The REPIPHYSIS™ prosthesis is indicated for cemented procedures where radical resection and replacement of the distal femur and/or proximal tibia is required with the following conditions: 1. patients suffering from severe arthropathy of the knee that does not respond to any conservative therapy or better alternative surgical treatment; 2. surgical intervention for severe trauma, revision knee arthroplasties, and/or oncology indications. 3. metastatic diseases (e.g., osteosarcomas, chondrosarcomas, gaint cell tumors, bone tumors)
Device Story
REPIPHYSIS™ Limb Salvage System is an expandable prosthesis for pediatric patients with malignant bone tumors to maintain limb length equality. System consists of distal hinge femur, tibial base, axial bushing, and axial pin. Implant features internal spring mechanism maintained by polyacetal locking component. Lengthening is non-invasive: external electromagnetic transmitter induces current in internal antenna; heat softens polyacetal lock; spring expands tubes. Procedure performed in fluoroscopy suite; patient awake; physician identifies antennae, applies transmitter for ~15 seconds. Radiographs verify 0.5-1.0 cm expansion. Benefits include avoiding repeated major surgeries for limb lengthening.
Clinical Evidence
No clinical data provided. Substantial equivalence supported by materials data and testing results.
Technological Characteristics
Prosthesis components include distal hinge femur, tibial base, axial bushing, and axial pin. Expansion mechanism utilizes polyacetal locking component and internal antenna. External electromagnetic transmitter used for non-invasive activation. Materials are standard, non-allergic, and tested for biological safety.
Indications for Use
Indicated for patients requiring radical resection and replacement of distal femur/proximal tibia due to severe knee arthropathy, severe trauma, revision knee arthroplasty, or oncology indications (e.g., osteosarcoma, chondrosarcoma, giant cell tumors).
Regulatory Classification
Identification
A knee joint femorotibial metal/polymer constrained cemented prosthesis is a device intended to be implanted to replace part of a knee joint. The device limits translation or rotation in one or more planes and has components that are linked together or affined. This generic type of device includes prostheses composed of a ball-and-socket joint located between a stemmed femoral and a stemmed tibial component and a runner and track joint between each pair of femoral and tibial condyles. The ball-and-socket joint is composed of a ball at the head of a column rising from the stemmed tibial component. The ball, the column, the tibial plateau, and the stem for fixation of the tibial component are made of an alloy, such as cobalt-chromium-molybdenum. The ball of the tibial component is held within the socket of the femoral component by the femoral component's flat outer surface. The flat outer surface of the tibial component abuts both a reciprocal flat surface within the cavity of the femoral component and flanges on the femoral component designed to prevent distal displacement. The stem of the femoral component is made of an alloy, such as cobalt-chromium-molybdenum, but the socket of the component is made of ultra-high molecular weight polyethylene. The femoral component has metallic runners which align with the ultra-high molecular weight polyethylene tracks that press-fit into the metallic tibial component. The generic class also includes devices whose upper and lower components are linked with a solid bolt passing through a journal bearing of greater radius, permitting some rotation in the transverse plane, a minimal arc of abduction/adduction. This generic type of device is limited to those prostheses intended for use with bone cement (§ 888.3027).
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo for Wright Medical Technology, Inc. The logo consists of a stylized "W" made up of three black triangles, followed by the company name in bold, sans-serif font. Below the company name is the address "5677 Airline Road, Arlington, TN 38002" and the phone number "901-867-9971".
