# WELCH ALLYN SPOT VITAL SIGN MONITOR WITH MP506 (K022163)

_Welch Allyn, Inc. · DQA · Sep 17, 2002 · Cardiovascular · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K022163

## Device Facts

- **Applicant:** Welch Allyn, Inc.
- **Product Code:** [DQA](/productcode/DQA.md)
- **Decision Date:** Sep 17, 2002
- **Decision:** SESE
- **Submission Type:** Special
- **Regulation:** 21 CFR 870.2700
- **Device Class:** Class 2
- **Review Panel:** Cardiovascular
- **Attributes:** Pediatric

## Indications for Use

The Welch Allyn Spot Vital Signs is intended for measurement of blood pressure, pulse rate, temperature and oxygen saturation (SpO2) of adult and pediatric patients. The device is not designed, sold or intended for use except as indicated. The Welch Allyn Spot Check Device is not designed for use with neonates. To ensure pediatric blood pressure accuracy and safety, note that the Welch Allyn small cuff (5200-03) and the small One Piece cuff (5200-13) are the smallest cuffs approved for use with young children and infants. The circumference of the child's arm must fit within the range markings on the cuff. The Welch Allyn Spot Check Device should not be used on patients who are linked to heart/lung machines. The Welch Allyn Spot Check Device is not designed for use of axillary temperature option above three years of age in children. The Welch Allyn Spot Check Device is not intended to monitor patient's vital signs. The Welch Allyn Spot Check Device is not defibrillator proof.

## Device Story

Spot vital signs measurement device; provides one-time readings for blood pressure, pulse, temperature, and SpO2. Inputs: non-invasive BP cuff, thermometry probe, pulse oximetry sensor. Outputs: LCD display of vital signs; optional thermal printer output via infrared port. Used in clinical settings by physicians, physician assistants, or nurses for high-traffic patient assessment. Device lacks continuous monitoring, programmable alarms, or memory. Benefits: cost-effective, portable, rapid spot-check capability for adult and pediatric patients.

## Clinical Evidence

Bench testing only. Compliance with recognized standards including EN60601-1, EN60601-1-1, EN60601-1-2, EN60601-1-4, AAMI SP10, ASTM 1112-86, and EN 865 demonstrated for the MP506 pulse oximetry module and base unit.

## Technological Characteristics

Non-invasive BP measurement, SureTemp thermometry, Nellcor SpO2 pulse oximetry. Infrared port for external printer. Rechargeable battery. Standards: EN60601-1, EN60601-1-1, EN60601-1-2, EN60601-1-4, AAMI SP10, ASTM 1112-86, EN 865. LCD display. Portable form factor with mounting accessories.

## Regulatory Identification

An oximeter is a device used to transmit radiation at a known wavelength(s) through blood and to measure the blood oxygen saturation based on the amount of reflected or scattered radiation. It may be used alone or in conjunction with a fiberoptic oximeter catheter.

## Predicate Devices

- Welch Allyn Spot Vital Signs ([K002530](/device/K002530.md))
- OxiMax Pulse Oximetry System w/N-595 Pulse Oximeter ([K012891](/device/K012891.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

K 022163

SEP 1 7 2002

# V. 510(k) Summary

[As described in CFR 807.92]

| Submitted by:        | Welch Allyn Inc.<br>4341 State Street Road<br>Skaneateles Falls, NY 13153                    |
|----------------------|----------------------------------------------------------------------------------------------|
| Contact Person:      | David Klementowski<br>Corporate Regulatory Affairs Manager                                   |
| Date Prepared:       | 20 May 2002                                                                                  |
| Proprietary Name:    | Welch Allyn Spot Vital Sign Monitor with MP506                                               |
| Common Name:         | Vital Signs measurement device                                                               |
| Classification Name: | Class II 870.1130 Noninvasive Blood Pressure<br>Measurement System                           |
| Predicate Devices:   | Welch Allyn Spot Vital Signs<br>Welch Allyn, Inc.<br>510(k) Document Control Number K002530  |
|                      | OxiMax Pulse Oximetry System w/N-595 Pulse Oximeter<br>Nellcor Puritan Bennett, Incorporated |

Noneor Parkan Delhott, Incorporator
510(k) Document Control Number K012891

{1}------------------------------------------------

## Description of the Device:

The Welch Allyn Spot Vital Signs is not a monitor, but a one-time vital signs measurement device. This product will not have continuous monitoring capability with timed cycle intervals, memory, or any various programmable alarm features. The device is intended to provide the physician, physician's assistant, or nurse, facing high patient traffic or multiple tasks, a cost effective method to determine a one-time vital signs reading on the spot. The base unit will have non-invasive blood pressure (BP) measurement. Options will also be offered such as SureTemp® thermometry, Nellcor® pulse oximetry (SpO2), mounting bracket, and rolling stand. The device may be interfaced with an external printer via an infrared port.

