K021782 · Aap Implantate AG · JDS · Aug 26, 2002 · Orthopedic
Device Facts
Record ID
K021782
Device Name
BNF
Applicant
Aap Implantate AG
Product Code
JDS · Orthopedic
Decision Date
Aug 26, 2002
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 888.3030
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
aap Biorigid Nail Femur (BNF) is intended to use in fractures of the femoral bone. The indications are: Types of Fractures: Simple closed fractures Comminuted fractures Open fractures of first, second or third degree Crush fractures Antegrade Indications: Femur shaft fractures Distal femur shaft fractures Pathological fractures, e.g. tumor osteolysis Periprosthetic femur fracture in the middle to third part with knee replacement in situ Retrograde Indications: Supracondylary fractures Simultaneous nailing of femur and tibia Distal femur fracture with osteosynthesis or hip prosthesis in situ Periprosthetic femur fracture at knee arthroplasty Borderline Indications: Percondylar femur fracture Pseudarthrosis (only reamed method) Osteoporosis
Device Story
Biorigid Nail Femur (BNF) is an intramedullary fixation rod system for stabilizing femoral fractures or correcting deformities. Device consists of a full-material titanium alloy nail, interlocking screws (5.3 mm diameter), interlocking nuts, protection caps, and CondyLock components. Nail features central holes to resist rotational forces and grooves for interlocking along the nail length, allowing for both antegrade and retrograde implantation techniques. Used by orthopedic surgeons in clinical settings for reamed or unreamed intramedullary nailing. Provides mechanical stabilization of bone segments to facilitate healing. Distributed non-sterile; requires decontamination and steam sterilization by the end-user prior to surgery.
Clinical Evidence
No clinical data provided. Substantial equivalence is based on bench testing and comparison of technological characteristics to predicate devices.
Technological Characteristics
Material: Titanium Alloy (Ti 6Al 4V E.L.I., ASTM F136). Design: Full-material intramedullary nail with central holes and grooves for interlocking. Energy source: None (mechanical). Connectivity: None. Sterilization: Non-sterile; requires user-performed steam sterilization per AAMI/ISO 11737 guidelines.
Indications for Use
Indicated for patients with femoral bone fractures, including simple closed, comminuted, open (1st-3rd degree), and crush fractures. Specific applications include antegrade or retrograde nailing for femur shaft, distal femur, supracondylar, periprosthetic, and pathological fractures (e.g., tumor osteolysis), as well as pseudarthrosis and osteoporosis.
Regulatory Classification
Identification
Single/multiple component metallic bone fixation appliances and accessories are devices intended to be implanted consisting of one or more metallic components and their metallic fasteners. The devices contain a plate, a nail/plate combination, or a blade/plate combination that are made of alloys, such as cobalt-chromium-molybdenum, stainless steel, and titanium, that are intended to be held in position with fasteners, such as screws and nails, or bolts, nuts, and washers. These devices are used for fixation of fractures of the proximal or distal end of long bones, such as intracapsular, intertrochanteric, intercervical, supracondylar, or condylar fractures of the femur; for fusion of a joint; or for surgical procedures that involve cutting a bone. The devices may be implanted or attached through the skin so that a pulling force (traction) may be applied to the skeletal system.
{0}------------------------------------------------
## AUG 2 6 2002 510(k) SUMMARY OF SAFETY AND EFFECTIVENESS
Submitters Name:
aap Implantate AG Lorenzweg 5 12099 Berlin Germany Ph.: +49 30 750 19 0 Fax: +49 30 750 19 111
Biorigid Nail Femur (BNF)
Intramedullary Fixation Rod
Biorigid Femoral Nail System
aap Biorigid Nail Femur (BNF)
cd1782
page 1 of 2
Contact Name: Name of Device: Classification Name:
Common/Usual Name:
Proprietary Name:
Classification:
전 - - - ---------------------------------------------------------------------------------------------------------------------------------------------------------------------
CFR Chapter I, Title 21 § 888.3020,
# 87 J DS #, # 87 H SB #, # 87 H TN #
Class II, Intramedullary Fixation Rod,
Dipl .- Ing. Christian Abel, Director Quality Management
Performance Standards:
Devices are manufactured according to cGMP's, applicable ASTM requirements, and applicable harmonised standards ISO 9001 / EN 46001.
Material Composition:
aap Biorigid Nail Femur components are manufactured of Titanium Alloy (Ti 6Al 4V E.L.I. = ASTM F136)
Intended Use:
aap Biorigid Nail Femur (BNF) is intended to use in fractures of the femoral bone.
