K022168 · Mennen Medical , Ltd. · DQA · Aug 1, 2002 · Cardiovascular
Device Facts
Record ID
K022168
Device Name
ENVOY
Applicant
Mennen Medical , Ltd.
Product Code
DQA · Cardiovascular
Decision Date
Aug 1, 2002
Decision
SESE
Submission Type
Special
Regulation
21 CFR 870.2700
Device Class
Class 2
Attributes
Pediatric
Indications for Use
The Envoy is intended for use as a multiparameter physiological patient monitoring system. The Envoy can monitor ECG/heart rate, two invasive blood pressure channels, two temperature channels, pulse oximetry, respiration, non-invasive blood pressure and EtCO2. This effectively allows the Envoy to monitor a wide-range of adult, pediatric and neonatal patient conditions, in many different areas of the hospital. Functions include display of multiparameter waveforms, vital signs, alarm & status messages, The Mennen Medical Envoy is intended for sale as a system for monitoring and recording patient information or any in-hospital application requiring patient monitoring. The following are examples of intended clinical applications: · Critical Care Patients · Cardiac Step-down/Telemetry Units · Emergency Departments · Intra-operative (Anesthesia) Monitoring · Post Anesthesia Care
Device Story
Envoy is a multiparameter physiological patient monitor for hospital use; monitors ECG, blood pressure, temperature, respiration, EtCO2, and pulse oximetry. Modification adds Masimo SET SpO2 module. Module uses red/infrared light absorption to measure arterial oxygen saturation and pulse rate. Masimo SET technology processes signals to maintain accuracy during patient motion, low perfusion, ambient light, and electrosurgical interference. System consists of main processing unit, color monitor, and module rack. Healthcare providers view waveforms and numeric data on display; alarms alert staff to patient status changes. Device aids clinical decision-making by providing continuous, real-time vital sign monitoring.
Clinical Evidence
Bench testing only. Performance and safety testing conducted to ensure compliance with functional requirements and industry standards (UL 2601-1, IEC 60601-1, AAMI/EC 11, AAMI/EC 13, AAMI/ES 1, ISO 9919).
Technological Characteristics
Multiparameter monitor with plug-in modules. SpO2 module uses red/infrared light absorption (photoplethysmography). Standards: UL 2601-1, IEC 60601-1, AAMI/EC 11, AAMI/EC 13, AAMI/ES 1, ISO 9919. Connectivity: module rack interface. Software-driven.
Indications for Use
Indicated for adult, pediatric, and neonatal patients in hospital settings (critical care, telemetry, ED, OR, PACU) requiring multiparameter monitoring of ECG, heart rate, invasive/non-invasive blood pressure, temperature, pulse oximetry, respiration, and EtCO2.
Regulatory Classification
Identification
An oximeter is a device used to transmit radiation at a known wavelength(s) through blood and to measure the blood oxygen saturation based on the amount of reflected or scattered radiation. It may be used alone or in conjunction with a fiberoptic oximeter catheter.
{0}------------------------------------------------
AUG 01 2002
K022468
Food and Drug Administration ENVOY Patient Monitor - Device Modification: Special 510(k) for new SpO2 module (Masimo)
Image /page/0/Picture/4 description: The image shows the logo for Mennen Medical. The logo consists of a lowercase letter 'm' inside of a black circle. To the right of the logo is the word 'mennenmedical' in lowercase letters.
Mennen Medical Ltd., 4 Hayarden Street, Yavne PO Box 102, Rehovot 76100 Israel Tel .: +972-8-9323333
Fax: +972-8-9328510
Date revised 24nd July, 2002
#### Topic: 510(k) Safety and Effectiveness Summary as per 21 CFR Section 807.92(c) Envoy Patient Monitor - Device Modification: Special 510k for new SpO2 module (Masimo)
#### Establishment Name, Registration Number and Address:
| Name: | Mennen Medical Ltd. |
|---------------------|-----------------------------------------|
| Registration Number | 9611022 |
| Operator Number: | 9011766 |
| Address: | 4 Hayarden Street, Yavne, 81228, Israel |
| Postal Address: | PO Box 102,<br>Rehovot, 76100, Israel |
Tel: +972-8-9323333 Fax: +972-8-9328510
(See Note on page 11 re the new address of Mennen Medical Ltd.)
