← Product Code [NBW](/productcode/NBW) · K021827

# MODIFICATION TO ACCU-CHEK ACTIVE SYSTEM (K021827)

_Roche Diagnostics Corp. · NBW · Jul 3, 2002 · Clinical Chemistry · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K021827

## Device Facts

- **Applicant:** Roche Diagnostics Corp.
- **Product Code:** [NBW](/productcode/NBW.md)
- **Decision Date:** Jul 3, 2002
- **Decision:** SESE
- **Submission Type:** Special
- **Regulation:** 21 CFR 862.1345
- **Device Class:** Class 2
- **Review Panel:** Clinical Chemistry
- **Attributes:** Pediatric

## Indications for Use

The Accu-Chek Active system is designed to quantitatively measure the concentration of glucose in capillary whole blood. The device is indicated for professional use and over-the-counter sale. The Accu-Chek Active system is indicated for use with capillary whole blood samples drawn from the fingertips, forearm, upper arm, thigh, calf and palm.

## Device Story

Accu-Chek Active System is a blood glucose monitoring device for professional and OTC use. It utilizes photometric measurement of a glucose dehydrogenase chemical reaction on test strips. The meter includes a third LCD to detect insufficient blood sample volume (under-dosing). Users apply capillary whole blood from fingertips or alternate sites (forearm, upper arm, thigh, calf, palm) to test strips. The device provides quantitative glucose concentration readings. It assists patients and clinicians in managing blood glucose levels. AST is cautioned against during rapid glucose changes or for patients with hypoglycemia unawareness.

## Clinical Evidence

Clinical study conducted at manufacturing facility comparing AST (palm) results to fingertip results. Sample size N=65 for each palm site (near thumb/small finger). Results showed high correlation (r=0.990 and 0.988) and slope values near 1.0 (1.031 and 1.019). Study confirms system accuracy for palm sampling is consistent with previously cleared alternate sites.

## Technological Characteristics

Photometric measurement system using glucose dehydrogenase reagent chemistry. Features a third LCD for under-dosed test strip detection. Supports alternate site testing (AST) including fingertips, forearm, upper arm, thigh, calf, and palm. Standalone meter device.

## Regulatory Identification

A glucose test system is a device intended to measure glucose quantitatively in blood and other body fluids. Glucose measurements are used in the diagnosis and treatment of carbohydrate metabolism disorders including diabetes mellitus, neonatal hypoglycemia, and idiopathic hypoglycemia, and of pancreatic islet cell carcinoma.

## Special Controls

*Classification.* Class II (special controls). The device, when it is solely intended for use as a drink to test glucose tolerance, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 862.9.

## Predicate Devices

- Accu-Chek Active System ([K012324](/device/K012324.md))
- Accu-Chek Active System ([K021448](/device/K021448.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

3 2002 JUL

K021827

# 510(k) Summary

ে

| Introduction                              | According to the requirements of 21 CFR 807.92, the following information<br>provides sufficient detail to understand the basis for a determination of<br>substantial equivalence.                                                                                                                                                                                    |
|-------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| 1) Submitter<br>name, address,<br>contact | Roche Diagnostics Corporation<br>9115 Hague Rd.<br>Indianapolis, IN 46250<br>(317) 845-2000<br>Contact Person:  Mike Flis<br>Date Prepared:                                                                                                                                                                                                                           |
| 2) Device name                            | Proprietary name: Accu-Chek Active System<br>Classification name: SYSTEM, TEST, BLOOD GLUCOSE, OVER THE<br>COUNTER<br>(21 C.F.R. § 862.1345)(75NBW)                                                                                                                                                                                                                   |
| 3) Predicate<br>device                    | We claim substantial equivalence to the current legally marketed Accu-Chek<br>Active System (K012324 & K021448)                                                                                                                                                                                                                                                       |
| 4) Device<br>Description                  | Instrument Operating Principle -- photometry<br>Reagent Test Principle -- glucose dehydrogenase                                                                                                                                                                                                                                                                       |
| 5) Intended use                           | The Accu-Chek Active system is designed to quantitatively measure the<br>concentration of glucose in capillary whole blood. The device is indicated for<br>professional use and over-the-counter sale. The Accu-Chek Active system is<br>indicated for use with capillary whole blood samples drawn from the<br>fingertips, forearm, upper arm, thigh, calf and palm. |
|                                           | Continued on next page                                                                                                                                                                                                                                                                                                                                                |
| 6) Similarities                           | The proposed modification is relatively modest in scope. All of the following are claims and features unaffected by the proposed modification.                                                                                                                                                                                                                        |

