K021249 · Power Medical Interventions, Inc. · GDW · May 15, 2002 · General, Plastic Surgery
Device Facts
Record ID
K021249
Device Name
SURGASSIST FLEXSHAFT 2, MODEL FS2 14
Applicant
Power Medical Interventions, Inc.
Product Code
GDW · General, Plastic Surgery
Decision Date
May 15, 2002
Decision
SESE
Submission Type
Special
Regulation
21 CFR 878.4750
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The SurgASSIST™ System with Circular Stapler Digital Loading Unit™ (DLU), has applications throughout the alimentary tract for end-to-end and side-to-side anastomosis. The SurgASSIST™ System with Right Angle Linear Cutter DLU has applications in gastrointestinal, gynecological, general abdominal and thoracic surgical procedures for resection, transection, and creation of anastomosis.
Device Story
SurgASSIST™ System with FlexShaft 2 is a flexible, steerable surgical instrument interface connecting a PC (power source) to a Digital Loading Unit (DLU). The device features a 2-meter flexible shaft with a distal articulation section controlled by four braided cables and bevel gears. It transmits rotary motion from PC motors to the DLU for tissue clamping and staple formation. Hall-effect transistors at the distal end provide quadrature feedback to the PC to track shaft revolutions. The device includes a memory module (usage counter) to track DLU firings and attachment history. Used in surgical settings by clinicians, the system facilitates remote positioning of the DLU for minimally invasive procedures. The PC micro-controller communicates with the DLU chip to ensure proper operation. The device is non-sterile and supports multiple sterilization methods including steam, ETO, cold liquid, and gas plasma.
Clinical Evidence
No clinical data provided; bench testing only.
Technological Characteristics
Flexible shaft (2m length) with 10cm articulation section; 4-cable steering; hall-effect transistor quadrature tracking; 2 electrical contacts for DLU communication; memory module for usage tracking. Sterilization: Steam (flash), ETO, cold liquid, gas plasma. Non-sterile supply.
Indications for Use
Indicated for patients requiring resection, transection, or anastomosis in the alimentary tract, gastrointestinal, gynecological, general abdominal, or thoracic surgical procedures.
Regulatory Classification
Identification
An implantable staple is a staple-like device intended to connect internal tissues to aid healing. It is not absorbable.
Predicate Devices
SurgASSIST™ System with Circular Stapler Disposable Loading Unit (K003277)
SurgASSIST™ System with Right Angle Linear Cutter Digital Loading Unit™ (K012809)
Submission Summary (Full Text)
{0}------------------------------------------------
K021249 (Pilot\$)
MAY 1 5 2002
Power Medical Interventions, Inc. SurgASSIST™ System with FlexShaft 2 Special 510(k) Device Modification PreMarket Notification April 17, 2002
## Special 510(k) Device Modification PREMARKET NOTIFICATION SAFETY AND EFFECTIVENESS SUMMARY
# SurgASSIST™ System with FlexShaft 2
In Accordance with 21 CFR section 807.92 Power Medical Interventions, Inc., is submitting the following safety and effectiveness summary.
- 1) Submitter Information:
Power Medical Interventions, Inc. 110 Union Square Drive New Hope, PA 18938 215-862-4450 Ph 215-862-1009 Fax
Applicant:
Barbara J. Whitman
April 17, 2002 Date of Notification:
- 2) Name of Device:
| Trade Name: | SurgASSIST ™ System |
|-------------|---------------------|
| | With FlexShaft 2 |
Common Name: FlexShaft 2, Flexible Shaft
Classification Name: Staple, Implantable, GDW; Stapler, GAG
#### 3) Predicate Devices:
- a) SurgASSIST™ System with Circular Stapler Disposable Loading Unit with Titagium Implantable Staple, Power Medical Interventions, Inc., K003277.