Asher Kassel, Director of Regulatory Affairs Contact person:
Food and Drug Administration To: Center for Devices and Radiological Health Document Mail Center (HFZ-401) 9200 Corporate Boulevard Rockville MD, 20850
Attn .: Document Control Clerk From: Asher Kassel, Director of Regulatory Affairs
#### Product Name:
ENVOY Proprietary: Common: Physiological Patient Monitor Mennen Medical Part Number: 550-010-000 (full system) 554-000-010 (CPU only)
New SpO2 module (Masimo) P/N: 551-139-000
{1}------------------------------------------------
ENVOY Patient Monitor - Device Modification: Special 510(k) for new SpO2 module (Masimo)
#### FDA Classification of Envoy Patient Monitor:
| Classification Name: | Arrhythmia Detector and Alarm |
|------------------------|-------------------------------|
| Classification Number: | 21 CFR 870.1025 |
| Classification: | Class III |
| Product Code: | 74 DSI |
#### FDA Classification of SpO2 Module:
| Classification Name: | Oximeter |
|------------------------|-----------------|
| Classification Number: | 21 CFR 870.2700 |
| Classification: | Class II |
| Product Code: | 74 DQA |
#### Performance Standards:
None promulgated
#### Voluntary Standards:
UL 2601-1, IEC 60601-1 for electrical safety for medical equipment AAMI/ EC 11 - Diagnostic electrocardiograph devices (1991) AAMI/ EC 13 - Cardiac monitors, heart-rate meters, alarms (1992) AAMI/ ES 1 - Safe current limits for electromedical apparatus (1993) IEC 60601-1: General Requirement for Safety for Medical Electrical Systems - part 1, (1988); Amendment 1 -1991-11 Amendment 2 -- 1995-03 IEC 60602-2-27: Medical electrical equipment, Part 2, (1994) Requirements for safety of electrocardiograph monitoring equipment. ISO 9919: Pulse oximeters for medical use requirements (1991)
#### Predicate Device:
MENNEN MEDICAL ENVOY PATIENT MONITOR (K001120).
{2}------------------------------------------------
ENVOY Patient Monitor - Device Modification: Special 510(k) for new SpO2 module (Masimo)
#### Device Description - Envoy Patient Monitor:
The Envoy is a multiparameter physiological patient monitor, capable of monitoring:
- · ECG/Heart Rate
- invasive blood pressure
- · non-invasive blood pressure
- · respiration
- · pulse oximetry
- two temperature channels
- cardiac output
- eTCO2
The Envoy bedside patient monitor consists of a main processing unit, a mountable color monitor, and a module rack housing the various Mennen Medical plug-in vital signs modules. The modules monitor the patient's vital signs. Up to six internal modules can be plugged into a module rack. The Envoy can accommodate two module racks. The vital sign data derived from the modules by the Envoy are presented on the monitor as waveform and numeric displays. The Envoy vital signs modules acquire vital signs data from the patient, and display their waveforms and alarms indications on the Envoy display unit. Vital signs and waveform information are displayed simultaneously on the Envoy Display Unit. Up to 8 traces can be displayed at any one time.
The Envoy is a reusable, software driven, patient monitor, intended for use as part of a physiological monitoring system in a hospital environment. As such it is not a life supporting. nor life sustaining device; nor is it implantable and therefore sterility is not a consideration.
The Envoy is not a kit, does not contain any drug or biological products and is not for prescription use.
#### Functional Description of the new SpO2 module (OEM Masimo technology):
The SpO2 module is the source of all pulse oximetry data in the Envoy bedside monitor.
