The PORGES™ Ureteral catheters are used for ureteral catheterisation: - drainage catheters: straight catheters with bevel, olive or cylindrical tip, elbowed catheter with olive tip - . catheters for retrograde ureteropyelography (R.U.P.): Chevassu, cone or Braasch - interventional catheters: flush, open straight, open elbowed. The appropriate guide-wire size for each catheter is shown on the box.
Device Story
Single-lumen ureteral catheter; medical-grade radiopaque polymer; sizes 03 FR to 12 FR; optional removable metal or polymer stylet for straightening during insertion; silicone stop ring on stylet for tip positioning. Used by physicians for ureteral drainage, retrograde ureteropyelography, or interventional procedures via Seldinger guide-wire. Catheter provides access to ureter; facilitates fluid drainage or contrast medium delivery. Benefits include improved procedural access and patient management during urological interventions.
Clinical Evidence
Bench testing only. No clinical data. Substantial equivalence demonstrated through mechanical/physical testing including flow rate, tensile strength, tear strength of bulbs, and passage of contrast medium. Biocompatibility testing performed per ISO 10993-1.
Technological Characteristics
Single-lumen catheter; medical-grade radiopaque thermoplastic (PVC, polyamide, polyurethane, or PEBA); graduated every cm along 50 cm; sizes 03-12 FR. Features include various tip configurations (bevel, olive, cylindrical, Chevassu, cone, Braasch, open straight/elbowed). Optional removable stylet. Sterile, single-use. Biocompatibility per ISO 10993-1.
Indications for Use
Indicated for patients requiring ureteral catheterization for drainage, retrograde ureteropyelography (R.U.P.), or interventional procedures (e.g., placement over a guide-wire).
Regulatory Classification
Identification
A urological catheter and accessories is a flexible tubular device that is inserted through the urethra and used to pass fluids to or from the urinary tract. This generic type of device includes radiopaque urological catheters, ureteral catheters, urethral catheters, coudé catheters, balloon retention type catheters, straight catheters, upper urinary tract catheters, double lumen female urethrographic catheters, disposable ureteral catheters, male urethrographic catheters, and urological catheter accessories including ureteral catheter stylets, ureteral catheter adapters, ureteral catheter holders, ureteral catheter stylets, ureteral catheterization trays, and the gastro-urological irrigation tray (for urological use).
C.R. Bard Inc. Urological Division's currently marketed ureteral catheter line
Submission Summary (Full Text)
{0}------------------------------------------------
# OCT 1 7 2002
Section J: 510(k) Sum
#### Ref. US021A41.DOC
Origin : Regulatory Affairs
Image /page/0/Picture/4 description: The image is a logo for Porges. The logo consists of a stylized symbol above the word "PORGES". The symbol is a circle with a line extending down from the center, ending in a small circle. The word "PORGES" is in a bold, sans-serif font.
Origin : Regulatory Affairs
This 510(k) summary of safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990.
#### J.1. Submitter's information
| Submitter's name: | PORGES S.A. |
|-------------------------|------------------------------------------------------------------------|
| Submitter's address: | Centre d'Affaires La Boursidière<br>92357 Le Plessis Robinson - France |
| Contact person: | Mr Bernard ISMAEL<br>Regulatory Affairs Manager |
| Contact person's phone: | + 33 1 46 01 32 06 |
| Contact person's fax: | + 33 1 46 01 32 56 |
| Contact person's email: | bernard.ismael@porges.com |
| Date of preparation: | February, 2002 |
#### J.2. Device name
| Classification name: | Catheter, ureteral, gastro-urolog |
|----------------------|-----------------------------------|
| Common / Usual name: | Ureteral catheters |
| Proprietary name: | PORGES™ Ureteral catheters |
#### J.3. Predicate devices
The PORGES™ Ureteral catheters are substantially equivalent to the PORGES™ pre-amendment range of ureteral catheters, C.R. Bard Inc. Urological Division's currently marketed ureteral catheter line and Bard Flexible Tip Ureteral catheter (K950300). PORGES S.A., is not aware of any information suggesting that C.R. Bard Inc. ureteral catheters are not legally marketed. Ureteral catheters may also be referic catheters, both denominations being equally used within the medical profession.
#### J.4. Description of the Device
The PORGES™ Ureteral catheters consist of a single lumen catheter in medical grade radiopaque polymer, graduated every cm along 50 cm, and fitted with either a metal or a polymer removable stylet or no stylet. The stylet allows straightening of the catheter for easier insertion. A wax knob is added on the external end of the stylet for easy retreat of the coudé tip catheters are fitted with a silicone stop ring on the stylet between the catheter and the wax knob to hold the renal tip of the stylet away from the catheter's renal tip.
The ureteral catheters are available with different types of renal tips and in sizes ranging from 03 FR to 12 FR.
The preamendment devices have been designed for R.U.P. (retrograde ureteropyelography).
PORGES has validated the following uses for the new range of ureteral catheters :
- . drainage
- R.U.P. (= retrograde uretero pyelography) .
- interventional ("per operating") use with a Seldinger guide wire (not supplied)
The predicate devices Bard® Flexible Tip ureteral catheters (K950300) and C.R. Bard's currently marketed devices are designed for the same intended use.
The PORGES™ Ureteral catheters are supplied sterile and for single use only.
#### J.5. Intended use of the Device
The PORGES™ Ureteral catheters are used for ureteral catheterisation:
- drainage catheters: straight catheters with bevel, olive or cylindrical tip, elbowed catheter with olive tip ●
- . catheters for retrograde ureteropyelography (R.U.P.): Chevassu, cone or Braasch
interventional catheters: flush, open straight, open elbowed. The appropriate guide-wire size for each catheter is shown on the box.