The Welch Allyn Spot Vital Signs is designed to non-invasively measure systolic and diastolic blood pressure, pulse rate, temperature and oxygen saturation (SpO2) for adult and pediatric patients. The Welch Allyn Spot Vital Signs also calculates Mean Arterial Pressure (MAP). All blood pressure, pulse, temperature and SpO2 values are displayed on a large, easy-to-read LCD display, and may be printed via an external thermal printer, as desired. The rechargeable battery and wide variety of mounting accessories make the Welch Allyn Spot Vital Signs convenient for many locations.

## Intended Use

The Welch Allyn Spot Check Device Check Device is intended for measurement of blood pressure, pulse rate, temperature and oxygen saturation (SpO2) of adult and pediatric patients. The device is not designed, sold or intended for use except as indicated.

The Welch Allyn Spot Check Device is not designed for use with neonates. To ensure pediatric blood pressure accuracy and safety, note that the Welch Allyn small cuff (5200-03) and the small One Piece Cuff (5200-13) are the smallest cuffs approved for use with young children and infants. The circumference of the child's arm must fit within the range markings on the cuff.

## The Welch Allyn Spot Check Device should not be used on patients who are linked to heart/lung machines.

The Welch Allyn Spot Check Device is not designed for use of axillary temperature option above three years of age in children.

The Welch Allyn Spot Check Device is not intended to monitor patient's vital signs.

The Welch Allyn Spot Check Device is not defibrillator proof.

{2}------------------------------------------------

Action Taken to Comply with Section 514 of the Act

The agency has recognized the following standards:

- a) EN60601-1
- b) EN60601-1-1
- c) EN60601-1-2
- d) EN60601-1-4
- e) AAMI SP10
- f) ASTM 1112-86
- g) EN 865

The Welch Allyn Spot Vital Sign with MP506 Pulse Oximeter OEM module meets the requirements called out in these standards. Evidence of compliance is on file at Welch Allyn and is available for review upon demand.

{3}------------------------------------------------

### DEPARTMENT OF HEALTH & HUMAN SERVICES

Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle, there is a stylized symbol that resembles three overlapping human profiles or abstract shapes, possibly representing people or services provided by the department.

#### Public Health Service

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

Mr. David Klementowski Corporate Regulatory Affairs Manager Welch Allyn, Incorporated 4341 State Street Road Skaneateles Falls, New York 13153-0220

Re: K022163

Trade/Device Name: Welch Allyn Spot Vital Sign Monitor with MP506 Regulation Number: 870.2700 Regulation Name: Oximeter Regulatory Class: II Product Code: DQA Dated: August 13, 2002 Received: August 20, 2002

Dear Mr. Klementowski:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28. 1976. the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting vour device can be found in the Code of Federal Regulations. Title 21. Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies.

{4}------------------------------------------------

Page 2 - Mr. Klementowski

You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 21 CFR Part 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4646. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html

Sincerely yours,

Timothy A. Ulatowski

Timothy Ulatowski Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{5}------------------------------------------------

## III. Indications for Use Statement

11 11 11 11

.

| 510(k) Number:       | Unknown K022163                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                     |
|----------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Device Name:         | Welch Allyn Spot Vital Signs                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                        |
| Indications for use: | The Welch Allyn Spot Vital Signs is intended for<br>measurement of blood pressure, pulse rate, temperature and<br>oxygen saturation (SpO2) of adult and pediatric patients.<br>The device is not designed, sold or intended for use except<br>as indicated.<br>* The Welch Allyn Spot Check Device is not designed for<br>use with neonates. To ensure pediatric blood pressure<br>accuracy and safety, note that the Welch Allyn small cuff<br>(5200-03) and the small One Piece cuff (5200-13) are the<br>smallest cuffs approved for use with young children and<br>infants. The circumference of the child's arm must fit<br>within the range markings on the cuff.<br>* The Welch Allyn Spot Check Device should not be used<br>on patients who are linked to heart/lung machines.<br>* The Welch Allyn Spot Check Device is not designed for<br>use of axillary temperature option above three years of age<br>in children.<br>* The Welch Allyn Spot Check Device is not intended to<br>monitor patient's vital signs.<br>* The Welch Allyn Spot Check Device is not defibrillator<br>proof. |

1. 1. 1.

## (Please Do Not Write Below This Line - Continue On Another Page If Needed)

Concurrence of CDRH, Office of Device Evaluation (ODE)

| Prescription Use     | <div style="text-align:center;">✓</div> |
|----------------------|-----------------------------------------|
| (Per 21 CFR 801.109) | Or Over-The-Counter Use                 |

 __________________________
   

 (Division Sign-Off)
   

 Division of Anesthesiology, General Hospital,
   

 Infection Control, Dental Devices

510(k) Number: IL022163

---

**Source:** [https://fda-staging.innolitics.com/device/K022163](https://fda-staging.innolitics.com/device/K022163)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com