SORIN BIOMEDICA SMARXT BCD VANGUARD SURFACE MODIFIED CARDIOPLEGIA SYSTEM
Applicant
Cobe Cardiovascular, Inc.
Product Code
DTR · Cardiovascular
Decision Date
Jul 2, 2002
Decision
SESE
Submission Type
Special
Regulation
21 CFR 870.4240
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The SMAR, T® BCD Vanguard™ Blood Cardioplegia System is intended to mix, cool, warm and deliver oxygenated blood and cardioplegic solution for periods up to six hours. The device also allows monitoring of temperature and pressure, traps bubbles, and removes air. Blood and cardioplegic solution are delivered to the patient by a single 100 % occlusive roller pump, through the extension line and appropriate cannula.
Device Story
Cardioplegia heat exchanger with integral bubble trap; used during cardiopulmonary bypass to mix, cool, warm, and deliver blood/cardioplegic solution. Input: oxygenated blood and cardioplegic solution. Operation: single 100% occlusive roller pump delivers mixture through extension lines and cannula. Features: hydrophobic membrane for air venting, one-way valve, pressure relief valve, filter screen, and ports for temperature/pressure monitoring. Surface-modifying material (Mimesys) coats heat exchanger to improve blood compatibility. Used in clinical settings (e.g., OR) by perfusionists/surgeons. Output: temperature and pressure data via monitoring ports; delivered cardioplegic mixture to patient. Benefits: improved blood compatibility, air removal, and precise thermal/pressure management during cardiac procedures.
Clinical Evidence
Bench testing only. In-vitro tests performed to demonstrate substantial equivalence to the predicate device.
Indicated for patients requiring cardiopulmonary bypass; device mixes, cools, warms, and delivers oxygenated blood and cardioplegic solution for up to six hours; provides bubble trapping, air removal, and temperature/pressure monitoring.
Regulatory Classification
Identification
A cardiopulmonary bypass heat exchanger is a device, consisting of a heat exchange system used in extracorporeal circulation to warm or cool the blood or perfusion fluid flowing through the device.
Predicate Devices
Sorin Biomedica BCD Vanguard™ Blood Cardioplegia System (K934847)
Reference Devices
SMAR, T Surface Modified Tubing and Connectors (K981613)
Submission Summary (Full Text)
{0}------------------------------------------------
021830
PAGE 1 OF 2
Special 510(k) Pre-Market Notification: SMAR,T® BCD Vanguard™ Blood Cardioplegia System COBE Cardiovascular, Inc.
16
# JUL - 2 2002
| SUBMITTER: | COBE Cardiovascular, Inc.<br>14401 W. 65th Way<br>Arvada, CO 80004 |
|----------------------|-------------------------------------------------------------------------------------|
| CONTACT PERSON: | Shawn Riedel<br>Phone:(303) 467-6521<br>Fax: (303) 467-6525 |
| DATE PREPARED: | 28 May 2002 |
| DEVICE TRADE NAME: | Sorin Biomedica SMARₓT® BCD Vanguard™<br>Surface Modified Blood Cardioplegia System |
| COMMON/USUAL NAME: | Cardioplegia Heat Exchanger |
| CLASSIFICATION NAME: | Cardiopulmonary Bypass Heat Exchanger |
| PREDICATE DEVICE: | Sorin Biomedica BCD Vanguard™ Blood Cardioplegia System |
DEVICE DESCRIPTION:
The SMAR, T BCD Vanguard System is a cardioplegia heat exchanger with integral bubble trap, available with various tubing ratio configurations connected to the heat exchanger assembly. Surface-modifying materials have been added to the primary blood contact surfaces of the heat exchanger, tubing and connectors to improve the blood compatibility of the system. The product is sterilized by ethylene oxide, is for single use only, and has nonpyrogenic fluid pathways.
The heat exchanger assembly (heat exchanger with bubble trap) contains a hydrophobic membrane, a one-way valve, a pressure relief valve, a filter screen, a temperature monitoring port and a pressure monitoring port. The housing is designed such that air entering the device rises to the membrane and is vented to the atmosphere. The one-way valve keeps air from entering the device. The pressure relief valve is designed to relieve excess pressure in the event of outlet line clamping, and has tubing attached to it to vent the solution to the cardiotomy reservoir. The temperature monitoring port is used to measure the temperature of the blood cardionlegia mixture, and the pressure monitoring line is connected to either a pressure manometer or a transducer to measure pressure.
