SMITH & NEPHEW IMAGE SURGICAL INSTRUMENTS FOR TRAUMA APPLICATIONS
K021815 · Smith & Nephew, Inc. · HAW · Oct 17, 2002 · Neurology
Device Facts
Record ID
K021815
Device Name
SMITH & NEPHEW IMAGE SURGICAL INSTRUMENTS FOR TRAUMA APPLICATIONS
Applicant
Smith & Nephew, Inc.
Product Code
HAW · Neurology
Decision Date
Oct 17, 2002
Decision
SESE
Submission Type
Abbreviated
Regulation
21 CFR 882.4560
Device Class
Class 2
Indications for Use
Smith & Nephew Image Guided Surgical Instruments for Trauma Applications are intended to be used with the Universal Accessories to assist the surgeon in precisely locating anatomical structures in either open or percutaneous procedures. Image Guided Surgical Instruments for Trauma Applications are indicated for use in surgical trauma procedures, in which the use of stereotactic surgery may be appropriate, and where a reference to a rigid anatomical structure such as a long bone (femur, tibia, humerus, radius, ulna, fibula), pelvic bone (including acetabular), calcaneus, and talus bone can be identified relative to a CT or MR based model or fluoroscopy images of the anatomy.
Device Story
Modified surgical instruments equipped with arrays containing infrared LEDs or passive spheres; arrays interface with external IGS Platform System for real-time tracking. System uses reference frames rigidly attached to patient anatomy to track position; software compensates for movement of anatomy or IGS platform to maintain navigation accuracy. Used in OR by surgeons for open or percutaneous trauma procedures; assists in precise localization of anatomical structures relative to CT, MR, or fluoroscopic images. Benefits include improved surgical precision and navigation during complex orthopedic and trauma interventions.
Clinical Evidence
No clinical data provided. Substantial equivalence is based on technological characteristics and intended use similarities to predicate devices.
Technological Characteristics
Stereotactic surgical instruments modified to mount infrared LED arrays or passive spheres. Sensing principle: optical tracking via infrared emission/reflection. System is networked with an IGS Platform. Components include sterile/non-sterile single-use passive spheres. Class II device.
Indications for Use
Indicated for patients undergoing surgical trauma procedures where stereotactic navigation is appropriate, specifically involving rigid anatomical structures including long bones (femur, tibia, humerus, radius, ulna, fibula), pelvic/acetabular bones, calcaneus, and talus. Used for procedures like fracture reduction, intramedullary nailing, plating, screw placement, and high tibia osteotomy.
Regulatory Classification
Identification
A stereotaxic instrument is a device consisting of a rigid frame with a calibrated guide mechanism for precisely positioning probes or other devices within a patient's brain, spinal cord, or other part of the nervous system.
Image Guided Surgical Instruments for Knees (K020415)
Knee Module For The StealthStation System (K020256)
StealthStation System Passive Instrument Option (K973545)
Submission Summary (Full Text)
{0}------------------------------------------------
Kozlfls
# 510(k) Summary Image Guided Surgical Instruments For Trauma Applications and Universal Accessories
# OCT 1 7 2002
| Submitter's name: | Smith & Nephew, Inc., Orthopaedic Division |
|-----------------------------------|----------------------------------------------------------------------------------------------------------|
| Submitter's address: | 1450 Brooks Road, Memphis, TN 38116 |
| Submitter's telephone number: | 901-399-6707 |
| Contact person: | Gino J. Rouss |
| Date summary prepared: | May 30, 2002 |
| Trade or proprietary device name: | Smith & Nephew Image Guided Surgical<br>Instruments for Trauma Applications and<br>Universal Accessories |
| Common or usual name: | Stereotaxic Instrument |
| Classification name: | Stereotaxic Instrument |
| Device Class: | Class II |
| Device Product Code: | HAW |
| Panel Code: | Neurology/84 |
# Subject device description:
The Smith & Nephew Image Guided Instruments for Trauma Applications are instruments that have been modified to allow image-guided arrays (Fighters) to be fixed onto the instruments. The image-guided arrays can use either infrared LEDs (light emitting diodes) or universal accessories (passive spheres) to transmit or reflect the infrared LEDs (light emitting diodes) that are emitted by an IGS Platform System. Along with commercially available software, this will allow the instruments to be recognized and tracked in real time in the surgical field. The universal accessories (passive spheres) will be available as either sterile or non-sterile and are for single use only. Each package will contain a multiple number of spheres. The infrared LEDs (light emitting diodes) or universal accessories (passive spheres) do not come in contact with the open wound during surgical procedures. The image-guided arrays that are affixed to the instruments work in conjunction with reference frames that are rigidly attached to the anatomy. Each reference frame is also fitted with either infrared LEDs (light emitting diodes) or universal accessories (passive spheres) to transmit or reflect the infrared LEDs (light emitting diodes) that are emitted by the IGS Platform System. The reference frames will allow the IGS Platform System to continuously track the position of the anatomy during navigation. If any movement of the IGS Platform System or anatomy is detected, the system can compensate for it, thereby maintaining accurate navigation.
