K021251 · Philips Medical Systems · DSI · May 7, 2002 · Cardiovascular
Device Facts
Record ID
K021251
Device Name
ST/AR ST AND ARRHYTHMIA SOFTWARE, RELEASE E.I.
Applicant
Philips Medical Systems
Product Code
DSI · Cardiovascular
Decision Date
May 7, 2002
Decision
SESE
Submission Type
Special
Regulation
21 CFR 870.1025
Device Class
Class 2
Attributes
Pediatric
Indications for Use
Where the clinician decides to monitor cardiac arrhythmia of adult, pediatric, and neonatal patients and/or ST segment of adult patients to gain information for treatment, to monitor adequacy of treatment, or to exclude causes of symptoms. The intended use of the ST/AR cardiotach is to monitor a neonatal, pediatric, or adult patient's ECG for heart rate and produce events/alarms for one or two ECG leads. The cardiotach function is capable of monitoring both paced and non-paced patients. The intended use of the ST/AR arrhythmia analysis algorithm is to monitor a neonatal, pediatric, or adult patient ECG's for heart rate and ventricular arrhythmias and produce events/alarms for one or two ECG leads. The arrhythmia analysis algorithm is capable of monitoring both paced and non-paced patients. The intended use of the ST/AR ST analysis algorithm is to monitor an adult patient's ECG for ST segment elevation or depression and produce events/alarms for all possible ECG leads. The ST analysis algorithm is capable of monitoring paced and non-paced adult patients.
Device Story
Philips ST/AR ST and Arrhythmia Software (Release E.1) processes ECG signals to monitor heart rate, arrhythmias, and ST segments. System provides cardiotach functionality (heart rate monitoring) and ST segment analysis, with or without arrhythmia analysis. Used in clinical settings by healthcare professionals to monitor paced and non-paced patients. Software analyzes one or two ECG leads for cardiotach/arrhythmia and all possible leads for ST segment analysis. Outputs include events and alarms displayed to clinicians to assist in treatment decisions and patient management.
Clinical Evidence
Bench testing only. Verification and validation activities included system-level tests, performance tests, and hazard analysis. Results demonstrated that Release E.1 meets all reliability requirements and performance claims established for the predicate devices.
Technological Characteristics
Software-based ECG analysis system. Operates as a component of patient monitoring systems. Provides cardiotach, arrhythmia detection, and ST segment analysis. Supports paced and non-paced patient monitoring. Connectivity and hardware platform are consistent with previously cleared predicate versions.
Indications for Use
Indicated for adult, pediatric, and neonatal patients requiring cardiac arrhythmia monitoring and/or adult patients requiring ST segment monitoring to inform treatment, monitor treatment adequacy, or exclude symptom causes. ST algorithm excludes analysis of ventricularly paced or ventricular ectopic beats.
Regulatory Classification
Identification
The arrhythmia detector and alarm device monitors an electrocardiogram and is designed to produce a visible or audible signal or alarm when atrial or ventricular arrhythmia, such as premature contraction or ventricular fibrillation, occurs.
Special Controls
*Classification.* Class II (special controls). The guidance document entitled “Class II Special Controls Guidance Document: Arrhythmia Detector and Alarm” will serve as the special control. See § 870.1 for the availability of this guidance document.
{0}------------------------------------------------
## MAY 0 7 2002
## 9.0 510 (k) Summary
This summary of 510(k) safety and effectiveness information is submitted in accordance with the requirements of the Safe Medical Devices Act of 1990 and 21 CFR §807.92.
1. The submitter of this premarket notification is: Dave Osborn Regulatory Affairs Engineer Cardiac & Monitoring Systems Group Philips Medical Systems 3000 Minuteman Road Andover, MA 01810-1085
> 978 659 3178 Tel: 978 685 5624 Fax: Email: dosborn@hsgmed.com
This summary was prepared on 17 April, 2002
- 2. The name of this device is the Philips ST/AR ST and Arrhythmia Software, Release E.1. Classification names are as follows:
| Classification | ProCode | Description |
|----------------|---------|----------------------------------------|
| 870.1025, III | 74 MLD | Monitor, ST Alarm |
| 870.1025, III | 74 DSI | Arrhythmia Detector and Alarm |
| None | 74 MHX | Physiological Monitor, Patient Monitor |
- The new device is substantially equivalent to the previously 3 . cleared ST/AR ST and Arrhythmia Software device marketed pursuant to K964122, K991773, K001348, K003621 and K014261.
