K021215 · Bio-Logic Systems Corp. · GWJ · May 15, 2002 · Neurology
Device Facts
Record ID
K021215
Device Name
BIO-LOGIC ABAER CUB
Applicant
Bio-Logic Systems Corp.
Product Code
GWJ · Neurology
Decision Date
May 15, 2002
Decision
SESE
Submission Type
Special
Regulation
21 CFR 882.1900
Device Class
Class 2
Attributes
Pediatric
Indications for Use
The Bio-logic Evoked Potential (EP) product family is indicated for use in the recording and analysis of human physiological data necessary for the diagnosis of auditory and hearing-related disorders. This product, the ABaer Cub, like it's predicate device, the ABaer, is especially indicated for use in the screening of infants to determine hearing loss. The Bio-logic EP System can be used for patients of all ages, from children to adults, including infants and geriatric patients. It is especially indicated for use in testing individuals for whom behavioral audiometric results are deemed unreliable, such as infants, young children, and cognitively impaired or uncooperative adults. The use of the Bio-logic EP family of products is to be performed under the prescription and supervision of a physician or other trained health care professional.
Device Story
ABaer Cub is a portable, battery-powered evoked response system for auditory screening. It records human physiological data to assist in diagnosing hearing disorders. The device uses a Digital Signal Processor (DSP) for real-time signal acquisition and processing, employing the Point Optimized Variance Ratio (POVR) algorithm to monitor signal-to-noise ratios and optimize signal quality. Input consists of physiological signals; output includes automated Pass/Refer screening recommendations and raw data for clinician review. Operated by nurses, audiologists, or physicians in clinical settings, the system interfaces with a PDA (e.g., HP Jornada) running Windows CE for display and control. The device enhances portability compared to the predicate by integrating battery power and PDA-based host control. Clinicians use the output to support diagnostic decisions; the system allows manual review/override of automated results.
Clinical Evidence
No clinical data provided. Safety and effectiveness established through design controls, hazard/risk analysis (Fault Tree Analysis per EN-1441), and bench testing of hardware/software modifications to ensure performance equivalence to the predicate device.
Technological Characteristics
Portable, battery-powered evoked response auditory stimulator. Features a Digital Signal Processor (DSP) for real-time data acquisition and POVR algorithm execution. Host interface via PDA (Windows CE). Patient isolation methods consistent with predicate. Materials and hardware design are substantially similar to the predicate, with added battery charging circuitry and enclosure modifications for portability. Software developed in C++.
Indications for Use
Indicated for recording/analysis of human physiological data for diagnosis of auditory/hearing disorders. Especially indicated for infant hearing screening. Suitable for all ages, including children, adults, geriatric patients, and individuals where behavioral audiometry is unreliable (e.g., cognitively impaired or uncooperative patients).
Regulatory Classification
Identification
An evoked response auditory stimulator is a device that produces a sound stimulus for use in evoked response measurements or electroencephalogram activation.
Predicate Devices
ABAER / Navigator Pro Evoked Potential device (K994149)
Navigator and Traveler Evoked Potential Product (K930328)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for Bio-logic Systems Corp. The logo is in black and white. The word "Bio-logic" is in a bold font, with a registered trademark symbol to the right of the word. Below the word "Bio-logic" is the text "SYSTEMS CORP" in a smaller font.
One Bio-logic Plaza Mundelein, IL 60060-3700 1-800-323-8326 Fax: 847-949-8615 www.bio-logic.com
Kozlers
## SECTION 2: SUMMARY AND CERTIFICATION
## 510(K) SUMMARY
Safety and effectiveness information concerning the ABaer Cub device modification to the Biologic Evoked Potential product family is summarized below.
Because this is not a CLASS III device, the special certification defined for this section is not required.
Bio-logic Systems Corp PREPARED BY: One Bio-logic Plaza Mundelein, IL 60060
TELEPHONE: (847)-949-5200
CONTACT PERSON: Norman E. Brunner
#### DATE ON WHICH THE SUMMARY WAS PREPARED: April 12, 2002
NAME OF DEVICE: Bio-logic ABaer Cub.
