K021748 · Endius, Inc. · HRX · Jun 26, 2002 · Orthopedic
Device Facts
Record ID
K021748
Device Name
ENDIUS ATAVI SYSTEM
Applicant
Endius, Inc.
Product Code
HRX · Orthopedic
Decision Date
Jun 26, 2002
Decision
SESE
Submission Type
Special
Regulation
21 CFR 888.1100
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Endius Atavi System is indicated for use in instrumented posterolateral fusion procedures where the TriFix Spinal Instrumentation System is utilized. The Endius Atavi System is also intended to be used for posterior endoscopic access to the lumbar spine for various endoscopic spinal procedures such as discectomy, nucleotomy, non-instrumented posterolateral fusion procedures.
Device Story
Endius Atavi System provides endoscopic access to lumbar spine. Components include tissue dilators, retracting device for access maintenance, endoscope, light source, light guide, camera control unit, and camera head. Used in OR by surgeons to visualize surgical site during spinal procedures. System facilitates minimally invasive access; allows surgeon to perform discectomy, nucleotomy, or fusion under endoscopic visualization. Benefits include reduced tissue trauma compared to open surgery.
Clinical Evidence
No clinical data provided. Substantial equivalence supported by bench performance testing and certification of compliance to design control requirements.
Technological Characteristics
Endoscopic spinal access system comprising tissue dilators, retractors, endoscope, light source, light guide, and camera system. Class II device. Operates as an endoscopic visualization and access tool. No specific materials or software algorithms detailed.
Indications for Use
Indicated for patients undergoing posterior endoscopic lumbar spine procedures, including discectomy, nucleotomy, and posterolateral fusion (instrumented with TriFix system or non-instrumented).
Regulatory Classification
Identification
An arthroscope is an electrically powered endoscope intended to make visible the interior of a joint. The arthroscope and accessories also is intended to perform surgery within a joint.
{0}------------------------------------------------
JUN 2 6 2002
K021742
## Section 7 - 510(k) Summary
| 7.1<br>Statement | Pursuant to §513(i)(3)(A) of the Food, Drug, and Cosmetic Act, Endius, Inc.<br>is required to submit with this Premarket Notification either an "... adequate<br>summary of any information respecting safety and effectiveness or state that<br>such information will be made available upon request of any person." Endius,<br>Inc. chooses to submit a summary of information respecting safety and<br>effectiveness. According to §513(i)(3)(B), "Any summary under<br>subparagraph (A) respecting a device shall contain detailed information<br>regarding data concerning adverse health effects..."<br><br>The summary regarding the adverse health effects of the modified device,<br>Endius Atavi System is provided below. |
|----------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| 7.2<br>Submitter | Endius, Inc.<br>23 West Bacon Street<br>Plainville, MA. 02762 (USA) |
| 7.3<br>Company<br>Contact | Christine Kuntz-Nassif<br>Director, Regulatory Affairs<br>Endius, Inc.<br>508-643-0983 |
| 7.4<br>Device Name/<br>Classification | Proprietary Name: Endius Atavi System<br>Common Name: Endoscopic Spinal Access System<br>Classification Name: Endoscope and Accessories<br><br>The devices in the Endoscopic Atavi System can be classified as class II,<br>876.1500 Endoscope and Accessories. The primary device in the system is the<br>endoscope. The accessory equipment is needed to gain access for placement<br>of the endoscope, to support the endoscope in position, or to work with the<br>endoscope for the purpose of visualization. |
| 7.5<br>Predicate<br>Legally<br>Marketed<br>Devices | Endius Endoscopic Access System: K991794, K994425, K002437 |
| 7.6 Device<br>Description | The Endius Atavi System includes instruments used to access the spine by<br>dilation of the overlying tissues, as well as a retracting device that is used to<br>maintain the access. The visualization components of the system include, an<br>endoscope, a light source, light guide, a camera control unit, and a camera<br>head. |
| 7.7 Indications<br>for Use | The Endius Atavi System is indicated for use in instrumented posterolateral<br>fusion procedures where the TriFix Spinal Instrumentation System is utilized.<br>The Endius Atavi System is also intended to be used for posterior endoscopic<br>access to the lumbar spine for various endoscopic spinal procedures such as<br>discectomy, nucleotomy, non-instrumented posterolateral fusion procedures. |
| 7.7<br>Safety and<br>Performance | This submission is a Special 510(k): Device Modification as described in<br>FDA's guidance document entitled "The New 510(k) Paradigm - Alternate<br>Demonstrating Substantial Equivalence in Premarket<br>Approaches to<br>Notifications."<br>In support of this 510(k), Endius, Inc. has provided<br>certification of compliance to 21 CFR 820.30 Design Control requirements,<br>and the results of validation testing (performance testing) for the device<br>modification. |
{1}------------------------------------------------
ﻢ
s
{2}------------------------------------------------
7.8 Substantial Equivalence Based on the indications for use, technological characteristics, and comparison to predicate devices, the proposed Endius Atavi System has been shown to be substantially equivalent to predicate devices under the Federal Food, Drug and Cosmetic Act.
Applicant
Date 5/24/02
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized eagle or bird design, composed of three curved lines that suggest the shape of a bird in flight. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" are arranged in a circular pattern around the bird symbol.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
Ms. Christine Kuntz-Nassif Director, Regulatory Affairs Endius, Inc. 23 West Bacon Street Plainville, MA 02762
JUN 2 6 2002
Re: K021748
Trade/Device Name: Endius® Atavi™ System Regulation Number: 888.1100 Regulation Name: Arthroscope Regulatory Class: II Product Code: HRX Dated: May 24, 2002 Received: May 28, 2002
Dear Ms. Kuntz-Nassif:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{4}------------------------------------------------
## Page 2 - Ms. Christine Kuntz-Nassif
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 21 CFR Part 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4659. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html
Sincerely yours.
Stypt Rhodes
a M. Witten, Ph.D., M.D. of Celi Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
510(k) Number (if known):
021748
Device Name: Endius® Atavi™ System
## Indications for Use:
The Endius Atavi System is indicated for use in instrumented posterolateral fusion procedures where the TriFix Spinal Instrumentation System is utilized.
The Endius Atavi System is also intended to be used for posterior endoscopic access to the lumbar spine for various endoscopic spinal procedures such as discectomy, nucleotomy, noninstrumented posterolateral fusion procedures.
## (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
Image /page/5/Picture/7 description: This image shows a document related to the Concurrence of CDRH, Office of Device Evaluation (ODE). It indicates that the document was posted on July 1, 1998, with an optional format date of 3-10-98. The document includes a division sign-off for the Division of General, Restorative, and Neurological Devices. The 510(k) number is listed as K021748, and the document number is 000002.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.