K021489
Page 1 of 4
# DEC (1 4 2002
# 510(K) SUMMARY OF SAFETY AND EFFECTIVENESS
In accordance with the Food and Drug Administration Rule to implement provisions of the Safe Medical Devices Act of 1990 and in conformance with 21 CRF 807, this information serves as a Summary of Safety and Effectiveness for the use of the REPIPHYSIS™ Limb Salvage System.
| Submitted By: | Wright Medical Technology, Inc. |
|-------------------------------------|----------------------------------------------------------------------------------------------------|
| Date: | May 7, 2002 |
| Contact Person: | Ehab M. Esmail<br>Manager, Regulatory Affairs |
| Proprietary Name: | REPIPHYSIS™ Limb Salvage System |
| Common Name: | Limb Salvage System |
| Classification Name and Reference: | 21 CFR 888.3510 Prosthesis, Knee, Femorotibial,<br>Constrained, Cemented, Metal/Polymer – Class II |
| Device Product Code and Panel Code: | Orthopedics/87/KRO |
### DEVICE INFORMATION
### A. INTENDED USE
The REPIPHYSIS™ prosthesis is indicated for cemented procedures where radical resection and replacement of the distal femur and/or proximal tibia is required with the following conditions:
- 1. patients suffering from severe arthropathy of the knee that does not respond to any conservative therapy or better alternative surgical treatment;
- 2. surgical intervention for severe trauma, revision knee arthroplasties, and/or oncology indications.
- 3. metastatic diseases (e.g., osteosarcomas, chondrosarcomas, gaint cell tumors, bone tumors)
The REPIPHYSIS™ Limb Salvage System components are for single use only.
Image /page/0/Picture/13 description: The image shows the logo for Wright Medical Technology, Inc. The logo features a stylized "W" symbol above the word "WRIGHT" in a bold, serif typeface. Below the name is the text "MEDICAL TECHNOLOGY, INC." in a smaller, sans-serif font. The overall design is simple and professional.
Image /page/0/Picture/14 description: The image shows a black and white drawing of the Earth. The Earth is partially illuminated, with the continents of Australia and Asia visible. There is a curved line above the Earth, which may represent the atmosphere or a satellite orbit.
{1}------------------------------------------------
K02148a
Page 2 of 4
#### B. DEVICE DESCRIPTION
#### BACKROUND DESCRIPTION
Malignant bone tumors in children often occur in the long bones, at or near the joints (shoulder, hip, and knee). Removal of the tumor often requires removal of part or the entire joint. Because the joint is where the growth of the limb occurs, one of the major problems in saving the limb in young growing children involves the ability to maintain equal lengths of the operated arm or leg with the unoperated side. Many attempts have been made to design a prosthesis, which could be easily expanded either without surgerv or through a very minor surgical procedure. Most of these designs have had mechanical difficulty or the lengthening procedure becomes a major surgical effort. The REPIPHYSISTM Limb Salvage System the complications that are known to occur with repeated operations. Once the prosthesis is fully implanted, the noninvasive lengthening of the prosthesis is activated by an external electromagnetic field.
The REPIPHYSISTM Limb Salvage System (Wright Medical Technology, Arlington, Tennessee) was previously named the Phenix Growing Prosthesis (Phenix Medical, Paris, France). The Phenix Growing Prosthesis was designed in France in the early 1980s. More than 100 patients have been treated with Phenix Growing Prosthesis in the last 20 years.
#### IMPLANT DESCRIPTION
The REPIPHYSIS™ Prosthesis is custom assembled for each patient. The REPIPHYSIS™ is implanted the same way as any other joint replacement device using non-allergic, standard materials that have been tested in biological situations. The REPIPHYSIS™ Limb Salvage Prosthesis Kit consists of the following components:
#### Distal Hinge Femur
The femoral housing is the main body of the distal hinge femoral portion of the implant. The most distal end of the femoral housing forms the outer components of the femoraltibial hinge. It internally houses the following various components of the expansion mechanism.
#### Tibial Base
The tibial base provides fixation to the proximal tibia. It consists of a cylindrical stem portion, which extends into the proximal tibial medullary canal. The more proximal portion of the tibial base has a flat plate that rests on the patient's tibial plateau region. On the underside of this plate are two small tabs that protrude down into the tibial plateau for rotational stability. On the top of this plate is the inner component of the femoral-tibial hinge mechanism.
#### Axial Bushing
The axial bushing goes inside the center hinge portion of the tibial base. This bushing is held in place by the two outer hinge features of the femoral housing. This bushing supplies the wear surface through which forces are transmitted between the femoral and tibial components and around which rotation occurs.
