoring device for professional and home use. It uses photometric analysis of a chemical reaction (glucose dehydrogenase) to measure blood glucose concentration. Input: capillary or venous whole blood sample (3.5 µL). Operation: user inserts test drum into meter; meter automatically reads code from drum. Output: quantitative glucose concentration (10-600 mg/dL) displayed to user. Device supports data transmission to external devices. Benefits: enables patient self-monitoring of blood glucose at multiple body sites (fingertips, forearm, upper arm, thigh, calf, palm), facilitating diabetes management. Healthcare providers use results to assess glycemic control and adjust treatment plans.
Clinical Evidence
Bench testing and clinical correlation study. Study demonstrated good correlation (r > 0.90) between AST results (calf, thigh, upper arm, forearm) and fingerstick results under steady-state conditions. Accuracy of AST blood from calf, thigh, and upper arm was shown to be unchanged from forearm results.
Technological Characteristics
Photometric glucose monitoring system. Reagent: Glucose-dye-oxidoreductase (recombinant from E. Coli). Sample volume: 3.5 µL. Hematocrit range: 25-65%. Power: two 1.5V AAA alkaline batteries. Dimensions: 4" L x 2" W x 1.5" H. Features automatic code reading from test drum. Connectivity: data transmission to external devices. Operating conditions: 50°F to 104°F.
Indications for Use
Indicated for professional and lay person use to quantitatively measure glucose in capillary and venous whole blood. Suitable for testing samples from fingertips, forearm, upper arm, thigh, calf, and palm. Not suitable for neonate samples.
Regulatory Classification
Identification
A glucose test system is a device intended to measure glucose quantitatively in blood and other body fluids. Glucose measurements are used in the diagnosis and treatment of carbohydrate metabolism disorders including diabetes mellitus, neonatal hypoglycemia, and idiopathic hypoglycemia, and of pancreatic islet cell carcinoma.
Special Controls
*Classification.* Class II (special controls). The device, when it is solely intended for use as a drink to test glucose tolerance, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 862.9.
{0}------------------------------------------------
# JUL 2 3 2002
| 510(k) Summary | K022171 |
|----------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Introduction | According to the requirements of 21 CFR 807.92, the following information<br>provides sufficient detail to understand the basis for a determination of<br>substantial equivalence. |
| 1) Submitter<br>name, address,<br>contact | Roche Diagnostics Corporation<br>9115 Hague Rd.<br>Indianapolis, IN 46250<br>(317) 845-2000<br>Contact Person: Scott Thiel<br>Date Prepared: July 2, 2002 |
| 2) Device name | Proprietary name: Accu-Chek Compact System<br>Classification name: Glucose Dehydrogenase, Glucose<br>(21 C.F.R. § 862.1345)(75LFR) |
| 3) Predicate<br>device | We claim substantial equivalence to the current legally marketed Accu-Chek<br>Compact System (K004010). |
| 4) Device<br>Description | Instrument Operating Principle -- photometry<br>Reagent Test Principle -- glucose dehydrogenase |
| 5) Intended use | The Accu-Chek Compact system is designed to quantitatively measure the<br>concentration of glucose in capillary and venous whole blood. The device is<br>indicated for professional use and over-the-counter sale. |
| | The Accu-Chek Compact system is indicated for lay person use with capillary<br>whole blood samples drawn from the fingertips, forearm, upper arm, thigh,<br>calf, and palm. |
| | Continued on next page |
| Feature/Claim | Detail |
| Intended use | The Accu-Chek Compact system is designed to quantitatively measure<br>the concentration of glucose in whole blood. The test device is indicated<br>for professional use and over-the-counter sale. This device is not suitable<br>for testing neonate samples. |
| Test principle | Glucose dehydrogenase chemical reaction. The instrument measures the<br>extent of color change (photometric) caused by the presence of glucose in<br>the sample. The amount of color change is related to the glucose<br>concentration in the blood sample. |
| Monitor coding<br>procedure | The code is automatically read from the test drum upon insertion of the<br>test drum into the meter. |
| Test strip storage<br>conditions | Store at room temperature between +36° F(+2° C) and +86° F(+30° C). |
| Test strip operating<br>conditions | Between +50° F(+10° C) and +104° F(+40° C). |
| Quality control<br>acceptable range | The mean is strip lot specific and will be determined individually. The<br>range of the controls is within ± 15 mg/dL or ± 15% compared to the<br>determined mean. |
| Labeling<br>instructions<br>regarding expected<br>results | The normal fasting adult blood glucose range for a non-diabetic is 70-105<br>mg/dL. One to two hours after meals, normal blood glucose levels should<br>be less than 140 mg/dL. Doctors will determine the range that is<br>appropriate for the patients. |
| Labeling<br>instructions<br>regarding response<br>to unusual results | Run a quality control test, if the result is outside the acceptable QC<br>recovery range contact Roche Diagnostics's Accu-Chek Customer Care<br>center; if result is within the acceptable range, review proper testing<br>procedure and repeat blood glucose test with a new test strip. |
| Acceptable sample<br>types | Capillary and venous whole blood samples |
19
{1}------------------------------------------------
## 510(k) Summary, Continued
The proposed modification is relatively modest in scope. All of the following 6) Similarities are claims and features unaffected by the proposed modification.