Evoked Response System. COMMON NAME:
CLASSIFICATION NAME: Evoked Response Auditory Stimulator (per CFR 882-1900).
PREDICATE DEVICE: ABAER / Navigator Pro Evoked Potential device, reference 510(k) #K994149.
#### DESCRIPTION OF THE DEVICE:
The Bio-logic Evoked Potential family of products is intended to be used for the recording and analysis of human physiological data for the purpose of neurological diagnosis and treatment of sensory disorders. The predicate device referenced above is the latest in a series of this type marketed by Bio-logic. Other related devices comprising the Evoked Potential family include:
- 510(k) #K803226 -- Bio-logic Evoked Response Stimulators. 1.
- 510(k) #K842543 Bio-logic Evoked Potential System. 2.
- 510(k) #K844992 Bio-logic Portable Evoked Response System. 3.
- 510(k) #K862690 Bio-logic Traveler LT System. 4.
- 5. 510(k) #K930328 - Navigator and Traveler Evoked Potential Product.
{1}------------------------------------------------
02/2/5
Section 2
The ABaer Predicate Device performs Evoked Potential screening, recording and analysis functions, provides one channel of data recording, and includes the Point Optimized Variance Ratio (POVR) algorithm for optimizing signal quality, implementing the screening function and enhancing speed of test completion. This new ABaer Cub device performs these same functions in essentially the same ways, but has hardware and software modifications and improvements over the Predicate device, primarily to enhance portability and ease of use.
The Predicate Device describes the use of the POVR algorithm to monitor the signal-to-noise ratio of the averaged response signal every 256 stimulation-response cycles. This modified device, ABaer Cub, incorporates the same POVR algorithm for the same purpose, to yield highquality test results in the shortest possible time. In addition, the feature modifications for this new device include the use of a Personal Digital Assistant (PDA) or "Pocket PC" which can be hand-carried separately or mounted right on the ABaer box, resulting in a completely batterypowered, portable hand-held test instrument. The Predicate Device uses a desktop or laptop computer instead of the PDA. The electronic hardware is very similar in design to the ABaer hardware used in the Predicate Device. The enclosure has been made slightly larger (deeper) to allow space to incorporate a rechargeable battery inside the unit. New circuitry for charging the battery has also been added to the unit, and the small LCD display in the ABaer has been removed from the ABaer Cub. The functions of the hardware are substantially the same as those of the Predicate Device hardware. The ABaer predicate device communicates test information and results to a separate host computer (desktop PC or laptop) through a serial communications port. In the ABaer Cub, this host computer is now a PDA or "pocket PC" (for instance, the Hewlett-Packard Jornada) which can be affixed right on the ABaer unit or hand-carried separately. Because both the PDA and the ABaer unit are battery powered, this significantly improves the portability and usability of the device. The ABaer Cub software for control of this device is a simplified subset of the ABaer Predicate Device software, with some GUI changes necessary to enhance the graphics display for use with the smaller PDA LCD screen. The host software for the Predicate Device is Windows 95/98/ME based, whereas the host software for the ABaer Cub is Windows CE based. The C++ programming language is used in both cases. Together, these minor hardware and software changes implement the same functionality and perform the same intended use as the Predicate Device, but with improved portability and easeof-use.
#### INTENDED USE:
The Bio-logic Evoked Potential (EP) product family is indicated for use in the recording and analysis of human physiological data necessary for the diagnosis of auditory and hearing-related disorders.
This product, the ABaer Cub, like it's predicate device, the ABaer, is especially indicated for use in the screening of infants to determine hearing loss.
The Bio-logic EP System can be used for patients of all ages, from children to adults, including infants and geriatric patients. It is especially indicated for use in testing individuals for whom behavioral audiometric results are deemed unreliable, such as infants, young children, and cognitively impaired or uncooperative adults. The use of the Bio-logic EP family of products is to be performed under the prescription and supervision of a physician or other trained health care professional.