SpO2 is measured to determine a monitored patient's arterial oxygen saturation and pulse rate. Arterial oxygen saturation is the percentage of oxygenated hemoglobin in relation to the total hemoglobin. For example, if 95 percent of the hemoglobin molecules in the arterial red blood cells combine with oxygen, the blood has an oxygen saturation of 95 percent. The Sp02 numeric value represents the percentage of hemoglobin molecules that have combined with oxygen molecules to form oxyhemoglobin.
The SpO2 module is housed in the Envoy module rack, where it occupies a single slot.
The Envoy bedside monitor measures oxygen saturation using the Pulse Oximetry method. This continuous, non-invasive method measures the light absorption in the patient's tissue (for example, a finger of the hand or foot) to a receiver on the other side.
{3}------------------------------------------------
#### ENVOY Intended Use:
The Envoy is intended for use as a multiparameter physiological patient monitoring system.
The Envoy can monitor ECG/heart rate, two invasive blood pressure channels, two temperature channels, pulse oximetry, respiration, non-invasive blood pressure and EtCO2. This effectively allows the Envoy to monitor a wide-range of adult, pediatric and neonatal patient conditions, in many different areas of the hospital.
Functions include display of multiparameter waveforms, vital signs, alarm & status messages,
The Mennen Medical Envoy is intended for sale as a system for monitoring and recording patient information or any in-hospital application requiring patient monitoring.
The following are examples of intended clinical applications:
- · Critical Care Patients
- · Cardiac Step-down/Telemetry Units
- · Emergency Departments
- · Intra-operative (Anesthesia) Monitoring
- · Post Anesthesia Care
{4}------------------------------------------------
ENVOY Patient Monitor – Device Modification: Special 510(k) for new SpO2 module (Masimo)
# Summary of the technological characteristics of the new SpO2 module (OEM Masimo technology):
The following tables summarize data on the Mennen Medical Envoy Envoy SpO2 Masimo module (modified device):
| Envoy<br>SpO2 Masimo module | |
|-----------------------------|---------------------------------------------------------------------------------------------|
| Part Number: | 551-139-000 |
| Monitored<br>Parameters: | SpO2 Oxygen saturation Pulse waveform Pulse rate volume |
| Module size: | Single slot<br>Height: 10.0cm (4.0 in)<br>Width: 4.0 cm (1.6 in)<br>Depth: 14.0 cm (5.5 in) |
| Envoy<br>SpO2 Masimo module | |
|----------------------------------|--------------------------------------------------------------------------------------|
| Saturation range: | 1% to 100% Sp02 |
| SpO2 Accuracy: | Adults: 70 % - 100 % ± 2 digits<br>Neonates: 70 % - 100 % ± 3 digits |
| Pulse Rate range: | 25 to 240 BPM ±3 BPM |
| Low saturation<br>alarm limits: | 50% to 100% |
| Data Stotrage<br>(in mainframe): | O2 saturation and Pulse Rate<br>Alarm event markers<br>Photoplethysmography waveform |
| | |
{5}------------------------------------------------
ENVOY Patient Monitor - Device Modification: Special 510(k) for new SpO2 module (Masimo)
| Envoy | |
|--------------------|------------------------------------------------------------|
| SpO2 Masimo module | |
| Features: | Red & Infrared light Tissues absorption method |
| | Waveform display |
| | Pulse rate derived from SpO2 with independent alarm limits |
| | Pulse sound distinct in tone and pitch from QRS tone |
| | Variation in the pulse sound pitch with SpO2 value |
| | Signal strength display |
| Technical Alarms: | "Cable out" and "Relocate probe" |
The major difference between the Envoy SpO2 Nellcor module and the Envoy SpO2 Masimo module is the addition of Masimo proprietary SET technology. SET signal processing is a Masimo developed technology that provides accurate monitoring during:
- Patient Motion or Movement ●
- Low Perfusion (low signal amplitude) .
- Intense Ambient Light (lighting or sunlight) .
- Electro surgical Instrument Interference .
