← Product Code [MSD](/productcode/MSD) · K022000

# POURCHEZ RETRO TWIN LUMEN CHRONIC HEMODIALYSIS CATHETER WITH SEPARATED TIPS (WITH AND WITHOUT SIDE HOLES) (K022000)

_Spire Biomedical, Inc. · MSD · Jul 19, 2002 · Gastroenterology, Urology · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K022000

## Device Facts

- **Applicant:** Spire Biomedical, Inc.
- **Product Code:** [MSD](/productcode/MSD.md)
- **Decision Date:** Jul 19, 2002
- **Decision:** SESE
- **Submission Type:** Special
- **Regulation:** 21 CFR 876.5540
- **Device Class:** Class 2
- **Review Panel:** Gastroenterology, Urology
- **Attributes:** Therapeutic

## Indications for Use

Spire Biomedical, Inc.'s Pourchez RetrO™ Silicone Twin Lumen Chronic Hemodialysis Catheter with Separated Tips is designed for chronic hemodialysis and apheresis. It is a radiopaque silicone catheter designed for percutaneous insertion or insertion via a cutdown.

## Device Story

Flexible radiopaque silicone catheter; twin-lumen design with separated distal tips to reduce recirculation; oval body; color-coded proximal extension adapters (red/arterial, blue/venous); includes polyester cuff for anchoring. Available in five implant lengths; optional side holes. Used by clinicians for chronic hemodialysis and apheresis access via jugular or subclavian vein. Device facilitates blood flow between patient and dialysis machine. Repair kit provided for extension/adapter replacement.

## Clinical Evidence

Bench testing only. Performance data demonstrated equivalent hemodialysis catheter performance to predicate devices and compliance with established ISO standards.

## Technological Characteristics

Radiopaque silicone construction; twin-lumen design; polyester cuff; color-coded locking connectors; five implant lengths; 52cm overall length with 5cm-50cm markings. Sterilization method not specified. No software or electronic components.

## Regulatory Identification

A blood access device and accessories is a device intended to provide access to a patient's blood for hemodialysis or other chronic uses. When used in hemodialysis, it is part of an artificial kidney system for the treatment of patients with renal failure or toxemic conditions and provides access to a patient's blood for hemodialysis. The device includes implanted blood access devices, nonimplanted blood access devices, and accessories for both the implanted and nonimplanted blood access devices.(1) The implanted blood access device is a prescription device and consists of various flexible or rigid tubes, such as catheters, or cannulae, which are surgically implanted in appropriate blood vessels, may come through the skin, and are intended to remain in the body for 30 days or more. This generic type of device includes various catheters, shunts, and connectors specifically designed to provide access to blood. Examples include single and double lumen catheters with cuff(s), fully subcutaneous port-catheter systems, and A-V shunt cannulae (with vessel tips). The implanted blood access device may also contain coatings or additives which may provide additional functionality to the device. (2) The nonimplanted blood access device consists of various flexible or rigid tubes, such as catheters, cannulae or hollow needles, which are inserted into appropriate blood vessels or a vascular graft prosthesis (§§ 870.3450 and 870.3460), and are intended to remain in the body for less than 30 days. This generic type of device includes fistula needles, the single needle dialysis set (coaxial flow needle), and the single needle dialysis set (alternating flow needle). (3) Accessories common to either type include the shunt adaptor, cannula clamp, shunt connector, shunt stabilizer, vessel dilator, disconnect forceps, shunt guard, crimp plier, tube plier, crimp ring, joint ring, fistula adaptor, and declotting tray (including contents).