- The modification is a software-based change that provides 4 . Cardiotach functionality without arrhythmia analysis and ST Segment Analysis without arrhythmia analysis.
- The new device has the same Indications for Use as the legally 5. marketed predicate device. Where the clinician decides to monitor cardiac arrhythmia of adult, pediatric, and neonatal patients and/or ST segment of adult patients to gain information for treatment, to monitor adequacy of treatment, or to exclude causes of symptoms.
- The new device has the same technological characteristics as the 6. legally marketed predicate device.
- 7. Verification, validation, and testing activities establish the performance, functionality, and reliability characteristics of the new device with respect to the predicate. Testing involved
44
{1}------------------------------------------------
K021251
p. 2/2
system level tests, performance tests, and safety testing from hazard analysis. Pass/Fail criteria were based on the specifications cleared for the predicate device and test results showed substantial equivalence. The results demonstrate that ST/AR Release E.1 meets all reliability requirements and performance claims.
{2}------------------------------------------------
Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo is circular, with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is an abstract symbol that resembles an eagle or bird in flight, composed of three curved lines.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
## MAY 0 7 2002
Mr. Dave Osborn Quality Program Manager Philips Medical Systems Cardiac and Monitoring Systems 3000 Minuteman Road Andover, MA 01810-1099
Re: K021251
Trade Name: Philips Medical Systems ST/AR and Arrhythmia Software, Release E.1. Regulation Name: Arrhythmia Detector and Alarm Regulation Number: 21 CFR 870.1025 Regulatory Class: Class III (three) Product Code: DSI Dated: April 18, 2002 Received: April 19, 2002
Dear Mr. Osborn:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. Iisting of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{3}------------------------------------------------
## Page 2 -- Mr. Dave Osborn
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must or car) x with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set Erth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 21 CFR Part 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4646. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html
Sincerely yours,
Darla Tule
Donna-Bea Tillman, Ph.D. Acting Director Division of Cardiovascular and Respiratory Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
Page / of /
510(k) Number (if known): ____________________________________________________________________________________________________________________________________________________
Device Name: Philips Medical Systems ST/AR Software, Release E.1.
Where the clinician decides to monitor cardiac Indications for Use: arrhythmia of adult, pediatric, and neonatal patients and/or ST segment of adult patients to gain information for treatment, to monitor adequacy of treatment, or to exclude causes of symptoms.
The intended use of the ST/AR cardiotach is to monitor a neonatal, pediatric, or adult patient's ECG for heart rate and produce events/alarms for one or two ECG leads. The cardiotach function is capable of monitoring both paced and non-paced patients.
The intended use of the ST/AR arrhythmia analysis algorithm is to monitor a neonatal, pediatric, or adult patient ECG's for heart rate and ventricular arrhythmias and produce events/alarms for one or two ECG leads. The arrhythmia analysis algorithm is capable of monitoring both paced and non-paced patients.
The intended use of the ST/AR ST analysis algorithm is to monitor an adult patient's ECG for ST segment elevation or depression and produce events/alarms for all possible ECG leads. The ST analysis algorithm is capable of monitoring paced and non-paced adult patients.
The ST algorithm does not analyze ventricularly paced Note: or ventricular ectopic beats.
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
| Prescription Use <span style="text-decoration:underline;"><b>✓</b></span> | OR | Over-The-Counter |
|---------------------------------------------------------------------------|----|------------------|
|---------------------------------------------------------------------------|----|------------------|
Use
(Per 21 CFR 801.109) (Optional Format 1-2-96)
Jada tu
Division of Cardiovascular & Respiratory Devices
510(k) Number K061251
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.