K021486 · Focus Technologies, Inc. · MXJ · Aug 1, 2002 · Microbiology
Device Facts
Record ID
K021486
Device Name
HERPESELECT 2 ELISA IGG, MODEL ELO920G
Applicant
Focus Technologies, Inc.
Product Code
MXJ · Microbiology
Decision Date
Aug 1, 2002
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 866.3305
Device Class
Class 2
Indications for Use
Focus Technologies' HerpeSelect® 2 ELISA IgG test is intended for qualitatively detecting the presence or absence of human IgG class antibodies to HSV-2 in human sera. In conjunction with the Focus HerpeSelect® I ELISA IgG, the test is indicated for testing sexually active adults or expectant mothers for aiding in the presumptive digenosis of HSV infection. The assay can be used manually or in conjunction with an automated system as outlined in the package insert. The user is responsible for assay performance characteristics when an automated system is used. The performance of this assay has not been established for use in a pediatric population, for neonatal screening, or for testing of immunocompromised patients.
Device Story
HerpeSelect® 2 ELISA IgG is an in vitro diagnostic assay for detecting HSV-2 IgG antibodies. Polystyrene microwells are coated with recombinant gG-2 antigen; diluted serum samples and controls are incubated to allow antibody-antigen binding. After washing, peroxidase-conjugated anti-human IgG is added, followed by enzyme substrate and chromogen to produce a color change. The reaction is stopped and quantified via spectrophotometric optical density (OD) measurement. The device is used in clinical laboratories, either manually or with an automated system. Healthcare providers use the resulting OD readings, compared against reference cut-off values, to determine the presence or absence of HSV-2 antibodies. This aids in the presumptive diagnosis of HSV infection, supporting clinical decision-making for sexually active adults and expectant mothers.
Clinical Evidence
Clinical performance evaluated using archived, masked sera from sexually active adults (n=246) and expectant mothers (n=241) compared to Western blot reference. Sexually active adults: 96.1% sensitivity, 97.0% specificity. Expectant mothers: 100% sensitivity, 96.1% specificity. Additional studies included culture-positive patients (n=63), a low-prevalence population (n=81), and cross-reactivity testing with taxonomically related viruses (CMV, EBV, HHV6, VZV). Reproducibility studies (intra-assay, inter-assay, inter-lot, inter-laboratory) and manual vs. automated method agreement (98.1% overall) were conducted.
Technological Characteristics
Polystyrene microwells coated with recombinant gG-2 antigen. Enzyme-linked immunosorbent assay (ELISA) principle using peroxidase-conjugated anti-human IgG. Detection via spectrophotometric optical density (OD) measurement. Compatible with manual or automated laboratory instrumentation. No specific ASTM material standards cited.
Indications for Use
Indicated for qualitative detection of human IgG antibodies to HSV-2 in human sera. Used in conjunction with HerpeSelect® 1 ELISA IgG to aid in presumptive diagnosis of HSV infection in sexually active adults and expectant mothers. Not indicated for pediatric populations, neonatal screening, or immunocompromised patients.
Regulatory Classification
Identification
Herpes simplex virus serological assays are devices that consist of antigens and antisera used in various serological tests to identify antibodies to herpes simplex virus in serum. Additionally, some of the assays consist of herpes simplex virus antisera conjugated with a fluorescent dye (immunofluorescent assays) used to identify herpes simplex virus directly from clinical specimens or tissue culture isolates derived from clinical specimens. The identification aids in the diagnosis of diseases caused by herpes simplex viruses and provides epidemiological information on these diseases. Herpes simplex viral infections range from common and mild lesions of the skin and mucous membranes to a severe form of encephalitis (inflammation of the brain). Neonatal herpes virus infections range from a mild infection to a severe generalized disease with a fatal outcome.
Special Controls
*Classification.* Class II (special controls). The device is classified as class II (special controls). The special control for the device is FDA's revised guidance document entitled “Class II Special Controls Guidance Document: Herpes Simplex Virus Types 1 and 2 Serological Assays.” For availability of the guidance revised document, see § 866.1(e).
