K021633 · Cti Pet Systems, Inc. · KPS · Aug 14, 2002 · Radiology
Device Facts
Record ID
K021633
Device Name
ECAT VISION PET SCANNER, MODELS 3000 & 5000
Applicant
Cti Pet Systems, Inc.
Product Code
KPS · Radiology
Decision Date
Aug 14, 2002
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.1200
Device Class
Class 2
Indications for Use
The CPS/Siemens ECAT Vision 3000 and Vision 5000 PET systems are positron emission tomography (PET) scanners that are intended to be utilized by appropriately trained health care professionals to image and measure the distribution of injected positron emitting radiopharmaceuticals in humans for the purpose of determining various metabolic and physiologic functions within the human body.
Device Story
ECAT Vision 3000 (3-head) and 5000 (5-head) are PET scanners using rotating flat panel detector heads and LSO crystal technology. Device acquires PET data via coincidence point source transmission scanning; processes data through ACS III system; reconstructs images using FOREJ rebinning and 2D Attenuation Normalization Weighted OSEM (ANW-OSEM) algorithms. Operated by trained healthcare professionals in clinical settings. System includes patient handling system (max weight 182 kg). Operating console runs Siemens Syngo software on Windows 2000/NT environment for acquisition, display, diagnostic, application, archival, and retrieval. Output consists of cross-sectional images used by clinicians to determine metabolic and physiologic functions. Benefits include improved visualization and diagnostic support for oncology applications.
Clinical Evidence
No clinical data provided; substantial equivalence is based on technological characteristics and design similarities to legally marketed predicate devices.
Technological Characteristics
PET scanner with 3 (Vision 3000) or 5 (Vision 5000) rotating flat panel detector heads. Uses LSO crystal detector technology. Includes patient handling system (max 182 kg). ACS III acquisition system. Reconstruction via dedicated computer. Connectivity: networked console running Siemens Syngo software on Windows 2000/NT. Software includes acquisition, display, and archival tools.
Indications for Use
Indicated for use by trained healthcare professionals to detect location and distribution of injected positron-emitting radionuclides in humans and produce cross-sectional images via computer reconstruction for metabolic and physiologic function assessment.
Regulatory Classification
Identification
An emission computed tomography system is a device intended to detect the location and distribution of gamma ray- and positron-emitting radionuclides in the body and produce cross-sectional images through computer reconstruction of the data. This generic type of device may include signal analysis and display equipment, patient and equipment supports, radionuclide anatomical markers, component parts, and accessories.
{0}------------------------------------------------
K021635
# 510(k) SUMMARY
## [As required by 21 CFR 807.87(h)]
### Identification of Submitter Submitter:
Telephone No: Fax No: Date of preparation:
#### Identification of the Product
Device Proprietary Name: Common Name: Classification Name:
William Skremsky Senior Regulatory Affairs Specialist CTI PET Systems, Inc. 810 Innovation Drive Knoxville, TN 37932 (865) 218-2522 (865) 218-3000 May 10, 2002
ECAT Vision 3000 and ECAT Vision 5000 Positron Emission Tomography (PET) Scanner Emission Computed Tomography System per 21 CFR 892.1200
#### Marketed Devices to Which Equivalence is Claimed
| Device | Manufacturer | 510(k) Number |
|---------------|-----------------------|---------------|
| ECAT EXACT | CTI PET Systems (CPS) | K962797 |
| E.CAM LSO 311 | CTI PET Systems (CPS) | K981027 |
| ECAT PET/CT | CTI PET Systems (CPS) | K002715 |
#### Device Description
The ECAT Vision 3000 and Vision 5000 are new positron emission tomography (PET) scanner systems. The ECAT Vision series PET scanners utilize rotating flat panel detector heads, LSO crystal detector technology and coincidence point source transmission scan capability. The ECAT Vision series scanners are produced and marketed under two distinct models, the ECAT Vision 3000 and the ECAT Vision 5000. The two models of the ECAT Vision are configured with either: 1) three detector heads in the ECAT Vision 3000, or 2) five detector heads in the ECAT Vision 5000.
A Patient Handling System (PHS) is provided with the Vision series scanners, which is a slightly modified version of the PHS included with the CPS ECAT EXACT series PET scanners (K962797). Vertical bed travel, horizontal scanning range and all other PHS operating characteristics are the same for the PHS provided with the Vision series scanners as those of the PHS provided with the ECAT EXACT series scanners. The maximum patient weight specification for the Vision series PHS, is 182 kg (400 lbs).
