# GYNEFLEX (K021115)

_Naissance Holdings, LC · HIR · Dec 20, 2002 · Obstetrics/Gynecology · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K021115

## Device Facts

- **Applicant:** Naissance Holdings, LC
- **Product Code:** [HIR](/productcode/HIR.md)
- **Decision Date:** Dec 20, 2002
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 884.1425
- **Device Class:** Class 2
- **Review Panel:** Obstetrics/Gynecology
- **Attributes:** Therapeutic

## Indications for Use

The GyneFlex is recommended for the strengthening of the perineal pelvic floor muscles by offering resistance to an individual's voluntary contractions of these muscles. It seeks to correct, through exercise, urinary incontinence.

## Device Story

GyneFlex is a V-shaped, reusable pelvic floor muscle exerciser; constructed of polymer plastic. Device inserted into vagina; provides graduated resistance during voluntary pelvic muscle contractions. Used as adjunct to Kegel exercises to train proper muscle contraction and treat urinary incontinence. Operated by patient; no external power or electronic components. Provides physical resistance feedback to user; helps strengthen perineal muscles.

## Clinical Evidence

No clinical data. Bench testing only; biocompatibility testing performed per ISO 10993 standards confirming material is nontoxic and well-tolerated by mucosal membranes.

## Technological Characteristics

V-shaped, reusable polymer plastic exerciser. Provides six ranges of mechanical resistance. No energy source. Biocompatible per ISO 10993.

## Regulatory Identification

A perineometer is a device consisting of a fluid-filled sack for intravaginal use that is attached to an external manometer. The devices measure the strength of the perineal muscles by offering resistence to a patient's voluntary contractions of these muscles and is used to diagnose and to correct, through exercise, uninary incontinence or sexual dysfunction.

## Predicate Devices

- Colonial Medical Supply Pelvic Muscle Therapy ([K002830](/device/K002830.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

### ATTACHMENT 2

### 510(k) SUMMARY

### K021115

### (Revised)

# GyneFlex

Common/Classification Name: Perineometer, 21 CFR 884.1425

Naissance Holdings, L.C. 4905 34th Street South St. Petersburg, FL 33711

Contact: Suzanne B. Sloan Nancy E. Taylor

Prepared: April 4, 2002 Revised: December 9, 2002

#### LEGALLY MARKETED DEVICE A.

The Colonial Medical Supply Pelvic Muscle Therapy was cleared for marketing on December 27, 2000, in premarket notification K002830. Naissance Holdings, L.C., submits this premarket notification for GyneFlex.

#### DEVICE DESCRIPTION B.

The GyneFlex is a pelvic floor muscle exerciser. The product has six ranges of resistance. Other than variation in resistance, each exerciser is identical, including identical indications and labeling.

It is a "v" sh aped device that provides resistance after insertion into the vagina when the patient contracts her pelvic muscles. The material is made of a polymer plastic specifically formulated to provide specific ranges of resistance for use as an adjunct to Kegel exercises.

#### C. INTENDED USE

Naissance Holdings, LLC Revised 510(k) for K021115 December 9, 2002

{1}------------------------------------------------

The GyneFlex is recommended for the strengthening of the perineal muscles by offering resistance to an individual's vo luntary contractions of these muscles. It seeks to correct, through exercise, urinary incontinence.

#### SUBSTANTIAL EQUIVALENCE SUMMARY D.

The GyneFlex is substantially equivalent to the Colonial Medical Supply Pelvic Muscle Therapy, a perineometer, which was cleared for marketing on December 27, 2000, in premarket notification K002830.

The GyneFlex has indications for use that are similar to those as the legally marketed predicate device. Both the GyneFlex and the predicate device have identical indications concerning the strengthening of the pelvic floor muscles, which has been found to help women with incontinence, is consistent with 21 C.F.R. 884.1425. GyneFlex provides graduated resistance in specified ranges that help train the user in proper and effective pelvic floor muscle contractions and exercises, which is similar to the biofeedback provided by the predicate device.

#### ய் TECHNOLOGICAL CHARACTERISTICS

The GyneFlex is a single, reusable, polymer plastic "v" shaped exerciser that exceeds the guidelines set forth in ISO 10993.

#### F. TESTING

Testing results indicate that the material is biocompatible, nontoxic and well tolerated by mucosal membranes.

#### G. CONCLUSIONS

This premarket submission has demonstrated that the GyneFlex is substantially

equivalent to a device previously cleared for marketing by FDA.

Naissance Holdings, LLC Revised 510(k) for K021115 December 9, 2002

{2}------------------------------------------------

# DEPARTMENT OF HEALTH & HUMAN SERVICES

Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized symbol that resembles a bird or an abstract human figure.

Public Health Service

DFC 2 0 2002

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

Naissance Holdings, L.C. % Ms. Eleanor A. Kolton Greenburg Traurig, LLP Attorneys at Law 800 Connecticut Avenue, Suite 500 WASHINGTON D.C. 20006

Re: K021115

Trade/Device Name: GyneFlex Regulation Number: 21 CFR 884.1425 Regulation Name: Perineometer Regulatory Class: II Product Code: 85 HIR Dated: October 24, 2002 Received: October 28, 2002

Dear Ms. Kolton:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (sections 531-542 of the Act); 21 CFR 1000-1050.

{3}------------------------------------------------

This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at one of the following numbers, based on the regulation number at the top of this letter:

| 8xx.1xxx                         | (301) 594-4591 |
|----------------------------------|----------------|
| 876.2xxx, 3xxx, 4xxx, 5xxx       | (301) 594-4616 |
| 884.2xxx, 3xxx, 4xxx, 5xxx, 6xxx | (301) 594-4616 |
| 892.2xxx, 3xxx, 4xxx, 5xxx       | (301) 594-4654 |
| Other                            | (301) 594-4692 |
|                                  |                |

Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21 CFR Part 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html.

Sincerely yours,

Henry C. Brogdon

Nancy C. Brogdon Director, Division of Reproductive, Abdominal, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

# ATTACHMENT 3

# STATEMENT OF INDICATIONS FOR USE

510(k) Number (if known): K021115

Device name: GyneFlex

Indications for Use:

The GyneFlex is recommended for the strengthening of the perineal pelvic floor muscles by offering resistance to an individual's voluntary contractions of these muscles. It seeks to correct, through exercise, urinary incontinence.

(PLEASE DO NOT WRITE BELOW THIS LINE -- CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Prescription Use (Per 21 CFR 801.109) OR

Over-the-Counter Use V

David A. Legman

(Division Sign-Off) Division of Reproductive, Abdominal, and Radiological Devices 510(k) Number _

Naissance Holdings, LLC Revised 510(k) for K021115 December 9, 2002

---

**Source:** [https://fda-staging.innolitics.com/device/K021115](https://fda-staging.innolitics.com/device/K021115)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).