← Product Code [NUC](/productcode/NUC) · K020827

# K-Y BRAND ULTRA GEL PERSONAL LUBRICANT (K020827)

_Personal Products Co., Div. of Mcneil Ppc, Inc. · NUC · Jun 12, 2002 · Obstetrics/Gynecology · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K020827

## Device Facts

- **Applicant:** Personal Products Co., Div. of Mcneil Ppc, Inc.
- **Product Code:** [NUC](/productcode/NUC.md)
- **Decision Date:** Jun 12, 2002
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 884.5300
- **Device Class:** Class 2
- **Review Panel:** Obstetrics/Gynecology

## Indications for Use

Patient lubricants are devices intended to lubricate a body orifice to facilitate entry of a diagnostic or therapeutic device. When used as an accessory to a condom, patient lubricants are deemed Class II Medical Devices. K-Y® Brand Ultra Gel™ is principally intended as personal lubricant to supplement the body's own natural lubricating fluids, to moisturize, relieve friction and to enhance the ease and comfort of intimate sexual activity. This lubricant may be safely applied to vaginal, anal or penile tissue for lubrication, and moisturization purposes. It is also compatible with latex condoms.

## Device Story

K-Y® Brand Ultra Gel™ is a non-sterile, water-based personal lubricant. It functions by supplementing natural body lubrication to reduce friction and enhance comfort during intimate sexual activity. The device is applied topically to vaginal, anal, or penile tissue by the user. It is compatible with latex condoms, tested per ASTM D 3492. The product is intended for over-the-counter use.

## Clinical Evidence

Clinical evidence includes human irritation studies (non-sensitizing, mildly irritating) and consumer-use testing comparing the formulation to a commercially marketed vaginal moisturizer. Consumer testing showed significantly higher ratings for texture, consistency, and moisturizing properties. Preclinical biocompatibility studies (GLP compliant) confirmed the formulation is non-irritating, lacks pharmacotoxic effects, and lacks systemic toxicity.

## Technological Characteristics

Water-based, non-sterile, high-viscosity liquid gel. Contains vitamin E. Compatible with latex condoms per ASTM D 3492. No exceptional technological characteristics; ingredients are water-soluble and similar to existing market lubricants.

## Regulatory Identification

A condom is a sheath which completely covers the penis with a closely fitting membrane. The condom is used for contraceptive and for prophylactic purposes (preventing transmission of sexually transmitted infections). The device may also be used to collect semen to aid in the diagnosis of infertility.

## Special Controls

*Classification.* (1) Class II (special controls) for condoms made of materials other than natural rubber latex, including natural membrane (skin) or synthetic.(2) Class II (special controls) for natural rubber latex condoms. The guidance document entitled “Class II Special Controls Guidance Document: Labeling for Natural Rubber Latex Condoms Classified Under 21 CFR 884.5300” will serve as the special control. See § 884.1(e) for the availability of this guidance document.

## Predicate Devices

- K-Y® Brand LIQUID Personal Lubricant

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

## 510(k) SUMMARY

| Submitter           | Personal Products Company Division of McNeil-PPC Inc.<br>199 Grandview Road<br>Skillman, New Jersey 08558-9418 |
|---------------------|----------------------------------------------------------------------------------------------------------------|
| Contact Person      | Marylou Panico<br>Mgr. Regulatory Affairs<br>(908) 904-3709 telephone<br>(908) 904-3748 telefax                |
| Date Prepared       | May 29, 2002                                                                                                   |
| Proprietary Name    | K-Y® Brand Ultra Gel™<br>Personal Lubricant                                                                    |
| Common Name         | Personal Lubricant                                                                                             |
| Classification Name | Condom 21CFR § 884.5300 Product Code 85HIS                                                                     |
| Predicate Device    | K-Y® Brand LIQUID Personal Lubricant                                                                           |

## Description of Device

K-Y® Brand Ultra Gel™ is a non-sterile water- based personal lubricant designed to supplement the body's own natural lubrication fluids. This formula is a clear, non-irritating, non-greasy, non-staining, high viscosity liquid gel. This highly lubricous product may be used with or without a latex condom during intimate sexual activity. K-Y® Brand Ultra Gel™ is not a contraceptive or spermicide. It is however, compatible with latex condoms as demonstrated in Condom Compatibility Testing conducted according the standards defined by ASTM D 3492.

