The Acuson Cypress ultrasound system with the 7L3 transducer, the 2MHz CW transducer, and additional calculations is a modification to the previously cleared Cypress system that has already been cleared by FDA for Fetal, Abdominal, Intra-Operative (Cardiac), Pediatric, Neonatal Cephalic, Cardiac (Adult), Cardiac (Pediatric), Trans-esophageal, Peripheral Vascular, Other (Intra-Luminal), and Other (Intra-Cardiac) Indications. No new intended uses are claimed for the modifications.
Device Story
Acuson Cypress is a compact, portable diagnostic ultrasound system featuring a fold-up keyboard, integrated LCD, and interchangeable transducers. It processes ultrasonic signals to generate real-time images and Doppler audio for clinical assessment. Used in clinical settings by healthcare professionals to visualize internal structures and blood flow; output aids in diagnostic decision-making across various specialties including cardiology and vascular medicine. Modifications include the addition of the 7L3 linear array transducer and an auxiliary 2MHz continuous wave (CW) Doppler transducer, expanding the system's diagnostic capabilities while maintaining existing clinical indications.
Clinical Evidence
Bench testing only. No clinical data provided.
Technological Characteristics
Compact, portable ultrasound platform with fold-up keyboard and integrated LCD. Supports 7L3 linear array and 2MHz CW Doppler transducers. Operates via ultrasonic pulsed echo and pulsed Doppler imaging. Class II device under 21 CFR 892.1550, 892.1560, and 892.1570.
Indications for Use
Indicated for diagnostic ultrasound imaging or fluid flow analysis in fetal, abdominal, intra-operative (cardiac), pediatric, neonatal cephalic, adult/pediatric cardiac, trans-esophageal, and peripheral vascular applications. Prescription use only.
Regulatory Classification
Identification
An ultrasonic pulsed doppler imaging system is a device that combines the features of continuous wave doppler-effect technology with pulsed-echo effect technology and is intended to determine stationary body tissue characteristics, such as depth or location of tissue interfaces or dynamic tissue characteristics such as velocity of blood or tissue motion. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.
{0}------------------------------------------------
K021497
# Section 8
# 510(K) SUMMARY OF SAFETY AND EFFECTIVENESS ACUSON CORPORATION'S CYPRESS ULTRASOUND SYSTEM
Acuson Corporation has not disclosed its intent to market this device modification and requests this notification be held CONFIDENTIAL by FDA, and not be released to any Freedom of Information request or addressed with any outside parties.
| Sponsor: | Acuson Corporation<br>1220 Charleston Road<br>Mountain View, CA 94039-7303 |
|------------------|--------------------------------------------------------------------------------------------------------|
| Contact Person: | Bob Leiker<br>Senior Regulatory Affairs Specialist<br>Telephone: (650) 694-5080<br>Fax: (650) 961-6158 |
| Submission Date: | May 8, 2002 |
Acuson Cypress System Device Name:
Diagnostic ultrasound system Common or Usual Name:
#### Classification:
Ultrasonic Pulsed Doppler Imaging System (90 IYN), class II (21 C.F.R. §892.1550) Ultrasonic Pulsed Echo Imaging System (90 IYO), class II (21 C.F.R. §892.1560) Diagnostic Ultrasound Transducer (90 ITX), class II (21 C.F.R. §892.1570)
## Predicate Devices:
K010950, 6/27/2001, cleared as the Acuson Cypress Ultrasound System.
K982800, 9/22/98, cleared as the Ecton Lynx Ultrasound System, marketed as the Acuson Cypress Ultrasound System with subsequent modifications.
K991872. 6/16/99, cleared as Ecton Lynx Ultrasound System, marketed Acuson Cypress Ultrasound System with subsequent modifications.
#### Device Description:
The Acuson Cypress Ultrasound System is an ultrasound imaging platform which is designed for use with a variety of internal and external transducers. The Acuson Cypress System is a compact and portable diagnostic ultrasound system with a fold-up keyboard, an integrated LCD type display, and interchangeable electronic transducers. The user interface includes a keyboard, an intuitive layout of specialized controls, color GUI display and Doppler audio.
{1}------------------------------------------------
#### Intended Use:
The Acuson Cypress ultrasound system with the 7L3 transducer, the 2MHz CW transducer, and additional calculations is a modification to the previously cleared Cypress system that has already been cleared by FDA for Fetal, Abdominal, Intra-Operative (Cardiac), Pediatric, Neonatal Cephalic, Cardiac (Adult), Cardiac (Pediatric), Trans-esophageal, Peripheral Vascular, Other (Intra-Luminal), and Other (Intra-Cardiac) Indications. No new intended uses are claimed for the modifications.
#### Technological Characteristics and Substantial Equivalence:
The modified Cypress System is substantially equivalent to the predicate devices with respect to intended use and indications for use, principles of operation, and technological characteristics of design construction, and materials. It is comparable in key safety and effectiveness features.
