K022085 · J & S Medical Assoc., Inc. · JJW · Jul 25, 2002 · Clinical Chemistry
Device Facts
Record ID
K022085
Device Name
J&S MEDICAL SENTRY URINE DIPSTICK CONTROL
Applicant
J & S Medical Assoc., Inc.
Product Code
JJW · Clinical Chemistry
Decision Date
Jul 25, 2002
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 862.1660
Device Class
Class 1
Indications for Use
J&S Medical Associates Sentry dipstick urine controls are intended to be used by laboratory technicians in order to verify the performance of various urine dipsticks as a part of the laboratory's quality control practices.
Device Story
Sentry Urine Dipstick Control consists of three liquid levels prepared with human and animal proteins and chemical additives. These controls interact with urine dipstick reactant pads to produce color changes mimicking normal or abnormal urine samples. Used in clinical laboratory settings by technicians to verify dipstick performance. Level 1 tests for leukocytes, nitrite, urobilinogen, protein, blood, ketone, bilirubin, glucose, hCG, and specific gravity. Level 2 tests for ascorbic acid, pH, and specific gravity. Level 3 tests for pH and specific gravity. Healthcare providers use the resulting color changes (visual or instrument-read) to confirm the accuracy and reliability of their urinalysis dipstick testing, ensuring valid clinical results for patient diagnosis.
Clinical Evidence
Bench testing only. Stability and performance were assessed by incubating controls at 4°C, ambient, 25°C, and 37°C for 65 days. Performance was verified using Quidel and Bayer dipsticks, evaluated both visually and via instruments (Behring Rapimat II, Bayer Clinitek 100). hCG stability in Level 1 was verified over 120 days using the Quidel QuickVue One-Step hCG Combo lateral flow device. Results confirmed shelf life of two years and open-vial stability of at least two months.
Technological Characteristics
Liquid-based quality control material containing human and animal proteins and chemical additives. Three levels with varying pH, specific gravity, and analyte concentrations. Compatible with various commercial urine dipsticks. Stability validated for two years (closed) and two months (open).
Indications for Use
Indicated for use by laboratory technicians to verify the performance of urine dipsticks as part of laboratory quality control practices.
Regulatory Classification
Identification
A quality control material (assayed and unassayed) for clinical chemistry is a device intended for medical purposes for use in a test system to estimate test precision and to detect systematic analytical deviations that may arise from reagent or analytical instrument variation. A quality control material (assayed and unassayed) may be used for proficiency testing in interlaboratory surveys. This generic type of device includes controls (assayed and unassayed) for blood gases, electrolytes, enzymes, multianalytes (all kinds), single (specified) analytes, or urinalysis controls.
Predicate Devices
The Dropper urine dipstick control (Quantimetrix Corporation, 510(k) #: Unknown)
Submission Summary (Full Text)
{0}------------------------------------------------
# JUL 2 5 2002
### 510(K) Summary
This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR 807.92
| The assigned 510(k) number is : | K022085 |
|---------------------------------|---------|
|---------------------------------|---------|
| | Submitter Information |
|--|-----------------------|
|--|-----------------------|
| Submitter's Name: | J&S Medical Associates |
|------------------------------------|-------------------------------------------------|
| Address: | 35 Tripp Street, Bldg.1<br>Framingham, MA 01702 |
| Contact Person: | Dr. Paul MacDonnell |
| Phone: | 1-800-229-6000 ext. 157 |
| Fax: | (508) 370-4554 |
| Establishment Registration Number: | 1218982 |
Date Prepared: 06/24/2002
#### 2) Identification of Device
| Device Name: | Urine Dipstick Control |
|--------------------------|---------------------------------------------|
| Proprietary/ Trade Name: | J&S Medical Sentry Urine Dipstick Control |
| Common Name: | Urinalysis Controls |
| Classification Name: | Urinalysis Controls (Assayed and Unassayed) |
| Device Classification: | 1 |
| Regulation Number: | 21 CFR 862.1660 |
| Panel: | Clinical Chemistry |
| Product Code: | JJW |
{1}------------------------------------------------
3) Identification of the Predicate Device:
Predicate Device Name: The Dropper urine dipstick control Quantimetrix Corporation Manufacturer: 510(k) #: Unknown
- 4.0 Description of the Device
The three Sentry controls are prepared with human and animal proteins and with various chemical additives. These chemicals interact with the dipstick reactant pads to produce specific color changes that mimic actual normal and/or abnormal urine samples.
