K021527 · Sectra-Imtec AB · KGI · Jun 7, 2002 · Radiology
Device Facts
Record ID
K021527
Device Name
SECTRA OSTEOPOROSIS PACKAGE
Applicant
Sectra-Imtec AB
Product Code
KGI · Radiology
Decision Date
Jun 7, 2002
Decision
SESE
Submission Type
Special
Regulation
21 CFR 892.1170
Device Class
Class 2
Attributes
Software as a Medical Device
Indications for Use
The device is intended for use to estimate bone mineral density ("BMD") in the forearm and to assess increased risk of osteoporotic fracture according to World Health Organization ("WHO") criteria. The device is indicated specifically for use to: (1) assist the physician in diagnosing subjects who already have been identified to be at risk for suffering from osteoporosis, together with other known risk factors (e.g., prior history of fractures, advanced age, low body weight, lack of physical exercise, lack of exposure to sunlight, insufficient dietary intake of vitamin D and/or calcium, and smoking); and (2) compare the BMD estimate with a reference population comprised of young normals and age-matched normals to compute T-scores and Z-scores, respectively.
Device Story
Sectra Osteoporosis Package is a software-based bone densitometer; estimates forearm bone mineral density (BMD) from digital computed radiography (CR) images. Operates within Sectra IDS5 Radiology Workstation; utilizes DICOM standard images. Physician-operated; provides BMD estimates, T-scores, and Z-scores by comparing patient data against reference populations. Assists in diagnosing osteoporosis in at-risk subjects. Features clinical and research modes; integrates with third-party PACS networks. Benefits patients by facilitating fracture risk assessment and clinical decision-making regarding osteoporosis management.
Clinical Evidence
Both in vitro and in vivo performance testing conducted to verify that design differences (direct digital image analysis vs. digitized X-ray) do not impact safety or efficacy. No specific clinical trial metrics (e.g., sensitivity, specificity) provided in the summary.
Technological Characteristics
Software-based bone densitometer; runs on Windows NT 4.0/2000 PCs. Utilizes DXR (Digital X-ray Radiogrammetry) techniques for BMD analysis. Complies with ACR/NEMA DICOM 3.0 standards. ISO 9001:2000 certified development. Integrates with Sectra IDS5 Radiology Workstation.
Indications for Use
Indicated for patients identified at risk for osteoporosis to estimate forearm bone mineral density (BMD) and assess fracture risk per WHO criteria. Used to assist physician diagnosis in conjunction with clinical risk factors (e.g., age, fracture history, lifestyle factors) and to compute T-scores and Z-scores via comparison to young and age-matched normal reference populations.
Regulatory Classification
Identification
A bone densitometer is a device intended for medical purposes to measure bone density and mineral content by x-ray or gamma ray transmission measurements through the bone and adjacent tissues. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.
{0}------------------------------------------------
# JUN - 7 2002
K021527
Sectra Document Number: 3-02.832-1.0
# 510(k) Summary of Safety & Effectiveness
(as required by 21 CFR 807.92c)
# Date Prepared:
May 8 2002
# Submitter's Information:
Sectra Imtec AB Teknikringen 2 SE-583 30 Linköping Sweden Phone: +1 46 13 23 52 00 Fax: +1 46 13 21 21 85
#### Trade Name, Common Name, Classification:
Trade name: Sectra Osteoporosis Package Bone Densitometer Common Names: Classification Name: Bone Densitometer (90 KGI) (21 C.F.R. § 892.1170)
# Predicate Device:
| Applicant: | Pronosco A/S |
|----------------|-------------------------------|
| 510(k) Number: | K002500 |
| Device: | X-Posure System Version 2 RAD |
#### Device Description:
The Sectra Osteoporosis Package is intended for use to estimate bone mineral densitv (BMD) in the forearm and to assess increased risk of osteoporotic fracture according to World Health Organization (WHO) criteria.
Sectra Osteoporosis Package features:
- indirect "DICOM capability" via the Sectra IDS5 Radiology workstation, i.e. it uses digital image data according to the DICOM standard.
- . single screen user interface reached via a right click on an image in a Sectra IDS5 Radiology Workstation.
- operates in two modes: Clinical mode and Research mode.
{1}------------------------------------------------
#### Indications for Use:
The device is intended for use to estimate bone mineral density ("BMD") in the forearm and to assess increased risk of osteoporotic fracture according to World Health Organization ("WHO") criteria. The device is indicated specifically for use to: (1) assist the physician in diagnosing subjects who already have been identified to be at risk for suffering from osteoporosis, together with other known risk factors (e.g., prior history of fractures, advanced age, low body weight, lack of physical exercise, lack of exposure to sunlight, insufficient dietary intake of vitamin D and/or calcium, and smoking); and (2) compare the BMD estimate with a reference population comprised of young normals and age-matched normals to compute T-scores and Z-scores, respectively.
