The X-SIZER Catheter System is indicated for the mechanical removal of thrombus in synthetic hemodialysis grafts.
Device Story
Single-use, disposable thrombectomy device for synthetic hemodialysis grafts. Consists of 135cm coaxial, dual-lumen hydrophilic-coated catheter connected to hand-held Control Module. Inner lumen contains torque cable in polyamide tube; distal tip features helical cutter. Outer lumen serves as vacuum path for debris removal. Motor (powered by 9V alkaline battery) rotates cutter at ~2,100 rpm. Physician operates device over coronary guidewire. Vacuum system in Control Module extracts thrombotic debris. Device enables mechanical thrombus removal, potentially restoring graft patency and reducing need for more invasive interventions.
Clinical Evidence
No human clinical data. Evidence consists of in vitro bench testing (functional operation, leak, battery life, corrosion, flow rate, kink resistance, tensile/torque/flexibility, and thrombus cutting) and in vivo animal model testing in thrombosed hemodialysis grafts. Animal study compared X-SIZER to Edwards Thrombex PMT, showing equivalent performance regarding vessel trauma, distal embolization, and thrombus removal efficacy.
Technological Characteristics
Coaxial, dual-lumen catheter (135cm). Materials: stainless steel, PEBAX, polyamide. Energy source: 9V alkaline battery-powered motor. Operation: mechanical rotary cutter (~2,100 rpm) with vacuum-assisted debris aspiration. Form factor: 1.5mm and 2.0mm diameter models. Sterile, single-use. Guidewire compatible (0.014").
Indications for Use
Indicated for mechanical removal of thrombus in synthetic hemodialysis grafts. Contraindicated in grafts inaccessible by 0.014" guidewire, grafts with insufficient diameter, patients with prolonged Prothrombin/Partial Prothrombin Time or other anticoagulation disorders, graft infection, significant cardiopulmonary vascular compromise, or allergy to procedure-related medications.
Regulatory Classification
Identification
An intraluminal artery stripper is a device used to perform an endarterectomy (removal of plaque deposits from arterisclerotic arteries.)
Predicate Devices
Edwards Lifesciences Thrombex PMT™ System (K993816)
{0}------------------------------------------------
## PREMARKET NOTIFICATION 510(k) SUMMARY (Per 21 CFR Part 807.92)
| Submitter's Name: | ev3 Inc. |
|------------------------|-------------------------------------------------------------------|
| Address: | 4600 Nathan Lane North<br>Plymouth, MN 55442 USA |
| Telephone No.: | 763-398-7000 |
| Facsimile No.: | 763-398-7200 |
| Contact Person: | Jill Munsinger |
| Date Summary Prepared: | June 9, 2004 |
| Trade Name: | X-SIZER® Catheter System |
| Common or Usual Name: | Thrombectomy Device |
| Classification Name: | Catheter, Peripheral, Atherectomy (MCW)<br>(21 CFR Part 870.4875) |
#### Legally marketed devices to which ev3 Inc. claims substantial equivalence:
| Edwards Lifesciences Thrombex PMT™ System | 510(k) No. K993816 |
|--------------------------------------------|--------------------|
| Solera Bacchus Thrombectomy Catheter (BTC) | 510(k) No. K003570 |
| Possis AngioJet® LF140 Catheter | 510(k) No. K960970 |
#### Description of the device that is the subject of this Premarket Notification:
The X-SIZER System is a single-use, disposable medical device intended for the selective removal of thrombus from blood vessels. The device consists of a 135cm coaxial, dual lumen hydrophilic coated outer catheter shaft connected to a hand-held Control Module. The inner lumen of the catheter is composed of a hollow torque cable encapsulated in a polyamide tube (flexible drive shaft). A helical cutter is contained within the cutter housing located at the distal tip of the catheter. The outer lumen of the catheter is the path in which the excised debris is removed by vacuum. In use, the cutter is rotated at approximately 2,100 pm by the motor, which is powered by a 9-volt alkaline battery. In addition to turning the helical cutter, the inner lumen provides the space through which a coronary guidewire is placed. The hand-held Control Module contains the motor, battery, circuitry, indicators, and various valves and manifolds for the vacuum removal of thrombotic debris. The X-SIZER System is available in the 1.5mm and 2.0mm diameter models.
#### Intended Use:
The X-SIZER Catheter System is indicated for the mechanical removal of thrombus in synthetic hemodialysis grafts.
