K020791 · Millennium Medical Technologies, Inc. · HWC · Jun 6, 2002 · Orthopedic
Device Facts
Record ID
K020791
Device Name
HBS HEADLESS BONE SCREW
Applicant
Millennium Medical Technologies, Inc.
Product Code
HWC · Orthopedic
Decision Date
Jun 6, 2002
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 888.3040
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
Headless bone screw system for the fixation of osseous fragments or fractures including: HBS - MINI: Scaphoid fractures, Lunate fractures, Capitate, Trapezial fractures, Metacarpal and metatarsal fractures, Phalangeal fractures, Radial head fractures, Ulnar styloid fractures, Osteo-chrondral, Small joint fusions. HBS - STANDARD: Scaphoid fractures, Carpal fractures & non-unions, Capitellum fractures, Metacarpal fractures, Phalangeal fractures, Distal radial fractures, Radial head fractures, Ulnar styloid fractures, Small joint fusions, Humeral head fractures, Glenoid fractures, Intercarpal fusions, Interphalangeal fractures, Metatarsal osteotomies, Tarsal fusions, Malleolar fractures, Patellar fractures, Osteo-chrondral fractures, Odontoid fractures, Mandibular fractures.
Device Story
HBS™ Headless Bone Screw system consists of cannulated titanium alloy screws in two series (Mini and Standard) for bone fixation. Screws feature headless, dual-threaded design to provide compression across fracture sites. System includes specialized instrumentation (drill guides, cannulated drills, taps, T-Drive screwdrivers) for surgical implantation. Used by physicians in clinical/surgical settings to stabilize fractures or perform joint fusions. Operates via mechanical compression of bone fragments; high/low compression modes available. Benefits include stable fixation of small bone fractures and osteotomies. Provided non-sterile; requires steam autoclave sterilization by user.
Clinical Evidence
Bench testing only. Biomechanical testing performed in high and low compression modes. High compression mode yielded values >200N; standard compression mode yielded 165N. Results exceed values obtained with similar predicate devices.
Indicated for fixation of osseous fragments or fractures in small bones (hands, feet) and selected fractures elsewhere in the body (e.g., humerus, mandible, patella) where medically indicated and bone mass is compatible.
Regulatory Classification
Identification
A smooth or threaded metallic bone fixation fastener is a device intended to be implanted that consists of a stiff wire segment or rod made of alloys, such as cobalt-chromium-molybdenum and stainless steel, and that may be smooth on the outside, fully or partially threaded, straight or U-shaped; and may be either blunt pointed, sharp pointed, or have a formed, slotted head on the end. It may be used for fixation of bone fractures, for bone reconstructions, as a guide pin for insertion of other implants, or it may be implanted through the skin so that a pulling force (traction) may be applied to the skeletal system.
Cannulated Navicular Screw, Onyx Medical Corp. (K931681)
Submission Summary (Full Text)
{0}------------------------------------------------
6 2002 JUN
020791/
page 1 of 5
December 14.
# 510(k) Premarket Notification Summary of Safety and Effectiveness Information
Device Name:
#### Trade Name: Common Name: Classification Name:
# HBS™ Headless Bone Screw
Bone Screw Smooth & threaded metallic bone fixation fasteners
# Establishment Name & Registration Number:
Millennium Medical Technologies, Inc. (MMT) Name: Number: Pending
#### Classification:
ProCode: HWC
# Performance Standards (Section 514 compliance):
Food and Drug Administration mandated Performance standards for bone screws are not in effect. MMT intends to comply with all voluntary Performance Standards applicable to the HBS™ Headless Bone Screw system.
At the present time, The device is produced according to ISO 9001-2000 regulations covering medical devices. In addition, the materials used to construct the device meet ISO-5832-3,1996(E).
## Special Controls:
All Class II devices are subject to Special Controls. No FDA mandated special controls are in effect at the present time.
# Labeling:
The labeling of the device includes the following statements:
Federal (United States) Law restricts this device to sale by or on the order of a Warning: physician only.
CAUTION: Mixing of dissimilar metals can accelerate the corrosion process, The components of this system must NOT be used with implants of other material composition. Components of the HBS™ Headless Bone Screw should NOT be used with screws or components from any other system or manufacturer.
# Equivalent Device(s):
HBS™ Headless Bone Screw may be directly contrasted with the following equivalent devices:
- Zimmer HBS™, Zimmer, Inc., K792022 .
- Cannulated Navicular Screw, Onyx Medical Corp. K931681 ●
The Zimmer HBS is essentially identical (clinically speaking) to the HBS™ Headless Bone Screw in terms of basic design, features and intended use. The Onyx device is also very similar in design except that the Onyx device is indicated only for Navicular fractures, where as the MMT device has broader indications for use.
{1}------------------------------------------------
K020791
page 2 of 5
#### Description of the Device:
The HBS™ (Headless Bone Screw) and the mini HBS™ are supplied as two series of cannulated bone screws varying in length and diameter. Both systems are intended for small bone applications and are thus considered a single system. Indeed, both systems are intended for use primarily in the hands and feet. The systems may be used for selected fractures elsewhere in the body so long as medically indicated and bone mass compatible.
Available screws and instrumentation are available as two individually boxed sets offered in the following sizes:
# MINI HBS™
- 1.5mm shaft diameter, 3.2mm proximal thread diameter, 2.5mm distal thread ● diameter.
- 10mm through 30mm length in one millimeter increments. .
#### Instrumentation:
- Handle for drill guide .
- . Drill guide
- Drill
- . Drill handle
- . T-Drive screw driver
- MBS Tray Mini .
#### STANDARD HBS™
- 3.0mm shaft diameter, 4.7mm proximal thread diameter, 4.0mm distal thread . diameter.
