The CLINITEK ATLAS PRO 12 Reagent Pak system is intended for use in at-risk patient groups to assist diagnosis in the following areas: kidney function, carbohydrate metabolism, urinary tract infections and liver function. The system also measures physical characteristics, including acid-base balance and urine concentration. Test results can be used along with other diagnostic information to rule out certain disease states. The CLINITEK ATLAS PRO 12 Reagent Pak system is for professional use in centralized laboratory locations such as in hospitals and reference laboratories.
Device Story
System consists of roll of 490 reagent strips for use with CLINITEK ATLAS Automated Urine Chemistry Analyzer; strips contain 11 reagent pads for protein (low/high), creatinine, occult blood, glucose, ketone, leukocytes, nitrite, pH, bilirubin, urobilinogen, plus 1 nonreactive pad for specimen color. Principle of operation: low-level protein assay uses dye-binding (high affinity dye); creatinine assay uses peroxidase-like activity of copper creatinine complexes. System calculates protein-to-creatinine ratio to minimize impact of exercise, orthostatic proteinuria, diuresis, and urine concentration on results. Used in centralized labs (hospitals/reference labs) by professional staff. Output provides quantitative/semi-quantitative diagnostic data to assist clinicians in assessing kidney, liver, and metabolic function; ratio results >150 mg/g considered abnormal. Benefits include improved discrimination of normal/abnormal protein levels using single-void specimens.
Clinical Evidence
Performance evaluated via internal and clinical studies in centralized laboratory settings. Studies compared system results against commonly used laboratory tests. Results demonstrated that the system provides clinical data comparable to existing methods in current clinical practice.
Technological Characteristics
System comprises firm plastic strips with 11 reagent areas and 1 color-sensing pad. Assays utilize dye-binding (albumin) and peroxidase-like catalytic reactions (creatinine). Designed for integration with CLINITEK ATLAS Automated Urine Chemistry Analyzer. Reagents include Bis(3',3"-diiodo-4',4"-dihydroxy-5',5"-dinitrophenyl)-3,4,5,6-tetrabromosulfonephthalein for albumin and diisopropylbenzene dihydroperoxide/3,3',5,5'-tetramethyl-benzidine for creatinine.
Indications for Use
Indicated for at-risk patient groups to assist in diagnosis of kidney function, carbohydrate metabolism, urinary tract infections, and liver function; measures acid-base balance and urine concentration; for professional use in centralized laboratory settings.
Regulatory Classification
Identification
A creatinine test system is a device intended to measure creatinine levels in plasma and urine. Creatinine measurements are used in the diagnosis and treatment of renal diseases, in monitoring renal dialysis, and as a calculation basis for measuring other urine analytes.
Predicate Devices
Multistix PRO™ 11 Reagent Strips
Multistix PRO™ 10LS Reagent Strips
Multistix PRO™ 7G Reagent Strips
CLINITEK® 50
CLINITEK® 100
CLINITEK® 200+
CLINITEK® 500 Urine Chemistry Analyzers
Submission Summary (Full Text)
{0}------------------------------------------------
'JUL
# 510(k) SAFETY AND EFFECTIVENESS SUMMARY
| Prepared: | April 25, 2002 |
|------------|----------------------------------------------------------------------------------------|
| Submitter: | Bayer Diagnostics |
| Address: | Bayer Diagnostics<br>511 Benedict Ave.<br>Tarrytown, NY 10591<br>phone: (914)-524-2446 |
| Contact: | Kenneth T. Edds, Ph.D.<br>Manager, Regulatory Affairs |
| Device: | |
|--------------------------|---------------------------------------------------------------|
| Trade/Proprietary Name: | CLINITEK ATLAS® PRO™ 12 Reagent<br>Pak System |
| Common/Usual Name: | Reagent Strips for Urinalysis<br>Test for Creatinine in Urine |
| Document Control Number: | K02/428 |
| Classification: | |
|-----------------|------------------------|
| Name: | Creatinine test system |
| Class: | II |
| CFR: | 21 CFR §862. 1225 |
| Code: | 75 JFY |
Predicate Devices:
Multistix PRO™ 11 Reagent Strips
Multistix PRO™ 10LS Reagent Strips
Multistix PRO™ 7G Reagent Strips
CLINITEK® 50, CLINITEK® 100, CLINITEK® 200+,
CLINITEK® 500 Urine Chemistry Analyzers
Manufacturer: Bayer Diagnostics
Page 1 of 3
{1}------------------------------------------------
### Device Description:
The CLINITEK ATLAS PRO 12 Reagent Pak is a roll of firm plastic to which are affixed 490 reagent strips. It is for use with the CLINITEK ATLAS® Automated Urine Chemistry Analyzer. Each reagent strip contains 11 separate reagent areas to test for low level protein (15 mg/dL albumin), high level protein (>30 mg/dL protein), creatinine, occult blood, glucose, ketone (acetoacetic acid), leukocytes, nitrite, pH, bilirubin and urobilinogen. A protein-to-creatinine ratio is also determined. In addition, each strip contains a nonreactive 12th pad that is used for the determination of the specimen color.
