← Product Code [JJX](/productcode/JJX) · K021624

# RNA GL4 GLUCOSE CALIBRATION VERIFICATION CONTROL, MODEL RNA CVC GL4 (K021624)

_Bionostics, Inc. · JJX · Jul 12, 2002 · Clinical Chemistry · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K021624

## Device Facts

- **Applicant:** Bionostics, Inc.
- **Product Code:** [JJX](/productcode/JJX.md)
- **Decision Date:** Jul 12, 2002
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 862.1660
- **Device Class:** Class 1
- **Review Panel:** Clinical Chemistry

## Indications for Use

RNA GL4 Glucose Calibration Verification Control is intended for use to verify the analytical performance of certain blood glucose meters as listed on the package insert.

## Device Story

RNA GL4 Glucose Calibration Verification Control is a seven-vial, viscosity-adjusted, aqueous liquid glucose control linearity set. It is designed for use with MediSense Point of Care blood glucose systems (PCx and i-STAT 1) using Precision PCx or Precision PCx Plus test strips. The control is packaged in plastic bottles with dropper tips for application to test strips and is colored red for visibility. It is used by laboratory personnel to perform periodic quality control checks, assessing linearity and calibration of the blood glucose system. By providing glucose values across the reportable range, it allows healthcare providers to verify the accuracy and performance of the meter, ensuring reliable patient glucose monitoring.

## Clinical Evidence

No clinical data. Bench testing only, including accelerated aging studies, precision testing, and correlation to reference methodology.

## Technological Characteristics

Viscosity-adjusted, aqueous glucose control solution. Seven specific glucose concentrations. Packaged in plastic bottles with dropper tips. Non-hazardous, contains no biological materials. Optimized to simulate whole blood response for specific blood glucose meters.

## Regulatory Identification

A quality control material (assayed and unassayed) for clinical chemistry is a device intended for medical purposes for use in a test system to estimate test precision and to detect systematic analytical deviations that may arise from reagent or analytical instrument variation. A quality control material (assayed and unassayed) may be used for proficiency testing in interlaboratory surveys. This generic type of device includes controls (assayed and unassayed) for blood gases, electrolytes, enzymes, multianalytes (all kinds), single (specified) analytes, or urinalysis controls.

## Predicate Devices

- Multi-Meter Glucose Calibration Verification Material ([K012430](/device/K012430.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

1621624

JUL 12 2002

# BIONOS

## 510(k) Summary*

- (a) (1) Submitter's name, address Bionostics, Inc. 7 Jackson Road Devens, MA 01432
Contact Person Kathleen Storro Director, QA & Regulatory Affairs (978) 772-7070 x 220

Date of preparation of this summary: 15 May 2002

| (2) Device trade or proprietary name: | RNA GL4 Glucose<br>Calibration Verification Control |
|---------------------------------------|-----------------------------------------------------|
|---------------------------------------|-----------------------------------------------------|

Device common or usual name or classification name:

Multi Analyte Control Solution, All Types (Assayed and Unassayed)

| PRODUCT NOMENCLATURE            | CLASSIFICATION<br>NUMBER | CLASS | PANEL     |
|---------------------------------|--------------------------|-------|-----------|
| SINGLE ANALYTE CONTROL SOLUTION | 862.1660 75 JJX          | 1     | CHEMISTRY |

#### (3) Substantial Equivalence

RNA GL4 Glucose Calibration Verification Control is substantially equivalent in function, safety and efficacy to currently marketed devices produced by Bionostics. In example:

Comparison of Multi-Meter Glucose Calibration Verification Material to predicate devices for substantial equivalency

| Characteristic    | Predicate Devices                                                                                                           | Modified Device                                                                                                             |
|-------------------|-----------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------|
| Name:             | Multi-Meter Glucose<br>Calibration Verification<br>Material                                                                 | RNA Medical Glucose<br>Calibration Verification<br>Control                                                                  |
| 510(k), Date:     | K012430, 08/27/01                                                                                                           | K002540, 09/18/00                                                                                                           |
| Number of levels: | 5*                                                                                                                          | 5*<br>*7 vials per kit, only 5 levels<br>for use on specific devices                                                        |
| Analytes:         | Glucose                                                                                                                     | Glucose                                                                                                                     |
| Container:        | plastic bottle                                                                                                              | plastic bottle                                                                                                              |
| Fill volume:      | 4 mL                                                                                                                        | 4 mL                                                                                                                        |
| Color:            | red                                                                                                                         | red                                                                                                                         |
| Matrix:           | Buffered, aqueous solution<br>of D-Glucose, viscosity<br>modifier, preservatives and<br>other, non-reactive<br>ingredients. | Buffered, aqueous solution<br>of D-Glucose, viscosity<br>modifier, preservatives and<br>other, non-reactive<br>ingredients. |

<sup>*</sup> This summary of safety and effectiveness is submitted in accordance with the requirements of SMDA 1990 and 21 CFR 807.92.