Image /page/1/Picture/13 description: The image shows the logo for Wright Medical Technology, Inc. The logo features a stylized "W" symbol above the word "WRIGHT" in a bold, sans-serif font. Below "WRIGHT" is the text "MEDICAL TECHNOLOGY, INC." in a smaller, sans-serif font, underlined by a thin line.
Image /page/1/Picture/14 description: The image shows a black and white drawing of a celestial body, possibly a planet or moon. The surface has a rough texture with dark and light patches, suggesting craters or other geological features. A curved line at the top indicates the edge of the sphere, and a small, dark shape is visible at the bottom, possibly a shadow or another object.
-510(K) SUMMARY
{2}------------------------------------------------
K021489
#### Axial Pin
The axial pin is introduced through the outer hinge portion of the distal femoral housing and into the axial bushing, which is contained in the center hinge portion of the tibial base. It extends through the other side of the outer hinge mechanism of the femoral housing. The axial pin mates with the outer hinge portions of the distal femoral housing and the inner diameter of the axial bushing. The axial pin is the pivot for the femoraltibial hinge mechanism.
#### INSTRUMENT DESCRIPTION
The REPIPHYSIS™ prosthesis is implanted the same way as any other Limb Salvage System.
In addition the REPIPHYSIS™ prosthesis uses an external electro-magnetic transmitter to activate the implant lengthening mechanism. The REPIPHYSIS Transmitter Unit operates by producing an electrical current in an inductor coil, which in turn generates a strong electro-magnetic field in the vicinity of the coil. When an electrically conductive material is placed within this electro-magnetic field, an electrical current is induced in the conductive material. If the proper conditions are established, this electrical current is sufficient to cause significant heating in this material.
### OPERATION DESCRIPTION
The REPIPHYSIS™ prosthesis is implanted the same way as any other Limb Salvage System.
The lengthening technique for the REPIPH YSIS™ prosthesis is briefly described below:
The REPIPHYSIS™ prosthesis consists of two tubes with a spring mechanism contained in the larger tube (Exhibit 11: Photo of the Femoral Housing and sub-components). It is maintained compressed by a polyacetal locking mechanism. One tube is connected to the stem of the implant, and the second comprises the hinge portion of the hinge prosthesis. The uninvolved side of the joint is minimally resurfaced with a press-fit stem to attempt to preserve function of the non-involved growth plate. Expansion is achieved via exposure to an external electromagnetic field around the extremity. The coil transmits an electromagnetic field around the implant in the vicinity of the v-shaped flange of the spring housing tube. This generates an electrical current within the flange of the spring housing tube and is focused on the receiving antennae within the implant itself. This antenna is heated by the electromagnetic field. The heated element softens the surrounding polyacetal locking mechanism, which allows the spring expansion. The spring expansion pushes the two tubes apart from each other, thus lengthening the overall length of the leg. As skeletal growth occurs on the contralateral side, a discrepancy in the overall length of the two legs occurs. Follow-up scanograms quantify the discrepancy. Once a .5- to 1.0-cm discrepancy has occurred, the patient is taken to the fluoroscopy suite and remains awake without sedation or anesthesia. The receiving antennae within the implant are identified under fluoroscopy guidance and the skin is marked. The electromagnetic device is then applied over the receiving antennae, and the patient
Image /page/2/Picture/10 description: The image shows the logo for Wright Medical Technology, Inc. The logo features a stylized "W" symbol above the word "WRIGHT" in a bold, sans-serif font. Below the word "WRIGHT" is the text "MEDICAL TECHNOLOGY, INC." in a smaller, sans-serif font. A thin line underlines the text "MEDICAL TECHNOLOGY, INC."
Image /page/2/Picture/11 description: The image shows a black and white drawing of a planet. The planet has a rough, textured surface with dark and light areas. There is a ring around the planet, and a small object is visible below it. The drawing is simple and lacks detail, giving it a somewhat abstract appearance.