## Special Controls

*Classification.* (1) Class II (special controls) for the implanted blood access device. The special controls for this device are:(i) Components of the device that come into human contact must be demonstrated to be biocompatible. Material names and specific designation numbers must be provided.
(ii) Performance data must demonstrate that the device performs as intended under anticipated conditions of use. The following performance characteristics must be tested:
(A) Pressure versus flow rates for both arterial and venous lumens, from the minimum flow rate to the maximum flow rate in 100 milliliter per minute increments, must be established. The fluid and its viscosity used during testing must be stated.
(B) Recirculation rates for both forward and reverse flow configurations must be established, along with the protocol used to perform the assay, which must be provided.
(C) Priming volumes must be established.
(D) Tensile testing of joints and materials must be conducted. The minimum acceptance criteria must be adequate for its intended use.
(E) Air leakage testing and liquid leakage testing must be conducted.
(F) Testing of the repeated clamping of the extensions of the catheter that simulates use over the life of the device must be conducted, and retested for leakage.
(G) Mechanical hemolysis testing must be conducted for new or altered device designs that affect the blood flow pattern.
(H) Chemical tolerance of the device to repeated exposure to commonly used disinfection agents must be established.
(iii) Performance data must demonstrate the sterility of the device.
(iv) Performance data must support the shelf life of the device for continued sterility, package integrity, and functionality over the requested shelf life that must include tensile, repeated clamping, and leakage testing.
(v) Labeling of implanted blood access devices for hemodialysis must include the following:
(A) Labeling must provide arterial and venous pressure versus flow rates, either in tabular or graphical format. The fluid and its viscosity used during testing must be stated.
(B) Labeling must specify the forward and reverse recirculation rates.
(C) Labeling must provide the arterial and venous priming volumes.
(D) Labeling must specify an expiration date.
(E) Labeling must identify any disinfecting agents that cannot be used to clean any components of the device.
(F) Any contraindicated disinfecting agents due to material incompatibility must be identified by printing a warning on the catheter. Alternatively, contraindicated disinfecting agents must be identified by a label affixed to the patient's medical record and with written instructions provided directly to the patient.
(G) Labeling must include a patient implant card.
(H) The labeling must contain comprehensive instructions for the following:
(
*1* ) Preparation and insertion of the device, including recommended site of insertion, method of insertion, and a reference on the proper location for tip placement;(
*2* ) Proper care and maintenance of the device and device exit site;(
*3* ) Removal of the device;(
*4* ) Anticoagulation;(
*5* ) Management of obstruction and thrombus formation; and(
*6* ) Qualifications for clinical providers performing the insertion, maintenance, and removal of the devices.(vi) In addition to Special Controls in paragraphs (b)(1)(i) through (v) of this section, implanted blood access devices that include subcutaneous ports must include the following:
(A) Labeling must include the recommended type of needle for access as well as detailed instructions for care and maintenance of the port, subcutaneous pocket, and skin overlying the port.
(B) Performance testing must include results on repeated use of the ports that simulates use over the intended life of the device.
(C) Clinical performance testing must demonstrate safe and effective use and capture any adverse events observed during clinical use.
(vii) In addition to Special Controls in paragraphs (b)(1)(i) through (v) of this section, implanted blood access devices with coatings or additives must include the following:
(A) A description and material characterization of the coating or additive material, the purpose of the coating or additive, duration of effectiveness, and how and where the coating is applied.
(B) An identification in the labeling of any coatings or additives and a summary of the results of performance testing for any coating or material with special characteristics, such as decreased thrombus formation or antimicrobial properties.
(C) A Warning Statement in the labeling for potential allergic reactions including anaphylaxis if the coating or additive contains known allergens.
(D) Performance data must demonstrate efficacy of the coating or additive and the duration of effectiveness.
(viii) The following must be included for A-V shunt cannulae (with vessel tips):
(A) The device must comply with Special Controls in paragraphs (b)(1)(i) through (v) of this section with the exception of paragraphs (b)(1)(ii)(B), (b)(1)(ii)(C), (b)(1)(v)(B), and (b)(1)(v)(C), which do not apply.
(B) Labeling must include Warning Statements to address the potential for vascular access steal syndrome, arterial stenosis, arterial thrombosis, and hemorrhage including exsanguination given that the device accesses the arterial circulation.
(C) Clinical performance testing must demonstrate safe and effective use and capture any adverse events observed during clinical use.
(2) Class II (performance standards) for the nonimplanted blood access device.
(3) Class II (performance standards) for accessories for both the implanted and the nonimplanted blood access devices not listed in paragraph (b)(4) of this section.
(4) Class I for the cannula clamp, disconnect forceps, crimp plier, tube plier, crimp ring, and joint ring, accessories for both the implanted and nonimplanted blood access device. The devices subject to this paragraph (b)(4) are exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.