{1}------------------------------------------------
## Contraindications:
The X-SIZER Catheter System is contraindicated under the following circumstances and clinical conditions:
- . Grafts that cannot be accessed with an 0.014" diameter guidewire.
- Grafts with reference diameters by visual estimation less than the minimum . recommended diameters.
- A Prolonged Prothrombin Time (PPT) or Partial Prothrombin Time (PTT) or . other documented anticoagulation state or disorder.
- . Graft Infection
- Significant cardiopulmonary vascular compromise. .
- Allergy or other contraindication to procedure related medications. .
## Technological Characteristics:
The X-SIZER Catheter System has the same technological characteristics as the predicate Edwards Lifesciences Thrombex PMT System, the Solera Bacchus Thrombectomv Catheter (BTC), and the Possis AngioJet LF140 Catheter. All of these devices, except the Possis AngioJet, use mechanical rotary motion to remove thrombotic materials from dialysis grafts. The X-SIZER Catheter System, the Thrombex PMT System, and the Possis AngioJet LF140 Catheter are constructed of stainless steel and PEBAX materials. All of the devices are designed for use over a guidewire and are packaged as sterile, single use devices.
## Performance Data:
The X-SIZER Catheter System was subjected to in vitro bench tests including: basic functional operation, leak tests, battery life test, corrosion resistance, motor current and trackability test, hydrophilic coating test, flow rate test, device output test, kink resistance test, tensile tests, bond tests, joint tests, torque tests, flexibility (deflection) tests, shaft torque resistance, cutter/housing binding, and bench top thrombus cutting (extraction).
Comparative performance testing was performed with the X-SIZER Catheter System and a predicate device. the Edwards Lifesciences Thrombex PMT System, in a bench model simulating synthetic hemodialysis grafts. The X-SIZER Catheter System performed equivalently to the Thrombex PMT System when assessing clot extraction and safety within a synthetic graft material.
Additionally, the X-SIZER Catheter System and the predicate device, the Thrombex PMT System, were evaluated in an in vivo animal model utilizing thrombosed hemodialysis grafts. The X-SIZER System performed equivalently to the Thrombex PMT System when assessing vessel trauma, distal embolization, and overall device performance for the removal of thrombus in occluded dialysis grafts.
The results of the in vitro and in vivo tests performed demonstrate the device is as safe and effective as the legally marketed predicate devices.
{2}------------------------------------------------
ev3 Inc. X-SIZER 510(k) Amendment K021096/A7
### Summary of Substantial Equivalence:
The X-SIZER Catheter System is substantially equivalent to the predicate Edwards Lifesciences Thrombex PMT System, the Solera Bacchus Thrombectomy Catheter (BTC), and the Possis AngioJet LF140 Catheter. The indications for use, principles of operation, methods of manufacture and materials are substantially equivalent to these legally marketed devices.
.
{3}------------------------------------------------
Public Health Service
Image /page/3/Picture/2 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo is circular and contains the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is an abstract symbol that resembles an eagle or a stylized human figure.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
SEP - 8 2004
ev3. Inc. c/o Ms. Brenda Johnson Senior Regulatory Affairs Specialist 4600 Nathan Lane North Plymouth, MN 55442
Re: K021096 X-SIZER Regulation Number: 21 CFR 870.4875 Regulation Name: Catheter, Peripheral, Atherectomy Regulatory Class: Class II Product Code: MCW Dated: June 9, 2004 Received: June 10, 2004
Dear Ms. Johnson:
We have reviewed your Section 510(k) premarket notification of intent to market the device we nave roviewed your be determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate for use stated in the encreationery to ttg_mg actment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, devices that have been recuire approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The r ou may, aterers, missus of the Act include requirements for annual registration, listing of general controlo provided the practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it If your device is elassince (600 at controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{4}------------------------------------------------
Page 2 -- Ms. Brenda Johnson
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (301) 594-4646. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html
Sincerely yours,
Durina D. Vochner
(A Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
# Indications for Use
510(k) Number (if known): _K021096
Device Name:_X-SIZER® Catheter System
Indications For Use:
The X-SIZER Catheter System is indicated for the mechanical removal of thrombus in synthetic hemodialysis grafts.
Prescription Use __ X (Part 21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use _ (21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
R. Lochner
(Division Sign. Off) (Division Of Cardiovascular Devices
510(K) Number_K02109 (609
Page 1 of 1
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.