K022133 · Becton, Dickinson & CO · GQG · Dec 10, 2002 · Microbiology
Device Facts
Record ID
K022133
Device Name
BD DIRECTIGEN EZ RSV KIT
Applicant
Becton, Dickinson & CO
Product Code
GQG · Microbiology
Decision Date
Dec 10, 2002
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 866.3480
Device Class
Class 1
Attributes
Pediatric
Indications for Use
The BD Directigen™ EZ RSV test is a rapid chromatographic immunoassay for the direct and qualitative detection of Respiratory Syncytial Virus (RSV) antigen in nasopharyngeal washes, nasopharyngeal aspirates, nasopharyngeal swabs, and nasopharyngeal swab/washes from patients suspected of having a viral respiratory infection. This test is intended for in vitro diagnostic use to aid in the diagnosis of RSV infections in neonatal and pediatric patients under the age of 20. It is recommended that negative test results be confirmed by cell culture.
Device Story
Rapid chromatographic immunoassay for qualitative detection of RSV A and B antigens in respiratory specimens (nasopharyngeal washes, aspirates, swabs). Extracted specimen added to test device; RSV antigens bind to antibody-colloidal gold conjugate; complex migrates to reaction area; captured by antibody line on membrane. Positive result: two reddish-purple lines (Test and Control); negative: one line (Control); uninterpretable: no Control line. Used in clinical settings; performed by healthcare professionals. Provides rapid visual qualitative result to aid clinical diagnosis of RSV infection. Benefits patient through timely identification of viral etiology, facilitating appropriate clinical management.
Clinical Evidence
Multicenter clinical trial (2001-2002 season) evaluated 1,176 specimens (nasopharyngeal washes, aspirates, swabs) compared to viral cell culture. Overall sensitivity 80%, specificity 92%. Discrepant analysis performed using PCR on 75/82 culture-negative/test-positive specimens; 47/75 were PCR positive. Analytical studies included LOD for 5 strains, cross-reactivity panel (99 microorganisms), and interference testing (blood, medications). Reproducibility 99.1% across three sites.
Technological Characteristics
Chromatographic immunoassay; lateral flow test strip; antibody-colloidal gold conjugate; membrane-bound antibody capture; visual readout. In vitro diagnostic use. No electronic components or software.
Indications for Use
Indicated for neonatal and pediatric patients under age 20 suspected of having a viral respiratory infection to aid in the diagnosis of RSV.
Regulatory Classification
Identification
Respiratory syncytial virus serological reagents are devices that consist of antigens and antisera used in serological tests to identify antibodies to respiratory syncytial virus in serum. Additionally, some of these reagents consist of respiratory syncytial virus antisera conjugated with a fluorescent dye (immunofluorescent reagents) and used to identify respiratory syncytial viruses from clinical specimens or from tissue culture isolates derived from clinical specimens. The identification aids in the diagnosis of respiratory syncytial virus infections and provides epidemiological information on diseases caused by these viruses. Respiratory syncytial viruses cause a number of respiratory tract infections, including the common cold, pharyngitis, and infantile bronchopneumonia.
{0}------------------------------------------------
K022133
#### 510(k) SUMMARY
DEC 1 0 2002
| SUBMITTED BY: | BECTON, DICKINSON AND COMPANY<br>7 LOVETON CIRCLE<br>SPARKS, MD 21152<br>Phone: 410-316-4988<br>Fax: 410-316-4499 |
|------------------------|-------------------------------------------------------------------------------------------------------------------------------------------|
| CONTACT NAME: | Colleen A. Kistler, Regulatory Affairs Specialist |
| DATE PREPARED: | June 28, 2002 |
| DEVICE TRADE NAME: | BD Directigen <sup>TM</sup> EZ RSV |
| DEVICE COMMON NAME: | Antigen, CF, respiratory syncytial virus |
| DEVICE CLASSIFICATION: | 21 CFR§866.3480 |
| PREDICATE DEVICES: | Cell Culture<br>Direct Fluorescent Antibody (DFA)<br>Directigen <sup>TM</sup> RSV (K882629) |
## INTENDED USE:
The BD Directigen™ EZ RSV test is a rapid chromatographic immunoassay for the direct and qualitative detection of Respiratory Syncytial Virus (RSV) antigen in nasopharyngeal washes, nasopharyngeal aspirates, nasopharyngeal swabs, nasopharyngeal swab/washes, lower respiratory specimens (sputum, tracheal aspirates and bronchoalveolar layage), and other swabs (pharyngeal, lower nasal, nose/throat) from patients suspected of having a viral respiratory infection.