## Intended Use

Patient lubricants are devices intended to lubricate a body orifice to facilitate entry of a diagnostic or therapeutic device. When used as an accessory to a condom, patient lubricants are deemed Class II Medical Devices.

K-Y® Brand Ultra Gel™ is principally intended as personal lubricant to supplement the body's own natural lubricating fluids, to moisturize, relieve friction and to enhance the ease and comfort of intimate sexual activity. This lubricant may be safely applied to vaginal, anal or penile tissue for lubrication, and moisturization purposes. It is also compatible with latex condoms.

## Technological Characteristics

K-Y® Brand Ultra Gel™ formula is proprietary. The product however, has no exceptional technological characteristics, consisting mainly of water-soluble ingredients similar to other lubricants currently on the US market.

{1}------------------------------------------------

#### 510(k) SUMMARY (continued)

#### Substantial Equivalence

This product has been shown in laboratory testing to be substantially equivalent to its predecessor, the currently marketed K-Y® Brand LIQUID Personal Lubricant. Both devices have the same intended use and similar formulations. One differentiating feature is that K-Y® Brand Ultra Gel™ contains vitamin E and is more lubricious, as demonstrated in laboratory coefficient of friction testing.

#### Preclinical Testing of Formulation

Preclinical biocompatibility studies on K-Y® Brand Ultra Gel™ Personal Lubricant were conducted by outside laboratories, in compliance with Good Laboratory Practices (GLPs). These studies demonstrated that this formulation was non-initating, did not produce any significant pharmacotoxic effects and was not associated with system toxicity.

#### Clinical Testing

In Human Irritation studies under conservative occluded and repetitive conditions, this formulation was non-sensitizing and only mildly irritating.

Consumer-Use Testing conducted to evaluate performance of the proposed formulation as compared to another commercially marketed vaginal moisturizer, demonstrated that the proposed formulation rated significantly higher for texture/consistency and moisturizing properties.

Laboratory, Preclinical and Clinical testing conducted on K-Y® Brand Ultra Gel™ Personal Lubricant has provided scientific evidence that this product is safe for its intended use and substantially equivalent to the predicate device, K-Y® Brand LIQUID Personal Lubricant.

{2}------------------------------------------------

Image /page/2/Picture/1 description: The image is a black and white logo for the U.S. Department of Health & Human Services. The logo features a stylized eagle or bird-like figure with three curved lines forming its body and wings. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" is arranged in a circular pattern around the bird figure.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

FEB 2 4 2014

Ms. Marylou Panico Manager, Regulatory Affairs Personal Products Company Division of McNeil-PPC, Inc. 199 Grandview Road SKILLMAN NJ 08558

Re: K020827

> Trade/Device Name: K-Y® Brand Marque Ultra Gel™ Personal Lubricant Regulation Number: 21 CFR §884.5300 Regulation Name: Condom Regulatory Class: II Product Code: NUC Dated (Date on orig SE ltr): March 12, 2002 Received (Date on orig SE ltr): March 14, 2002

Dear Ms. Panico:

This letter corrects our substantially equivalent letter of June 12, 2002.

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. Iisting of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be

{3}------------------------------------------------

Page 2 -

found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office

of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours.

# Benjamin R. Fisher -S

Benjamin R. Fisher, Ph.D. Director Division of Reproductive, Gastro-Renal, and Urological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

# 510(k) INDICATIONS FOR USE FORM (Replica of FDA Form)

510(k) Number (if known):

Device Name: K-Y Brand ® Ultra Gel™ Personal Lubricant Indications For Use: Personal Lubricant compatible with latex condoms, helps enhance intimacy

(PLEASE DO NOT WRITE BELOW THIS LINE CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Prescription Use OR Over-the-Counter-Use V (per 21 CFR 801.109)

--

Girard G. Geyson

JUN 12 2002

02-0827

00-000003

---

**Source:** [https://fda-staging.innolitics.com/device/K020827](https://fda-staging.innolitics.com/device/K020827)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