Predicate Devices
HerpeSelect®2 ELISA IgG (manual option)
Submission Summary (Full Text)
{0}------------------------------------------------
1 2002
Image /page/0/Picture/1 description: The image shows the logo for Focus Technologies. The word "FOCUS" is written in large, bold, sans-serif letters. Below the word "FOCUS" is the word "technologies" written in a smaller, sans-serif font. The two crescent moon shapes are placed in the middle of the word FOCUS, replacing the letter "O".
510(k) Summary of Safety and Effectiveness HerpeSelect®2 ELISA IgG Catalog No. EL0920G Prepared July 23, 2002 Page 1 of 9
021486
| Applicant | Focus Technologies, Inc.<br>10703 Progress Way<br>Cypress, California 90630<br>USA |
|-----------------------------------|-------------------------------------------------------------------------------------------------|
| Establishment<br>Registration No. | 2023365 |
| Contact Person | Michael J. Wagner, Esq.<br>tel (714) 220-1900<br>fax (714) 995-6921<br>mwagner@focusanswers.com |
| Summary Date | July 23, 2002 |
| Proprietary Name | HerpeSelect®2 ELISA IgG (automated option) |
| Generic Name | HSV-2 ELISA IgG |
| Classification | Herpes Simplex Virus Serological Reagents<br>21 CFR §866.3305<br>Class III |
| Predicate Device | HerpeSelect®2 ELISA IgG (manual option) |
### Device Description
In the HerpeSelect 2 ELISA IgG assay, the polystyrene microwells are coated with recombinant gG-2 antigen. Diluted serum samples and controls are incubated in the wells to allow specific antibody present in the samples to react with the antigen. Nonspecific reactants are removed by washing, and peroxidase-conjugated anti-human IgG is added and reacts with specific IgG. Excess conjugate is removed by washing. Enzyme substrate and chromogen are added, and the color is allowed to develop. After adding the Stop Reagent, the resultant color change is quantified by a spectrophotometric reading of optical density (OD). Sample optical density readings are compared with reference cut-off OD readings to determine results.
#### Intended Use
Focus Technologies' HerpeSelect® 2 ELISA IgG test is intended for qualitatively detecting the presence or absence of human IgG class antibodies to HSV-2 in human sera. In conjunction with the Focus HerpeSelect® I ELISA IgG, the test is indicated for testing sexually active adults or expectant mothers for aiding in the presumptive digenosis of HSV infection. The assay can be used manually or in conjunction with an automated system as outlined in the package insert. The user is responsible for assay performance characteristics when an automated system is used. The performance of this assay has not been established for use in a pediatric population, for neonatal screening, or for testing of immunocompromised patients.
{1}------------------------------------------------
Image /page/1/Picture/1 description: The image shows the logo for Focus Technologies. The logo consists of the word "FOCUS" in a bold, sans-serif font, with a stylized crescent moon shape incorporated into the "O". Below the word "FOCUS" is the word "technologies" in a smaller, sans-serif font with more spacing between the letters. The logo is black and white.
510(k) Summary of Safety and Effectiveness HerpeSelect®2 ELISA IgG Catalog No. EL0920G Prepared July 23, 2002 Page 2 of 9
## EXPECTED VALUES
An outside investigator assessed the device with masked, archived and unselected sera from 1) sexually active adults over the age of 14 (n = 246), and 2) from expectant mothers (n = 241). The reference method was a HSV-2 Western blot from a Pacific Northwest university. The observed prevalences and the hypothetical predictive values for the two populations are shown in the tables below. The positive value will decrease proportionally to the prevalence of HSV infection as reflected in the table below. The calculations are based on HerpeSelect 2 ELISA IgG having
I) a hypothetical sensitivity of 96.1% & a hypothetical specificity of 97.0% (sexually active adults), and
2) a hypothetical sensitivity of 100% and a hypothetical specificity of 96.1% (expectant mothers).
| Population | HSV-2<br>Serostatus | Observed Prevalence | |
|--------------------------|---------------------|---------------------|-------------|
| | | WB | Focus ELISA |
| Sexually Active Adults * | neg | 68.5% | 67.2% |
| | + | 31.5% | 32.4% |
| Expectant Mothers † | neg | 75.6% | 72.3% |
| | + | 24.4% | 27.3% |
### Observed Prevalence with Sexually Active Adults & Expectant Mothers
* Excludes 5 atypical Western blots and 1 ELISA equivocal.
