K021763 · Rs Medical · IPF · Jun 13, 2002 · Physical Medicine
Device Facts
Record ID
K021763
Device Name
RS2MI MUSCLE STIMULATOR FAMILY
Applicant
Rs Medical
Product Code
IPF · Physical Medicine
Decision Date
Jun 13, 2002
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 890.5850
Device Class
Class 2
Attributes
Therapeutic, 3rd-Party Reviewed
Indications for Use
Muscle Stimulation, Interferential and Non-Interferential Modality - Relax muscle spasms . - Prevent or retard disuse atrophy . - Maintain or increase range of motion ● - Increase local blood circulation . - Re-educate muscle . - Immediate post-surgical stimulation of calf muscles to prevent venous thrombosis . Pain Relief, Interferential Current Modality - Relieve acute pain . - Relieve and manage chronic pain .
Device Story
RS2mi family consists of four configurable muscle/interferential stimulators (RS2m, RS2i, RS2c, RS2s). Device delivers electrical stimulation via output cables and electrode pads to patient. Pulse mode: 57.5V peak, 115mA peak, 415uSec pulse width, 71Hz frequency, bi-phasic pulses with on/off ramp. Interferential mode: 100mA peak, symmetric balanced sine waves, zero net charge; operates in 4-pad (true) or 2-pad (pre-mixed) modes with fixed or variable frequency. Device housed in plastic enclosure with LED indicators. Operated by clinician or patient under prescription. Output intensity controlled by pulse width. Benefits include pain relief, muscle re-education, and prevention of atrophy/thrombosis.
Clinical Evidence
Bench testing only. No clinical data provided.
Technological Characteristics
Plastic enclosure; LED patient indicators; output cables; electrode pads; AC charging adapter. Pulse mode: 57.5V peak, 115mA peak, 415uSec pulse width, 71Hz frequency. Interferential mode: 100mA peak, symmetric balanced sine wave, zero net charge. 2-channel output. Software-configured modality emulation.
Indications for Use
Indicated for patients requiring muscle stimulation to relax spasms, prevent disuse atrophy, maintain/increase range of motion, increase local blood circulation, re-educate muscle, or provide post-surgical calf stimulation to prevent venous thrombosis. Also indicated for relief and management of acute and chronic pain via interferential current.
Regulatory Classification
Identification
A powered muscle stimulator is an electrically powered device intended for medical purposes that repeatedly contracts muscles by passing electrical currents through electrodes contacting the affected body area.
{0}------------------------------------------------
Premarket Notification RS Medical RS2mi (Rev 2)
## 510(k) SUMMARY RS2mi family May 15, 2002
This summary is provided in accordance with the Safe Medical Devices Act of 1990 (SMDA). The information provided in the 510(k), premarket notification was in accordance with 21 CFR 807.87.
## 1. Applicant, Official Correspondent and Owner of 510(k)
RS Medical, Inc. 14401 SE First St. Vancouver, WA 98684-6981
Attn: Mike McGraw, Vice President, Product Development Telephone: (360) 896-2503 Fax: (306) 896-2566
## Submitter of 510(k) and Third Party Reviewer
Underwriters Laboratories Inc. 2600 NW Lake Road Camas, WA 98607-8542 Telephone: (360) 817-5515 Fax: (360) 817-6133
## 2. Name of Device
Trade/Proprietary Name: RS2mi Muscle Stimulator family
Common/Usual Name: Muscle and Interferential Current Stimulator
Classification Name: 21 CFR 890.5850 "Powered Muscle Stimulator", Class II.
## 3. Legally Market Predicate Devices
The RS2mi family is substantially equivalent to its legally marketed predecessor the RS-4M Plus (K000114) muscle stimulator.
{1}------------------------------------------------
Premarket Notification RS Medical RS2mi (Rev 2)
## 4. Indications for Use
## Muscle Stimulation, Interferential and Non-Interferential Modality
- Relax muscle spasms .
- Prevent or retard disuse atrophy .
- Maintain or increase range of motion ●
- Increase local blood circulation .
- Re-educate muscle .
- Immediate post-surgical stimulation of calf muscles to prevent venous thrombosis .
## Pain Relief, Interferential Current Modality
- Relieve acute pain .
- Relieve and manage chronic pain .