# DEVICE DESCRIPTION:
The BD Directigen™ EZ RSV test is a chromatographic assay to qualitatively detect RSV antigen in samples extracted from respiratory specimens. When extracted specimens are added to the test device, RSV A and/or B antigens bind to the antibody-colloidal gold conjugate in the test strip forming an antigenantibody complex. This complex migrates across the test strip to the reaction area and is captured by the line of RSV antibody on the membrane. Excess conjugate binds to a second line of inactivated RSV antigen that serves as a functional control. A positive result is indicated by the appearance of two reddish purple lines in the read window, one line next to the Test "T" and the other next to the Control "C". The absence of a reddish purple line next to the "T" and the presence of a reddish purple line next to the "C" indicate a negative result. The test is considered uninterpretable is no visible reddish purple line is present next to the "C".
{1}------------------------------------------------
## DEVICE COMPARISON:
The BD Directigen™ EZ RSV was compared to viral cell culture, direct fluorescent antibody (DFA) tests, and the Directigen™ RSV test. Although there are some differences between the BD Directigen™ EZ RSV test and the predicate devices, these differences do not present new issues of safety and effectiveness. The impact of these differences on the safety and effectiveness of the BD Directigen™ EZ RSV test was assessed using accepted scientific methods.
# SUMMARY OF PERFORMANCE DATA:
## Analytical Sensitivity
The limit of detection (LOD) for the BD Directigen™ EZ RSV test was established for a total of five RSV strains; two RSV A and three RSV B strains.
| Viral Strain | Viral Type | LOD (TCID50) |
|------------------|------------|--------------|
| RSV (Long) | A | 3.95 X 103 |
| RSV (A-2) | A | 5.95 X 103 |
| RSV (9320) | B | 4.05 X 102 |
| RSV (Washington) | B | 7.03 X 103 |
| RSV (Wild) | B | 5.56 X 102 |
## Cross Reactivity
A panel of 99 microorganisms (including bacteria, yeasts and viruses) were cultured and tested at appropriate concentrations in triplicate with the BD Directigen™ EZ RSV test. None of the microorganisms tested in the panel were shown to cross react with the BD Directigen™ EZ RSV test.
## Interfering Substances
A variety of substances were tested with the BD Directigen™ EZ RSV test at concentration levels comparable to or greater than levels that may be present in patient respiratory samples. Substances tested included blood, mouthwashes, throat drops, nasal sprays, cold medications and prescription medications. Each substance was tested in triplicate. None of the substances evaluated were shown to interfere with the performance of the BD Directigen™ EZ RSV test.
## Reproducibility
The reproducibility of the BD Directigen EZ RSV test was evaluated at three sites. The overall reproducibility for the BD Directigen RSV EZ test was 99.1%.
{2}------------------------------------------------
# Clinical Performance:
The overall sensitivity and specificity of the BD Directigen™ EZ RSV test for RSV when compared to culture were 80% and 92%, respectively. The performance characteristics of the BD Directigen™ EZ RSV as compared to cell culture for each specimen type are shown in Table 1.