† Excludes 3 atypical Western blots and 1 ELISA equivocal.
| Prevalence vs. Hypothetical Predictive Values | | | | |
|-----------------------------------------------|------------------------|-------|-------------------|-------|
| Prevalence | Sexually Active Adults | | Expectant Mothers | |
| | PPV | NPV | PPV | NPV |
| 50% | 97.0% | 97.0% | 96.2% | 96.1% |
| 40% | 95.5% | 98.0% | 94.5% | 97.4% |
| 30% | 93.2% | 98.7% | 91.7% | 98.3% |
| 25% | 91.4% | 99.0% | 89.5% | 98.7% |
| 20% | 88.9% | 99.2% | 86.5% | 99.0% |
| 15% | 85.0% | 99.5% | 81.9% | 99.3% |
| 10% | 78.1% | 99.7% | 74.0% | 99.6% |
| 5% | 62.8% | 99.8% | 57.4% | 99.8% |
## Prevalence vs. Hynothetical Predictive Values
Note: Sexually active adult and expectant mother populations in different geographic areas may produce different frequency distributions from the table above. Each laboratory should establish frequency distributions for their specific patient populations.
#### PERFORMANCE CHARACTERISTICS
{2}------------------------------------------------
Image /page/2/Picture/1 description: The image shows the logo for Focus Technologies. The logo is in black and white and features the word "FOCUS" in large, bold letters. Below the word "FOCUS" is the word "technologies" in a smaller, sans-serif font. The two crescent shapes are nested inside the letter O.
510(k) Summary of Safety and Effectiveness HerpeSelect®2 ELISA IgG Catalog No. EL0920G Prepared July 23, 2002 Page 3 of 9
## Relative Sensitivity and Relative Specificity with Expectant Mothers †
An outside investigator assessed the device's relative sensitivity and relative specificity with sera from expectant mothers (n = 241). The sera were sequentially submitted to the laboratory, archived, and masked. The reference method was a HSV-2 Western blot (WB) from a Pacific Northwest university. Of 3 atypical Western blots, HerpeSelect® 2 ELISA IgG (EL) was 1 equivocal and 2 negatives. Of 58 Western blot positives, HerpeSelect® 2 ELISA IgG was 58 positive. Of 180 Western blot negatives, HerpeSelect® 2 ELISA IgG was 172 negatives, 7 positives, and 1 equivocal.
| Characteristic | % (EL/WB)* | 95% CI |
|--------------------------------------|-----------------|------------|
| Sensitivity relative to Western blot | 100% (58/58) | 93.8-100% |
| Specificity relative to Western blot | 96.1% (172/179) | 92.1-98.4% |
Relative Sensitivity and Relative Specificity with Expectant Mothers (n = 241) †
* Excludes three atypical Western blots and one ELISA equivocal
+ The word "relative" refers to comparing this assay's results with those of a similar assay. No attempt was made to correlate the assay results to disease presence or absence. No judgment can be made on the similar assay's accuracy in predicting disease. Assay performed by manual method.
#### Relative Sensitivity and Relative Specificity with Sexually Active Adults †
An outside investigator assessed the device's relative sensitivity and relative specificity with sera from sexually active adults over the age of 14 (n = 246). The sera were sequentially submitted to the laboratory, archived, and masked. The reference method was a HSV-2 Western blot from a Pacific Northwest university. Of 5 atypical Western blots, HerpeSelect® 2 ELISA IgG was 2 equivocal, 2 negative and 1 positive. Of 76 Western blot positives, HerpeSelect® 2 ELISA IgG was 73 positive. Of 165 Western blot negatives, HerpeSelect® 2 ELISA IgG was 159 negative, 5 positive, and 1 equivocal.