## Technological Characteristics:
The Sectra Osteoporosis Package will run on the Windows NT 4.0 and Windows 2000 operating systems for PCs (as a minimum and depending upon system configuration).
#### Performance Data:
The subject device is developed according to ISO 9001:2000 and complies with ACR/NEMA Digital Imaging Communications in Medicine version 3.0.
## Conclusion:
The Sectra Osteoporosis Package is substantially equivalent to the previously cleared X-Posure System Version 2 RAD (K002500). The intended use and indications for use are identical. The principal difference in technological features from the predicate devices is the image source. That is, the Sectra Osteoporosis Package analyzes computed radiography ("CR") images obtained directly in digital format. This is in contrast to the predicate devise, which is limited to analyzing either conventional X-ray images alone that must first be separately digitized by using a scanner. This difference in technical characteristics between the subject device and the predicate device does not raise any new questions of safety or effectiveness, because the question of whether the images are of sufficient quality to permit BMD analysis is common to both devices. Also, both devices use essentially the same DXR techniques to analyze the given images. Both in vitro and in vivo performance testing has been conducted to verify that this design difference does not impact either safety or efficacy.
Additionally, the Sectra Osteoporosis Package can receive digital DICOM images that have been stored in a third party's network or PACS system, while the predicate device did not have this option. However, the additional ability to retrieve digital images stored on these other information systems has been validated by the company and in no way affects how the software analyzes BMD or the quality of this analysis.
{2}------------------------------------------------
The other difference is that the Sectra Osteoporosis Package is integrated with a Sectra IDS5 Radiology Workstation giving the operator direct access to the BMD analysis. However, essentially the Sectra Osteoporosis Package is the same as the predicate device, its stand-alone counterpart, and presents no new safety or effectiveness concerns (as discussed above). Furthermore, as validated by the company, the packaging of these systems together in no way impacts the device's safety and effectiveness. Therefore, the Sectra Osteoporosis Package and the predicate device are substantially equivalent.
Peter Ander
Peter Andersson Regulatory Affairs Officer Sectra Imtec AB Teknikringen 2 SE-58330 Linköping Sweden
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image is a black and white logo for the Department of Health & Human Services - USA. The logo consists of a circular border with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" written around it. Inside the circle is an abstract symbol that resembles an eagle or bird in flight, with three curved lines representing its wings.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
JUN = 7, 2002 - 1
Sectra Imtec AB % Mr. Carl Alletto OTech, Inc. 1100 Lakeview Blvd. DENTON TX 76208 Re: K021527
Trade/Device Name: Sectra Osteoporosis Package Bone Densitometer Regulation Number: 21 CFR 892.1170 Regulation Name: Bone Densitometer Regulatory Class: II Product Code: 90 KGI Dated: May 8, 2002 Received: May 10, 2002
Dear Mr. Alletto:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (sections 531-542 of the Act); 21 CFR 1000-1050.
{4}------------------------------------------------
This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at one of the following numbers, based on the regulation number at the top of this letter:
| 8xx.1xxx | (301) 594-4591 |
|----------------------------------|----------------|
| 876.2xxx, 3xxx, 4xxx, 5xxx | (301) 594-4616 |
| 884.2xxx, 3xxx, 4xxx, 5xxx, 6xxx | (301) 594-4616 |
| 892.2xxx, 3xxx, 4xxx, 5xxx | (301) 594-4654 |
| Other | (301) 594-4692 |
Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers. International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html.
Sincerely yours,
Nancy C. Brogdon
Nancy C. Brogdon Director, Division of Reproductive, Abdominal, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
KO21527
Page 1 of -1
510(k) Number:
Device Name: Sectra Osteoporosis Package
Indications for Use:
The device is intended for use to estimate bone mineral density ("BMD") in the forearm and to assess increased risk of osteoporotic fracture according to World Health Organization ("WHO") criteria.
The device is indicated specifically for use to:
- (1) assist the physician in diagnosing subjects who already have been identified to be at risk for suffering from osteoporosis, together with other known risk factors (e.g., prior history of fractures, advanced age, low body weight, lack of physical exercise, lack of exposure to sunlight, insufficient dietary intake of vitamin D and/or calcium, and smoking); and
- (2) compare the BMD estimate with a reference population comprised of young normals and age-matched normals to compute T-scores and Z-scores, respectively.
# (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
| Prescription Use<br>(Per 21 CFR 801.109) | ✓ |
|-----------------------------------------------------------------------------------------|--------------------------|
| OR | |
| Over -The-Counter Use | |
| | (Optional Format 1-2-96) |
| (Division Sign-Off)<br>Division of Reproductive, Abdominal,<br>and Radiological Devices | |
| 510(k) Number | K021527 |
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.