K020796 · GE Medical Systems · JAK · Mar 26, 2002 · Radiology
Device Facts
Record ID
K020796
Device Name
CARD IQ ANALYSIS II
Applicant
GE Medical Systems
Product Code
JAK · Radiology
Decision Date
Mar 26, 2002
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.1750
Device Class
Class 2
Attributes
Software as a Medical Device, 3rd-Party Reviewed
Indications for Use
CardIQ Analysis II is a post processing software option for Advantage Windows Workstation (AW) Platform. This product can be used for the analysis of CT angiography images and for the assessment of plaque densities with the coronary arteries. It provides a number of display, measurement and batch filming features. The product can be used to aid trained physicians for visualizing and assessing cardiac anatomy, and coronary vessels.
Device Story
CardIQ Analysis II is a post-processing software option for the Advantage Windows (AW) Workstation. It processes CT angiography images to generate visualizations of cardiac structures, including MIP (Maximum Intensity Projection), reformatted, and volume-rendered views. The software enables clinicians to perform measurements of vessel diameter, Hounsfield Units within coronary arteries, and plaque density. Outputs include processed images, measurements, and reports, which can be saved to the workstation or printed. Used in clinical settings by trained physicians to aid in the visualization and assessment of cardiac anatomy and coronary vessels, supporting clinical decision-making regarding cardiac health.
Clinical Evidence
No clinical data. Safety and effectiveness are supported by software development, verification, and validation processes, including hazard analysis and adherence to industry and international standards.
Technological Characteristics
Post-processing software for CT angiography; operates on Advantage Windows (AW) Workstation platform. Features include image display (MIP, volume rendering), vessel diameter measurement, and plaque density assessment. Software-based; no specific hardware materials or energy sources described. Verification follows industry and international standards.
Indications for Use
Indicated for trained physicians to visualize and assess cardiac anatomy and coronary vessels using CT angiography images, including assessment of plaque densities within coronary arteries.
Regulatory Classification
Identification
A computed tomography x-ray system is a diagnostic x-ray system intended to produce cross-sectional images of the body by computer reconstruction of x-ray transmission data from the same axial plane taken at different angles. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows a date, "MAR 2 6 2002". The month is March, the day is the 6th, and the year is 2002. The text is in a bold, sans-serif font.
Image /page/0/Picture/1 description: The image shows the General Electric (GE) logo. The logo consists of the letters "GE" intertwined and enclosed within a circular shape. The letters and the circle are depicted in black, creating a strong contrast against the white background. The logo is a well-known symbol associated with the General Electric company.
K020796
(8)
GE Medical Sv General Electric Company P O Box 414 Milwaukee, WI 53201
# 510(k) SUMMARY OF SAFETY AND EFFECTIVENESS
This 510(k) summary of safety and effectiveness information is submitted in accordance with the requirements of 21 CFR Part 807.87(h).
Submitter Larry A. Kroger, Ph.D. Senior Regulatory Program Manager Telephone: (262) 544-3894, FAX: (262) 544-3863 Date Prepared: February 8, 2002
# PRODUCT IDENTIFICATION
| Name: | CardIQ Analysis II |
|------------------------|---------------------------------------------------------------------------------------|
| Classification Name: | Accessory to Computed Tomography System |
| Classification Panel: | 892 - Radiology |
| Classification Number: | 892.1750 |
| Manufacturer : | General Electric Medical Systems<br>283, rue de la Miniere<br>78533 Buc Cedex, FRANCE |
| Distributor: | General Electric Medical Systems, Milwaukee, WI |
| Marketed Devices | CardIQ Analysis II is substantially equivalent to the device listed below: |
| Model: | CardIQ Option 1.0 |
| Manufacturer: | General Electric Medical Systems, Milwaukee, WI |
| 510(k) #: | K003408 |
### Device Description:
CardIQ Analysis II is a post processing software option that can be used in the analysis of CT angiography images to display structures of the heart in a MIP, reformat or volume rendering view. The software has the ability to measure the diameter of the vessel, Hounsfield Units within connay arteries, or plaque density within a vessel. Measurements, densities and images can all be priced to reports or saved to the AW workstation.
{1}------------------------------------------------
#### Indications for Use:
CardIQ Analysis II is a post processing software option for Advantage Windows Workstation (AW) Platform. This product can be used for the analysis of CT angiography images and for the assessment of plaque densities with the coronary arteries. It provides a number of display, measurement and batch filming features. The product can be used to aid trained physicians for visualizing and assessing cardiac anatomy, and coronary vessels.
