← Product Code [JDK](/productcode/JDK) · K021719

# FORTITUDE CEMENT RESTRICTOR (K021719)

_Spinal Concepts, Inc. · JDK · Aug 15, 2002 · General, Plastic Surgery · SESU_

**Canonical URL:** https://fda-staging.innolitics.com/device/K021719

## Device Facts

- **Applicant:** Spinal Concepts, Inc.
- **Product Code:** [JDK](/productcode/JDK.md)
- **Decision Date:** Aug 15, 2002
- **Decision:** SESU
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 878.3300
- **Device Class:** Class 2
- **Review Panel:** General, Plastic Surgery
- **Attributes:** Therapeutic

## Indications for Use

The Fortitude Cement Restrictor System is intended for use as a cement restrictor in orthopedic surgeries such as those involving the femoral canal and tibial plateau in hip stem and total knee replacement. This device is not intended for use in spinal applications.

## Device Story

Fortitude Cement Restrictor is a hollow, titanium alloy device with teeth on two opposing flat sides; available in various sizes, straight and tapered styles. Used in orthopedic surgery to hold bone cement (PMMA) in the distal diaphyseal canal during cemented arthroplasty (hip stem and total knee replacement). Implanted by orthopedic surgeons in clinical settings. Device acts as a physical barrier to prevent cement migration beyond the desired site, ensuring proper cement mantle formation. Benefits include improved fixation of arthroplasty components.

## Clinical Evidence

No clinical data. Bench testing only.

## Technological Characteristics

Hollow titanium alloy device; features teeth on two opposing flat sides for fixation in the canal. Available in straight and tapered configurations. Mechanical device; no energy source, connectivity, or software.

## Regulatory Identification

Surgical mesh is a metallic or polymeric screen intended to be implanted to reinforce soft tissue or bone where weakness exists. Examples of surgical mesh are metallic and polymeric mesh for hernia repair, and acetabular and cement restrictor mesh used during orthopedic surgery.

## Predicate Devices

- Medtronic Sofamor Danek Cement Restrictor(s) ([K012255](/device/K012255.md) and [K013014](/device/K013014.md))
- Signus RABEA 1M Cement Restrictor ([K990345](/device/K990345.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

# AUG 1 5 2002

# Safety and Effectiveness Information

Submitted By:

David M. Hooper, PhD Manager, Clinical & Regulatory Affairs Spinal Concepts, Inc. 12012 Technology Blvd., Suite 100 Austin, TX 78727 512-918-2700 May 22, 2001

#### Device:

Trade Name: Proposed Classification Name: Product Code:

Fortitude Cement Restrictor Prosthesis. Hip. Cement Restrictor JDK (21CFR 878.3300)

musddy

#### Predicate Devices:

The Fortitude Cement Restrictor is similar in terms of intended use, materials of construction, and technological characteristics to the predicate devices reviewed, the Medtronic Sofamor Danek Cement Restrictor(s) (K012255 and K013014) and the Signus RABEA 1M Cement Restrictor (K990345).

### Device Description:

The Fortitude Cement Restrictor is a hollow, titanium allov device with teeth on two opposing flat sides. The device comes in various sizes and is offered in straight and tapered styles. The device is used to hold bone cement (PMMA) in the distal diaphyseal canal in patients requiring a cemented arthroplasty device.

#### Intended Use:

The Fortitude Cement Restrictor System is intended for use as a cement restrictor in orthopedic surgeries such as those involving the femoral canal and tibial plateau in hip stem and total knee replacement.

This device is not intended for use in spinal applications.

### Substantial Equivalence:

The Fortitude Cement Restrictor was demonstrated to be substantially equivalent to previously cleared devices such as the Medtronic Sofamor Danek Cement Restrictor(s) (K012255 and K013014) and the Signus RABEA110 Cement Restrictor (K990345). A Design Review for the device was provided in this submission.

{1}------------------------------------------------

Image /page/1/Picture/1 description: The image is a black and white circular seal for the Department of Health & Human Services USA. The seal features the department's name in a circular arrangement around the perimeter. In the center of the seal is a stylized image of an eagle with its wings spread, symbolizing the department's mission to protect and promote the health and well-being of the people in the United States.

Public Health Service

AUG 1 5 2002

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

David Hooper, Ph.D. Manager, Clinical & Regulatory Affairs Spinal Concepts Incorporated 12012 Technology Blvd, Suite 100 Austin, Texas 78727

Re: K021719 Fortitude Cement Restrictor Regulation Number: 21 CFR 878.3300 Regulation Name: Surgical mesh Regulatory Class: II Product Code: JDK Dated: May 22, 2002 Received: May 23, 2002

Dear Dr. Hooper:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval (PMA) application. You may, therefore, market the device, subject to the general controls provisions of the Act and the limitations described below. The general controls provisions of the Act include requirements for annual registration. listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

The Office of Device Evaluation has determined that there is a reasonable likelihood that this device will be used for an intended use not identified in the proposed labeling and that such use could cause harm. Therefore, in accordance with Section 513(i)(1)(E) of the Act, the following limitation must appear in the Warnings section of the device's package insert and also as a Warning on the product label:

WARNING: THIS DEVICE IS NOT INTENDED FOR ANY SPINAL INDICATIONS.

# THE SAFETY AND EFFECTIVENESS OF THIS DEVICE WHEN IMPLANTED IN THE SPINE HAVE NOT BEEN ESTABLISHED.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can

{2}------------------------------------------------

Page 2 - David Hooper, Ph.D.

be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and permits your device to proceed to the market. This letter will allow you to begin marketing your device as described in your 510(k) premarket notification if the limitation statement above is added to your labeling, as described.

Please note that the above labeling limitations are required by Section 513(i)(1)(E) of the Act. Therefore, a new 510(k) is required before these limitations are modified in any way or removed from the device's labeling.

If vou desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 21 CFR Part 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4659. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address: http://www.fda.gov/cdrh/dsma/dsmamain.html

Sincerely yours

Bernard Statland, M.D., Ph.D. Director Office of Device Evaluation Center for Devices and Radiological Health

{3}------------------------------------------------

## INIDICATIONS FOR USE STATEMENT

Page 1 of 1

510(k) Number (if known): _ K ○ 2 1 7 / 9

Device Name: Spinal Concepts, Inc. Fortitude Cement Restrictor

Indications for Use:

The Fortitude Cement Restrictor System is intended for use as a cement restrictor in orthopedic surgeries such as those involving the femoral canal and tibial plateau in hip stem and total knee replacement.

This device is not intended for use in spinal applications.

for Mark N. Mikkelsen
(Division Sign-Off)

ivision Sign-Off) Division of General, Restorative and Neurological Devices

510(k) Number K02179

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Prescription Usc: (Per 21 CFR 801.109) OR

Over-The-Counter: (Optional Format 1-2-96)

---

**Source:** [https://fda-staging.innolitics.com/device/K021719](https://fda-staging.innolitics.com/device/K021719)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
