FIRST CHECK HOME DRUG TEST FOR MDMA (ECSTASY)

K020869 · Worldwide Medical Corp. · DKZ · Jun 11, 2002 · Clinical Toxicology

Device Facts

Record IDK020869
Device NameFIRST CHECK HOME DRUG TEST FOR MDMA (ECSTASY)
ApplicantWorldwide Medical Corp.
Product CodeDKZ · Clinical Toxicology
Decision DateJun 11, 2002
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 862.3100
Device ClassClass 2

Indications for Use

The First Check® Home Drug Test for MDMA (Ecstasy) is a screening test for the rapid detection of MDMA and its metabolites in human urine at a cut-off level of 500 ng/ml. The test is intended for consumer use as the first step in a two step process to provide consumers, including but not limited to concerned parents, with information concerning the presence or absence of MDMA or its metabolites in a urine sample. Such information is beneficial to consumer efforts to comply with applicable laws and/or societal expectations. Information regarding second step confirmatory testing is provided.

Device Story

First Check® Home Drug Test for MDMA (Ecstasy) is a rapid, one-step chromatographic immunoassay for qualitative detection of MDMA and metabolites in human urine. Device utilizes antibody/target drug/antibody complex formation to produce visual color results. Intended for home use by lay users (e.g., parents) to obtain immediate, preliminary information regarding potential MDMA use. User applies urine sample to test card well; results interpreted visually. Provides preliminary screening; positive results require secondary confirmatory testing. Benefits include immediate information for consumers regarding drug presence.

Clinical Evidence

Consumer study conducted with lay users demonstrated excellent overall performance. Analytical studies of the identical predicate (K011133) confirmed specificity for MDMA and metabolites, with no cross-reactivity for structurally related compounds (d-amphetamine, l-amphetamine, d-methamphetamine, l-methamphetamine) at 10-100 ug/ml. Cutoff concentration established at 500 ng/ml.

Technological Characteristics

One-step rapid chromatographic immunoassay. Principle: recognition and formation of specific antibody/target drug/antibody complexes. Visual colorimetric readout. Cutoff: 500 ng/ml. Standalone test card format. No electronic components or software.

Indications for Use

Indicated for consumers, including parents, for the rapid, qualitative screening of MDMA and its metabolites in human urine at a 500 ng/ml cutoff. Intended as a preliminary step requiring follow-up confirmatory testing.

Regulatory Classification

Identification

An amphetamine test system is a device intended to measure amphetamine, a central nervous system stimulating drug, in plasma and urine. Measurements obtained by this device are used in the diagnosis and treatment of amphetamine use or overdose and in monitoring levels of amphetamine to ensure appropriate therapy.

Special Controls

*Classification.* Class II (special controls). An amphetamine test system is not exempt if it is intended for any use other than employment or insurance testing or is intended for Federal drug testing programs. The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 862.9, provided the test system is intended for employment and insurance testing and includes a statement in the labeling that the device is intended solely for use in employment and insurance testing, and does not include devices intended for Federal drug testing programs (*e.g.,* programs run by the Substance Abuse and Mental Health Services Administration (SAMHSA), the Department of Transportation (DOT), and the U.S. military).