## 5. Device Description and Substantial Equivalence
The RS2mi device family consists of a hardware/software system that can be setup as one of four devices: an RS2m which is a pulse mode muscle stimulator only, an RS2i which is an interferential stimulator only, an RS2c which can act as either a pulse mode muscle stimulator or an interferential stimulator, or an RS2s which acts as a sequential stimulator by first providing an interferential treatment segment followed by a pulse mode muscle stimulation segment. The software system is setup to emulate one of these four devices and the device faceplate is setup to indicate to the user the features of that device only. To provide flexibility, the faceplates can be interchanged by RS Medical to create the type of device needed in the marketplace. Throughout this document the name RS2mi is used to indicate this 'family' of four separate devices.
The RS4M+ incorporates all of these features in one unit, the primary difference is that the predicate device has 4 channels, while the RS2mi has only 2 channels
The RS2mi device family, like a number of legally marketed predicate devices, incorporates traditional muscle stimulation and interferential current stimulation modalities into one unit. Only one modality may be operated at a time. The RS2mi is housed in a plastic enclosure. The front of the enclosure houses LED patient indicators. The accessories provided with the RS2mi include the output cables, the electrode pads, and the AC Charging Adapter.
Both the RS4M+ and the RS2mi use the same materials and texture for the enclosure. Only the enclosure color (TBD) is different. The RS4M+ uses a 2-line LCD Patient Indicator Display. The RS2mi uses LEDs. The RS2mi is designed to use the same output cables, electrode pads, and the AC Charging Adapter as the predicate device
{2}------------------------------------------------
The RS2mi pulse mode muscle stimulation modality operates at a specified 57.5 volts peak (+/-10% into a 500 ohm load) and 115 mA peak (+/-10% into a 500 ohm load) with a maximum pulse width of 415 uSec.( +/-10%) and a cycle frequency of 71 Hz(+/-5%). The pulses are bi-phasic. Intensity levels are controlled via pulse width while maintaining the pulse voltage within the specified peak. The waveform includes an on/off ramp, which slowly increases the pulse width to the desired setting.
The RS4M+ is the same.
The RS2mi interferential modality operates at a specified maximum of 100 mA peak (+/-10% into a 500 ohm load). The carrier and interferential signals are simulated sine wave symmetric, balanced outputs with zero net charge. The interferential modality can operate in a true interferential mode (4 pad mode) or the signals can be pre-mixed and only the pre-mixed signals sent to the patient (2 pad mode). The interference signal frequency can be fixed (Continuous) or varied based on three selections (Variable).
The RS4M+ is the same.
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image is a circular seal for the Department of Health & Human Services - USA. The seal features the department's name in a circular arrangement around the perimeter. In the center of the seal is a stylized eagle emblem, composed of three profiles facing to the right.
Public Health Service
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
JUN 1 3 2002
Mrs. Molly Bettger RS Medical C/O: Underwriters Laboratories, Inc. 2600 NW Lake Road Camas, WA 98607-8542
Re: K021763
Trade/Device Name: RS2mi muscle stimulator family, models: RS2m, RS2i, RS2c, and RS2s Regulation Number: 21 CFR 890.5850 Regulation Name: Powered muscle stimulator Regulatory Class: Class II Product Code: 89 IPF and 84 LIH Dated: May 20, 2002 · Received: May 29, 2002
Dear Mrs. Bettger:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set
{4}------------------------------------------------
#### Page 2 - Mrs. Bettger
forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 21 CFR Part 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html
Sincerely yours,
Stept
Celia M. Witten, Ph.D., M.D.
Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
Premarket Notification RS Medical RS-2mi
K021763
# Premarket Notification
# Indications For Use
Device Name: RS2mi Muscle Stimulator family
#### Indications For Use:
## Muscle Stimulation, Interferential and Non-Interferential Modality
- . Relax muscle spasms
- Prevent or retard disuse atrophy .
- Maintain or increase range of motion .
- Increase local blood circulation .
- Re-educate muscle .
- Immediate post-surgical stimulation of calf muscles to prevent venous � thrombosis
## Pain Relief, Interferential Current Modality
- Relieve acute pain .
- Relieve and manage chronic pain .
## (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Prescription Use: (Per 21 CFR 801.109)
(Optional Format 1-2-96)
OR
Over-The-Counter:
Htyk Purdy
(Division Sign-Off) Division of General, Restorative and Neurological Devices
1021763 510(k) Number __
- 8 -
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.