The indications are:
| • Types of Fractures: | Simple closed fractures<br>Comminuted fractures<br>Open fractures of first, second or third degree<br>Crush fractures |
|---------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| • Antegrade Indications: | Femur shaft fractures<br>Distal femur shaft fractures<br>Pathological fractures, e.g. tumor osteolysis<br>Periprosthetic femur fracture in the middle to third part<br>with knee replacement in situ |
| • Retrograde Indications: | Supracondylary fractures<br>Simultaneous nailing of femur and tibia<br>Distal femur fracture with osteosynthesis or hip<br>prosthesis in situ<br>Periprosthetic femur fracture at knee arthroplasty |
| • Borderline Indications: | Percondylar femur fracture<br>Pseudarthrosis (only reamed method)<br>Osteoporosis |
{1}------------------------------------------------
Device Description:
용품 | ------------------------------------------------------------------------------------------------------------------------------------------------------------------------
Jofa Kod1782
Intramedullary rods are used for the stabilisation of fractures or correction of deformity of ong bones with or without interlocking screws. The connection between the nail and the interlocking screw is realised with central holes or grooves. Reamed and unreamed intramedullary nailing and for the femoral bone the antegrade and retrograde entrance are standard methods today.
agp Bioriqid Nail Femur (BNF) is the consistent development of the Biorigid Nail Tibia for the femoral bone. The BNF is a full material nail. The interlocking of the nail happens with solid Ø 5.3 mm screws through central holes and grooves.
The central holes on the nail protect against rotational forces. The grooves are designed for interlocking possibility over the whole length of the nail and a defined reduction of the rigidity because of the full material nail.
The BNF permits the execution of different implantation techniques and interlocking with only one nail. Therefore it meets the different injury manners and necessity for individual solutions.
aap Biorigid Nail Femur (BNF) incorporates
- Nail in different diameters .
- Interlocking screws in different lenghts ●
- Interlocking nut ●
- Protection cap in different lengths ●
- CondyLock ●
- Accessories, like targeting devices etc. ●
Predicate Devices for Substantial Equivalence:
aap Implantate AG 's Biorigid Nail Femur (BNF) is similar in size, material and intended use to the Howmedica Osteonics Corp. T2 Femoral Nail System (K010801 / K011622 / K014220), Synthes Ti Distal Femoral Nail (K970733), Synthes Proximal Femoral Nail (K970097), Synthes Titanium Unreamed Femoral Nail (K923597), Smith & Nephew Titanium Nail System (K981529), Smith & Nephew Intramedullary Nail System (K983942), DePuy ACE AIM Titanium Supracondylar Nail (K974781), Biomet Titanium Intramedullary Nails (K982953), Biomet Titanium Retrograde Femoral Nail (K013923)
Comparision of Technological Characteristics: aap BNF is substantially equivalent to the predicate devices with respect to physical/technical and material characteristics.
Sterilisation Information:
The devices are distributed in non sterile condition.
The instruments and implants provided in non-sterile condition must be decontaminated. cleaned and sterilised prior to each surgery. All packaqing materials must be removed.
Recommendations for sterilisation are contained in the package insert.
Note: These devices are sterilised by end users utilizing the approved/outlined quidelines found in the AAMI Guideline "Good Hospital Practice: Steam Sterilisation and Sterility Assurance" and in ANSI/AAMI/ISO 11737 guidelines to achieve the acceptable Sterility Assurance Level (SAL).
{2}------------------------------------------------
## DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo is circular and contains the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized symbol that resembles three abstract human profiles or faces stacked on top of each other.
## Public Health Service
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
**AUG 26 2002**
AAP Implantate AG Dipl.-Ing. Christian Abel Director Quality Management Lorenzweg 5 12099 Berlin Germany
Re: K021782
Trade/Device Name: aap Biorigid Nail Femur (BNF) Regulation Number: 888.3020 Regulation Name: Intramedullary Fixation Rod Regulatory Class: Class II Product Code: JDS Dated: May 14, 2002 Received: May 30, 2002
Dear Mr. Abel:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{3}------------------------------------------------
Page 2 - Mr. Christian Abel
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 21 CFR Part 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4659. Additionally, for questions on the promotion and advertising of vour device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html
Sincerely yours.
Celia M. Witten, Ph.D., M.D. Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
Page 1 of 1
510(k) Number (if known):_K021782
Device Name: BNF
Indications For Use:
Types of Fractures: Simple Closed Fractures Comminuted Fractures Open Fractures of First, Second or Third Degree Crush Fractures
Antegrade Indications: Femur Shaft Fractures Distal Femur Shaft Fractures Pathological Fractures (e.g. Tumor Osteolysis) Periprosthetic Femur Fracture in the Middle of Third Part with Knee Replacement in Situ
Retrograde Indications: Supracondylary Fractures Simultaneous Nailing of Femur and Tibia Distal Femur Fracture with Osteosynthesis or Hip Prosthesis in Situ Periprosthetic Femur Fracture at Knee Arthosplasty
Borderline Indications: Percondylar Femur Fracture, (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
| (Division Sign-Off) | |
|-----------------------------------------------------------|---------|
| Division of General, Restorative and Neurological Devices | |
| 510(k) Number | K021782 |
| Prescription Use | OR | Over-The-Counter Use |
|------------------|----|----------------------|
|------------------|----|----------------------|
(Per 21 CFR 801.109)
Over-The-Counter Use*_*_
(Optional Formal 1-2-1
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.