The acquisition computer system is an ACS III system, similar to that used for the ECAT PET/CT system (K002715). Image reconstruction for the ECAT Vision scanners is performed using a dedicated reconstruction computer. The reconstruction algorithm is FOREJ rebinning and 2D
{1}------------------------------------------------
# 510(k) Summary ECAT VISION PET Scanner p. 2
Attenuation Normalization Weighted OSEM (ANW-OSEM). The reconstruction system interfaces with the acquisition raid disk system for access to the basic acquired data.
The operating console computer is an updated version of the existing ECAT PET/CT operating console. The console runs Siemens Syngo software, with CPS ECAT PET acquisition and application software overlaying the Siemens software. This computer is based on an Intel Pentium system running Microsoft's Windows 2000, or later, operating system. Operating in a syngo-based, NT environment, the ECAT Vision operating software provides acquisition, display, diagnostic, application, system, archival, and retrieval support capabilities. Initially, whole-body oncology applications and visualization tools are being provided.
## Indications for Use
The CPS/Siemens ECAT Vision 3000 and Vision 5000 PET systems are positron emission tomography (PET) scanners that are intended to be utilized by appropriately trained health care professionals to image and measure the distribution of injected positron emitting radiopharmaceuticals in humans for the purpose of determining various metabolic and physiologic functions within the human body.
## Comparison with Predicate Devices
The ECAT Vision series PET scanners leverage much of the already proven technology, as well as many components and features of currently produced CPS tomographic systems. The flat panel detectors, LSO crystal detector technology, and coincidence point source transmission scanning design concepts were borrowed from the E.CAM LSO 311 PET/SPECT system (K981027). The patient handling system (PHS) provided with ECAT Vision scanners is a slightly modified version of the PHS used on the CPS ECAT EXACT and ECAT EXACT HR+ PET systems (K962797). The Advanced Computational System (ACS III), used to store and process acquired PET data into sinograms, as well as the reconstruction computer, control console, and the ECAT PET/syngo software are updated versions of those components used in the ECAT PET/CT system (K002715).
{2}------------------------------------------------
Image /page/2/Picture/0 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is an abstract symbol resembling a stylized caduceus or a representation of the human form, with three wavy lines forming the body and head.
# DEPARTMENT OF HEALTH & HUMAN SERVICES
ood and Drug Administrat Corporate Boulevard ville MD 20850
AUG 1 4 2002
Mr. William Skremsky Senior Regulatory Affairs Specialist CTI, Inc. 810 Innovation Drive KNOXVILLE TN 37932
## Re: K021633
Trade/Device Name: ECAT Vision PET Scanner Regulation Number: 21 CFR 892.1200 Regulation Name: Emission computed tomography system
Regulatory Class: II Product Code: 90 KPS Dated: May 10, 2002 Received: May 17, 2002
Dear Mr. Skremsky:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (sections 531-542 of the Act); 21 CFR 1000-1050.
{3}------------------------------------------------
Page 2
This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at one of the following numbers, based on the regulation number at the top of this letter:
| 8xx.1xxx | (301) 594-4591 |
|----------------------------------|----------------|
| 876.2xxx, 3xxx, 4xxx, 5xxx | (301) 594-4616 |
| 884.2xxx, 3xxx, 4xxx, 5xxx, 6xxx | (301) 594-4616 |
| 892.2xxx, 3xxx, 4xxx, 5xxx | (301) 594-4654 |
| Other | (301) 594-4692 |
Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers. International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html.
Sincerely yours.
Nancy C. Snigdon
Nancy C. Brogdon Director, Division of Reproductive, Abdominal, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
Page 1 of 1
K121632 510(k) Number (if known):
Device Name: ECAT Vision PET Scanner
# Indications for Use:
CPS ECAT Vision 3000 and ECAT Vision 5000 positron emission tomography (PET) scanners are intended to be utilized by appropriately trained health care professionals to detect the location and distribution of injected positron-emitting radionuclides in the body and produce cross- sectional images through computer reconstruction of the data.
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Prescription Use
(Per 21 CFR 801.109)
OR
Over-The-Counter Use
(Optional Format 1-2-96)
David A. Heymann
(Division Sign-Off) Division of Reproductive, Abdominal, and Radiological Devices 510(k) Number ________________________________________________________________________________________________________________________________________________________________
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.