{2}------------------------------------------------
Image /page/2/Picture/1 description: The image shows the seal of the U.S. Department of Health & Human Services. The seal features a stylized caduceus, a symbol often associated with medicine and healthcare. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" are arranged around the symbol in a circular fashion. The seal is black and white.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
9 2002 JUL
Mr. Bob Leiker Senior Regulatory Affairs Specialist Acuson Corporation 1220 Charleston Road MOUNTAIN VIEW CA 94039
Re: K021497
Trade Name: The Acuson Cypress System Regulation Number: 21 CFR 892.1550 Regulation Name: Ultrasonic pulsed doppler imaging system Regulation Number: 21 CFR 892.1560 Regulation Name: Ultrasound pulsed echo imaging system Regulation Number: 21 CFR 892.1570 Regulaton Name: Diagnostic ultrasound transducer Regulatory Class: II Product Code: 90 IYN, IYO and ITX Dated: June 11, 2002 Received: June 12, 2002
#### Dear Mr. Leiker:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to Mav 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
This determination of substantial equivalence applies to the following transducers intended for use with the Acuson Cypress System, as described in your premarket notification:
Transducer Model Number
6 MHz Linear Array External Model 7L3 2 MHz Continuous Wave Doppler External, Model Auxiliary CW
{3}------------------------------------------------
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
This determination of substantial equivalence is granted on the condition that prior to shipping the first device, you submit a postclearance special report. This report should contain complete information, including acoustic output measurements based on production line devices, requested in Appendix G, (enclosed) of the Center's September 30, 1997 "Information for Manufacturers Seeking Marketing Clearance of Diagnostic Ultrasound Systems and Transducers." If the special report is incomplete or contains unacceptable values (e.g., acoustic output greater than approved levels), then the 510(k) clearance may not apply to the production units which as a result may be considered adulterated or misbranded.
The special report should reference the manufacturer's 510(k) number. It should be clearly and prominently marked "ADD-TO-FILE" and should be submitted in duplicate to:
> Food and Drug Administration Center for Devices and Radiological Health Document Mail Center (HFZ-401) 9200 Corporate Boulevard Rockville, Maryland 20850
This letter will allow you to begin marketing your device as described in your premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus permits your device to proceed to market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801, please contact the Office of Compliance at (301) 594-4591. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or at (301) 443-6597 or at its Internet address "http://www.fda.gov/cdrh/dsmamain.html".
{4}------------------------------------------------
Page 3 – Mr. Leiker
If you have any questions regarding the content of this letter, please contact Rodrigo C. Perez at (301) 594-1212.
Sincerely yours,
Diriel A. dozerson
for Nancy C. Brogdon
Director, Division of Reproductive, Abdominal and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure(s)
{5}------------------------------------------------
# Diagnostic Ultrasound Indications for Use For The Acuson Cypress System
Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
| Mode of Operation | | | | | | | | | | |
|-----------------------------------------------------------------------------------|---|---|---|-----|-----|------------------|---------------------------------|------------------------------|-----------------------|--------------------------------|
| Clinical Application | A | B | M | PWD | CWD | Color<br>Doppler | Power<br>(Amplitude)<br>Doppler | Color<br>Velocity<br>Imaging | Combined<br>(Specify) | (Other)<br>Harmonic<br>Imaging |
| Opthalmic | | | | | | | | | | |
| Fetal | | P | P | P | P | P | P | | | P |
| Abdominal | | P | P | P | P | P | P | | | P |
| IntraOperative<br>(Cardiac) | | P | P | P | P | P | P | | | P |
| IntraOperative<br>Neurological | | | | | | | | | | |
| Pediatric | | P | P | P | P | P | P | | | P |
| Small Organ<br>(Specify) | | | | | | | | | | |
| Neonatal Cephalic | | P | P | P | P | P | P | | | P |
| Adult Cephalic | | | | | | | | | | |
| Cardiac (Adult) | | P | P | P | P | P | P | | | P |
| Cardiac (Pediatric) | | P | P | P | P | P | P | | | P |
| Transesophageal | | P | P | P | P | P | P | | | P |
| Trans-Rectal | | | | | | | | | | |
| Trans-Vaginal | | | | | | | | | | |
| Trans-Urethral | | | | | | | | | | |
| Intra-Vascular | | | | | | | | | | |
| Peripheral Vascular | | P | P | P | P | P | P | | | P |
| Laparascopic | | | | | | | | | | |
| Muscular-Skeletal<br>Conventional | | | | | | | | | | |
| Muscular-Skeletal<br>Superficial | | | | | | | | | | |
| Other (Intra-Luminal) | | P | P | P | P | P | P | | | P |
| Other (Intra-Cardiac) | | P | P | P | P | P | P | | | P |
| N = new indication<br>P = previously cleared by FDA<br>E = added under Appendix E | | | | | | | | | | |
Mode of Oneration
E = added under Appendix E
Additional Comments:
Concurrence of CDRH, Office of Device Evaluation (ODE)
Prescription Use (Per 21 CFR 801.109)
Daniel A. Bergman
Division Sign Off
(Division Sign-Off) Division of Reproductive, Abdominal and Radiological Deyic 510(k) Number .
