MODIFICATION TO SPIRAL RADIUS 90-D SST SYSTEM

K021825 · Unitedstates Surgical · KWP · Jul 1, 2002 · Orthopedic

Device Facts

Record IDK021825
Device NameMODIFICATION TO SPIRAL RADIUS 90-D SST SYSTEM
ApplicantUnitedstates Surgical
Product CodeKWP · Orthopedic
Decision DateJul 1, 2002
DecisionSESE
Submission TypeSpecial
Regulation21 CFR 888.3050
Device ClassClass 2
AttributesTherapeutic

Indications for Use

The components of the Spiral Radius 90-DTM SST System are indicated for spinal fixation. When used as an anterolateral/anterior system consisting of rods and screws, the levels of attachment are the lumbar, thoracic and cervical spine. The points of attachment and methods are screw fixation to the anterolateral vertebral bodies of the lumbar and thoracic spine (T6-L5) and the anterior vertebral bodies of the cervical spine. The Indications are degenerative disc disease defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies, spondylolisthesis, fracture, spinal stenosis, deformities (i.e., scoliosis, kyphosis, lordosis), tumor, pseudoarthrosis, or failed previous fusion (pseudoarthrosis). When used as a nonpedicle posterior system consisting of hooks, crosslinks and sacral/iliac screws the levels of attachments are the lumbar, thoracic and cervical spine and sacrum and ilium. Intended uses include hook and sacral screw fixation to the lumbar spine, noncervical spine and to the T1-S1 spine; and hook and sacral/iliac screw fixation to the noncervical spine. The indications are degenerative disc disease defined as back pain of discogenic origin with degenerative of the disc confirmed by history and radiographic studies, spondylolisthesis, fracture, spinal stenosis, deformities (i.e., scoliosis, kyphosis, lordosis), tumors, pseudoarthrosis, or failed previous fusion (pseudoarthrosis). When used as a pedicle screw system, in the non-cervical spine of skeletally mature patients. The Spiral Radius 90-D™ SST System is indicated for patients: (a) having severe spondylolisthesis (Grades 3 and 4) at the L5-S1 joint: (b) who are receiving fusions using autogenous bone graft only: (c) who are having the device fixed or attached to the lumbar and sacral spine (L3 and (d) who are having the device removed after the development of solid fusion mass. In addition, when used as a pedicle screw system, in the non-cervical spine of skeletally mature patients, the Spiral Radius 90-D™ SST System is intended to provide immobilization and stabilization of spinal segments, as an adjunct to fusion, in the treatment of the following acute and chronic instabilities of the thoracic, lumbar, and sacral spine: degenerative spondylolisthesis with objective evidence of neurologic impairment, fracture, dislocation, scoliosis, spinal tumor, and failed previous fusion (pseudoarthrosis).

Device Story

Spiral Radius 90-D SST System is a spinal fixation construct; consists of stainless steel rods, hooks, screws, and cross-bars. Used by surgeons in clinical/OR settings to provide immobilization and stabilization of spinal segments as an adjunct to fusion. Components attached to vertebral bodies or pedicles to treat instabilities; accommodate varying patient morphology via multiple sizes. Provides rigid construct to support bone healing; benefits patient by stabilizing spine during fusion process.

Clinical Evidence

No clinical data provided; bench testing only.

Technological Characteristics

Construct consists of stainless steel rods, hooks, screws, and cross-bars. Mechanical fixation system for spinal stabilization. No software or electronic components.

Indications for Use

Indicated for spinal fixation in patients with degenerative disc disease, spondylolisthesis, fracture, spinal stenosis, deformities (scoliosis, kyphosis, lordosis), tumor, or pseudoarthrosis. Pedicle screw use restricted to skeletally mature patients with severe spondylolisthesis (Grades 3-4) at L5-S1 or acute/chronic instabilities of thoracic, lumbar, and sacral spine.