Relative Sensitivity and Relative Specificity with Sexually Active Adults (n = 246) †
| Characteristic | % (EL/WB)* | 95% CI |
|--------------------------------------|-----------------|------------|
| Sensitivity relative to Western blot | 96.1% (73/76) | 88.9-99.2% |
| Specificity relative to Western blot | 97.0% (159/164) | 93.0-99.0% |
* Excludes five atypical Western blots and one ELISA equivocal
* The word "relative" refers to comparing this assay's results with those of a similar assay. No attempt was made to correlate the assay results to disease presence or absence. No judgment can be made on the similar assay's accuracy in predicting disease. Assay performed by manual method.
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the logo for Focus Technologies. The logo is black and white and features the word "FOCUS" in large, bold letters. Below the word "FOCUS" is the word "technologies" in smaller letters. The two crescent moon shapes are used to create the letter O in the word FOCUS.
510(k) Summary of Safety and Effectiveness HerpeSelect®2 ELISA IgG Catalog No. EL0920G Prepared July 23, 2002 Page 4 of 9
#### Relative Sensitivity with Culture Positives +
An outside investigator assessed the device's relative sensitivity using sera from culture positive patients (n = 63). Reference methods included culture (infection) and a HSV-2 Western blot (antibody) from a Pacific Northwest university. Of 5 atypical Western blots, HerpeSelect®2 ELISA IgG was 2 equivocal, 2 negative and 1 positive. Of 63 culture positives, HerpeSelect® 2 ELISA IgG was 61 positive and 2 negative, and Western blot was 62 positive and 1 negative. Of 62 Western blot positives, HerpeSelect® 2 ELISA IgG was 61 positive and 1 negative.
| Characteristic | % (EL/WB or<br>Culture) | 95% CI |
|--------------------------------------|-------------------------|------------|
| Sensitivity relative to culture | 96.8% (61/63)* | 89.0-99.6% |
| Sensitivity relative to Western blot | 98.4% (61/62)* | 91.3-100% |
*Of the 2 ELISA negatives, one was WB positive and the other WB negative.
+ The word "relative" refers to comparing this assay's results with those of a similar assay. No attempt was made to correlate the assay results to disease presence or absence. No judgment can be made on the similar assay's accuracy in predicting disease. Assay performed by manual method.
### Agreement with CDC Panel +
The following information is from a serum panel obtained from the CDC and tested by Focus Technologies. The results are presented as a means to convey further information on the performance of this assay with a masked, characterized serum panel. This does not imply an endorsement of the assay by the CDC. The panel consists of 37% positive and 63% negative samples. The HerpeSelect® 2 ELISA IgG demonstrated 100% total agreement with the Of the results obtained by Focus Technologies, there was 100% agreement with the positive CDC results. specimens and 100% agreement with the negative specimens.
+ Assay performed by manual method.
#### Relative Specificity with a Low Prevalence Population *
An outside investigator assessed the device's relative specificity using sera from a population of college students claiming to lack sexual experience (n = 81), and having a published HSV-2 antibody prevalence of 2% (4/186) ** The laboratory reference method was a HSV-2 Western blot from a Pacific Northwest university. One atypical Western blot was an HerpeSelect® 2 ELISA IgG negative. Of 78 Western blot negatives, HerpeSelect® 2 ELISA IgG was 77 negative and 1 positive. Of 2 Western blot negatives, HerpeSelect® 2 ELISA IgG was 2 positive.
| Characteristic | % (EL/WB)* | 95% CI |
|---------------------------------------|---------------|-----------|
| Specificity relative to Western blot† | 98.7% (77/78) | 93.1-100% |
| Sensitivity relative to Western blot† | 100% (2/2) | 15.8-100% |
Relative Specificity with a Low Prevalence Population (n = 81) +
* Excludes one atypical Western blot.
<sup>**</sup> Corey, L. A. Wald, New Developments in the Biology of Genital Herpes, in Clinical Management of Herpes Viruses, p.46.
<sup>+</sup> The word "relative" refers to comparing this assay's results with those of a similar assay. No attempt was made to correlate the assay results to disease presence or absence. No judgment can be made on the similar assay's accuracy in predicting disease. Assay performed by manual method.
{4}------------------------------------------------
Image /page/4/Picture/1 description: The image shows the logo for Focus Technologies. The logo is black and white and features the word "FOCUS" in large, bold letters. Below the word "FOCUS" is the word "technologies" in smaller letters. The two crescent moon shapes are used in place of the letter "O" in the word "FOCUS".