## Predicate Devices

- Pourchez XpressO™ ([K013160](/device/K013160.md) and [K021212](/device/K021212.md))
- Kendall Tandem-Cath™ ([K002902](/device/K002902.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

K022000 Page 1 of 2

Image /page/0/Picture/1 description: The image shows the logo for Spire Biomedical. The logo consists of the word "spire" in a bold, sans-serif font, with a stylized sun or starburst symbol above the "s". Below "spire" is the word "Biomedical" in a smaller, italicized font. The entire logo is in black and white.

Spire Biomedical, Inc. · One Patriots Park · Bedford, MA 01730-2396 (781) 275-6001 • (781) 275-6010 fax

JUL 1 9 2002

#### SECTION 7

# 510(K) Summary

Pourchez RetrO™ Twin Lumen Chronic Hemodialysis Catheter with Separated Tips (with and without side holes)

- Date: June 18, 2002
- Spire Biomedical. Inc. Submitter: One Patriots Park Bedford, MA 01730-2396 Phone: (781) 275-6001 Fax: (781) 275-6010
- Contact Person: Donald Fickett Director of RA/QA Spire Biomedical. Inc. Phone: (781) 275-6001 x221 Fax: (781) 275-6010 e-mail: dfickett@spirecorp.com

#### Device Names:

| Trade Name:  | Pourchez RetrO™ Twin Lumen Chronic Hemodialysis Catheter<br>with Separated Tips |
|--------------|---------------------------------------------------------------------------------|
| Common Name: | Catheter, Intravascular, Long-Term                                              |

Common Name: Catheter, Intravascular, Long-Term

Classification Name: Catheter, Hemodialysis, Implant (Long-Term)

#### Legally Marketed Devices to Which Substantial Equivalence is Claimed:

- 1) Spire Biomedical, Inc. Pourchez XpressO™ (K013160 and K021212)
- 2) Kendall Tandem-Cath™ (for dynamic and static flow comparison K002902)

Device Description: The Pourchez RetrO™ Twin Lumen Chronic Hemodialysis Catheter with Separated Tips is a flexible radiopaque silicone catheter. The distal end has two tips separated over a predetermined distance. The distal arterial and venous lumens are round and are staggered to reduce recirculation. The body is oval, and the proximal end, which is connected to the catheter body by color-coded extension adapters, has two distinctive lumens with color-coded adapters (red for arterial and blue for venous). The catheter is available in five different implant lengths. It is also available with and without side holes on the distal ends. The catheter has a polyester cuff located at one of five different implant lengths from the distal end.

A RetrO catheter repair kit will also be offered to replace damaged extensions and/or extension adapters.

{1}------------------------------------------------

Ko222000 Page 2 of 2

Image /page/1/Picture/1 description: The image shows the logo for Spire Biomedical. The logo consists of the word "spire" in a bold, sans-serif font, with the word "Biomedical" in a smaller, italicized font underneath. Above the word "spire" is a circular graphic that resembles a sun or a stylized flower.

Spire Biomedical, Inc. • One Patriots Park • Bedford, MA 01730-2396 781) 275-6001 • (781) 275-6010 fax

# 510(K) Summary (Continued)

Pourchez RetrO™ Twin Lumen Chronic Hemodialysis Catheter with Separated Tips (with and without side holes)

#### Intended Use

The indicated use and indications of the Pourchez RetrO™ catheter have not changed. It is designed for chronic (long-term) hemodialysis and apheresis. It is designed for percutaneous insertion or insertion via cutdown into the jugular or subclavian vein.