## Intended Use:
The CLINITEK ATLAS PRO 12 Reagent Pak system is intended for use in at-risk patient groups to assist diagnosis in the following areas: kidney function, carbohydrate metabolism, urinary tract infections and liver function. The system also measures physical characteristics, including acid-base balance and urine concentration. Test results can be used along with other diagnostic information to rule out certain disease states. The CLINITEK ATLAS PRO 12 Reagent Pak system is for professional use in centralized laboratory locations such as in hospitals and reference laboratories.
Technological Characteristics:
The CLINITEK ATLAS PRO 12 Reagent Pak system includes a new "Protein Low" reagent that allows "low level" detection of urine albumin at 15 mg/dL in addition to the current protein reagent used for detection of "high level" protein at 30 mg/dL. A new creatinine reagent is used for the estimation of the protein-to-creatinine ratio of the urine specimen based on all three tests. Use of the protein-to-creatinine ratio can assist in the diagnosis of kidney function by minimizing the impact of changes in the protein result due to exercisc, orthostatic proteinuria, diuresis and urine concentration. The ratio allows for the use of single-void specimens in the discrimination of normal and abnormal levels of protein. Ratio results of >150 mg/g are considered to be abnormal.
The assay for low level protein (albumin) is based on dye binding using the high affinity dye, Bis(3',3"-diiodo-4',4"-dihydroxy-5',5"-dinitrophenyl)-3,4,5,6tetrabromosulfonephthalein). At a constant pH, the development of a blue color is owing to the presence of albumin; the intensity of the color is proportional to the albumin concentration. The assay for creatinine is based on the peroxidase-like activity of copper creatinine complexes that catalyze the reaction of diisopropylbenzene dihydroperoxide and 3,3',5,5'-tetramethyl-benzidine.
{2}------------------------------------------------
Assessment of Performance:
The performance of the CLINITEK ATLAS PRO 12 Reagent Pak System was studied internally and in clinical settings. The studies demonstrated that typical users in centralized laboratories could obtain clinical test results that are comparable to commonly used laboratory tests.
Conclusion:
The CLINITEK ATLAS PRO 12 Reagent Pak System has been developed for use with the CLINITEK ATLAS Automated Urine Chemistry Analyzer. New tests for the determination of low levels of albumin and urine creatinine have been added to the CLINITEK ATLAS Automated Urine Chemistry Analyzer menu to allow for a determination of the protein-to-creatinine ratio of the urine specimen. Studies show that the product provides clinical results comparable to other test methods in current clinical practice.
{3}------------------------------------------------
#### DEPARTMENT OF HEALTH & HUMAN SERVICES
Public Health Service
Image /page/3/Picture/2 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle-like symbol with three curved lines representing the body and wings. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" is arranged in a circular fashion around the symbol.
Food and Drug Administration 2098 Gaither Road Rockville MD 20850
5 2002 JUL
Kenneth T. Edds, Ph.D. Manager, Regulatory Affairs Bayer Diagnostics 511 Benedict Avenue Tarrytown, NY 10591
k021428 Re:
Trade/Device Name: Clinitek Atlas® Pro™ 12 Reagent Pak System Regulation Number: 21 CFR 862.1225 Regulation Name: Creatinine test system Regulatory Class: Class II Product Code: JFY; JIP; JIP; JIL; LJX Dated: April 25, 2002 Received: May 3, 2002
Dear Dr. Edds:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{4}------------------------------------------------
#### Page 2 -
This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and ' additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4588. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsma/dsmamain.html".
Sincerely yours,
Steven Butman
Steven I. Gutman, M.D., M.B.A. Director Division of Clinical Laboratory-Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
510(k) Number (if known): K091428
Device Name:
#### CLINITEK ATLAS® PRO™ 12 Reagent Pak System
#### Indications for Use:
The CLINITEK ATLAS PRO 12 Reagent Pak is a roll of firm plastic to which are affixed 490 reagent strips. It is for use with the CLINITEK ATLAS® Automated Urine Chemistry Analyzer. Each reagent strip contains 11 separate reagent areas to test for low level protein (15 mg/dL albumin), high level protein (>30 mg/dL protein), creatinine, occult blood, glucose, ketone (acetoacetic acid), leukocytes, nitrite, pH, bilirubin and urobilinogen. A protein-tocreatinine ratio is also determined. In addition, each strip contains a nonreactive 12" pad that is used for the determination of the specimen color.
The CLINITEK ATLAS PRO 12 Reagent Pak system is intended for use in at-risk patient groups to assist diagnosis in the following areas: kidney function, carbohydrate metabolism, urinary tract infections and liver function. The system also measures physical characteristics, including acid-base balance and urine concentration. Test results can be used along with other diagnostic information to rule out certain disease states. The CLINITEK ATLAS PRO Reagent Pak system is for professional use in centralized laboratory locations such as in hospitals and reference laboratories.
# (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
| | <img alt="Signature" src="signature.jpg"/> |
|-----------------------------------------|--------------------------------------------|
| (Division Sign-Off) | |
| Division of Clinical Laboratory Devices | |
| 510(k) Number | K021428 |
| Prescription Use<br>(Per 21 CFR 801.109) | OR | Over-The-Counter Use<br>(Optional Format 1-2-96) |
|------------------------------------------|----|--------------------------------------------------|
|------------------------------------------|----|--------------------------------------------------|
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.