## 510(k) Summary of Safety and Effectiveness HerpeSelect®2 ELISA IgG Catalog No. EL0920G Prepared July 23, 2002 Page 5 of 9
Type Specificity with HSV-1 Western Blot Positives †
An outside investigator assessed the device's type specificity using HSV-1 Western blot positive and HSV-2 Western Blot negative sera from the above described populations (n = 287): expectant mothers, sexually active adults, low prevalence persons, and HSV-1 culture positives. Of 287 HSV-1 Western blot positive and HSV-2 Western blot negative samples, HerpeSelect® 2 ELISA IgG was 276 negatives, 1 equivocal and 10 positives.
| Characteristic | % (EL/WB)* | 95% CI |
|------------------------------------------------|-----------------|------------|
| Type-specificity relative to Western blot | 96.5% (276/286) | 93.7-98.3% |
| Type cross-reactivity relative to Western blot | 3.5% (10/286) | 1.7-6.3% |
| | | Type Specificity with HSV-1 Western Blot Positives (n = 287) † | |
|--|--|----------------------------------------------------------------|--|
| | | | |
* Excludes one equivocal ELISA result.
+ The word "relative" refers to comparing this assay's results with those of a similar assay. No attempt was made to correlate the assay results to disease presence or absence. No judgment can be made on the similar assay's accuracy in predicting disease. Assay performed by manual method. Assay performed by manual method.
### Cross-reactivity with Taxonomically Related Viruses *
Focus Technologies assessed the device's cross-reactivity using sera (n = 27) from 1) HSV sero-negative by another manufacturer's FDA cleared HSV ELISAs, and 2) IFA IgG positive for taxonomically similar viruses including CMV, EBV VCA, HHV6 and VZV. Discrepants between the FDA cleared HSV ELISAs and the HerpeSelect" 2 ELISA IgG were analyzed using a type specific Western blot from a major university located in the Northwestern United States.
| Cross-reactivity with Taxonomically Related Viruses (n = 27) † | | |
|----------------------------------------------------------------|--------------------------|------------|
| IFA IgG Pos | % Agreement<br>Negative* | 95% CI |
| CMV | 91.7% (11/12) | 61.5-99.8% |
| EBV VCA | 90.9% (20/22) | 70.8-98.9% |
| HHV6 | 90.9% (20/22) | 70.8-98.9% |
| VZV | 90.5% (19/21) | 69.6-98.8% |
| Total | 90.9% (70/77) | 82.2-96.3% |
## Cross-reactivity with Taxonomically Related Viruses (n
* Excludes 3 Western blot positives, and one discrepant that was not analyzed with the Western blot because of insufficient volume
+ Assay performed by manual method.
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the logo for Focus Technologies. The logo consists of the word "FOCUS" in large, bold, sans-serif letters, with a crescent moon shape replacing the letter "O". Below the word "FOCUS" is the word "technologies" in a smaller, sans-serif font. The logo is black and white.
#### Intra-assay & Inter-assay Reproducibility †
An internal investigator assessed the device's intra-assay reproducibility by assaying seven samples in duplicate, twice a day, for twenty days, for a total of forty runs. Two sets of samples were masked duplicates.
#### Inter-lot Reproducibility †
An internal investigator assessed the device's inter-lot reproducibility. Five samples were run on three separate days with three separate lots. For one lot, the samples were run in duplicate with the other two lots. Each of the three lots had a different lot of Antigen Wells.
### Inter-laboratory Reproducibility †
An internal investigator and two off-site laboratories assessed the device's inter-laboratory reproducibility. Each of the three laboratories ran seven samples in triplicate on three different days. Three points were excluded because an incorrect sample (instead of sample 27) was run one day.