Continued on next page
{2}------------------------------------------------
# 510(k) Summary, Continued
Similarities,
Contd.
| Reportable range | 10-600 mg/dL |
|-----------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Hematocrit range | 25 – 65% |
| Warnings and precautions | For in vitro diagnostic use only. |
| Reagent active ingredients | Glucose-dye-oxidoreductase* Bis-(2-hydroxyethyl)-(4-hydroximinocyclohexa-2,5-dienylidene)-ammonium chloride 2,18-Phosphomolybdic acid *(from A. Calcoaceticus, recombinant from E. Coli) |
| Minimum sample volume | 3.5 µL |
| Under-dosed test strip detection method | Yes |
| Dosing test strips outside of meter | No |
| Meter physical dimensions | 4" L x 2" W x 1 1/2" H |
| Batteries required | Two 1.5 volt AAA alkaline batteries |
| Data transmission to external devices | Yes |
and the comments of the comments of the contraction of the contribution of the contribution of the contribution of the contribution of the contribution of the contribution of
Continued on next page
{3}------------------------------------------------
## 510(k) Summary, Continued
#### Difference
| Feature | Accu-Chek Compact System<br>(modified) | Accu-Chek Compact System<br>(predicate) |
|---------------------------------------------------------------------|------------------------------------------------------|-----------------------------------------|
| Self-testing<br>capillary whole<br>blood sample<br>collection sites | Finger, forearm, calf, thigh, upper<br>arm, and palm | Finger |
#### Performance Characteristics
The results of a study conducted at our manufacturing facility demonstrated consistent quality performance of this product. This study demonstrated good correlation (r> 0.90) between AST results and finger results under steady state conditions. With these data it is proved that the system accuracy with AST blood from calf, thigh and upper arm is unchanged from forearm.
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" arranged around the perimeter. Inside the circle is an abstract symbol that resembles three stylized human profiles facing to the right, stacked on top of each other.
#### DEPARTMENT OF HEALTH & HUMAN SERVICES
Food and Drug Administration 2098 Gaither, Road Rockville MD 20850
### JUL 2 3 2002
Mr. Scott Thiel Regulatory Affairs Specialist Roche Diagnostics Corporation 9115 Hague Road P.O. Box 50457 Indianapolis, IN 46250-0457
Re: k022171
> Trade/Device Name: Accu-Chek Compact Test System Regulation Number: 21 CFR 862.1345 Regulation Name: Glucose test system Regulatory Class: Class II Product Code: NBW: LFR Dated: July 2, 2002 Received: July 3, 2002
Dear Mr. Thiel:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. Iisting of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations. Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{5}------------------------------------------------
Page 2 -
.
This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and ' additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4588. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.gov/cdrb/dsmamain.html".
Sincerely yours,
Steven Butman
Steven I. Gutman, M.D., M.B.A. Director Division of Clinical Laboratory-Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{6}------------------------------------------------
### Indications for Use Statement
## 510(k) Number (if known): Device Name: Accu-Chek Compact Test System
Indications for Use:
The Accu-Chek Compact system is designed to quantitatively measure the concentration of glucose in capillary and venous whole blood. The device is indicated for professional use and over-the-counter sale.
The Accu-Chek Compact system is indicated for lay person use with capillary whole blood samples drawn from the fingertips, forearm, upper arm, thigh, calf, and palm.
#### (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Prescription Use (Per 21 CFR 801.109)
OR
Over-The-Counter Use
(Optional Format 1-2-96)
Sean Cooper
ivision Sign-Off livision of Clinical Laboratory D 510(K) Number
-
-
-
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.