AXYA MEDICAL, INC, MODEL 100 SONIC SCALPEL ULTRASONIC SURGICAL SYSTEM

K021929 · Axya Medical, Inc. · LFL · Sep 10, 2002 · SU

Device Facts

Record IDK021929
Device NameAXYA MEDICAL, INC, MODEL 100 SONIC SCALPEL ULTRASONIC SURGICAL SYSTEM
ApplicantAxya Medical, Inc.
Product CodeLFL · SU
Decision DateSep 10, 2002
DecisionSESE
Submission TypeTraditional
Device ClassClass U
AttributesTherapeutic

Indications for Use

The Axya Model 100 Sonic Scalpel™ Ultrasonic Surgical System is indicated for use in endoscopic and open surgical procedures for the cutting and coagulation of soft tissue structures.

Device Story

System comprises ultrasonic generator, reusable handpiece with transducer, and disposable titanium shears; shears feature fixed and movable blades for cutting, coagulation, and blunt dissection. Used in endoscopic and open surgical procedures; operated by surgeons. Generator drives transducer to create ultrasonic energy at blade tip; mechanical vibration facilitates tissue cutting and coagulation. Device provides surgical utility similar to electrosurgical systems; benefits include controlled tissue interaction during procedures.

Clinical Evidence

In vitro studies evaluated blade excursion (amplitude) and temperature across power settings and operating conditions. In vivo study in laboratory animals evaluated efficacy, showing performance similar to typical electrosurgical systems.

Technological Characteristics

Ultrasonic surgical system; titanium instrument blades; reusable handpiece with ultrasonic transducer; disposable shears. Complies with IEC 601-1, IEC 601-1-2, CISPR11, and UL2601-1. Sterilization of single-patient-use shears.

Indications for Use

Indicated for use in endoscopic and open surgical procedures for the cutting and coagulation of soft tissue structures.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ # KOZI929 ## SUMMARY OF SAFETY AND EFFECTIVENESS #### General Company Information | Name: | Axya Medical, Inc. | |------------|--------------------------------------------------------| | Address: | 100 Cummings Center<br>Suite 444C<br>Beverly, MA 01915 | | Telephone: | (978) 232 - 9997 | | Fax: | (978) 232 - 9998 | #### General Device Information | Product Name: | Model 100 Sonic Scalpel™ Ultrasonic Surgical System | |-----------------|-------------------------------------------------------------| | Classification: | "Ultrasonic Surgical System", Product code: LFL<br>Class II | ### Predicate Devices Ultracision, Inc. (currently marketed by Ethicon, Inc.) Ultrasonic Scalpel [501(k) Number K895252] Olympus Optical Co. SonoSurg™ System [510(k) Number K972114]. ### Description The Sonic Scalpel™ Ultrasonic Surgical System consists of an Ultrasonic Generator / Co. Unit, a reusable handpiece that contains the ultrasonic transducer, and a family of disposable cutting / coagulation shears. The Sonic Scalpel™ shears are available with a range of shaft lengths, diameters and blade lengths as single-patient-use, sterile instruments. The shears are coupled to the reusable handpiece by means of a collar. The titanium instrument blade tip consists of one fixed blade and one movable blade. The tip may be rotated to facilitate the surgical approach. The shears are designed for cutting, coagulation and blunt dissection. {1}------------------------------------------------ ### Indications The Axva Model 100 Sonic Scalpel™ Ultrasonic Surgical System is indicated for use in endoscopic and open surgical procedures for the cutting and coagulation of soft tissue structures. #### Substantial Equivalence This submission supports the position that the Axya Model 100 Ultrasonic Surgical System is substantially equivalent to a number of previously cleared devices, including the Ultracision. Inc. (currently marketed by Ethicon, Inc.) Ultrasonic Scalpel [501(k) Number K895252] and the Olympus Optical Co. SonoSurg™ System [510(k) Number K9721141. These systems are designed, manufactured and tested to meet the requirements of IEC 601-1, IEC 601-1-2. CISPR11 and UL2601-1. The 510(k) Notice contains summaries of in vitro studies that were conducted to evaluate the shears blade excursion (amplitude) at various power settings and the blade temperature over a range of operating conditions. The data presented demonstrate that the blade amplitude and temperature range of the Axya Model 100 Sonic Scalpel Ultrasonic Surgical System is comparable to those parameters for the predicate Ultracision device. The Notice contains a summary report of an in vivo study that evaluated the efficacy of the Ultrasonic Surgical System in laboratory animals. The results of this testing indicate that the system produces results similar to the performance of typical electrosurgical systems. The single-patient-use shears component of the Ultrasonic Surgical System is provided sterile. Axya Medical. Inc. believes that the information provided establishes that similar legally marketed systems have been used for the same clinical applications as the Axya Model 100 Ultrasonic Surgical System. The materials from which the Axya device is fabricated have an established history of use in medical applications. {2}------------------------------------------------ Image /page/2/Picture/1 description: The image shows a logo for the Department of Health & Human Services. The logo features the words "DEPARTMENT OF HEALTH & H" arranged in a semi-circle on the left side of the image. To the right of the text is a symbol consisting of four curved lines that resemble a stylized human figure. Public Health Service Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 SEP 1 0 2002 Axya Medical, Inc. Howard L. Schrayer 100 Cummings Center, Suite 444 C Beverly, Massachusetts 01915 Re: K021929 Trade/Device Name: Axya, Model 100 Sonic Scalpel™ Ultrasonic Surgical System Regulatory Class: Unclassified Product Code: LFL Dated: June 11, 2002 Received: June 12, 2002 Dear Mr. Schrayer: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. {3}------------------------------------------------ Page 2 – Mr. Howard L. Schrayer This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 21 CFR Part 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4659. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97), Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html Sincerely yours, Celia M. Witten, Ph.D., M.D. Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {4}------------------------------------------------ 510(k) Number (if known): ____________________________________________________________________________________________________________________________________________________ Device Name: Axya, Model 100 Sonic Scalpel™ Ultrasonic Surgical System Indications For Use: The Axya Model 100 Sonic Scalpel™ Ultrasonic Surgical System is indicated for use in endoscopic and open surgical procedures for the cutting and coagulation of soft tissue structures. (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) Styt Purke (Division Sign-Off) Division of General, Restorative and Neurological Devices 510(k) Number_Ko21929 **Prescription Use** (Per 21 CFR 801.109) [checkmark] OR Over-The-Counter Use _________________________________________________________________________________________________________________________________________________________ (Optional Format 1-2-96)
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...