| Sample | Inter- & Intra-assay | | | Inter-lot | | Inter-Laboratory | | |
|--------|----------------------|------------------------|------------------------|---------------|--------------|------------------|------------------------|------------------------|
| | Index<br>Mean | Intra-<br>assay<br>%CV | Inter-<br>assay<br>%CV | Index<br>Mean | Index<br>%CV | Index<br>Mean | %CV of<br>Lab<br>Means | Mean of<br>Lab<br>%CVs |
| 21* | 0.2 | 20.5 | 15.9 | 0.3 | 52.4 | 0.2 | 19.6 | 17.3 |
| 26* | 0.2 | 12.2 | 12.4 | NA | NA | 0.3 | 33.1 | 20.7 |
| 22** | 1.2 | 6.3 | 6.2 | 1.2 | 5.1 | 1.2 | 3.9 | 7.8 |
| 27** | 1.2 | 5.2 | 6.3 | NA | NA | 1.1 | 14.1 | 8.8 |
| 23 | 1.8 | 4.7 | 5.5 | 1.8 | 5.4 | 1.7 | 5.2 | 7. |
| 24 | 3.4 | 3.2 | 7.9 | 3.2 | 16.7 | 2.8 | 11.0 | 10.8 |
| 25 | 8.2 | 3.0 | 6.9 | 8.0 | 7.4 | 6.8 | 18.6 | 4.5 |
Reproducibility †
* #21 & #26 are same material.
** #22 & #27 are same material.
ート Assay performed by manual method.
{6}------------------------------------------------
Image /page/6/Picture/1 description: The image shows the logo for Focus Technologies. The logo is in black and white and features the word "FOCUS" in large, bold letters. Below the word "FOCUS" is the word "technologies" in smaller, lowercase letters. The two crescent moon shapes are used in place of the letter "O" in the word "FOCUS".
## 510(k) Summary of Safety and Effectiveness HerpeSelect®2 ELISA IgG Catalog No. EL0920G Prepared July 23, 2002 Page 7 of 9
% Agreement between the Manual and Automated Methods (n = 257)
An internal and an external investigator compared % agreement between the HerpeSelect automated ws. the manual method as part of a CLIA validation for a major clinical laboratory located in Southern California. The external investigator sequentially selected and manually tested 257 sample was from an adult, and was submitted for HSV testing. 255 samples were from the US, and two samples from outside the US. Of the 257 samples, the manual method detected 175 negatives, 3 equivocals, and 79 positives. Of the 175 negatives by the manual method, the automated method agreed with 99.4% (174/175). Of the 3 equivocals by the manual method, the automated method agreed with 0% (0/3). Of the 79 positives by the manual method, the automated method agreed with 98.7% (78/79). Overall, the two methods agreed 98.1% (252/257). Of the five discrepants, two resolved in favor of the automated method, and the other three did not resolve.
| Interpretation* | % Agreement | 95% CI |
|-----------------|-----------------|------------|
| Negative | 99.4% (174/175) | 96.4-100% |
| Equivocal | 0.0% (0/3) | 0.0-70.8% |
| Positive | 98.7% (78/79) | 93.1-100% |
| Overall | 98.1% (252/257) | 95.5-99.4% |
### % Agreement between Manual and Automated Methods
* Interpretation by manual method.
{7}------------------------------------------------
Image /page/7/Picture/1 description: The image shows the logo for Focus Technologies. The logo is black and white and features the word "FOCUS" in large, bold letters. Below the word "FOCUS" is the word "technologies" in smaller letters. The two crescent shapes are used to create the letter "O" in the word "FOCUS".
## 510(k) Summary of Safety and Effectiveness HerpeSelect®2 ELISA IgG Catalog No. EL0920G Prepared July 23, 2002 Page 8 of 9
### Reproducibility Using an Automated Instrument
An internal investigator assessed the device's inter-assay reproducibility with an automated instrument. Ten samples were tested in triplicate on three different days. The manual and automated methods agreed 98.9% (89/90). One point from Sample 3 was an outlier (162 standard deviations from the mean).
| Sample | Mean Index | Intra-assay<br>%CV | Inter-assay<br>%CV |
|----------------------|------------|--------------------|--------------------|
| 6 | 0.06 | 27.2 | 20.8 |
| 3 (without outlier)* | 0.45 | 8.8 | 4.6 |
| 8 | 0.10 | 28.3 | 25.2 |
| 3 (with outlier)* | 1.18 | 51.4 | 105.4 |
| 10 | 1.39 | 6.0 | 5.5 |
| 9 | 1.87 | 4.8 | 2.5 |
| 5 | 5.23 | 2.8 | 0.9 |
| 1 | 6.38 | 1.3 | 3.5 |
| 4 | 7.14 | 3.0 | 3.3 |
| 7 | 7.97 | 6.4 | 10.8 |
| 2 | 8.21 | 0.8 | 4.6 |
*One point from Sample 3 was an outlier (162 standard deviations from the mean).