{1}------------------------------------------------
Section J: 510(k) Summa PORGES™ Ureteral catheters 510(k) submission
Ref. US021A41.DOC
Origin : Regulatory Affairs
Image /page/1/Picture/3 description: The image shows a logo with the word "PORGES" written in bold, block letters at the bottom. Above the word is a symbol that resembles a stylized key or a tuning fork. The symbol consists of a circle with a vertical line extending downwards from its center, ending in a small circle. The logo is enclosed in a rectangular border.
#### J.6. Technological characteristics
The PORGES™ Ureteral catheters have similar technological and performance characteristics to the predicate devices. The PORGES™ Ureteral catheters are similar as it concerns its materials to PORGES and C.R. Bard's currently marketed ureteral catheters and to the Bard Flexible tip ureteral catheter (K950300). All are made of medical grade thermoplastic material. The choice of either a polyvinyl chloride, polyamide, polyurethane or PEBA (polyether block amide) catheter or metal or polymer stylet may provide a slight difference in stiffness, which does not affect clinical properties, but may be preferred by practioners according to the method of use.
The polyvinyl chloride and polyamide mixtures which are used for the manufacturing of most PORGES Ureteral catheters are the same as those used since 1971 for the manufacturing of the PORGES preamendment devices.
The straight open tip of the interventional ureters allows placement over a quide-wire. The straight open tip is similar in performance as the straight whistle tip of both Porges' and Bard's currently marketed catheters.
#### J.7. Test Summary
The PORGES™ Ureteral catheters referenced in this submission are held to the same design, manufacture, and performance specifications as the predicate devices. Substantial equivalence of the devices with respect to functional performance has been demonstrated by mechanical/physical testing. The following tests have been performed :
- Flow rate
- Tensile strength .
- Tear strength of bulbs (Chevassu and cone-tip catheters)
- Passage of contrast medium
The PORGES™ Ureteral catheters have successfully passed biocompatibility testing per ISO 10993-1.
#### J.8. Conclusion.
The biocompatibility and mechanical/physical tests confirm the safety and effectiveness for these devices is assured and these devices perform as well as or better than the predicate devices.
{2}------------------------------------------------
### DEPARTMENT OF HEALTH & HUMAN SERVICES
Public Health Service
Image /page/2/Picture/2 description: The image is a black and white logo for the U.S. Department of Health & Human Services. The logo features a stylized caduceus symbol, which is a staff with two snakes coiled around it. The caduceus is surrounded by the text "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" in a circular arrangement. The text is in all capital letters and is evenly spaced around the caduceus symbol.
## OCT 1 7 2002
Food and Drug Administratio 9200 Corporate Boulevard Rockville MD 20850
Re: K021856
Trade/Device Name: PORGES™ Ureteral Catheters Models ACN6xx, ACN5xx, AC5B, AC5C07, ACP5xx, ACP2xx, ACP3xx, ACP4xx, and ACP6xx Regulation Number: 21 CFR §876.5130 Regulation Name: Urological catheter and accessories Regulatory Class: II Product Code: 78 EYB Dated: September 24, 2002 Received: October 7, 2002
Dear Mr. ISMAEL:
Mr. Bernard ISMAEL
PORGES S.A.
FRANCE
Regulatory Affairs Manager
Centre d'Affaires La Boursidière 92357 Le Plessis-Robinson Cedex
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (sections 531-542 of the Act); 21 CFR 1000-1050.
{3}------------------------------------------------
Page 2
This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at one of the following numbers, based on the regulation number at the top of this letter:
| 8xx.1xxx | (301) 594-4591 |
|----------------------------------|----------------|
| 876.2xxx, 3xxx, 4xxx, 5xxx | (301) 594-4616 |
| 884.2xxx, 3xxx, 4xxx, 5xxx, 6xxx | (301) 594-4616 |
| 892.2xxx, 3xxx, 4xxx, 5xxx | (301) 594-4654 |
| Other | (301) 594-4692 |
Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers. International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html.
Sincerely yours,
Nancy C. Croydon.
Nancy C. Brogdon Director, Division of Reproductive, Abdominal, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
# Section F: Indications for Use Statement
PORGES™ Ureteral catheters 510(k) submission
Ref. US021A41.DOC
Origin : Requlatory Affairs
Image /page/4/Picture/4 description: The image is a logo for Porges. The logo features a stylized image of a key with a circular head. The word "PORGES" is written in bold, sans-serif letters below the key image. The logo is black and white and has a simple, modern design.
Page 1 of 1
510(k) Number (if known): K1021856
Device Name: PORGES™ Ureteral catheters
Indications for use:
The PORGES™ Ureteral catheters are used for ureteral catheterisation:
- drainage catheters: straight catheters with bevel, olive or cylindrical tip, elbowed catheter with olive tip
- catheters for retrograde ureteropyelography (R.U.P.): Chevassu, cone or Braasch -
- interventional catheters: flush, open straight, open elbowed. The appropriate quide-wire size for each catheter is shown on the box.
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
CONCURRENCE OF CDRH, OFFICE OF DEVICE EVALUATION (ODE)
(Optional Format 3-10-98)
Prescription Use
David A. Segerson
IDivision Sign-Off) Division of Seproductive, Abdominal, and Padiciogical Devices 510(k) Number ________________________________________________________________________________________________________________________________________________________________
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.