#### INDICATIONS FOR USE
The SMAR, T BCD Vanguard Blood Cardioplegia System is intended to mix, cool, warm and deliver oxygenated blood and cardioplegic solution for periods up to six hours. The device also allows monitoring of temperature and pressure, traps bubbles, and removes air. Blood and cardioplegic solution are delivered to the patient by a single 100 % occlusive roller pump, through the extension line and appropriate cannula.
{1}------------------------------------------------
Page 2 of 2 Special 510(k) Pre-Market Notification: SMAR T® BCD Vanguard" Blood Cardioplegia System COBE Cardiovascular, Inc.
#### 17
## STATEMENT OF TECHNICAL CHARACTERISTICS COMPARISON
The SMAR.T BCD Vanguard Surface Modified Blood Cardioplegia System has the same intended use as the Sorin BCD Vanguard Blood Cardioplegia System. Unless noted below the , materials, components, design, sterilization and manufacturing processes for the two devices are the same. The new device will be called the Sorin Biomedica SMAR T® BCD Vanguard™ Surface Modified Blood Cardioplegia System.
The changes in the two devices: 1) the SMAR, T BCD Vanguard contains a surface-modifying material. Mimesys which coats the heat exchanger (not including the bubble trap) of the System, 2) Substituting SMAR, T Surface Modified Tubing and Connectors (K981613) for various tubing and connectors within the circuit configurations and 3) Substituting High Impact Polystyrene (HIPS) for PolyEthylene Terephthalate Glycol (PETG) as the packaging material.
In-Vitro tests were performed to demonstrate that the SMAR-T BCD Vanguard Blood Cardioplegia System described in this submission is substantially equivalent to the BCD Vanguard Blood Cardioplegia System (K934847).
### Truthful and Accurate Statement
A statement attesting to the truthfulness and accuracy of the information contained in this submission is attached as Appendix 5.
### Further Information
In the event additional information is required, please contact:
Shawn Riedel Regulatory/Quality Manager COBE Cardiovascular, Inc. 14401 West 65th Way Arvada, Colorado 80004 USA Phone: (303) 467-6521 Fax: (303) 467-6525
{2}------------------------------------------------
Image /page/2/Picture/1 description: The image shows the seal of the Department of Health & Human Services USA. The seal is circular, with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" arranged around the perimeter. Inside the circle is an abstract image of an eagle with three human profiles incorporated into its design. The eagle's head and neck are formed by the profiles, creating a symbolic representation of health and human services.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
COBE Cardiovascular, Inc. c/o Mr. Shawn Riedel Regulatory/Quality Manager 14401 W. 65th Way Arvada, CO 80004-3599
JUL = 2 2002
Re: K021830
> Device Name: Sorin Biomedica SMART® BCD Vanguard™ Surface Modified Blood Cardioplegia System Regulation Number: 21 CFR 870.4240 Regulation Name: Cardiopulmonary bypass heat exchanger. Regulatory Class: Class II (two) Product Code: DTR Dated: June 3, 2002 Received: June 4, 2002
Dear Mr. Riedel:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. Iisting of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21
{3}------------------------------------------------
Page 2 - Mr. Shawn Riedel
CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 21 CFR Part 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4646. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97), Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html
· Sincerely yours,
Donna Dee Tillman, Ph.D.
onna-Bea Tillman, Ph.D. Acting Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
3
Special 510(k) Pre-Market Notification: SMARy1® BCD Vanguard™ Blood Cardioplegia System COBE Cardiovascular, Inc.
Indications For Use
510(k) Number (If known): K021832
Device Name:
Sorin Biomedica SMAR, T® BCD Vanguard™ Surface Modified Blood Cardioplegia System
Indications For Use:
The SMAR, T® BCD Vanguard™ Blood Cardioplegia System is intended to mix, cool, warm and deliver oxygenated blood and cardioplegic solution for periods up to six hours. The device also allows monitoring of temperature and pressure, traps bubbles, and removes air. Blood and cardioplegic solution are delivered to the patient by a single 100 % occlusive roller pump, through the extension line and appropriate cannula.
### PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Q. Carter
510(k) Number K021830
Image /page/4/Picture/13 description: The image shows the words "Prescription Use" followed by the text "(Per 21 CFR 801.109)". There is a large X drawn over the words "Prescription Use". The text is in a simple, sans-serif font.
OR
Over-The-Counter Use
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.