| | Culture/ EZ | | | | | Discrepant Resolution<br>for<br>Culture - / EZ +<br>Specimens | | | |
|-----------------------------|-------------|-----|-----|-----|-----|---------------------------------------------------------------|-----------------------------|--------------|---------|
| Specimen Type | n | +/+ | -/+ | +/- | -/- | Sensitivity (%)<br>(95% CI) | Specificity (%)<br>(95% CI) | # PCR tested | # PCR + |
| Nasopharyngeal<br>Wash | 358 | 130 | 30 | 20 | 178 | 86.7<br>80.2 - 91.6 | 85.6<br>80.1 - 90.1 | 30 | 22 |
| Nasopharyngeal<br>Aspirate | 405 | 90 | 24 | 27 | 264 | 76.9<br>68.2 - 84.2 | 91.7<br>87.6 - 94.6 | 182 | 12 |
| Nasopharyngeal<br>Swab/Wash | 162 | 28 | 6 | 11 | 117 | 71.8<br>55.1 - 85.0 | 95.1<br>89.7 – 98.2 | 6 | 2 |
| Nasopharyngeal<br>Swab | 286 | 20 | 20 | 10 | 236 | 66.7<br>47.2 - 82.7 | 92.2<br>88.2 - 95.2 | 20 | 11 |
| Other Swabs1 | 73 | 3 | 1 | 1 | 68 | 75.0<br>19.4 - 99.4 | 98.6<br>92.2 - 99.9 | 03 | 0 |
| Lower<br>Respiratory | 51 | 5 | 1 | 1 | 44 | 83.3<br>35.9 - 99.6 | 97.8<br>88.2 - 99.9 | 1 | 0 |
Table 1: Summary of the Performance of the BD Directigen EZ RSV (EZ) Test Compared to Culture for all Specimen Types
(+) = RSV positive (-) = RSV negative (#) = number
1 Refer to Limitations of the Procedure for additional information on these specimen types
2 Two specimens not submitted for PCR testing; four specimens had insufficient quantity for PCR testing 3 One specimen not submitted for PCR testing
There were 82 specimens that were culture negative, BD Directigen™ EZ RSV test positive. PCR testing was performed on 75 of the 82 specimens; a total of 47 of the 75 specimens were positive by PCR. The uninterpretable rate for the BD Directigen™ EZ RSV test was 0.0%.
Overall performance of the BD Directigen™ EZ RSV test is substantially equivalent' to viral cell culture and DFA tests that were in use prior to May 28, 1976 and to the BD Directigen™ RSV test.
<sup>1</sup> The term "substantial equivalence" as used in this 510(k) notification is limited to the definition of substantial equivalence as found in the Federal Food, Drug and Cosmetic Act, as amended and as applied under 21 CFR 807, Subpart E under which a device can be marketed without premarket approval or reclassification. A determination of substantial equivalency under this notification is not intended to have any bearing whatsoever on the resolution of patent infringement suits or any other patent matters. No statements related to, or in support of substantial equivalence herein shall be construed as an admission against interest under the US Patent Laws or their application by the courts.
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized caduceus symbol, with three intertwined snakes forming a triple helix. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular pattern around the symbol.
Food and Drug Administration 2098 Gaither Road Rockville MD 20850
2003 FEB 3
Ms. Colleen Kistler Regulatory Affairs BD Diagnostic Systems Becton, Dickenson and Company 7 Loveton Circle Sparks, MD 21152
Re: K022133
Trade/Device Name: BD Directigen™ EZ RSV Kit Regulation Number: 21 CFR 866.3480 Regulation Name: Respiratory Syncytial Virus Serological Reagents Regulatory Class: Class I Product Code: GQG Dated: September 23, 2002 Received: September 24, 2002
Dear Ms. Kistler:
This letter corrects our substantially equivalent letter of December 10, 2002, regarding the BD Directigen™ EZ RSV Kit. The revised Indications for use and 510(k) summary are enclosed.
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in Title 21, Code of Federal Regulations (CFR), Parts 800 to 895. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); and good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820).
{4}------------------------------------------------
#### Page 2 -
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific information about the application of labeling requirements to your device, or questions on the promotion and advertising of your device, please contact the Office of In Vitro Diagnostic Device Evaluation and Safety at (301) 594-3084. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html.
Sincerely yours,
Steven Sutman
Steven I. Gutman, M.D., M.B.A. Director Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
Page 1 of 1
510(k) Number (if known):
BD Directigen™ EZ RSV Test Device Name:
Indications for Use:
The BD Directigen™ EZ RSV test is a rapid chromatographic immunoassay for the direct and qualitative detection of Respiratory Syncytial Virus (RSV) antigen in nasopharyngeal washes, nasopharyngeal aspirates, nasopharyngeal swabs, and nasopharyngeal swab/washes from patients suspected of having a viral . respiratory infection. This test is intended for in vitro diagnostic use to aid in the diagnosis of RSV infections in neonatal and pediatric patients under the age of 20. It is recommended that negative test results be confirmed by cell culture.