{1}------------------------------------------------

#### Description of the new device (4)

RNA GL4 Glucose Calibration Verification Control is a seven-vial, viscosityadjusted, aqueous liquid glucose control linearity set which provides two, five-level CVC sets, optimized for use with MediSense Point of Care blood glucose systems PCx and i-STAT 1 using Precision PCx or Precision PCx Plus test strips. RNA GL4 Glucose Calibration Verification Control provides a convenient method of performing periodic QC checks for laboratories selecting to measure liquid QC material as a part of their quality assurance program. The product is packaged in plastic bottles with dropper tips for application of the solution to test strips. The control has a red color to help users see the solution while dispensing onto a test strip.

RNA GL4 Glucose Calibration Verification Control contains glucose values at the lower and upper limits of reportable range as well as three points within the range and therefore, may be used to assess the linearity and calibration, or verify performance of the blood glucose systems listed on the package insert.

| Strip Type                                          | Solution A           | Solution B           | Solution C            | Solution D            | Solution E            | Solution F            | Solution G            |
|-----------------------------------------------------|----------------------|----------------------|-----------------------|-----------------------|-----------------------|-----------------------|-----------------------|
| <i>PCx Plus</i><br>strip lots beginning<br>with "6" | <b>40</b><br>Level 1 | --                   | <b>120</b><br>Level 2 | <b>220</b><br>Level 3 | <b>290</b><br>Level 4 | <b>370</b><br>Level 5 | --                    |
| <i>PCx</i><br>strip lots beginning<br>with "1"      | --                   | <b>40</b><br>Level 1 | <b>103</b><br>Level 2 | --                    | <b>249</b><br>Level 3 | <b>324</b><br>Level 4 | <b>440</b><br>Level 5 |

Target midpoint values for Precision PCx and Precision PCx Plus test strips

RNA GL4 Glucose Calibration Verification Control is a non-hazardous aqueous solution containing no biological materials.

#### (5) Intended use of the device

RNA GL4 Glucose Calibration Verification Control is intended for use to verify the analytical performance of certain blood glucose meters as listed on the package insert.

#### (6) Technological characteristics of the device.

This material consists of viscosity-adjusted, aqueous glucose control solutions prepared in seven specific glucose concentrations. The solutions have been optimized to simulate the response of whole blood for the blood glucose meters in combination with test strips, listed in the package insert.

{2}------------------------------------------------

### Summary of non-clinical tests submitted with the premarket notification for the (b) (1) device.

Tests were conducted to verify specific performance requirements:

- a) Accelerated aging studies on most labile analytes, together with real time experience with products with similar formulations and failure mode to support stability claim.
- b) Test precision
- c) Correlation to reference methodology
- Summary of clinical tests submitted with the premarket notification for the (b) (2) device.

N/A

#### Conclusions drawn from the clinical and non-clinical trials. (b) (3)

Comparison of technological characteristics, formulation and intended use to predicate devices listed in this summary support the claim of substantial equivalence.

{3}------------------------------------------------

Image /page/3/Picture/0 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circle with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the top half. Inside the circle is a stylized symbol that resembles three wavy lines or a caduceus without the staff.

Food and Drug Administration 2098 Gaither Road Rockville MD 20850

### JUL 1 2 2002

Ms. Kathleen Storro Director, QA and Regulatory Affairs Bionostics 7 Jackson Road Devens, MA 01432

Re: k021624

> Trade/Device Name: RNA GL4 Glucose Calibration Verification Control Regulation Number: 21 CFR 862.1660 Regulation Name: Quality control material (assayed and unassayed) Regulatory Class: Class I, reserved Product Code: JJX Dated: May 15, 2002 Received: May 16, 2002

Dear Ms. Storro:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807): labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

{4}------------------------------------------------

Page 2 -

.

This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and 1 additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4588. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsmaldsmamain.html".

Sincerely yours,

Steven Putman

Steven I. Gutman, M.D., M.B.A. Director Division of Clinical Laboratory-Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{5}------------------------------------------------

KO21624 510(k) Number:

RNA GL4 Glucose Calibration Verification Control Device Name:

Indications for Use:

RNA GLA Glucose Calibration Verification Control assayed controls are intended for use to verify the performance of multiple blood glucose meters as indicated on the package insert at their upper and lower ends of reportable range and at three points within the range. These controls can therefore be used to assess the linearity and calibration of the test system, or to verify the test systems' performance.

For In Vitro Diagnostic Use

Sean Cooper

(Division Sign-Off)
Division of Clinical Laboratory Devices

510(k) Number.

(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Prescription Use (Per 21 CFR 801.109)

OR

Over-The-Counter Use _

(Optional Format 1-2-96)

510(K) PREMARKET NOTIFICATION – GL4 Glucose CVC

---

**Source:** [https://fda-staging.innolitics.com/device/K021624](https://fda-staging.innolitics.com/device/K021624)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
