← Product Code [KTT](/productcode/KTT) · K021091

# EBI XFIX DFS SYSTEM (K021091)

_Ebi, L.P. · KTT · May 3, 2002 · Orthopedic · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K021091

## Device Facts

- **Applicant:** Ebi, L.P.
- **Product Code:** [KTT](/productcode/KTT.md)
- **Decision Date:** May 3, 2002
- **Decision:** SESE
- **Submission Type:** Special
- **Regulation:** 21 CFR 888.3030
- **Device Class:** Class 2
- **Review Panel:** Orthopedic
- **Attributes:** Therapeutic

## Indications for Use

The EBI® XFIX® DFS® System is a unilateral external fixation device intended for use in the treatment of bone conditions including leg lengthening, osteotomies, arthrodesis, fracture fixation, and other bone conditions amenable to treatment by use of the external fixation modality.

## Device Story

EBI XFIX DFS System is a unilateral external fixation device; consists of metallic bone screws and external fixation frame components. Bone screws are inserted through skin and soft tissue into bone; fixator frame attaches to screw shanks. Used by surgeons for stabilization in leg lengthening, osteotomies, arthrodesis, and fracture fixation. Modification to ankle clamp component. Device provides mechanical stabilization of bone segments to facilitate healing.

## Clinical Evidence

No clinical data provided; substantial equivalence based on technological characteristics and design comparison.

## Technological Characteristics

External fixation system; components manufactured from titanium, stainless steel, and aluminum. Unilateral frame configuration. Mechanical fixation principle. Non-sterile or sterile (implied by standard orthopedic practice). No software or electronic components.

## Regulatory Identification

Single/multiple component metallic bone fixation appliances and accessories are devices intended to be implanted consisting of one or more metallic components and their metallic fasteners. The devices contain a plate, a nail/plate combination, or a blade/plate combination that are made of alloys, such as cobalt-chromium-molybdenum, stainless steel, and titanium, that are intended to be held in position with fasteners, such as screws and nails, or bolts, nuts, and washers. These devices are used for fixation of fractures of the proximal or distal end of long bones, such as intracapsular, intertrochanteric, intercervical, supracondylar, or condylar fractures of the femur; for fusion of a joint; or for surgical procedures that involve cutting a bone. The devices may be implanted or attached through the skin so that a pulling force (traction) may be applied to the skeletal system.

## Predicate Devices

- EBI® XFIX® DFS® System ([K953406](/device/K953406.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

K021091

page 1 of 2

#### 510(k) Summary of Safety & Effectiveness

This 510(k) Summary of Safety and Effectiveness for the EBI XFIX® DFS® System is provided as required per Section 513(3) of the Food, Drug and Cosmetic Act.

- Contact Person: Frederic Testa, RAC 1. Submitter: EBI, L.P. Telephone: (973) 299-9300, x2208 100 Interpace Parkway Parsippany, NJ 07054
April 3, 2002 Date prepared:

| 2. Proprietary Name:  | EBI® XFIX® DFS® System                              |
|-----------------------|-----------------------------------------------------|
| Common Name:          | External Fixation Device                            |
| Classification Names: | Single/Multiple Component Metallic Bone<br>Fixation |
|                       | Appliances and Accessories, 21 CFR 888.303          |

## 3. Predicate or legally marketed devices that are substantially equivalent:

- + EBI® XFIX® DFS® System EBI, L.P. (K953406)
- 4. Description of the device: The EBI® XFIX® DFS® System has not changed the Indications for Use or fundamental scientific technology of the previously cleared system. The system consists of external fixation components and implantable bone screws. The EBI® XFIX® DFS® System is utilized in the following manner: bone screws are inserted through the patient's skin and soft tissue and into the bone. The fixator frame of the EBI® XFIX® DFS® System is attached to the shanks of the bone screws. This submission is for the modifications to the Ankle Clamp of the EBI® XFIX® DFS® System.
- The EBI® XFIX® DFS® System is intended for use in the treatment of bone 5. Intended Use: conditions including leg lengthening, osteotomies, arthrodesis, fracture fixation, and other bone conditions amenable to treatment by use of the external fixation modality.

{1}------------------------------------------------

Koo1091

page 2 of 2

- 6. Materials: The components of the System may be manufactured from materials such as titanium, stainless steel, and aluminum.
- 7. Comparison of the technological characteristics of the device to predicate devices: There are no significant differences between the EBI® XFIX® DFS® Fixation System and other currently marketed external fixation systems. It is substantially equivalent* to the predicate device in regards to intended use, materials, and function.

*Any statement made in conjunction with this submission regarding a determination of substantial equivalence to any other product is intended only to relate to whether the product can be lawfully marketed without pre-market approval or reclassification and is not intended to be interpreted as an admission or any other type of evidence in patent infringement litigation. [Establishment Registration and Premarket Notification Procedures, Final Regulation, Preamble, August 23, 1977, FR 42520 (Docket No. 76N-0355.)]

{2}------------------------------------------------

## DEPARTMENT OF HEALTH & HUMAN SERVICES

Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" around the perimeter. Inside the circle is a stylized image of an eagle with three wavy lines representing its wings.

#### Public Health Service

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

Mr. Jon Caparotta, RAC Manager, Regulatory Affairs EBI, L.P. 100 Interpace Parkway Parsippany, NJ 07054

Re: K021091

Trade/Device Name: EBX® XFIX® DFS® System Regulation Number: 888.3030 Regulation Name: Single/multiple component metallic bone fixation appliances and accessories Regulatory Class: II Product Code: KTT Dated: April 3, 2002 Received: April 4, 2002

MAY 0 3 2002

Dear Mr. Caparotta:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

{3}------------------------------------------------

### Page 2 - Mr. Jon Caparotta, RAC

This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 21 CFR Part 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4659. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html

Sincerely vours,

Mach N. Millener

Celia Witten, Ph.D., M.D. Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

# STATEMENT OF INDICATIONS FOR USE

Page 1 of __ 1

510(k) Number (if known): KO2 / O9 /

Device Name: EBI® XFIX® DFS® System

Indications For Use:

The EBI® XFIX® DFS® System is a unilateral external fixation device intended for use in the treatment of bone conditions including leg lengthening, osteotomies, arthrodesis, fracture fixation, and other bone conditions amenable to treatment by use of the external fixation modality.

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Prescription Use______________________________________________________________________________________________________________________________________________________________ (Per 21 CFR 801.109) OR

Over-The-Counter Use

(Optional Format 1-2-96)

fo Mark H. Milleson

Division of General, Restorative and Neurological Devices

510(k) Number_

---

**Source:** [https://fda-staging.innolitics.com/device/K021091](https://fda-staging.innolitics.com/device/K021091)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).