# (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Freddie M. Poole
vision Sign-Division of Clinical Laboratory Devices
510(k) Number KD22133
For Prescription Use ✓
(Optional Format 3-10-98)
{6}------------------------------------------------
## 510(k) SUMMARY
| SUBMITTED BY: | BECTON, DICKINSON AND COMPANY<br>7 LOVETON CIRCLE<br>SPARKS, MD 21152 |
|------------------------|--------------------------------------------------------------------------------|
| Phone: | 410-316-4988 |
| Fax: | 410-316-4499 |
| CONTACT NAME: | Colleen A. Kistler, Regulatory Affairs Specialist |
| DATE PREPARED: | December 10, 2002 |
| DEVICE TRADE NAME: | BD Directigen™ EZ RSV |
| DEVICE COMMON NAME: | Antigen, CF, respiratory syncytial virus |
| DEVICE CLASSIFICATION: | 21 CFR§866.3480 |
| PREDICATE DEVICES: | Cell Culture<br>Direct Fluorescent Antibody (DFA)<br>Directigen™ RSV (K882629) |
#### INTENDED USE:
The BD Directigen™ EZ RSV test is a rapid chromatographic immunoassay for the direct and qualitative detection of Respiratory Syncytial Virus (RSV) antigen in nasopharyngeal washes, nasopharyngeal aspirates, nasopharyngeal swabs, and nasopharyngeal swab/washes from patients suspected of having a viral respiratory infection. This test is intended for in vitro diagnostic use to aid in the diagnosis of RSV infections in neonatal and pediatric patients under the age of 20. It is recommended that negative test results be confirmed by cell culture.
## DEVICE DESCRIPTION:
The BD Directigen™ EZ RSV test is a chromatographic assay to qualitatively detect RSV antigen in samples extracted from respiratory specimens. When extracted specimens are added to the test device, RSV A and/or B antigens bind to anti-RSV conjugated to visualizing particles in the test strip. The antigenconjugate complex migrates across the test strip to the reaction area and is captured by the line of antibody on the membrane. A positive result is indicated by the appearance of two reddish purple lines in the read window, one line next to the Test "T" and the other next to the Control "C". The absence of a reddish purple line next to the "T" and the presence of a reddish purple line next to the
{7}------------------------------------------------
"C" indicate a negative result. The test is considered uninterpretable if no visible reddish purple line is present next to the "C".
# DEVICE COMPARISON:
The BD Directigen™ EZ RSV was compared to viral cell culture, direct fluorescent antibody (DFA) tests, and the Directigen™ RSV test. Although there are some differences between the BD Directigen™ EZ RSV test and the predicate devices, these differences do not present new issues of safety and effectiveness. The impact of these differences on the safety and effectiveness of the BD Directigen™ EZ RSV test was assessed using accepted scientific methods.
# SUMMARY OF PERFORMANCE DATA:
Several analytical studies and a clinical study were performed to establish the performance characteristics of the assay.
Analytical Studies: The limit of detection (LOD) for the BD Directigen™ EZ RSV test was established for two RSV A and three RSV B strains. A panel of 99 microorganisms (including bacteria, yeasts and viruses) were cultured and tested at appropriate concentrations in triplicate with the BD Directigen™ EZ RSV test. A variety of substances were tested with the BD Directigen™ EZ RSV test at concentration levels comparable to or greater than levels that may be present in patient respiratory samples. Substances tested included blood, mouthwashes. throat drops, nasal sprays, cold medications and prescription medications. Each substance was tested in triplicate. The reproducibility of the BD Directigen EZ RSV test was evaluated at three sites.
Clinical Study: A total of 1176 specimens, consisting of nasopharyngeal washes, nasopharyngeal aspirates, nasopharyngeal swabs, and nasopharyngeal swab/washes from patients suspected of having RSV were evaluated with the BD Directigen EZ RSV test in a multicenter trial during the 2001-2002 RSV season.
Based on the results of the analytical and clinical studies, the performance of the BD Directigen™ EZ RSV test is substantially equivalent¹ to viral cell culture and DFA tests that were in use prior to Mav 28, 1976 and to the BD Directigen™ RSV test.
<sup>1</sup> The term "substantial equivalence" as used in this 510(k) notification is limited to the definition of substantial equivalence as found in the Federal Food, Drug and Cosmetic Act, as amended and as applied under 21 CFR 807, Subpart E under which a device can be marketed without premarket approval or reclassification. A determination of substantial equivalency under this
{8}------------------------------------------------
notification is not intended to have any bearing whatsoever on the resolution of patent infringement suits or any other patent matters. No statements related to, or in support of substantial equivalence herein shall be construed as an admission against interest under the US Patent Laws or their application by the courts.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.