{2}------------------------------------------------
Special 510(k): ABaer Cub, Device Modification to Bio-logic Evoked Potential Product
## SAFETY AND EFFECTIVENESS SUMMARY
To establish the safety and effectiveness of this modification to the Bio-logic Evoked Potential software, the modification was designed and incorporated into the product in accordance with the Bio-logic internal Product Development procedures which are intended to meet ISO-9001, EN-46001 and FDA QSR Design Control specifications. A detailed Hazard/Risk analysis for the EP family of products was performed using the Fault Tree analysis (FTA) approach, and a detailed Risk Assessment was written in accordance with EN-1441, the International Standard for Hazard/Risk analysis. An addendum to this Hazard/Risk file was written based on a review of this new ABaer Cub design.
The ABaer Cub patient-connection hardware is basically the same as that of the ABaer. There are no newly-introduced hardware-related methods by which the patient can be harmed or injured through the use of this device. The same patient isolation methods are used in both products. Both devices utilize a medical-grade power supply, although the ABaer Cub will be used most often in its portable, battery-operated mode. Direct hardware control of all ABaer Cub functions is provided from the Digital Signal Processor (DSP) and its program code located inside the ABaer Cub package, just as it is in the ABaer. By distributing the hardware-specific functions to the DSP, the Windows-based host computer program has fewer real-time demands, resulting is high reliability and performance.
Nurses, audiologists, and physicians can use the ABaer Cub system. No knowledge of or training in Evoked Potentials is necessary. Anyone, even a user with limited computer skills, who has been trained to operate this equipment and trained in techniques for proper infant handling, can use the ABaer Cub effectively.
The ABAER Cub software does not make any final decisions that result in any direct forms of diagnosis or treatment. The Pass/Refer recommendations of the automated screening process in the ABaer Cub can be reviewed, if desired, by a qualified health care professional, and may be modified, overridden or deleted as determined by the qualified user. The program provides additional functionality to allow the qualified user to review all raw data collected, and to perform other data analysis to suit his or her requirements. This optional data review capability allows the Physician to view the same data that the ABaer Cub used to make its recommendations.
The chart on the following page provides a summary comparison of the technological characteristics of the new ABaer Cub device relative to the predicate ABaer device. This is to demonstrate that this new ABAER Cub device has no significant differences which would adversely affect product safety and effectiveness.
{3}------------------------------------------------
#### DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized image of three abstract shapes that resemble birds in flight.
A. ANSWERS TO QUESTIONS
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
## MAY 1 5 2002
Mr. Norman E. Brunner Vice-President of Research and Development Bio-logic Systems Corporation One Bio-logic Plaza Mundelein, IL 60060-3700
Re: K021215
Trade/Device Name: Bio-logic Abaer Cub Regulation Number: 882.1900 Regulation Name: Evoked response auditory stimulator Regulatory Class: II Product Code: GWJ Dated: April 16, 2002 Received: April 17, 2002
Dear Mr. Brunner:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{4}------------------------------------------------
Page 2 - Mr. Norman E. Brunner
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 21 CFR Part 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4659. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html
Sincerely yours,
Miriam C. Provost
for Celia M. Witten, Ph.D., M.D. Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
#### INDICATIONS FOR USE STATEMENT
510(k) Number (if known):
Not Assigned
## Device Name: ABaer Cub, Modification to Bio-logic Evoked Potential Product.
#### Indications For Use:
The Bio-logic Evoked Potential (EP) product family is indicated for use in the recording and analysis of human physiological The Dio 10g.ro Droxee diagnosis of auditory and hearing-related disorders.
This product, the ABaer Cub, like it's predicate device, the ABaer, is especially indicated for use in the screening of infants to determine hearing loss.
The Bio-logic EP System can be used for patients of all ages, from children to adults, including infants. It The DV-10glt LT System can oc use to parties of an agos behavioral audiometric results are deemed unreliable, such as is especially macated for use in essing marridaas for wisen other standable. The use of the Bio-logic EP family of products is to mans, young cindicit, and cogments of interest or of a physician or other trained health care professional.
# (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Miriam C. Provost
eral, Restorative and Neurological Devices
510(k) Number K021215
Prescription Use
(Per 21 CFR 801.109)
OR
Over-The-Counter Use
(Optional Format 1-2-96)
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.