K021626 · New Deal, S.A. · HRS · Aug 12, 2002 · Orthopedic
Device Facts
Record ID
K021626
Device Name
HALLU PLATES
Applicant
New Deal, S.A.
Product Code
HRS · Orthopedic
Decision Date
Aug 12, 2002
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 888.3030
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The HALLU® -PLATES are intended to be implanted for fixation of fractures, osteotomies or arthrodesis of the first metatarso-phalangeal joint. Examples include: -Hallux rigidus -Severe hallux valgus (IM angle > 20° and HV angle > 40°) -Deformity from rheumatoid arthritis -Failed previous surgical procedure -Traumatic arthritis -Neuromuscular instability.
Device Story
HALLU®-PLATES are low-profile titanium bone plates designed for first metatarso-phalangeal (MTP) joint arthrodesis; pre-bent for 10° valgus and 10° dorsiflexion. System includes two designs (HALLU®-C and HALLU®-S) in six sizes for anatomical fit. Implants are secured to bone using titanium SNAP-OFF® screws (2.7 mm and 3.0 mm diameters). Bone contact surface is sand-blasted to enhance stability. Used by orthopedic surgeons in clinical/surgical settings to stabilize bone during healing or fusion. Provides rigid internal fixation to correct deformities or treat fractures, potentially improving patient mobility and reducing pain associated with MTP joint instability or arthritis.
Clinical Evidence
Bench testing only. Testing included determination of bending strength and stiffness per French Standard ISO 9585 and ASTM F-382-99, and fatigue strength testing to ensure minimal risk of plate rupture.
Technological Characteristics
Titanium alloy (TiAl6V4) bone plates; low-profile; pre-bent for 10° valgus/10° dorsiflexion; sand-blasted bone contact surface. Fixation via 2.7 mm and 3.0 mm titanium SNAP-OFF® screws. Complies with ASTM F-382-99 and ISO 9585 for bending strength/stiffness.
Indications for Use
Indicated for patients requiring fixation of fractures, osteotomies, or arthrodesis of the first metatarso-phalangeal joint, including those with hallux rigidus, severe hallux valgus, rheumatoid arthritis deformity, failed prior surgery, traumatic arthritis, or neuromuscular instability.
Regulatory Classification
Identification
Single/multiple component metallic bone fixation appliances and accessories are devices intended to be implanted consisting of one or more metallic components and their metallic fasteners. The devices contain a plate, a nail/plate combination, or a blade/plate combination that are made of alloys, such as cobalt-chromium-molybdenum, stainless steel, and titanium, that are intended to be held in position with fasteners, such as screws and nails, or bolts, nuts, and washers. These devices are used for fixation of fractures of the proximal or distal end of long bones, such as intracapsular, intertrochanteric, intercervical, supracondylar, or condylar fractures of the femur; for fusion of a joint; or for surgical procedures that involve cutting a bone. The devices may be implanted or attached through the skin so that a pulling force (traction) may be applied to the skeletal system.
{0}------------------------------------------------
**AUG 1 2 2002**
K021626 p113
#### SUMMARY OF SAFETY AND EFFECTIVENESS 3.
- SPONSOR IDENTIFICATION: A. NewDeal SA Parc d'Activitiés Garigliano Rue de la Convention 38 200 VIENNE FRANCE
Telephone: (33) 4 74 78 15 15
Fax: (33) 4 74 78 15 16
#### B. ESTABLISHMENT REGISTRATION NUMBER: 9615741
## OFFICIAL CONTACT PERSON C. Norman F. Estrin, Ph.D., RAC President Estrin Consulting Group, Inc. 9109 Copenhaver Drive Potomac, MD 20854
Tel: (301) 279 -2899 Fax: (301) 294-0126
#### DATE OF PREPARATION OF THIS SUMMARY: April 11th, 2002 D.
#### HALLU®PLATES PROPRIETARY (TRADE) NAME: E.
- Metatarsal-phalangeal arthrodesis plate F. COMMON NAME:
#### CLASSIFICATION NAME AND REFERENCE G.
Plate, Fixation, Bone (21 CFR, Section 888. 3030)
- H. PROPOSED REGULATORY CLASS: Class II
- DEVICE PRODUCT CODE: I. 87 HRS
- 87 OR Orthopedic J. PANEL CODE:
- K. DESCRIPTION OF DEVICE:
The HALLU® -PLATES are available in two different designs :
- the HALLU®-C PLATE -
- the HALLU®-S PLATE -
The HALLU®-C PLATE and the HALLU® -S PLATE are low profile Titanium plates dedicated to first metatarso-phalangeal arthrodesis. Those implants are pre-bent for optimal anatomical adaptation (10° valgus and 10°dorsiflexion). Their fixation is provided by Titanium SNAP-OFF® screws available in a two diameters: 3.0 and 2.7 mm. They exist in different colors for size identification. The bone contact surface is sand blasted in order to maximize plate stability. The range of HALLU -C PLATE and
01 2005
{1}------------------------------------------------
K021624 p²/3
HALLUS S PLATE include 6 sizes ( 31eft/3right) for optimal anatomic fit. Moreover, the HALLU®-S PLATE has an anatomical design providing optimal bone coverage.