#### Conclusion of comparison of technological characteristics:
We consider the Envoy Sp2 Masimo module to be substantially equivalent to the Envoy SpO2 Nellcor module and we submit that any differences between the two modules:
- fall within the scope of a Special 510(k) Device Modification .
- do not raise any new issues of safety and effectiveness .
# Testing
The Envoy SpO2 Masimo module has been subject to extensive safety and performance testing. Final testing for the system included various performance tests designed to ensure that the device meets all functional requirements and performance specifications. Safety testing and EMC testing were performed by an independent testing laboratory to ensure that the device complies to applicable industry and safety standards.
{6}------------------------------------------------
### INDICATIONS FOR USE
The Envov is intended for use as a multiparameter monitoring system.
The Envoy can monitor ECG/heart rate, two invasive blood pressure channels, two temperature channels, pulse oximetry, respiration, non-invasive blood pressure and EtCO2. This effectively allows the Envoy to monitor a wide-range of adult, pediatric and neonatal patient conditions, in many different areas of the hospital.
Functions include display of multiparameter waveforms, vital signs, alarm & status messages.
The Mennen Medical Envoy is intended for sale as a system for remote monitoring and recording patient information or any in-hospital application requiring remote patient monitoring.
The following are examples of intended clinical applications:
- · Critical Care Patients
- · Cardiac Step-down/Telemetry Units
- · Emergency Departments
- · Intra-operative (Anesthesia) Monitoring
- · Post Anesthesia Care
{7}------------------------------------------------
ENVOY Patient Monitor – Device Modification: Special 510(k) for new SpO2 module (Masimo)
Signature:
Asher Kassel
Asher Kassel Director of Regulatory Affairs MENNEN MEDICAL LTD.
Tel: Fax: +972-8-9328510 E-mail: asher@mmi.co.il
{8}------------------------------------------------
Image /page/8/Picture/0 description: The image is a black and white logo for the Department of Health & Human Services USA. The logo features a stylized image of a human figure with three faces in profile, one behind the other. The figure is surrounded by a circular border with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" written around it.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
# AUG 01 2002
Mr. Asher Kassel Director of Regulatory Affairs Mennen Medical Limited 4 Hayarden Street, Yavne P.O. Box 102 Rehovot 76100 ISRAEL
Re: K022168
> Trade/Device Name: ENVOY Patient Monitor Regulation Number: 21 CFR 870.2700, 870.1025 Regulation Name: Oximeter, Arrhythmia detector and alarm Regulatory Class: III Product Code: DQA, DSI Dated: July 2, 2002 Received: July 3, 2002
Dear Mr. Kassel:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{9}------------------------------------------------
# Page 2 – Mr. Kassel
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 21 CFR Part 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4646. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.qov/cdrh/dsma/dsmamain.html
Sincerely,
B. Altrud
Timothy A. Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
{10}------------------------------------------------
#### INDICATIONS FOR USE
The Envoy is intended for use as a multiparameter monitoring system.
The Envoy can monitor ECG/heart rate, two invasive blood pressure channels, two temperature channels, pulse oximetry, respiration, non-invasive blood pressure and EtCO2. This effectively allows the Envoy to monitor a wide-range of adult, pediatric and neonatal patient conditions, in many different areas of the hospital.
Functions include display of multiparameter waveforms, vital signs, alarm & status messages.
The Mennen Medical Envoy is intended for sale as a system for remote monitoring and recording patient information or any in-hospital application requiring remote patient monitoring.
The following are examples of intended clinical applications:
- · Critical Care Patients
- · Cardiac Step-down/Telemetry Units
- Emergency Departments .
- Intra-operative (Anesthesia) Monitoring ●
- Post Anesthesia Care
H. Wishman
(Division Sign-Off)
ntal, Infection Control, eneral Hospital De 510(k) Number
Image /page/10/Picture/15 description: The image shows a check mark and the words "Prescription Use". The check mark is located to the left of the words. The check mark is above a horizontal line.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.