K021522 · Walter Lorenz Surgical, Inc. · DZL · May 21, 2002 · Dental
Device Facts
Record ID
K021522
Device Name
LORENZ 1.5 X 4 MM LACTOSORB TAP FREE SCREW
Applicant
Walter Lorenz Surgical, Inc.
Product Code
DZL · Dental
Decision Date
May 21, 2002
Decision
SESE
Submission Type
Special
Regulation
21 CFR 872.4880
Device Class
Class 2
Attributes
Therapeutic, Pediatric
Indications for Use
The Lactosorb® Tap Free Screw is indicated for use in the following midface and craniofacial procedures. - A. General Indication: trauma procedures of the midface or craniofacial skeleton Specific Indications: - Comminuted fractures of the naso-ethmoidal infraorbital areas 1. - Comminuted fractures of the frontal sinus wall 2. - Pediatric midface or craniofacial trauma 3. - 4. LeFort (I, II, II) fractures - 5. Orbital floor fractures - 6. Fractures of the maxilla, zygoma, zygomatic arch, orbital rim, nasal, ethmoid, and lacrimal bones. - Trauma of the craniofacial skeleton including: frontal, parietal, 7. temporal, sphenoid, and occipital bones. - B. General Indication: reconstructive procedures of the midface or craniofacial Skeleton - 1. Infant craniofacial surgery (i.e. craniosynostosis, congenital malformation, trauma, etc.) - 2. LeFort (1, II, II, II) osteotomies - 3. Tumor reconstruction in midface or craniofacial procedures - 4. Bone graft procedures in the midface or craniofacial skeleton - 5. Pediatric reconstructive procedures - Reconstructive procedures of the craniofacial skeleton including: 6. frontal, parietal, temporal, sphenoid, and occipital bones - 7. Craniotomy flap fixation This system is not designed for use in the mandible and/or full load bearing procedures.
Device Story
Resorbable bone screw; 1.5 mm x 4 mm dimensions; composed of 82% L-Lactide/18% Glycolide copolymer. Device features non-threaded pin portion, head, and helical groove on major diameter. Implanted via rotation into pre-drilled hole smaller than major diameter; fixation achieved through press-fit mechanism. Used by surgeons in clinical settings for midface/craniofacial skeletal fixation. Provides temporary stabilization of bone fragments during healing; material degrades via hydrolysis into lactic and glycolic acids, then metabolized via Krebs cycle. Benefits include elimination of secondary removal surgery due to resorbable nature.
Clinical Evidence
No clinical data provided. Substantial equivalence based on design, material, and functional comparison to predicate devices.
Technological Characteristics
Resorbable copolymer (82% L-Lactide/18% Glycolide); 1.5 mm x 4 mm dimensions; helical groove design; press-fit fixation; non-metallic; sterile.
Indications for Use
Indicated for pediatric and adult patients undergoing midface and craniofacial trauma repair or reconstructive surgery, including craniosynostosis, congenital malformations, tumor reconstruction, bone grafting, and craniotomy flap fixation. Contraindicated for use in the mandible or full load-bearing procedures.
Regulatory Classification
Identification
An intraosseous fixation screw or wire is a metal device intended to be inserted into fractured jaw bone segments to prevent their movement.
Predicate Devices
Lactosorb® 1.5 x 4 mm Direct Drive Screw (K971870)
{0}------------------------------------------------
MAY 2 1 2002
K021520
## Summary of Safety and Effectiveness
## Lorenz Lactosorb® 1.5 mm x 4 mm Tap Free Screw
Walter Lorenz Surgical, Inc. Submitter: 1520 Tradeport Drive Jacksonville, Florida 32218-2480
Establishment Registration: 1032347
Contact Person: Sheryl Malmberg Regulatory Affairs (904)-741-4400
Trade Name: Lorenz 1.5 x 4 mm Lactosorb® Tap Free Screw
Common or Usual Name: Resorbable Bone Screw
Classification Name: Screw, Fixation, Intraosseous (21 CFR 872.4880)
Device Classification: Class II
Device Product Code: DZL
Indications for Use:
The Lactosorb® Tap Free Screw is indicated for use in the following midface and craniofacial procedures.
- A. General Indication: trauma procedures of the midface or craniofacial skeleton Specific Indications:
- Comminuted fractures of the naso-ethmoidal infraorbital areas 1.
- Comminuted fractures of the frontal sinus wall 2.
- Pediatric midface or craniofacial trauma 3.
- 4. LeFort (I, II, II) fractures
- 5. Orbital floor fractures
- 6. Fractures of the maxilla, zygoma, zygomatic arch, orbital rim, nasal, ethmoid, and lacrimal bones.
- Trauma of the craniofacial skeleton including: frontal, parietal, 7. temporal, sphenoid, and occipital bones.
{1}------------------------------------------------
- B. General Indication: reconstructive procedures of the midface or craniofacial Skeleton
- 1. Infant craniofacial surgery (i.e. craniosynostosis, congenital malformation, trauma, etc.)