{3}------------------------------------------------
K021489
Page 4 of 4
activates the device for a period of approximately 15 seconds. A radiograph using magnification markers is then obtained of the level of the spring to determine how much lengthening has occurred. Once the desired amount of lengthening (typically 0.5 to 1cm) is achieved, the process is stopped. One or two activation processes are required to achieve the desired lengthening. The procedure is halted if the patient experiences any discomfort or anxiety. Follow-up radiographs are scheduled at 2 to 3-month intervals until the next lengthening is required.
### C. SUBSTANTIAL EQUIVALENCE INFORMATION
The intended use, material, type of interface, and design features of REPIPHYSISTM Limb Salvage System are substantially equivalent to the competitive devices previously cleared for market. The safety and effectiveness of the REPIPHYSIS™ Limb Salvage System are adequately supported by the substantial equivalence information, materials data, and testing results provided within the Premarket Notification.
Image /page/3/Picture/4 description: The image shows the logo for Wright Medical Technology, Inc. The logo features the word "WRIGHT" in large, bold, sans-serif font. Above the word "WRIGHT" is a stylized "W" symbol. Below the word "WRIGHT" are the words "MEDICAL TECHNOLOGY, INC." in a smaller font.
{4}------------------------------------------------
## DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/4/Picture/1 description: The image shows the seal of the Department of Health & Human Services (HHS) of the United States. The seal features a stylized caduceus, a symbol often associated with medicine and healthcare, with a double helix-like design. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" is arranged in a circular pattern around the caduceus.
#### Public Health Service
DEC 04 2002
200 Cornorate Roulevard
Ehab M. Esmail Manager, Regulatory Affairs Wright Medical Technology, Inc. 5677 Airline Road · Arlington, Tennessee 38002
#### Re: K021489
Trade/Device Name: REPIPHYSIS™ Limb Salvage System Regulation Number: 21 CFR 888.3510 Regulation Name: Knee joint femorotibial metal/polymer constrained cemented prosthesis Regulatory Class: Class II Product Code: KRO Dated: November 18, 2002 Received: November 19, 2002
Dear Mr. Esmail:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You mav, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{5}------------------------------------------------
Page 2 – Mr. Ehab M. Esmail
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (301) 594-4659. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594 4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html
Sincerely vours.
Celia M. Witten, Ph.D., M.D. Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{6}------------------------------------------------
Image /page/6/Picture/0 description: The image shows a logo with a large letter W on the left side. To the right of the W is the text "WRIG" on the first line and "MEDICAL TECHN" on the second line. Below that is the address "5677 AIRLINE" and "ARLINGTON,," followed by the phone number "901-867".
K021489
Page 1 of 1
# REPIPHYSIS™ Limb Salvage System
# INDICATIONS STATEMENT
Limb salvage surgery is indicated for cemented procedures where radical resection and replacement of the distal femur and/or proximal tibia is required with the following conditions:
- 1. patients suffering from severe arthropathy of the knee that does not respond to any conservative therapy or better alternative surgical treatment;
- 2. surgical intervention for severe trauma, revision knee arthroplasties, and/or oncology indications.
- metastatic diseases (e.g., osteosarcomas, chondrosarcomas, gaint cell tumors, bone 3. tumors)
for Mark A. Millburn
(Division Sign-Off)
Division of General. Restorative and Neurological Devices
510(k) Number K031489
Image /page/6/Picture/11 description: The image shows the logo for Wright Medical Technology, Inc. The logo features a stylized "W" symbol above the word "WRIGHT" in a bold, sans-serif font. Below the word "WRIGHT" are the words "MEDICAL TECHNOLOGY, INC." in a smaller, sans-serif font.
Image /page/6/Picture/12 description: The image shows a black and white photograph of a person's head and shoulders. The person appears to be wearing a dark garment, and their face is partially obscured by shadows. The background is blurred, but it seems to be an indoor setting. The image has a grainy texture, which suggests that it may have been taken with an older camera or that it has been digitally altered.
-INDICATIONS STATEMENT
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.