#### Comparison with Predicate:
CardIQ Analysis II is a software option for CT Scanners. Features of this software package are substantially equivalent to the following device:
| A MARK & MARK MANAGE & MARKER AN<br>------------------------------------------------------------------------------------------------------------------------------------------------------------------------------<br>of the first and a provinces and consistences and the desires of the desires of<br><br>I a since an an an extens a series<br><br>------------------------------------------------------------------------------------------------------------------------------------------------------------------------------ | .<br>. I wa can a sualle a second consumer consideration and<br>.<br>on the commend an experience of the control of the contributed to the collection of the collection of the collection of the collection of the collection of the collection of<br>ANNUAL CONNECT & CONSULER CONNECT CONSULER CO<br>------------------------------------------------------------------------------------------------------------------------------------------------------------------------------<br>A 114<br>------------------------------------------------------------------------------------------------------------------------------------------------------------------------------<br><br><br> |
|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| <br>------------------------------------------------------------------------------------------------------------------------------------------------------------------------------<br>------------------------------------------------------------------------------------------------------------------------------------------------------------------------------<br>A B S CONSULTION OF COLLECTION OF CONSULT OF CONSULT OF CONSULT OF CONSULT OF CONSULTION OF CONSULTION OF CHILIP OF CONSULTION OF CHILIP OF CHILING OF CHILIP OF CHILINOR OF C<br>-<br>Comments of the color of the collection of the contrôlean of the comments of the consisted to the consideration of<br>------------------------------------------------------------------------------------------------------------------------------------------------------------------------------ | Market and distribution to the August of the programments of the program and the program and the pro<br>------------------------------------------------------------------------------------------------------------------------------------------------------------------------------<br>I I E LEREN I SEARNHISMAN<br><br>------------------------------------------------------------------------------------------------------------------------------------------------------------------------------<br>-------------------------------------------------------------------<br><br>contrasses of the program and the consideration of charges of the comments of the comments of the comments of the comments of the comments of the comments of the comments of<br>---------------------------------------------------- |
#### Adverse Effects on Health :
The potential hazards are identified in a risk management summary (hazard analysis) and are controlled by:
- · Software Development, Validation and Verification Process to ensure performance to specifications, Federal Regulations and user requirements.
(C
- · Adherence to industry and international standards.
#### Conclusions:
CardIQ Analysis II does not result in any new potential safety risks and performs as well as devices currently on the market. GE considers features of the CardIO Analysis II to be equivalent to those of CT CardIQ Option 1.0 (K003408).
{2}------------------------------------------------
Image /page/2/Picture/1 description: The image shows the seal of the Department of Health & Human Services (HHS) of the United States. The seal features a stylized eagle with three stripes forming its wings, symbolizing strength and protection. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" is arranged in a circular pattern around the eagle, indicating the department's name and national affiliation. The seal is presented in black and white, emphasizing its official and governmental nature.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
MAR 2 6 2002
General Electric Medical Systems % Mr. Wolfram Gmelin Technical Manager TÜV Rheinland of North America 12 Commerce Road NEWTOWN CT 06470
Re: K020796
Trade/Device Name: Card IQ Analysis II CT Scanner Post-Processing Software Regulation Number: 21 CFR 892.1750 Regulation Name: Computed tomography x-ray system Regulatory Class: II Product Code: 90 JAK Dated: February 1, 2002 Received: March 12, 2002
Dear Mr. Gmelin:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (sections 531-542 of the Act); 21 CFR 1000-1050.
{3}------------------------------------------------
This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at one of the following numbers, based on the regulation number at the top of this letter:
| 8xx.1xxx | (301) 594-4591 |
|----------------------------------|----------------|
| 876.2xxx, 3xxx, 4xxx, 5xxx | (301) 594-4616 |
| 884.2xxx, 3xxx, 4xxx, 5xxx, 6xxx | (301) 594-4616 |
| 892.2xxx, 3xxx, 4xxx, 5xxx | (301) 594-4654 |
| Other | (301) 594-4692 |
Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html.
Sincerely yours,
Nancy C. Brogdon
Nancy C. Brogdon Director, Division of Reproductive, Abdominal, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
## STATEMENT OF INTENDED USE
510(k) Number (if known): _ Ko 2 079 6 _______________________________________________________________________________________________________________________________________
Device Name: CardIQ Analysis II
### Indications For Use:
CardIQ Analysis II is a post processing software option for Advantage Windows Workstation (AW) Platform. This product can be used for the analysis of CT angiography images and for the assessment of plaque densities with the coronary arteries. It provides a number of display, measurement and batch filming features. The product can be used to aid trained physicians for visualizing and assessing cardiac anatomy, and coronary vessels.
## (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
| Prescription Use | <span></span> | -OR- | Over-The-Counter Use | <span></span> |
|----------------------|---------------|------|----------------------|---------------|
| (Per 21 CFR 801.109) | | | | |
Nancy C. Brogdon
| (Division Sign-Off) | |
|---------------------------------------------------------------|---------|
| Division of Reproductive, Abdominal, and Radiological Devices | |
| 510(k) Number | K020796 |
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.