← Product Code [LZA](/productcode/LZA) · K021220

# NITRILE PATIENT EXAMINATION GLOVE POWDER-FREE (BLUE COLOR) TESTED FOR USE WITH CHEMOTHERAPY DRUGS (K021220)

_Siam Sempermed Corp., Ltd. · LZA · Jun 6, 2002 · General Hospital · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K021220

## Device Facts

- **Applicant:** Siam Sempermed Corp., Ltd.
- **Product Code:** [LZA](/productcode/LZA.md)
- **Decision Date:** Jun 6, 2002
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 880.6250
- **Device Class:** Class 1
- **Review Panel:** General Hospital

## Indications for Use

A patient examination glove is a disposable device intended for medical purposes that is worn on the examiners hand or finger to prevent contamination between patient and examiner.

## Device Story

Nitrile patient examination glove; powder-free; blue color; non-sterile. Designed for use by healthcare professionals to prevent cross-contamination between patient and examiner. Tested for permeation resistance against specific chemotherapy drugs. Provides barrier protection during drug preparation and administration. Manufactured via halogenation/siliconization and extensive water washing. Surface is free from glove powder. Evaluated for primary dermal irritation and sensitization; results indicate non-irritant status.

## Clinical Evidence

Bench testing only. Testing included physical property evaluation per ASTM D6319-00a, pinhole testing per ASTM D5151, and primary dermal irritation/sensitization studies (Buehler method) per 16 CFR 1500.41. Permeation testing performed for specific chemotherapy drugs.

## Technological Characteristics

Nitrile material; powder-free; blue color; non-sterile. Surface treated via halogenation and siliconization. Meets ASTM D6319-00a (dimensions, physical properties, residue) and ASTM D5151 (pinhole).

## Regulatory Identification

A non-powdered patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner. A non-powdered patient examination glove does not incorporate powder for purposes other than manufacturing. The final finished glove includes only residual powder from manufacturing.

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Siam Sempermed Corp., Ltd. D 110 Moo 8 Kanjanavanit Rd., Hat Yai, Songkhla, Thailand 90230 Tel: 66 074 291 648 to 9 Fax: 66 074 291 650

#### 510 (k) SUMMARY

4021220

Page 1 (2)

#### 1.0 APPLICANT:

Dr. POONSUK CHERDKIATGUMCHAI SIAM SEMPERMED CORPORATION., Ltd. 110 MOO 8 KANJANAVANIT ROAD PATHONG HATYAI SONGKHLA THAILAND 90230 TEL: 66 074 291 648 OR 291 649 FAX: 66 074 291 650

#### 2.0 CONTACT PERSON

Dr. POONSUK CHERDKIATGUMCHAI SIAM SEMPERMED CORPORATION., Ltd. 110 MOO 8 KANJANAVANIT ROAD PATHONG HATYAI SONGKHLA THAILAND 90230 TEL: 66 074 291 648 OR 291 649 FAX: 66 074 291 650

MR WILLIAM HARRIS SEMPERMED USA Inc. 30798 US Hwy. 19 N Palm Harbor, USA FL 34684 TEL: 727 787 7250 FAX: 727 787 7558

3.0 Device Class: I Product code: 80LZA

4.0 Specification: Nitrile patient examination glove , Powder Free Glove , Blue color , Non sterile

Class I 80LZA meets all of the requirements of ASTM standard D6319-00a"

5.0 Device Description: Nitrile Patient Examination glove , Powder free Glove, Blue color

non sterile

Glove were tested for permeation with Chemotherapy Drugs

- 6.0 Intended use: A patient examination glove is a disposable device intended for medical purposes that is worn on the examiners hand or finger to prevent contamination between patient and examiner.
- 7.0 Surface treatment: Halogenation / Siliconization and extensive washing in water. Outer surface : Free from glove powder
- 8.0 Primary Dermal Irritation in Rabbits Guinea Pig Sensitization (Buehler) : Consumer Product Testing Co. Experiment reference number : T99-0171-1

Conclusion : According to Federal Hazardous Substances Act Regulation , (16 CFR 1500.41), and under the conditions of this test , This test article is not a primary dermal irritant

This document and its contents are confidential. Do not discuss with or give access to people not designated.

{1}------------------------------------------------

## Siam Sempermed D Corp., Ltd. Page 2 (2)

## 510 (k) SUMMARY

## 9.0 QUALITY CHARACTERISTICS

| Dimensions            | Meet ASTM D 6319-00ael                     |
|-----------------------|--------------------------------------------|
| Physical Properties   | Meet ASTM D 6319-00ael                     |
| Residue Powder        | Meet ASTM D 6319-00ael                     |
| Freedom from pinholes | Meet ASTM D 6319-00ael<br>Meet ASTM D 5151 |

10. Conclusion: Siam Sempermed Nitrile patient examination glove , Powder Free Glove , Blue color , non sterile

meet the ASTM standard or equivalent standard meet pinhole FDA requirements meet labeling claims (see 5.0 and 6.0 above)

p. Quedlinburg

Dr. POONSUK CHERDKIATGUMCHAI Chief Quality Officer Nov 12,2001

This document and its contents are confidential. Do not discuss with or give access to people not designated.

{2}------------------------------------------------

Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo is a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" arranged around the perimeter. Inside the circle is a stylized image of an eagle with its wings spread.

Public Health Service

JUN 6 2002

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

Siam Sempermed Corporation Limited C/O Mr. John V. Calhoun Sempermed USA, Incorporated 30798 US Highway 19 North Palm Harbor. Florida 34684

Re: K021220

Trade/Device Name: Nitrile Patient Examination Glove Powder-Free (Blue Color) Tested for Use with Chemotherapy Drugs Regulation Number: 880.6250 Regulation Name: Patient Examination Glove Regulatory Class: I Product Code: LZA Dated: March 18, 2002 Received: March 22, 2002

Dear Mr. Calhoun:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies.

{3}------------------------------------------------

### Page 2 - Mr. Calhoun

You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 21 CFR Part 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4618. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html

Sincerely yours,

Susan Runser

Timothy A. Ulatowski Director Division of Dental, Infection Control and General Hospital Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

# INDICATIONS FOR USE

Applicant:

Siam Sempermed Corp., Ltd.

K 021220

510(k) Number:

Device Name:

Nitrile Patient Examination Glove Powder-Free (Blue Color) Tested for Use with Chemotherapy Drugs

Indications For Use:

A Patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hands or finger to prevent contamination between patient and examiner. (21CFR 880,6250)

Gloves to be used for protection against chemotherapy drug exposure during use with chemotherapy drug preparation and administration to patients. Chemotherapy drugs tested are: Amethopterin Hydrate, Cisplatin, Cyclophosphamide, Dacarbazine, Doxorubincin Hydrochloride, Etoposide, 5-Flourouracil, Paclitaxel, and Vincristine Sulfate.

Qiu S. Lim

(Division Sign-Off) Division of Dental, Infection Control, and General Hospital Devices 510(k) Number _

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

---

**Source:** [https://fda-staging.innolitics.com/device/K021220](https://fda-staging.innolitics.com/device/K021220)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com