Regulatory Classification

Identification

A spinal interlaminal fixation orthosis is a device intended to be implanted made of an alloy, such as stainless steel, that consists of various hooks and a posteriorly placed compression or distraction rod. The device is implanted, usually across three adjacent vertebrae, to straighten and immobilize the spine to allow bone grafts to unite and fuse the vertebrae together. The device is used primarily in the treatment of scoliosis (a lateral curvature of the spine), but it also may be used in the treatment of fracture or dislocation of the spine, grades 3 and 4 of spondylolisthesis (a dislocation of the spinal column), and lower back syndrome.

Submission Summary (Full Text)

{0}------------------------------------------------ # Attachment 4 1 2002 JUL - 510(k) Summary Q21825 ### Surgical Dynamics Spiral Radius 90-DTM SST System page 1 of 2 United States Surgical 150 Glover Avenue Norwalk, CT 06856 USA ### DEVICE DESCRIPTION The Spiral Radius 90-DJM SST System consists of stainless steel rods attached to the spinal column through the use of hooks. Screws may be used to attach rods to the spinal column in certain select patients and certain select levels (See Indications). Cross-bars may be used to connect rods to provide a more rigid construct, as well as screws to rods, and hooks to rods. Screws and hooks are provided in several sizes to accommodate varying patient morphology. ## INDICATIONS FOR USE The components of the Spiral Radius 90-D™ SST System are indicated for spinal fixation. When used as an anterolateral/anterior system consisting of rods and screws, the levels of attachment are the lumbar, thoracic and cervical spine. The points of attachment and methods are screw fixation to the anterolateral vertebral bodies of the lumbar and thoracic spine (T6-L5) and the anterior vertebral bodies of the cervical spine. The Indications are degenerative disc disease defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies, spondylolisthesis, fracture, spinal stenosis, deformities (i.e., scoliosis, kyphosis, lordosis), tumor, pseudoarthrosis, or failed previous fusion (pseudoarthrosis). When used as a nonpedicle posterior system consisting of hooks, crosslinks and sacral/iliac screws the levels of attachments are the lumbar, thoracic and cervical spine and sacrum and ilium. Intended uses include hook and sacral screw fixation to the lumbar spine, noncervical spine and to the T1-S1 spine; and hook and sacral/iliac screw fixation to the noncervical spine. The indications are degenerative disc disease defined as back pain of discogenic origin with degenerative of the disc confirmed by history and radiographic studies, spondylolisthesis, fracture, spinal stenosis, deformities (i.e., scoliosis, kyphosis, lordosis), tumors, pseudoarthrosis, or failed previous fusion (pseudoarthrosis). When used as a pedicle screw system, in the non-cervical spine of skeletally mature patients, The Spiral Radius 90-D™ SST System is indicated for patients: (a) having severe spondylolisthesis (Grades 3 and 4) at the L5-S1 joint; (b) who are receiving fusions using autogenous bone graft only; (c) who are having the device fixed or attached to the lumbar and sacral spine (L3 and (d) who are having the device removed after the development of solid fusion mass. In addition, when used as a pedicle screw system, in the non-cervical spine of skeletally mature patients, the Spiral Radius 90-DTM SST System is intended to provide immobilization and stabilization of spinal segments, as an adjunct to fusion, in the treatment of the following acute and chronic instabilities of the thoracic, lumbar, and sacral spine: degenerative spondylolisthesis with objective evidence of neurologic impairment, fracture, dislocation, scoliosis, spinal tumor, and failed previous fusion (pseudoarthrosis). {1}------------------------------------------------ ## SUBSTANTIAL EQUIVALENCE* K021825 page 2 of 2 The modified components of the Surgical Dynamics Spiral Radius 90-DTM vstem was claimed to be substantially equivalent* to the currently marketed version of the device. Information pertaining to this device was provided in the submission. ... * Any claim of substantial equivalence is made exclusively in regard to the U.S. Food, Drug and Cosmetic Act and should not be viewed in any other light. {2}------------------------------------------------ Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular border with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" arranged around the top half of the circle. Inside the circle is a stylized image of a bird or dove, composed of three curved lines that also suggest a human profile. Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 #### 1 2002 JUL Ms. Tina Rideout Regulatory Affairs Senior Associate United States Surgical Surgical Dynamics 150 Glover Avenue Norwalk, Connecticut 06856 Re: K021825 Trade/Device Name: Spiral Radius 90-D™ SST System Regulation Number: 21 CFR 888.3050, 888.3060 and 888.3070 Regulation Name: Spinal interlaminal fixation orthosis, Spinal intervertebral body fixation orthosis, Pedicle screw system, Spondylolisthesis spinal fixation device system Regulatory Class: II Product Code: KWP. KWO. MNH, MNI Dated: June 3, 2002 Received: June 4, 2002 Dear Ms. Rideout: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing {3}------------------------------------------------ . Page 2 – Ms. Tina Rideout (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 21 CFR Part 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594- . Additionally, for questions on the promotion and advertising of vour device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html Sincerely vours. Stephen Rhodes Celia M. Witten, Ph.D., M.D. Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {4}------------------------------------------------ ## Attachment 2 # Indications for Use Statement 510(k) Number ## Device Name: Spiral Radius 90-DTM SST System ### Indications For Use The components of the Spiral Radius 90-DTM SST System are indicated for spinal fixation. When used as an anterolateral/anterior system consisting of rods and screws, the levels of attachment are the lumbar, thoracic and cervical spine. The points of attachment and methods are screw fixation to the anterolateral vertebral bodies of the lumbar and thoracic spine (T6-L5) and the anterior vertebral bodies of the cervical spine. The Indications are degenerative disc disease defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies, spondylolisthesis, fracture, spinal stenosis, deformities (i.e., scoliosis, kyphosis, lordosis), tumor, pseudoarthrosis, or failed previous fusion (pseudoarthrosis). When used as a nonpedicle posterior system consisting of hooks, crosslinks and sacral/iliac screws the levels of attachments are the lumbar, thoracic and cervical spine and sacrum and ilium. Intended uses include hook and sacral screw fixation to the lumbar spine, noncervical spine and to the T1-S1 spine; and hook and sacral/iliac screw fixation to the noncervical spine. The indications are degenerative disc disease defined as back pain of discogenic origin with degenerative of the disc confirmed by history and radiographic studies, spondylolisthesis, fracture, spinal stenosis, deformities (i.e., scoliosis, kyphosis, lordosis), tumors, pseudoarthrosis, or failed previous fusion (pseudoarthrosis). When used as a pedicle screw system, in the non-cervical spine of skeletally mature patients. The Spiral Radius 90-D™ SST System is indicated for patients: (a) having severe spondylolisthesis (Grades 3 and 4) at the L5-S1 joint: (b) who are receiving fusions using autogenous bone graft only: (c) who are having the device fixed or attached to the lumbar and sacral spine (L3 and (d) who are having the device removed after the development of solid fusion mass. In addition, when used as a pedicle screw system, in the non-cervical spine of skeletally mature patients, the Spiral Radius 90-D™ SST System is intended to provide immobilization and stabilization of spinal segments, as an adjunct to fusion, in the treatment of the following acute and chronic instabilities of the thoracic, lumbar, and sacral spine: degenerative spondylolisthesis with objective evidence of neurologic impairment, fracture, dislocation, scoliosis, spinal tumor, and failed previous fusion (pseudoarthrosis). (Please do not write below this line - continue on another page if needed) | Concurrence of CDRH, Office of Device Evaluation (ODE) | |--------------------------------------------------------| |--------------------------------------------------------| | Prescription Use:<br>(Per 21 CFR 801.109) | X | OR Over-The-Counter Use: | |-------------------------------------------|---|--------------------------| |-------------------------------------------|---|--------------------------| | Special 510 (k): Premarket Notification<br>June 3, 2002 | | Division of General, Restorative<br>Page 1.8<br>and Neurological Devices | |---------------------------------------------------------|--|--------------------------------------------------------------------------| |---------------------------------------------------------|--|--------------------------------------------------------------------------| | 510(k) Number | K021875 | |---------------|---------| |---------------|---------|
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...