### 000016
{1}------------------------------------------------
K024034 (P.2 of 4)
Power Medical Interventions, Inc. SurgASSIST™ System with FlexShaft 2 Special 510(k) Device Modification PreMarket Notification April 17, 2002
- b) SurgASSIST™ System with Right Angle Linear Cutter Digital Loading Unit™, Power Medical Interventions, Inc., K012809.
#### 4) Device Description:
The FlexShaft 2 resembles a colonoscope and provides the mechanical and electrical interface between the PC and the DLU. The FlexShaft 2 has a rigid mechanical assembly at the proximal end that provides the mechanical interface between the FlexShaft 2 and PC. The remainder of the FlexShaft 2 is flexible throughout its length and incorporates an articulation section near the distal end that allows for remote positioning of the DLU. The distal tip of the FlexShaft 2 has features that allow for the attachment and removal of various configurations of a DLU.
The FlexShaft 2 contains two drive shafts that couple two motors in the PC to the two inputs of a DLU. Rotary motion provided by the motors located in the PC is delivered to the DLU through these drive shafts for various purposes such as clamping tissue or forming staples.
The distal assembly of the FlexShaft 2 contains a printed circuit board that has two pairs of hall-effect transistors mounted. Each pair of hall-effect transistors is arranged for quadrature output to enable tracking of the rotation of the two drive shafts at the distal end. This tracking provides the signal to the PC to insure that the appropriate number of shaft revolutions is delivered to the DLU.
Internal to the FlexShaft 2 are four braided (steering) cables. The purpose of these cables is to provide the means for articulating the steering section of the FlexShaft 2. Articulation is achieved when tension is applied to the cable or cables located in the desired articulation direction. Rotary motion, provided by motors in the PC, is translated into linear "cable pull" motion through a pairs of bevel gears that are coupled to capstans.
Since the DLU contains an electronic "chip" carrying its identification and usage status, it must be able to communicate with the micro controller within the PC. The FlexShaft 2 incorporates two electrical contacts that mate with corresponding contacts on the DLU when the DLU is attached to the FlexShaft 2.
Internal to the FlexShaft 2 is a usage counter. The purpose of this counter is to allow the storage of information relating to the number of DLU's fired and the number of FlexShaft 2 attachments to the PC. Each memory module has a unique serial number that will be used for FlexShaft 2 identification.
{2}------------------------------------------------
K021249 (P.3 of 4)
Power Medical Interventions, Inc. SurgASSIST™ System with FlexShaft 2 Special 510(k) Device Modification PreMarket Notification April 17, 2002
- 5) Indications For Use The FlexShaft 2 will have identical Indications For Use as the predicate device.
The SurgASSIST™ System with Circular Stapler DLU, has applications throughout the alimentary tract for end-to-end and sideto-side anastomosis.
The SurgASSIST™ System with Right Angle Linear Cutter Digital Loading Unit™ has applications in gastrointestinal, gynecological, general abdominal and thoracic surgical procedures for resection, transection, and creation of anastomosis.