{1}------------------------------------------------
K021f15
# A. Applicable 510(k)'s
| Image Guided Instruments for Trauma Applications | | | |
|--------------------------------------------------|---------------------------------------------------------|-------------|--------------------------|
| Manufacturer | Submission Name | Exhibit No. | FDA<br>Clearance<br>Date |
| Surgical Navigation<br>Technologies | StealthStation | 10 | 1-24-96 |
| Surgical Navigation<br>Technologies | StealthStation™ System - ENT Application Addendum | 11 | 1-21-98 |
| Surgical Navigation<br>Technologies | StealthStation® FluoroNav™ Module | 12 | 4-22-99 |
| Surgical Navigation<br>Technologies | Indications Modifications for the StealthStation System | 13 | 2-22-00 |
| Surgical Navigation<br>Technologies | StealthStation Generation 3 | 14 | 5-3-00 |
| BrainLAB AG | VectorVision® Trauma | 15 | 3-14-02 |
| BrainLAB AG | VectorVision2 | 16 | 5-19-99 |
| Smith & Nephew, Inc. | Image Guided Surgical Instruments for Knees | 17 | 2-8-02 |
| Surgical Navigation<br>Technologies | Knee Module For The StealthStation™ System | 18 | 1-25-02 |
| Surgical Navigation<br>Technologies | StealthStation™ System Passive Instrument Option | 19 | 9-16-97 |
# Subject device intended use:
Image Guided Surgical Instruments for Trauma Applications are intended to be used with the Universal Accessories to assist the surgeon in precisely locating anatomical structures in either open or percutaneous procedures. Image Guided Surgical Instruments for Trauma Applications are indicated for use in surgical trauma procedures, in which the use of stereotactic surgery may be appropriate, and where a reference to a rigid anatomical structure such as a long bone (femur, tibia, humerus, radius, ulna, fibula), pelvic bone (including acetabular), calcaneus, and talus bone can be identified relative to a CT or MR based model or fluoroscopy images of the anatomy.
# Example procedures include but are not limited to:
Spinal procedures and spinal implant procedures such as pedicle screw placement Pelvis and acetabular fracture treatment such as screw placement or ilio-sacral screw fixation Fracture treatment procedures, such as intramedullary nailing or plating or screwing or external fixation procedures in the tubular bones Fracture reduction procedures High Tibia Osteotomy
{2}------------------------------------------------
Kozifis
# Technological Characteristics:
Image Guided Surgical Instruments for Trauma Applications are similar to currently legally marketed Class II stereotactic instruments in that they incorporate infrared LED (light emitting diodes) or Universal Accessories (passive spheres) onto the instruments that allow the instruments to be recognized and tracked in real time in the surgical field.
{3}------------------------------------------------
# DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" around the perimeter. Inside the circle is a stylized image of three human profiles facing to the right, with flowing lines representing hair or clothing.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
# OCT 1 7 2002
Mr. Gino J. Rouss Clinical/Regulatory Affairs Specialist Smith & Nephew, Inc. Orthopaedic Division 1450 Brooks Road Memphis, Tennessee 38116
Re: K021815
Trade/Device Name: Smith & Nephew Image Surgical Instruments for Trauma Applications Regulation Number: 882.4560 Regulation Name: Stereotaxic instrument Regulatory Class: II Product Code: HAW Dated: July 18, 2002 Received: July 22, 2002
Dear Mr. Rouss:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{4}------------------------------------------------
Page 2 -- Mr. Gino J. Rouss
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 21 CFR Part 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4659. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html
Sincerely vours.
uriam C. Provost
for Celia M. Witten, Ph.D., M.D. Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
### 510(k) Number (if known): K021815
### Image Guided Surgical Instruments for Trauma Applications Device Name: and Universal Accessories
## Indications For Use:
Smith & Nephew Image Guided Surgical Instruments for Trauma Applications are intended to be used with the Universal Accessories to assist the surgeon in precisely locating anatomical structures in either open or percutaneous procedures. Image Guided Surgical Instruments for Trauma Applications are indicated for use in surgical trauma procedures, in which the use of stereotactic surgery may be appropriate, and where a reference to a rigid anatomical structure such as a long bone (femur, tibia, humerus, radius, ulna, fibula), pelvic bone (including acetabular), calcaneus, and talus bone can be identified relative to a CT or MR based model or fluoroscopy images of the anatomy.
## Example procedures include, but are not limited to:
Spinal procedures and spinal implant procedures such as pedicle screw placement Pelvis and acetabular fracture treatment such as screw placement or ilio-sacral screw fixation Fracture treatment procedures, such as intramedullary nailing or screwing or external fixation procedures in the tubular bones
## Additional procedures include:
Fracture reduction procedures
High Tibia Osteotomy
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Prescription Use
(Per 21 CFR 801.109)
OR
Over-The-Counter Use
(Optional Format 1-2-96)
Muriam C. Provost
eral, Restorative and Neurological Dev
510(k) Number K621815
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.