VERAVIEWEPOCS PANORAMIC X-RAY UNIT WITH CEPHALOMETRIC CAPABILITY
K021372 · J. Morita USA, Inc. · EAG · Jun 7, 2002 · Dental
Device Facts
Record ID
K021372
Device Name
VERAVIEWEPOCS PANORAMIC X-RAY UNIT WITH CEPHALOMETRIC CAPABILITY
Applicant
J. Morita USA, Inc.
Product Code
EAG · Dental
Decision Date
Jun 7, 2002
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 872.1830
Device Class
Class 2
Indications for Use
The Veraviewepocs Panoramic X-ray Unit with Cephalometric Capability that is the subject of this 510(k) is an extraoral source X-ray unit that is used for dental radiographic examination and diagnosis of teeth, jaw, oral structure, TM-joints and skull by exposing an X-ray image receptor to ionizing radiation.
Device Story
Veraviewepocs is an extraoral dental X-ray system. Device captures radiographic images of teeth, jaw, TM-joints, and skull. System includes cephalometric capability. Operation involves exposing an X-ray image receptor to ionizing radiation. Features an 'auto exposure' mode for cephalometric imaging; sensor behind cassette measures accumulated radiation to automatically terminate exposure based on patient size (fixed at 80 kV, 10mA). Used in dental clinical settings by trained professionals. Output is a radiographic image used by clinicians for diagnostic assessment of oral structures. Benefits include improved diagnostic visualization of dental and cranial anatomy.
Clinical Evidence
No clinical data. Evidence consists of bench testing, including image quality testing and verification of compliance with electrical and mechanical safety standards (UL2601-1, IEC 60601-1). Software validation was performed, including hazard analysis and code revision procedures.
Indicated for dental radiographic examination and diagnosis of teeth, jaw, oral structures, TM-joints, and skull in patients requiring extraoral X-ray imaging. Prescription use only.
Regulatory Classification
Identification
A cephalometer is a device used in dentistry during x-ray procedures. The device is intended to place and to hold a patient's head in a standard position during dental x-rays.
{0}------------------------------------------------
### 510(k) SUMMARY
# J. Morita Manufacturing Corporation's Veraviewepocs Panoramic X-ray Unit with Cephalometric Capability
## Name of Device and Name/Address of Sponsor
| | Trade or Proprietary Name: Veraviewepocs Panoramic X-ray Unit with Cephalometric |
|----------------------|------------------------------------------------------------------------------------|
| | Capability |
| Common Name: | Extraoral Source X-ray with Cephalometric Capability |
| Classification Name: | Extraoral Source X-ray System and Cephalometer |
| Product Code : | EHD (Extraoral Source X-ray System) |
| | and EAG (Cephalometer) |
J. Morita USA, Inc. 9 Mason Irvine, California USA 92618 Telephone: 949-581-9600 949-581-9688 Facsimile: Contact Person: Mr. Junichi Miyata, President Date Prepared: April 18, 2002
### Intended Use
The Veraviewepocs Panoramic X-ray Unit with Cephalometric Capability that is the subject of this 510(k) is an extraoral source X-ray unit that is used for dental radiographic examination and diagnosis of teeth, jaw, oral structure, TM-joints and skull by exposing an X-ray image receptor to ionizing radiation.
#### Technological Characteristics and Substantial Equivalence
The Veraviewepocs Panoramic X-ray Unit is an FDA-cleared extraoral source X-ray unit that it used for dental radiographic examination and diagnosis of teeth, jaw, and oral structure by exposing an X-ray image receptor to ionizing radiation (K#013955). The modified Veraviewepocs covered by this submission simply adds a cephalometric capability to the existing device.
In addition to manual settings, the modified Veraviewepocs also has an "auto exposure" feature in the cephalometric mode. Exposure time is automatically adjusted depending on the patient's size, and the device is set at 80 kV and 10mA. A sensor behind the cephalometric cassette measures the amount of x- radiation, and the exposure is terminated once the accumulated x-radiation reaches a certain level.