{1}------------------------------------------------

# 510(k) Summary, Continued

、

| Feature/Claim                                 | Detail                                                                                                                                                                                                                                                                                                                                                                                       |
|-----------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Intended use                                  | The Accu-Chek Active systems is designed to quantitatively measure the<br>concentration of glucose in whole blood. The test device is indicated for<br>professional use and over-the-counter sale. This device is not suitable for<br>testing neonate samples.                                                                                                                               |
| Test principle                                | Glucose dehydrogenase chemical reaction. The instrument measures the<br>extent of color change (photometric) caused by the presence of glucose in<br>the sample. The amount of color change is related to the glucose<br>concentration in the blood sample.                                                                                                                                  |
| Warnings and<br>precautions                   | For in vitro diagnostic use only.                                                                                                                                                                                                                                                                                                                                                            |
| Under-dosed test<br>strip detection<br>method | Yes. Includes a third LCD intended solely to detect whether the test pad<br>is sufficiently covered by the blood sample. This design more effectively<br>identifies under-dosed test strips.                                                                                                                                                                                                 |
| Dosing test strips<br>outside of meter        | Possible                                                                                                                                                                                                                                                                                                                                                                                     |
| AST precautionary<br>instructions             | AST is not recommended:<br>- during periods of rapid decreases or increases in blood glucose levels;<br>- for people with a history of recurrent hypoglyecmia, who suspect that<br>their blood glucose is extremely low, or who are unaware of<br>hypoglycemic events.<br>Studies show that AST results immediately before a meal and near<br>bedtime are more similar to fingertip results. |

### 7) Differences

| Feature                                 | Accu-Chek Active System<br>(modified)                | Accu-Chek Active System<br>(predicate)         |
|-----------------------------------------|------------------------------------------------------|------------------------------------------------|
| Self-testing sample<br>collection sites | Finger, forearm, calf, thigh, upper<br>arm, and palm | Finger, forearm, calf, thigh, and<br>upper arm |

Continued on next page

{2}------------------------------------------------

## 510(k) Summary, Continued

8) Data The results of a study conducted at our manufacturing facility demonstrated demonstrating consistent quality performance of this product. This study demonstrated good substantial correlation (r> 0.90) between AST results and finger results under steady state equivalence conditions. With these data it is proved that the system accuracy with AST blood from calf, thigh and upper arm is unchanged from forearm.

AST Sample versus Finger (Same Accu-Chek Active Meter)

| AST Location     | Palm (near thumb) | Palm (near small<br>finger) |
|------------------|-------------------|-----------------------------|
| N                | 65                | 65                          |
| Range (mg/dL)    | 31-371            | 31-371                      |
| Slope            | 1.031             | 1.019                       |
| Slope 95% CI     | (0.994, 1.067)    | (0.978, 1.059)              |
| Intercept        | -0.5              | 1.7                         |
| Intercept 95% CI | (-6.7, 5.6)       | (-5.1, 8.5)                 |
| Correlation      | 0.990             | 0.988                       |
| Std Error        | 11.0              | 12.1                        |

{3}------------------------------------------------

#### DEPARTMENT OF HEALTH & HUMAN SERVICES

Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo is a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" around the perimeter. Inside the circle is an abstract image of an eagle with three stripes representing the agency's mission to protect the health of all Americans and provide essential human services.

Public Health Service

்பட 3 2002 Food and Drug Administration 2098 Gaither Road Rockville MD 20850

Mr. Mike Flis Regulatory Affairs Principal Roche Diagnostics Corporation 9115 Hague Road P.O. Box 50457 Indianapolis, IN 46250-0457

k021827 Re:

> Trade/Device Name: Accu-Chek Active System Regulation Number: 21 CFR 862.1345 Regulation Name: Glucose test system Regulatory Class: Class II Product Code: NBW: LFR Dated: June 3, 2002 Received: June 4, 2002

Dear Mr. Flis:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

{4}------------------------------------------------

### Page 2 -

This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and ' additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4588. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsma/dsmamain.html".

. Sincerely yours.

Steven Putman

Steven I. Gutman, M.D., M.B.A. Director Division of Clinical Laboratory-Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{5}------------------------------------------------

### Indications for Use Statement

K021827 510(k) Number (if known): Device Name: Accu-Chek Active Test System

Indications for Use:

The Accu-Chek Active system is designed to quantitatively measure the concentration of glucose in capillary whole blood. The device is indicated for professional use and over-the-counter sale.

The Accu-Chek Active system is indicated for use with capillary whole blood samples drawn from the fingertips, forearm, upper arm, thigh, calf and palm.

## (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Han
Cooper

(Division Sign-Off)
Division of Clinical Laboratory Devices
510(k) Number 6921827

Prescription Use (Per 21 CFR 801.109) OR Over-The-Counter Use i

(Optional Format 1-2-96)

---

**Source:** [https://fda-staging.innolitics.com/device/K021827](https://fda-staging.innolitics.com/device/K021827)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
