K021064 · Scican · FLE · Aug 8, 2002 · General Hospital
Device Facts
Record ID
K021064
Device Name
QUANTIM 16 AUTOCLAVE
Applicant
Scican
Product Code
FLE · General Hospital
Decision Date
Aug 8, 2002
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 880.6880
Device Class
Class 2
Indications for Use
The QUANTIM 16 tabletop autoclave is a gravity steam autoclave designed to process medical and dental instruments to achieve successful sterilization in a clinical setting. It utilizes saturated steam at high pressures in order to attain an effective kill of infectious bio-organisms and prevent cross infection. The unit is intended to sterilize heat and moisture stable medical and dental instruments (including dental handpieces) which are commonly found in medical and dental offices, hospitals, clinics, and other facilities. The instruments must be suitable for steam sterilization at 132°C (270°F). Two additional cycles are also specified for (i) the sterilization of textiles and wrapped surgical packs and (ii) liquid loads contained in open top glass flasks. However, the QUANTIM 16 Table-top Autoclave is not intended nor recommended for sterilization of liquids intended for direct patient contact.
Device Story
Gravity steam autoclave; processes medical/dental instruments. Input: heat/moisture stable instruments placed on trays. Operation: small boiler beneath chamber generates saturated steam; high-pressure steam kills bio-organisms. Features four programmed cycles (Unwrapped, Wrapped, Packs, Liquids); automatic 30-minute vacuum drying cycle post-sterilization. Used in hospitals, clinics, and dental offices; operated by clinical staff. Output: sterilized instruments. Benefit: effective sterilization, reduced chamber build-up, extended chamber life.
Clinical Evidence
Bench testing only. Laboratory validation performed using Bacillus stearothermophilus spores per FDA 1993 Guidance for Sterilizers. Testing included 1/2 cycle, total kill end-point, and simulated use testing with medical/dental instruments and handpieces. Results demonstrated no detectable live spores post-processing, confirming sterilization to acceptable SAL.
Technological Characteristics
Gravity steam autoclave; utilizes small external boiler for steam generation. Four programmed cycles (121°C/132°C). Features automatic vacuum drying cycle. Materials and controls similar to predicate. Standalone unit.
Indications for Use
Indicated for sterilization of heat and moisture stable medical and dental instruments, including dental handpieces, textiles, wrapped surgical packs, and liquids in open-top flasks (not for direct patient contact) in clinical settings. Instruments must be compatible with 132°C steam sterilization.
Regulatory Classification
Identification
A steam sterilizer (autoclave) is a device that is intended for use by a health care provider to sterilize medical products by means of pressurized steam.
{0}------------------------------------------------
# AUG 0 8 2002
## Summary of Safety & Effectiveness Information
# K021064
#### 510(k) Summary:
| Submitted by: | SciCan, Division of Lux & Zwingenberger Ltd.<br>1440 Don Mills Road<br>Toronto, Ontario Canada<br>M3B 3P9 |
|----------------------|-----------------------------------------------------------------------------------------------------------|
| Contact Person: | Brenda Murphy - Regulatory Affairs Manager<br>(416) 446-2797 |
| Date of Preparation: | March 27, 2002 |
| Name of Device: | QUANTIM 16 Autoclave |
| Predicate Device: | Midmark M-11 Ultraclave Steam Sterilizer<br>510(k) 990189 |
#### Description of Device:
The QUANTIM 16 Autoclave is a gravity steam autoclave designed to process medical and dental instruments to achieve successful sterilization. The unit utilizes saturated steam at high pressures in order to attain an effective kill of infectious bio-organisms and prevent cross infection. Instruments are simply placed within the autoclave on the trays or carriers provided, the doors are closed, the cycle parameters selected and the start button depressed. The unit is then fully automatic for the complete sterilizing cycle.
#### Intended use:
The unit is intended to sterilize heat and moisture stable medical and dental instruments (including dental handpieces) within a hospital or clinical setting such as medical and/or dental surgeries. The instruments processed within the QUANTIM 16 Autoclave must be suitable for steam sterilization at 132℃ (270°F).
{1}------------------------------------------------
The sterilization cycles as to established times, temperatures and indicated uses for the QUANTIM 16 Autoclave are as follows:
| CYCLE | TEMPERATURE | TIME | INTENDED USE |
|-----------|----------------|---------|--------------------------------------------------------------------------------------------------------------------|
| UNWRAPPED | 132°C (270°F) | 3 min. | Flash cycle for solid instruments, hinged<br>instruments, dental handpieces, with drying |
| WRAPPED | 132°C (270°F) | 15 min. | Solid & hinged instruments, dental<br>handpieces, with drying |
| PACKS | 121°C (250°F). | 30 min. | Textiles & wrapped surgical packs with<br>drying |
| LIQUIDS | 121°C (250°F). | 30 min. | Liquids suitable for sterilization in open top<br>flasks, with drying<br>(not intended for direct patient contact) |
## Technological Characteristics Compared to the Predicate Device:
The OUANTIM 16 Autoclave uses similar technological characteristics of the predicate device. In this respect, the intended use, operating principle, general materials of construction, and controls are similar for both devices. Both units use saturated steam at high pressures and temperatures to achieve the complete destruction of micro-organisms.