- 2. LeFort (1, II, II, II) osteotomies
- 3. Tumor reconstruction in midface or craniofacial procedures
- 4. Bone graft procedures in the midface or craniofacial skeleton
- 5. Pediatric reconstructive procedures
- Reconstructive procedures of the craniofacial skeleton including: 6. frontal, parietal, temporal, sphenoid, and occipital bones
- 7. Craniotomy flap fixation
This system is not designed for use in the mandible and/or full load bearing procedures.
Device Description: The device is a screw having a non-threaded pin portion and a head portion. The Lactosorb® Tap Free Screw is designed with a slight helical groove that is cut into the major diameter. The head geometry, drive mechanism, major diameter, and tip geometry of the 1.5 x 4 mm Tap Free Screw is identical to that of the 1.5 x 4 mm Direct Drive Screw (K971870). The only difference between the two screws is the "thread" form.
Although the tap free screws are similar in design to the standard Lactosorb® direct drive screws (K971870), they function more like the Lactosorb® Push Screws (K002423). The push screws are designed with barbs that are pushed into a drilled hole that is slightly smaller than the diameter around the barbs. This press fit holds the push screw is place and provides adequate fixation. The tap-free screws are implanted by rotating the screws into a hole that is slightly smaller than the major diameter of the screws. The press fit between the implant and the hole provides the fixation.
Material: Lactosorb® (resorbable copolymer) - a polyester derivative of lactic and glycolic acids
Lactosorb® is made of 82% L-Lactide/18% Glycolide copolymer that degrades by hydrolysis into L-Lactic and glycolic acids. These hydrolytic products are then further degraded into carbon dioxide and water via the cellular Krebs cycle. Lactosorb® has been previously cleared by 510(k) notifications for use in bone plates (K992355, K992158, K971870, K960988, K955729) and bone screws (K002423, K981666, K960988) for cranial and maxillofacial use.
The modifications made to this device are not anticipated to significantly affect its safety and effectiveness.
## 000048
{2}------------------------------------------------
### DEPARTMENT OF HEALTH & HUMAN SERVICES
21 2002
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
Ms. Sheryl Malmberg Walter Lorenz Surgical, Incorporated 1520 Tradeport Drive Jacksonville, Florida 32218-2480
Re: K021522
Trade/Device Name: Lorenz 1.5 mm x 4.0 mm Lactosorb® Tap Free Screw Regulation Number: 872.4880 Regulation Name: Intraosseous Fixation Screw or Wire Regulatory Class: Class II Product Code: DZL Dated: May 08, 2002 Received: May 10, 2002
Dear Ms. Malmberg:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{3}------------------------------------------------
#### Page 2 - Ms. Sheryl Malmberg
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. or the Fee of all , I with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice and inoms (2) roquiceller as lectronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
This letter will allow you to begin marketing your device as described in your Section 110(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 21 CFR Part 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4613. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dspamain.html
Sincerely yours,
Timothy A. Ulatowski Director Division of Dental, Infection Control, and General Hospital Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
# STATEMENT OF INDICATIONS FOR USE
510(k) Number K021522
Device Name: Lorenz 1.5 mm x 4mm Lactosorb® Tap Free Screw
#### Indications for Use:
The Lactosorb® Tap Free Screw is indicated for use in the following midface and craniofacial procedures.
- A. General Indication: trauma procedures of the midface or craniofacial skeleton Specific Indications:
- 1. Comminuted fractures of the naso-ethmoidal infraorbital areas
- 2. Comminuted fractures of the frontal sinus wall
- 3. Pediatric midface or craniofacial trauma
- 4. LeFort (1, II, II) fractures
- 5. Orbital floor fractures
- 6. Fractures of the maxilla, zygoma, zygomatic arch, orbital rim, nasal, ethmoid, and lacrimal bones.
- 7. Trauma of the craniofacial skeleton including: frontal, parietal, temporal, sphenoid, and occipital bones.
- B. General Indication: reconstructive procedures of the midface or craniofacial Skeleton
Specific Indications:
- 1. Infant craniofacial surgery (i.e. craniosynostosis, congenital malformation, trauma, etc.)
- 2. LeFort (I, II, II) osteotomies
- 3. Tumor reconstruction in midface or craniofacial procedures
- 4. Bone graft procedures in the midface or craniofacial skeleton
- 5. Pediatric reconstructive procedures
- 6. Reconstructive procedures of the craniofacial skeleton including: frontal. parietal, temporal, sphenoid, and occipital bones
- 7. Craniotomy flap fixation
This system is not designed for use in the mandible and/or full load bearing procedures.
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED.)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Prescription Use (Per 21 CFR 801.109)
~~SORRump~~ OR Over-The-Counter-Use
(Division Sign-Off) ~~(Optional Format 1-2-96) ~~
Division of Dental, Infection Control,
and General Hospital Devices
510(k) Number K021522
000006
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.