- 4.1 Sentry Level 1 Control is designed to test positive for leukocytes, nitrite. urobilinogen, protein, blood, ketone, bilirubin, and glucose with dipsticks. In addition, this control will have a pH of about 7 with an elevated specific gravity and will test positive for hCG.
- 4.2 Sentry Level 2 Control is designed to test positive for ascorbic acid, have a pH of about 5, and a moderately elevated specific gravity.
- 4.3 Sentry Level 3 Control is designed to have a pH of about 7 with a low specific gravity.
- 5.0 Intended Use
The controls are intended to be used by laboratory technicians in order to verify the performance of urine dipsticks made by various companies.
- 6.0 Substantial Equivalence
- 6.1 The Sentry Urine Dipstick Controls (Levels 1, 2, and 3) and the Quantimetrix Controls (Level 1 and 2) are supplied as liquids and do not require further treatment before use.
- 6.2 The Sentry and Quantimetrix controls are used to verify the performance of various urine dipsticks.
- 6.3 The Sentry and the Quantimetrix Controls will have value assignments to account for variations in manufacturers' dipstick formulations.
{2}------------------------------------------------
#### 7.0 Assignment of Expiration date
- Samples of the three controls were incubated at 4º C, ambient (for open 7.1 vial room temperature stability), 25° C, and 37° C for 65 days.
- The stability of the three controls were assessed as follows: 7.2
- Solutions were tested at frequent intervals using Quidel dipsticks. 7.2.1 The results were recorded visually by comparing reactant pad color intensity to the color chart supplied by Quidel; the results also were measured with the Behring Rapimat II instrument. After 65 days incubation of the controls at 37°, all of the controls performed within specifications, indicating a shelf life of two years for the three controls (Table I).
- 7.2.2 The stability of the controls also was assessed by testing with Baver dipsticks visually and with the Bayer Clinitek 100 instrument. The stability data with the Bayer strips are similar to that with the Quidel strips indicating an expiration date of two vears for the three controls (Table II).
- The stability of hCG in Sentry Level 1 was assessed by testing the 7.2.3 solution at periodic intervals up to 120 at days 4°. 25°, and 37° C. When tested with Quidel QuickVue One-Step hCG Combo lateral flow device a positive response was detected after 120 days. An expiration date of two years can be assigned (Table III).
- 8.0 Open vial stability of Sentry Controls 1, 2, and 3
Samples of the Sentry Urine Dipstick Control Levels 1, 2, and 3 were stored at ambient temperature (19° C to 22° C). Quidel dipsticks and Bayer dipsticks were used to test the controls at periodic intervals for 65 days. At day 65, the controls reacted within specifications, indicating open vial shelf life of at least two months for the three controls (Tables IV, V)
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the seal of the Department of Health & Human Services (USA). The seal features the department's name in a circular arrangement around a stylized caduceus symbol. The caduceus is depicted with three intertwined strands, representing health, services, and human aspects.
### DEPARTMENT OF HEALTH & HUMAN SERVICES
Public Health Service
Food and Drug Administration 2098 Gaither Road Rockville MD 20850
# JUL 2 5 2002
Ms. Joslyn S. Murphy Ouality Assurance Manager J&S Medical Associates 35 Tripp Street, Bldg. 1 Framingham. MA 01702
Re: k022085
> Trade/Device Name: Sentry urine dipstick control Regulation Number: 21 CFR 862.1660 Regulation Name: Quality control material (assayed and unassayed) Regulatory Class: Class I Product Code: JJW Dated: June 24, 2002 Received: June 27, 2002
Dear Ms. Murphy:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{4}------------------------------------------------
Page 2 -
This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and ' additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4588. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers International and Consumer Assistance at its toll-free number (800) 638-2041 or. (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsma/dsmamain.html".
Sincerely yours,
Steven Butman
Steven I. Gutman, M.D., M.B.A. Director Division of Clinical Laboratory-Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
## INDICATIONS FOR USE STATEMENT
510(K) Number (if known):
Device Name: Sentry urine dipstick control
Indications for Use:
J&S Medical Associates Sentry dipstick urine controls are intended to be used by laboratory technicians in order to verify the performance of various urine dipsticks as a part of the laboratory's quality control practices.
Lein Cooper
(Division Sign-Off)
Division of Clinical Laboratory Devices Kho 510(k) Number_
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Description Use X
(Per 21 CFR 801.109)
OR
Over-The-Counter Use
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.