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/0 description: The image shows the logo for Worldwide Medical Corporation. The logo consists of the letters "WM" with a globe design incorporated into the letters. To the right of the logo is the text "WORLDWIDE MEDICAL" in bold, block letters, with the word "CORPORATION" underneath in smaller, less bold letters. K 0208 13 Spectrum Pointe Drive Lake Forest, CA 92630 Tel 949.598 TEST (8378) Fax 949 598 8757 www.wwmed.com Section D ## 510(k) Summary [As required by 21 CFR 807.92] - I. Submitter: - Name: Worldwide Medical Corporation A.. JUN 1 1 2002 - Address: 13 Spectrum Pointe Drive, Lake Forest, California 92630 B. - Phone: 949/598-8378 C. Phone and Fax Numbers: Phone: 949/598-8757 Fax: - Contact Person: Francisco J. Rojas, Ph.D. D. - II. Date of Preparation of this Summary: March 14, 2002 - Trade Name: First Check® Home Drug Test for MDMA (Ecstasy) III. - IV. Common Name: At home drugs of abuse rapid screening test for MDMA (Ecstasy) in urine. - V. Classification Name: Immunoassay for the qualitative detection of drugs of abuse in urine. - VI. The Marketed Products to Which Equivalence is claimed: The First Check® Home Drug Test for MDMA (Ecstasy) that is the subject of this submission is identical to the Applied Biotech SureStep™ MDMA (Ecstasy) Drug Screen Test in terms of use, product design, performance characteristics, materials of construction, and manufacturing process. It is also substantially equivalent to the Phamatech At Home™ Drug Test, and other commercially available drug screening tests that qualitatively measure the presence of target drugs or metabolites by visual color one-step immunoassay technology. - VII. Statement of Intended Use Compared to Other Products: The intended use of the First Check® Home Drug Test for MDMA (Ecstasy) is substantially equivalent to the listed products; it is a preliminary, rapid screening test for the detection of MDMA (Ecstasy) and its metabolites in urine. This product is intended to be the first step in a two step process to provide consumers, including but not limited to concerned parents, with information regarding the presence of MDMA (and its metabolites) in a urine sample. Information regarding the second step, confirmatory testing, is provided. б {1}------------------------------------------------ Image /page/1/Picture/0 description: The image shows the logo for Worldwide Medical Corporation. The logo consists of the letters "WM" with a globe encircling the letters. To the right of the logo is the text "WORLDWIDE MEDICAL" in bold, block letters, with the word "CORPORATION" underneath in smaller, thinner letters. The logo is black and white. 13 Spectrum Pointe Drive Lake Forest, CA 92630 Tel. 949.598.TEST (8378) Fax 949 598:8757 www.wwmed.com - Discussion of Technological Characteristics: The First Check® Home Drug Test VIII. for MDMA (Ecstasy), like other commercially available drug screening tests, qualitatively measures the presence or absence of MDMA and its metabolites in urine, using a one step, rapid chromatographic immunoassay which operates under the principle of recognition and formation of specific antibody/target drug/antibody complexes. Examples of predicate devices include the First Check® Home Drug Tests using a single or a multi-drug display. The analytical studies of the identical Applied Biotech SureStep™ MDMA (Ecstasy) Drug Screen Test (K011133) indicate that the drug test reacts specifically with MDMA and its metabolites. The structurally related compounds d- amphetamine, 1amphetamine, d- methamphetamine, and 1-methamphetamine at concentrations of 10 to 100 ug/ml produce negative test results. A consumer study using the First Check® Home Drug Test for MDMA (Ecstasy) demonstrates that the test exhibits excellent overall performance in the hands of lay users. The data supports the conclusion that the consumer can use the First Check® Home Drug Test for MDMA (Ecstasy) to obtain immediate, preliminary information regarding the possible use of MDMA. The cutoff concentration for MDMA is 500 ng/ml. - IX. Safety and Effectiveness: Because the First Check® Home Drug Test for MDMA (Ecstasy) is identical to the Applied Biotech SureStep™ MDMA (Ecstasy) Drug Screen Test that is legally marketed under K011133, and because no special skills, training, education, or licensure are required to transfer a dropper level of urine sample into the test card well, there is no issue regarding the safety or effectiveness of the product to perform its intended function, i.e., to screen urine for the presence or absence of MDMA and its metabolite. Because the labeling of the First Check® Home Drug Test for MDMA (Ecstasy) is substantially equivalent to a variety of rapid screening tests currently in commercial distribution, including the Phamatech At Home™ Drug Test, and there have been no reports of consumer inability to follow instructions or interpret results during the twenty-four months in which the First Check® product line has been purchased by the general public and used in quantities exceeding 200,000 tests, it is concluded that the product can be used effectively by the lay user. 7 {2}------------------------------------------------ DEPARTMENT OF HEALTH & HUMAN SERVICES Image /page/2/Picture/1 description: The image is a black and white logo for the U.S. Department of Health & Human Services. The logo features a stylized image of an eagle with three lines representing its wings and head. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular pattern around the eagle image. Food and Drug Administration 2098 Gaither Road Rockville MD 20850 JUN 1 1 2002 Francisco J. Rojas, Ph.D. . Chief Scientific Officer Worldwide Medical Corporation 13 Spectrum Pointe Drive Lake Forest, CA 92630 k020869 Re: Trade/Device Name: First Check® Home Drug Test for MDMA (Ecstasy) Regulation Number: 21 CFR 862.3100 Regulation Name: Amphetamine test system Regulatory Class: Class II Product Code: DKZ Dated: May 2, 2002 Received: May 3, 2002 Dear Dr. Rojas: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. {3}------------------------------------------------ Page 2 - This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and ' additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4588. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.gov/cdrb/dsmaldsmamain.html". Sincerely yours, Steven Putman Steven I. Gutman, M.D., M.B.A. Director Division of Clinical Laboratory-Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {4}------------------------------------------------ Image /page/4/Picture/0 description: The image shows the logo for Worldwide Medical Corporation. The logo consists of a stylized globe with the letters "WM" superimposed on it. To the right of the logo is the text "WORLDWIDE MEDICAL" in bold, sans-serif font, with the word "CORPORATION" underneath in a smaller font. 13 Spectrum Pointe Drive Lake Forest, CA 92630 Tel. 949.598 TEST (8378) Fax 949,598-8757 www.wwmed.com Section A Page 1 of 1 510 (k) Number (if known): k020869 Device Name: First Check® Home Drug Test for MDMA (Ecstasy) Indications for Use: The First Check® Home Drug Test for MDMA (Ecstasy) is a screening test for the rapid detection of MDMA and its metabolites in human urine at a cut-off level of 500 ng/ml. The test is intended for consumer use as the first step in a two step process to provide consumers, including but not limited to concerned parents, with information concerning the presence or absence of MDMA or its metabolites in a urine sample. Such information is beneficial to consumer efforts to comply with applicable laws and/or societal expectations. Information regarding second step confirmatory testing is provided. (Division Sign-Off) Division of Clinical Laboratory Devices 510(k) Number K020869 (PLEASE DO NOT WRITE BELOW THIS LINE- CONTINUE ON ANOTHER PAGE IF NEEDED.) Concurrence of CDRH, Office of Device Evaluation (ODE) I (Optional Format 3-10-98)
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...