1
{1}------------------------------------------------
The modified Veraviewepocs is substantially equivalent, for purposes of the FDA *s medical device regulations, to the predicate device, Instrumentarium Corporation Imaging Division's Orthopantomograph OP100, and particularly the upgraded ORTHOCEPH OC 100 for cephalometric radiographv (K#930338, 973642, and 001439). The modified Veraviewepocs has the same general intended use. similar principles of operation, and similar technological characteristics as the previously cleared predicate device. Although there are minor difference in the characteristics of the modified Veraviewepocs and the predicate device, these differences do not raise new questions of safety or efficacy.
The software used in the modified Veraviewepocs has been successfully validated by Morita. The software validation report describes the development process for the device's software/firmware: the software change control and code revision procedures; the system and software requirements; the software handling and storage procedures; a hazard analysis; and a software/firmware certification that the company followed the above-described procedures and policies.
The modified Veraviewepocs was tested to ensure compliance with UL2601-1 and IEC 60601-1, and and it complied with the applicable requirements. The modified Veraviewepocs will be tested and will comply with the applicable requirements of 21 C.F.R. Subchapter J prior to marketing. The modified Veraviewepocs also passed the image quality testing.
The modified Veraviewepocs complies with the applicable thermal, mechanical, and electrical safety requirements of UL2601-1 and IEC 60601-1, and will comply with the applicable requirements of 21 C.F.R. Subchapter J prior to marketing.
The modified Veraviewepocs uses biocompatible metals and plastics on any body contacting surfaces, such as the temple stabilizers and covers, ear rods, chin rests, patient handles, and front/rear head stabilizers. The metals and plastics have been widely used in other medical applications in which the metal or plastic is in body contact, including oral contact.
40100525.doc
{2}------------------------------------------------
Image /page/2/Picture/1 description: The image shows the seal of the Department of Health & Human Services - USA. The seal features a stylized eagle with three lines forming its body and head. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular pattern around the eagle.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
7 2002 .IUN
J. Morita USA, Inc. % Mr. Keith A Barritt Fish & Richardson P.C. 601 Thirteenth Street N.W. WASHINGTON DC 20005 Re: K021372
Trade/Device Name: Veraviewepocs Panoramic X-Ray Unit with Cephalometric Capability Regulation Number: 21 CFR 872.1830 Regulation Name: Cephalometer Regulatory Class: II Product Code: 76 EAG Dated: April 29, 2002 Received: April 30, 2002
Dear Mr. Barritt:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. Iisting of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA). it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (sections 531-542 of the Act); 21 CFR 1000-1050.
{3}------------------------------------------------
· This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at one of the following numbers, based on the regulation number at the top of this letter:
| 8xx.1xxx | (301) 594-4591 |
|----------------------------------|----------------|
| 876.2xxx, 3xxx, 4xxx, 5xxx | (301) 594-4616 |
| 884.2xxx, 3xxx, 4xxx, 5xxx, 6xxx | (301) 594-4616 |
| 892.2xxx, 3xxx, 4xxx, 5xxx | (301) 594-4654 |
| Other | (301) 594-4692 |
Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html.
Sincerely yours,
Nancy C. Brogdon
Nancy C. Brogdon Director, Division of Reproductive, Abdominal, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
U.S. Food and Drug Administration - Center for Devices and Radiological Health
Page 1 of 1
510(k) Number (if known): K021372
Device Name: Veraviewepocs Panoramic X-ray Unit with Cephalometric Capability
Indications for Use:
The Veraviewepocs is an extraoral source X-ray unit that is used for dental radiographic examination and diagnosis of teeth, jaw, and oral structure by exposing an X-ray image receptor to ionizing radiation.
## (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
(Optional Format 3-10-98)
*Prescription Use*
David G. Lawson
510(k) Nun
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.