- 6) Comparison to Predicate Devices
The following table compares the subject FlexShaft 2 device to the previously cleared predicate FlexShaft device:
| Features & Description | SurgASSIST™ w/ FlexShaft 2 (FS214) | Predicate<br>SurgASSIST™ w/ the original<br>FlexShaft (FS14) |
|-----------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------|
| Name | SurgASSIST™ System with FlexShaft 2 | SurgASSIST System with FlexShaft<br>component |
| Manufacturer of Record | Power Medical Interventions, Inc. | Power Medical Interventions, Inc. |
| Contract Manufacturer | Gore & Associates, Inc.<br>Newark, DE | Design Standards Corporation<br>Charlestown, NH |
| 510(k) Clearance Numbers | Subject of this Notification | K003277 |
| Product Codes | FS214 | FS14 |
| Intended use | Applications throughout the alimentary<br>tract for end to end, end to side, and<br>side to side anastomosis. Also,<br>applications in gastrointestinal,<br>gynecological, general abdominal and<br>thoracic surgical procedures for<br>resection, transection, and creation of<br>anastomoses. | Applications throughout the alimentary<br>tract for end to end, end to side, and<br>side to side anastomosis |
| FDA Class (System) | II | II |
| Physical Characteristics | ------------------------------------------------ | ------------------------------------------------ |
| Length of FlexShaft | 2 meters ± 3cm | 2 meters ± 3cm |
| Length of Articulation Section | 10cm ± 2cm | 10cm ± 2cm |
| Diameter of Articulation<br>Section | 15mm ± .25mm | 15mm ± .25mm |
| Articulated tip Steering | 4 Braided Cables | 4 Braided Cables |
| Diameter of flexible<br>non-articulation<br>section | 13.5mm ± .25mm | 13.5mm ± .25mm |
| Weight | 2.15 lbs ± .25 lbs | 2.15 lbs ± .25 lbs |
| Color -<br>Articulation Sheath | Black | Black |
| Color -<br>Insertion Tube | Black | Black |
| Color -<br>Proximal End Receptacle | Black | Black |
| Color -<br>Distal End Receptacle | Stainless Steel | Stainless Steel |
| Color -<br>Lettering | White | White |
| Internal Power | None | None |
| Electrical Contacts | 2 | 2 |
| Software Containing | Yes - Usage counter memory module<br>relating to the # of DLUs fired. | No |
| How Supplied | Non-Sterile, plastic wrapped in single<br>unit corrugated box | Non-Sterile, plastic wrapped in single<br>unit corrugated box |
| Recommended Method of<br>Sterilization | Steam (Flash) Sterilization; ETO; Cold<br>liquid sterilant (FDA cleared); and Gas<br>plasma sterilization | ETO; Cold liquid sterliant (FDA cleared) |
## FlexShaft Product Features Comparison Chart
{3}------------------------------------------------
K021249 (4 of 4)
# FlexShaft Product Features Comparison Chart - continued from previous page
{4}------------------------------------------------
DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/4/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized design featuring three abstract human profiles facing to the right, with flowing lines beneath them.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
# AY 1 5 2002
Ms. Barbara J. Whitman Regulatory Affairs Specialist Power Medical Interventions, Inc. 110 Union Square Drive New Hope, PA 18938
Re: K021249
Trade/Device Name: SurgASSIST™ System with FlexShaft 2 Regulation Number: 878.4750 Regulation Name: Implantable staple Regulatory Class: II Product Code: GDW Dated: April 18, 2002 Received: April 19, 2002
Dear Ms. Whitman:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate for abound in the May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). Tour coosinede Prov (110) the device, subject to the general controls provisions of the Act. The I ou may , and severy in the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{5}------------------------------------------------
Page 2 – Ms. Barbara J. Whitman
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 21 CFR Part 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4659. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html
Sincerely yours,
Miriam C. Provost
for Celia M. Witten, Ph.D., M.D. Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{6}------------------------------------------------
031249
Power Medical Interventions, Inc. SurgASSIST™ System with FlexShaft 2 Power Modification PreMarket Notification PreMarket Notification April 17, 2002
Power Medical Interventions, Inc. New Hope, PA 18938
510(k) No. K 021249
Device Name:
SurgASSIST™ System with FlexShaft 2
INDICATIONS FOR USE: (Note: The intended use for this product modification will be substantially identical to that of Power Medical Interventions, Inc. immediate 510(k) Notification's, K003277 & K012809).
The SurgASSIST™ System with Circular Stapler Digital Loading Unit™ (DLU), has applications throughout the alimentary tract for end-to-end and side-to-side anastomosis.
The SurgASSIST™ System with Right Angle Linear Cutter DLU has applications in gastrointestinal, gynecological, general abdominal and thoracic surgical procedures for resection, transection, and creation of anastomosis.
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Prescription Use
Per 21CFR §801.109
Over-The-Counter Use OR
Muriam C. Provost
(Division Sign-Off) Division of General, Restorative and Neurological Devices
510(k) Number K021249
000014
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.