{8}------------------------------------------------
Image /page/8/Picture/1 description: The image shows the logo for Focus Technologies. The word "FOCUS" is written in large, bold, black letters. Below the word "FOCUS" is the word "technologies" written in a smaller, thinner font. The two crescent moon shapes are used to create the letter "O" in the word "FOCUS".
## 510(k) Summary of Safety and Effectiveness HerpeSelect®2 ELISA IgG Catalog No. EL0920G Prepared July 23, 2002 Page 9 of 9
## Stability after Opening Reagents
An internal investigator assessed stability after the reagents had been opened and used with an automated instrument. The kit was used in the inter-assay reproducibility study (above), re-closed, stored at 2-8C for at least 30 days, and then used again to re-test the same samples. There was 100% agreement with the index when the reagents were opened.
| Sample | Index when<br>Opened | After at Least<br>30 Days (Run 1) | After at Least<br>30 Days (Run 2) |
|---------------------|----------------------|-----------------------------------|-----------------------------------|
| 6 | 0.06 | 0.06 | 0.07 |
| 3 (without outlier) | 0.45* | 0.43 | 0.48 |
| 8 | 0.10 | 0.14 | 0.10 |
| 10 | 1.39 | 1.60 | 1.76 |
| 9 | 1.87 | 2.03 | 2.11 |
| 5 | 5.23 | 5.13 | 5.10 |
| 1 | 6.38 | 6.79 | 6.57 |
| 4 | 7.14 | 7.67 | 7.26 |
| 7 | 7.97 | 8.16 | 8.29 |
| 2 | 8.21 | 8.38 | 8.35 |
## Stability after Opening Reagent
*One point from Sample 3 was an outlier, it was 162 SDs from the mean. The values without the outlier are used here.
{9}------------------------------------------------
Image /page/9/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo features a stylized eagle with three lines representing its wings. The eagle is positioned to the right of the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA", which is arranged in a circular fashion around the left side of the logo.
Food and Drug Administration 2098 Gaither Road Rockville MD 20850
## AUG 0 1 2002
Michael J. Wagner, Esq. Senior Regulatory Affairs Specialist Focus Technologies, Inc. 10703 Progress Way Cypress, CA 90630
Re: K021486 Trade/Device Name: HerpeSelect ™ 2 ELISA IgG Regulation Number: 21 CFR 866.3305 Regulation Name: Herpes Simplex Virus Serological Reagents Regulatory Class: Class III Product Code: MXJ Dated: May 2, 2002 Received: May 3, 2002
Dear Mr. Wagner:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. Iisting of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{10}------------------------------------------------
Page 2 -
This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and ' additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4588. Additionally, for questions on the promotion and advertising of your device. please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.gov/cdrb/dsma/dsmamain.html".
Sincerely yours,
Steven Butman
Steven I. Gutman, M.D., M.B.A. Director Division of Clinical Laboratory-Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{11}------------------------------------------------
510(k) Number (if known): K021486
Device Name:
HerpeSelect® 2 ELISA IgG
Indications for Use:
Focus Technologies' HerpeSelect® 2 ELISA IgG test is intended for qualitatively detecting the presence or absence of human IgG class antibodies to HSV-2 in human sera. In conjunction with the Focus HerpeSelect® 1 ELISA IgG, the test is indicated for testing sexually active adults or expectant mothers for aiding in the presumptive diagnosis of HSV infection. The assay can be used manually or in conjunction with an automated system as outlined in the package insert. The user is responsible for assay performance characteristics when an automated system is used. The performance of this assay has not been established for use in a pediatric population, for neonatal screening, or for testing of immunocompromised patients.
(PLEASE DO NOT WRITE BELOW THIS LINE CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE)
Woody Duboer
0(k) Number
For PRESCRIPTION USE_X
(Optional Format 3-10-98)
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.