- The HALLU® -PLATES are intended to be implanted for L. INTENDED USE: fixation of fractures, osteotomies or arthrodesis of the first metatarso-phalangeal joint.
### INDICATIONS M. The HALLU®-PLATES are intended to be implanted for FOR USE: fixation of fractures, osteotomies or arthrodesis of the first metatarso-phalangeal joint. Examples include:
-Hallux rigidus
-Severe hallux valgus (IM angle > 20° and HV angle > 40°)
-Deformity from rheumatoid arthritis
-Failed previous surgical procedure
-Traumatic arthritis
-Neuromuscular instability.
#### N. PREDICATE
DEVICE:
The HALLU - PLATES are substantially equivalent to the Howmedica Luhr plate (K935448) and Howmedica Profyle® Hand and Small Fragment System (K961497) , the Acumed Congruent bone plate system (K012655) and the Synthes modular foot system (K001941).
#### COMPARISON OF TECHNOLOGICAL CHARACTERISTICS: 0.
The HALLU® - PLATES, the Howmedica Luhr and
Howmedica Profyle plates, the Acumed plate and Synthes
modular foot system are intended for internal bone fixation for bone fractures or reconstruction.
Luhr® HALLU®- PLATES, the Howmedica The and Howmedica Profyle® plates, the Acumed plate and Synthes modular foot system are all indicated for bone fractures or reconstruction in the foot.
The HALLU® - PLATES, the Howmedica Profyle " plates and the Acumed plate are in Titanium alloy TIAI6V4 whereas the Synthes modular foot system is manufactured in Stainless steel 316 L and the Howmedica Luhr in Vitallium alloy .
{2}------------------------------------------------
All those systems are fixed with screws except for the Synthes modular foot system attached to bone via buttress pins.
KOZIG 26
Test 1 : Determination of bending strength and stiffness of SUMMARY OF STUDIES: bone plates was compared with requirement of the French Standard ISO 9585 and ASTM standard F-382-99 and found to have a bending stiffness and strength in compliance with the selected standards.
> Test 2 : Determination of fatigue strength of bones plates were conducted and have shown the risk of rupture of the plate is minimal.
> > () . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
{3}------------------------------------------------
# DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" around the perimeter. Inside the circle is an abstract symbol resembling three stylized human profiles or birds in flight, stacked on top of each other.
## Public Health Service
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
NewDeal SA c/o Norman F. Estrin, PhD, RAC President Estrin Consulting Group, Inc. 9109 Copenhaver Drive Potomac, Maryland 20854
AUG 1 2 2002
Re: K021626
Trade/Device Name: HALLU®-Plates System Regulation Number: 21 CFR §888.3030 Regulation Name: Single/multiple component metallic bone fixation appliances and accessories Regulatory Class: II Product Code: HRS Dated: May 14, 2002 Received: May 17, 2002
Dear Dr. Estrin;
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{4}------------------------------------------------
Page 2 - Norman F. Estrin, PhD, RAC
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 21 CFR Part 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4659. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97), Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html
Sincerely yours,
Mark H-Milliman
Celia M. Witten, Ph.D., M.D. Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health
{5}------------------------------------------------
# Page 1 of 1
# 510(k) Number (if known): 《ՀՀՀՆՆՀՀՀ
HALLU®-PLATES SYSTEM Device Name: Indications for Use:
> The HALLU® -PLATES are intended to be implanted for fixation of fractures, osteotomies or arthrodesis of the first metatarso-phalangeal joint. Examples include:
> > -Hallux rigidus
-Severe hallux valgus (IM angle > 20° and HV angle > 40°)
-Deformity from rheumatoid arthritis
-Failed previous surgical procedure
-Traumatic arthritis
-Neuromuscular instability.
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
| Prescription Use<br>(Per 21 CFR 801.109) | OR | Over-the-Counter Use<br>(Optional Format 1-2-96) |
|------------------------------------------|----|--------------------------------------------------|
|------------------------------------------|----|--------------------------------------------------|
for
(Division Sign-Off)
Division of General, Restorative
and Neurological Devices
| 510(k) Number | K021626 |
|---------------|---------|
|---------------|---------|
004E
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.