The Midmark M-11 Sterilizer generates the steam inside the sterilization chamber by use of an electric heating element.
In an innovative process, the QUANTIM 16 actually generates the steam in a small boiler located beneath the main chamber and which then quickly fills the chamber with saturated steam. This process serves to reduce chamber build-up and extend the chamber life significantly.
The QUANTIM 16 Autoclave and the Midmark M-11 Sterilizer both have four (4) programmed cycles as described in the above table. At the end of every sterilization cycle, the QUANTIM 16 unit automatically begins a vacuum drying cvcle of approximately 30 minutes. While the load is drying, the door may be opened at any time by pressing the button. This action terminates the dry cycle and automatically opens the door when it is safe.
In comparison, the Midmark-11 unit exhausts upon completion of the sterilization cycle, and the door automatically opens slightly to allow air drying of the instruments. However, the goods may also be removed at any time during the drying cycle.
{2}------------------------------------------------
#### Performance Data:
The actual performance levels in relation to sterility claims of the QUANTIM 16 Autoclave were demonstrated through laboratory testing using the commonly recognized test organism. Bacillus stearothermophilus. Using the validated culture medium and validated spores, ½ cycle testing, total kill end point testing, and simulated use testing with dental handpieces and medical and dental instruments were performed as outlined in the FDA's March 1993, Guidance of Premarket Notification [510(k)] Submission for Sterilizers Intended for Use in Health Care Facilities
### Test Results
Testing which was conducted under generally recognized and accepted laboratory practices by subcontracted parties was designed to be complete and challenging in relation to the performance and safety of the Quantim 16 unit for its intended use.
SciCan considers that the testing results demonstrated the Quantim 16 Autoclave to be safe and effective and to be substantially equivalent to the predicate device.
The testing results indicated that no live spores could be detected on any of the test instruments following simulated use testing. It was determined that the Quantim 16 unit met the objective of sterilization of the defined load(s) -- including dental hand pieces - to the acceptable SAL.
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the seal of the Department of Health & Human Services - USA. The seal features a stylized caduceus, a symbol often associated with medicine and healthcare, consisting of a staff with two snakes coiled around it. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular pattern around the caduceus.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
# AUG 0 8 2002
Ms. Brenda Murphy Manager, Regulatory Affairs SciCan, Division of Lux & Zwingenberger 1440 Don Mills Road Toronto. Ontario Canada M3B 3P9
Re: K021064
> Trade/Device Name: Ouantim 16 Autoclave Regulation Number: 21 CFR 880.6880 Regulation Name: Steam Sterilizer Regulatory Class: II Product Code: FLE Dated: June 28, 2002 Received: July 3. 2002
Dear Ms. Murphy:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{4}------------------------------------------------
Page 2 - Ms. Murphy
:
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 21 CFR Part 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4618. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.qov/cdrh/dsmamain.html
Sincerely,
Timothy A. Ulatowski
Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
# INDICATIONS FOR USE
: : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : :
.
| 510(k) Number: | K021064 |
|----------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Device Name: | QUANTIM 16 Table-top Autoclave |
| Indications for use: | The QUANTIM 16 tabletop autoclave is a gravity steam autoclave<br>designed to process medical and dental instruments to achieve<br>successful sterilization in a clinical setting. It utilizes saturated<br>steam at high pressures in order to attain an effective kill of<br>infectious bio-organisms and prevent cross infection.<br>The unit is intended to sterilize heat and moisture stable medical<br>and dental instruments (including dental handpieces) which are<br>commonly found in medical and dental offices, hospitals, clinics,<br>and other facilities. The instruments must be suitable for steam<br>sterilization at 132°C (270°F). Two additional cycles are also<br>specified for (i) the sterilization of textiles and wrapped surgical<br>packs and (ii) liquid loads contained in open top glass flasks.<br>However, the QUANTIM 16 Table-top Autoclave is not intended<br>nor recommended for sterilization of liquids intended for direct<br>patient contact. |
The sterilization cycles as to established time, temperatures, load configurations, drying time and indicated uses for the QUANTIM 16 Table-top Autoclave are as follows:
| CYCLE | STERILIZATION<br>TEMPERATURE | MAXIMUM<br>LOAD | STERILIZATION<br>TIME | DRYING<br>TIME | INTENDED USE |
|-----------|------------------------------|-----------------|-----------------------|-----------------------------|----------------------------------------------------------------------------------------------------------|
| Unwrapped | 132°C (270°F) | 2kg (4.4lbs) | 3 min. | 30 min. | Flash cycle for solid<br>and hinged<br>instruments, dental<br>handpieces |
| Wrapped | 132°C (270°F) | 1kg (2.2lbs) | 15 min. | 30 min. | Solid and hinged<br>instruments, dental<br>handpieces |
| Packs | 121°C (250°F) | 1kg (2.2lbs) | 30 min. | 30 min. | Textiles and wrapped<br>surgical packs |
| Liquids | 121°C (250°F) | 2kg (4.4lbs) | 30 min. | N/A (no<br>drying<br>phase) | Liquids suitable for<br>sterilization in open top<br>flasks (not intended for<br>direct patient contact) |
Qiu S. Lin
(Division Sign-Off) Division of Dental, Infection Control, and General Hospital Devices
510(k) Number K021064
1-3
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.