# HSA-SOLUTION, MODEL 10034/10064 (K021896)

_Vitrolife Sweden AB · MQL · Sep 6, 2002 · Obstetrics/Gynecology · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K021896

## Device Facts

- **Applicant:** Vitrolife Sweden AB
- **Product Code:** [MQL](/productcode/MQL.md)
- **Decision Date:** Sep 6, 2002
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 884.6180
- **Device Class:** Class 2
- **Review Panel:** Obstetrics/Gynecology

## Indications for Use

HSA-solution contains Human Serum Albumin solution (100 mg/mL) and is intended for use in assisted reproductive procedures which include gamete and embryo manipulation. These procedures include the use of HSA-solution as a supplement for culture medium. Not for use as an injectable product.

## Device Story

HSA-solution is a human serum albumin solution (100 mg/mL) used as a supplement for culture media in assisted reproductive technology (ART) procedures. It is utilized in clinical laboratory settings by embryologists or trained laboratory personnel to support the manipulation and culture of gametes and embryos. The device is not for injection. It functions by providing necessary protein supplementation to the culture environment, maintaining the physiological conditions required for gamete and embryo development outside the body.

## Clinical Evidence

No clinical data provided; substantial equivalence is based on technological characteristics and intended use identical to legally marketed predicate devices.

## Technological Characteristics

Human Serum Albumin solution (100 mg/mL). Classified under 21 C.F.R. § 884.6180 as a reproductive media and supplement. Product code: MQL.

## Regulatory Identification

Reproductive media and supplement are products that are used for assisted reproduction procedures. Media include liquid and powder versions of various substances that come in direct physical contact with human gametes or embryos (including water, acid solutions used to treat gametes or embryos, rinsing solutions, sperm separation media, supplements, or oil used to cover the media) for the purposes of preparation, maintenance, transfer or storage. Supplements are specific reagents added to media to enhance specific properties of the media (e.g., proteins, sera, antibiotics, etc.).

## Special Controls

*Classification.* Class II (special controls) (mouse embryo assay information, endotoxin testing, sterilization validation, design specifications, labeling requirements, biocompatibility testing, and clinical testing). The device, when it is phosphate-buffered saline used for washing, and short-term handling and manipulation of gametes and embryos; culture oil used as an overlay for culture media containing gametes and embryos; and water for assisted reproduction applications, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 884.9.

## Predicate Devices

- Human Serum Albumin™ (K98354)

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

## SEP 6 2002

## IX. PREMARKET NOTIFICATION SUMMARY

| Submitted by:                  | Vitrolife Sweden AB<br>Mölndalsvägen 30<br>SE-412 63 Gothenburg<br>SWEDEN                                                                                                                                                                                                                                             |
|--------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Contact Person:                | Mr. Eiler Anderson<br>Vitrolife Sweden AB<br>Mölndalsvägen 30<br>SE-412 63 Gothenburg<br>SWEDEN                                                                                                                                                                                                                       |
| Date Prepared:                 | June 7, 2002                                                                                                                                                                                                                                                                                                          |
| Trade Name:                    | HSA-solution™                                                                                                                                                                                                                                                                                                         |
| Common Name:                   | Assisted Reproduction Media                                                                                                                                                                                                                                                                                           |
| Classification Name:           | Reproductive Media and Supplements<br>(21 C.F.R. § 884.6180)                                                                                                                                                                                                                                                          |
| Predicate Device:              | Human Serum Albumin™ (510(k) # K98354)                                                                                                                                                                                                                                                                                |
| Description of the Device:     | Human Serum Albumin solution.<br>For supplementation of gamete and embryo<br>media.                                                                                                                                                                                                                                   |
| Intended Use:                  | HSA-solution contains Human Serum Albumin<br>solution (100 mg/mL) and is intended for use in<br>assisted reproductive procedures which include<br>gamete and embryo manipulation. These<br>procedures include the use of HSA-solution as a<br>supplement for culture medium. Not for use as an<br>injectable product. |
| Technological Characteristics: | The technological characteristics of HSA-<br>solution™ are identical to other legally marketed<br>culture media classified under 21 C.F.R. §<br>884.6180, Reproductive Media and Supplements.                                                                                                                         |

{1}------------------------------------------------

Image /page/1/Picture/1 description: The image is a black and white seal for the Department of Health & Human Services. The seal is circular with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" around the perimeter. In the center of the seal is a stylized image of three human profiles facing right, with flowing lines representing hair or clothing.

11 11 11 11

SEP 6 200

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 -

Vitrolife Sweden AB % Mr. Gary L. Yingling Kirkpatrick & Lockhart 1800 Massachusetts Avenue, NW WASHINGTON DC 20036-1800 Re: K021896

Trade/Device Name: HSA-solution™ Regulation Number: 21 CFR 884.6180 Regulation Name: Reproductive media and supplements Regulatory Class: II Product Code: 85 MQL Dated: June 7, 2002 Received: June 10, 2002

Dear Mr. Yingling:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (sections 531-542 of the Act); 21 CFR 1000-1050.

{2}------------------------------------------------

This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at one of the following numbers, based on the regulation number at the top of this letter:

| 8xx.1xxx                         | (301) 594-4591 |
|----------------------------------|----------------|
| 876.2xxx, 3xxx, 4xxx, 5xxx       | (301) 594-4616 |
| 884.2xxx, 3xxx, 4xxx, 5xxx, 6xxx | (301) 594-4616 |
| 892.2xxx, 3xxx, 4xxx, 5xxx       | (301) 594-4654 |
| Other                            | (301) 594-4692 |

Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html.

Sincerely yours,

Henry C. Grigdon

Nancy C. Brogdon Director, Division of Reproductive, Abdominal, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{3}------------------------------------------------

## X. - INDICATIONS FOR USE STATEMENT

510(k)_Number:

KO21896

Device Name:

HSA-solution™ Assisted Reproduction Media

Indications For Use:

HSA-solution contains Human Serum Albumin solution (100 mg/mL) and is intended for use in assisted reproductive procedures which includc gamete and embryo manipulation. These procedures include the use of HSA-solution as a supplement for culture medium. Not for use as an injectable product.

Rate Phillips ncs

(Division Sign-Off) Division of Reproductive, Abdominal, and Radiological Devices 5 10(k) Number _

(PLEASE DO NOT WRITE BELOW THIS LINE – CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE)

Prescription Use (Per 21 C.F.R. § 801.109)

OR

Over-the Counter Use_

---

**Source:** [https://fda-staging.innolitics.com/device/K021896](https://fda-staging.innolitics.com/device/K021896)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com