#### Technological Characteristics Comparison to Predicate Devices

The Pourchez RetrO™ Catheter has the same materials of construction, intended use, implant lengths, number of lumens, cross-sectional lumen area, insertion method and insertion sites, blood recirculation and mechanical hemolysis rates, and sterilization methods as the Pourchez XpressO™ Catheter.

Each catheter is 52cm overall with the usable lengths marked every 5cm to 50cm. The extensions are connected to the catheter body using color-coded (red for arterial and blue for venous) locking connectors made of same material used for the luer adapters.

Additionally, the Pourchez RetrO™ catheter has similar flow rates and priming volumes as the Kendall Tandem™ catheter.

#### Performance Data

A series of tests were performed to demonstrate substantial equivalence to predicate devices or conformation to established ISO standards for hemodialysis catheters. In all cases, the Pourchez RetrO™ catheter demonstrated equivalent performance to the predicate devices and acceptance criteria established by the appropriate standard.

{2}------------------------------------------------

Image /page/2/Picture/1 description: The image is a black and white logo for the Department of Health & Human Services USA. The logo features a stylized image of three human profiles facing to the right, with flowing lines extending from the bottom of the profiles. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" is arranged in a circular pattern around the image.

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

## JUL 1 9 2002

Mr. Donald Fickett Director of RA/OA Spire Biomedical, Inc. One Patriots Park BEDFORD MA 01730-2396

Re: K022000

Trade/Device Name: Pourchez RetrO™ Twin Lumen Chronic Hemodialysis Catheter with Separated Tips Regulation Number: 21 CFR 876.5540 Regulation Name: Blood access device and accessories Regulatory Class: III Product Code: 78 MSD Dated: June 18, 2002 Received: June 19, 2002

Dear Mr. Fickett:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. However, vou are responsible to determine that the medical devices you use as components in the kit have either been determined as substantially equivalent under the premarket notification process (Section 510(k) of the act), or were legally on the market prior to May 28, 1976, the enactment date of the Medical Device Amendments. Please note: If you purchase your device components in bulk (i.e., unfinished) and further process (e.g., sterilize) you must submit a new 510(k) before including these components in your kit. 'The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, and labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{3}------------------------------------------------

Page 2 - Mr. Donald Fickett

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration (21 CFR Part 807); listing (21 CFR Part 807), labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on the labeling regulation (21 CFR Part 801 and additionally Part 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4610. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification"(21 CFR Part 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers, International and Consumer Assistance at its toll free number (800) 638-2041 or (301) 443-6597, or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html.

Sincerely yours,

David A. Glynn

for

Nancy C. Brogdon Director, Division of Reproductive, Abdominal, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

KO22000

Image /page/4/Picture/1 description: The image shows the logo for Spire Biomedical. The word "spire" is written in a bold, sans-serif font. Above the word "spire" is a circular design that resembles a sun or a stylized flower. Below the word "spire" is the word "Biomedical" in a smaller, italicized font.

Spire Biomedical, Inc. · One Patriots Park · Bedford, MA 01730-2396 (781) 275-6001 • (781) 275-6010 fax

### APPENDIX B - Indications for Use Statement

- Pourchez RetrO™ Twin Lumen Chronic Hemodialysis Catheter Device Name: with Separated Tips Indications for Use: Spire Biomedical, Inc.'s Pourchez RetrO™ Silicone Twin Lumen
- Chronic Hemodialysis Catheter with Separated Tips is designed for chronic hemodialysis and apheresis. It is a radiopaque silicone catheter designed for percutaneous insertion or insertion via a cutdown.

#### (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

*Prescription Use* ✓

(Optional Format 3-10-98)

David h. Syson

(Division Sign-Off) Division of Reproductive, Ab and Radinlogical Devices 5 : ປິ້າ Number

---

**Source:** [https://fda-staging.innolitics.com/device/K022000](https://fda-staging.innolitics.com/device/K022000)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).