{6}------------------------------------------------
# Diagnostic Ultrasound Indications for Use For The Acuson Cypress System
#### 6 MHz Linear Array External Transducer, Model Name: 7L3
Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
| Mode of Operation | | | | | | | | | | |
|-----------------------------------|---|---|---|-----|-----|------------------|---------------------------------|------------------------------|-----------------------|--------------------------------|
| Clinical Application | A | B | M | PWD | CWD | Color<br>Doppler | Power<br>(Amplitude)<br>Doppler | Color<br>Velocity<br>Imaging | Combined<br>(Specify) | (Other)<br>Harmonic<br>Imaging |
| Opthalmic | | | | | | | | | | |
| Fetal | | N | N | N | | N | N | | | N |
| Abdominal | | N | N | N | | N | N | | | N |
| IntraOperative<br>(Cardiac) | | N | N | N | | N | N | | | N |
| IntraOperative<br>Neurological | | | | | | | | | | |
| Pediatric | | N | N | N | | N | N | | | N |
| Small Organ<br>(Specify) | | | | | | | | | | |
| Neonatal Cephalic | | | | | | | | | | |
| Adult Cephalic | | | | | | | | | | |
| Cardiac (Adult) | | N | N | N | | N | N | | | N |
| Cardiac (Pediatric) | | N | N | N | | N | N | | | N |
| Transesophageal | | | | | | | | | | |
| Trans-Rectal | | | | | | | | | | |
| Trans-Vaginal | | | | | | | | | | |
| Trans-Urethral | | | | | | | | | | |
| Intra-Vascular | | | | | | | | | | |
| Peripheral Vascular | | N | N | N | | N | N | | | N |
| Laparascopic | | | | | | | | | | |
| Muscular-Skeletal<br>Conventional | | | | | | | | | | |
| Muscular-Skeletal<br>Superficial | | | | | | | | | | |
| Other (Intra-Luminal) | | | | | | | | | | |
| Other (Intra-Cardiac) | | | | | | | | | | |
Mode of Operation
N = new indication
P = previously cleared by FDA
E = added under Appendix E
Additional Comments: The transducer being added is new, but the indications are not new to the Cypress Ultrasound System.
Concurrence of CDRH, Office of Device Evaluation (ODE)
(Division Sige 00
(Division Sign-Off)
Division of Reproductive, Abdominal,
and Radiological Devices
510(k) Number K021497
Prescription Use (Per 21 CFR 801.109)
,
{7}------------------------------------------------
## Diagnostic Ultrasound Indications for Use For The Acuson Cypress System
## 2 MHz Continuous Wave Doppler External Transducer, Model Name: Auxiliary CW
Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
| Mode of Operation | | | | | | | | | | |
|-----------------------------------|---|-------------------------------|---|----------------------------|-----|------------------|---------------------------------|------------------------------|-----------------------|--------------------------------|
| Clinical Application | A | B | M | PWD | CWD | Color<br>Doppler | Power<br>(Amplitude)<br>Doppler | Color<br>Velocity<br>Imaging | Combined<br>(Specify) | (Other)<br>Harmonic<br>Imaging |
| Opthalmic | | | | | | | | | | |
| Fetal | | | | | N | | | | | |
| Abdominal | | | | | N | | | | | |
| IntraOperative<br>(Cardiac) | | | | | | | | | | |
| IntraOperative<br>Neurological | | | | | | | | | | |
| Pediatric | | | | | N | | | | | |
| Small Organ<br>(Specify) | | | | | | | | | | |
| Neonatal Cephalic | | | | | | | | | | |
| Adult Cephalic | | | | | | | | | | |
| Cardiac (Adult) | | | | | N | | | | | |
| Cardiac (Pediatric) | | | | | N | | | | | |
| Transesophageal | | | | | | | | | | |
| Trans-Rectal | | | | | | | | | | |
| Trans-Vaginal | | | | | | | | | | |
| Trans-Urethral | | | | | | | | | | |
| Intra-Vascular | | | | | | | | | | |
| Peripheral Vascular | | | | | N | | | | | |
| Laparascopic | | | | | | | | | | |
| Muscular-Skeletal<br>Conventional | | | | | | | | | | |
| Muscular-Skeletal<br>Superficial | | | | | | | | | | |
| Other (Intra-Luminal) | | | | | | | | | | |
| Other (Intra-Cardiac) | | | | | | | | | | |
| N = new indication | | B = previously cleared by FDA | | E = added under Appendix E | | | | | | |
Moda of Oneration
Additional Comments: The transducer being added is new, but the indications are not new to the Cypress Ultrasound System.
Concurrence of CDRH, Office of D
H, Office of Device Evaluation (ODE)
(Division Sign-Off)
Prescription Use (Per 21 CFR 801.109)
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.