- 10mm through 30mm length in one millimeter increments. ●
#### Instrumentation:
- Guide wire, 1mm ◆
- Guiding handle for drill guide ●
- . Drill guide
- Measuring sleeve
- Cannulated drill .
- . Cortical Drill
- . Tap, standard compression
- . Tap, high compression
- . Cleaning wire
- . Wire Dispencer
- Drill handle .
- T-Drive screw driver
- Reduction Sleeve
- MBS Tray Standard .
- Container 300 x 300 x 140 .
{2}------------------------------------------------
020791
page 3 of 5
## Intended Use: MINI
- Scaphoid fractures ●
- . Lunate fractures
- Capitate .
- Trapezial fractures ●
- Metacarpal and metatarsal fractures ●
- Phalangeal fractures .
- . Radial head fractures
- Ulnar styloid fractures ●
- Osteo-chrondral ●
- Small joint fusions ●
# Intended Use: STANDARD
- Scaphoid fractures t
- Carpal fractures & non-unions .
- . Capitellum fractures
- Metacarpal fractures .
- Phalangeal fractures .
- . Distal radial fractures
- Radial head fractures .
- Ulnar styloid fractures .
- � Small joint fusions
- . Humeral head fractures
- . Glenoid fractures
- Intercarpal fusions .
- Interphalangeal fractures .
- Metatarsal osteotomies .
- Tarsal fusions .
- Malleolar fractures .
- Patellar fractures .
- Osteo-chrondral fractures .
- . Odontoid fractures
- Mandibular fractures .
MMT supplies instrumentation designed specifically to interface with this implant system. Drawings of necessary specialized instruments are included in Appendix II with the system engineering drawings.
The screws are made from titanium alloy. Material standard(s) are as specified in the table below:
| be the consinues and consideration of the county of the first and a<br>BACTI TITY | A BOOK BOOK AND LEAST LEAST LEAST LEAST LEAST LEAST LEAST LEAST LEAST LEAST LEAST LEAST LEAST LEAST LEAST LEAST LEAST LEAST LEAST LEAST LEAST LEAST LEAST LEAST LEAST LEAST LE<br>CHANAGE ALL PRODUCTION A CONSULTION CALL CREATING<br>------------------------------------------------------------------------------------------------------------------------------------------------------------------------------<br>------------------------------------------------------------------------------------------------------------------------------------------------------------------------------<br>Cama<br>C | ------------------------------------------------------------------------------------------------------------------------------------------------------------------------------<br>------------------------------------------------------------------------------------------------------------------------------------------------------------------------------<br>------------------------------------------------------------------------------------------------------------------------------------------------------------------------------<br>***<br>Conn<br>C |
|-----------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | ------------------------------------------------------------------------------------------------------------------------------------------------------------------------------<br>www.management.com/sports/article/ship/submit/shop/<br>Comment of Children and | BARREAL BALLERS AND AND AND<br>--------------------------------------------------------------------<br>Comments of the property of the |
# Summary of Biomechanical Testing:
Fixation characteristics of the screws were determined in both high compression and low compression modes. High compression mode vielded values in excess of 200N. Standard compression mode yielded values of 165N. In both instances, these values are in excess of values obtained with similar devices.
{3}------------------------------------------------
020791
page 4 of 5
# Applicant / Sponsor Name / Address:
Millennium Medical Technologies, Inc. 460 St. Michaels Drive, Suite 901 Santa Fe, New Mexico, 87505 505.988.7595 -- 505.988.7234 fax
# Contact Person:
Mr. Fred Kolb Millennium Medical Technologies, Inc. 460 St. Michaels Drive, Suite 901 Santa Fe, New Mexico, 87505 505.988.7595 - 505.988.7234 fax
#### Submission Correspondent:
Mr. David W. Schlerf Buckman Company, Inc. 200 Gregory Lane, Suite C-100 Pleasant Hill, CA 94523-3389 925.356.2640 - 925.356.2654 fax
#### Manufacturing Facility:
At the present time, the devices are contract manufactured in Europe and will be imported into the United States. The contract manufacturer is a recognized, registered manufacturer of implantable medical devices. The HBS screws are manufactured for MMT for distribution in the U.S.A.
#### Performance Standards:
There are no applicable FDA mandated performance standards for this device. However, voluntary standards such as ASTM, various in-house Standard Operating Procedures and vendor qualification procedures are in place and utilized in the production of the screws.
# St rilization, Packaging & Storage Information:
Sterilization. The HBS™ Headless Bone Screws and instruments are provided non-sterile and must be sterilized prior to use. The devices are clean and have been processed to remove debris and manufacturing residue.
The recommended sterilization process for the instruments is high temperature steam autoclave sterilization. The recommended sterilization cycle is: saturated steam at 270° F for 30 minutes. This is a typical or usual steam sterilization cycle used by hospitals for surgical devices and instruments. Use of this cycle produces a Sterility Assurance Level (SAL) of at least 10-6. Validation of the recommended sterilization cycle is completed. The following validated sterilization cycle is on file at MMT.
| Method: | Steam |
|----------------|---------------|
| Cycle: | Gravity |
| Temperature: | 270°F (134°C) |
| Exposure Time: | 30 minutes |
Packaging. Materials used in the production of the device are typical medical grade tubes, peel-type pouches of the generic mylar/non-woven sandwich variety, etc. All packages containing implants or instruments should be intact upon receipt.
{4}------------------------------------------------
K020791
page 5 of 5
# Comparison Table:
| FEATURE | MMT HBS™ | Zimmer HBS™ | Onyx Navicular | SE? |
|---------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------|----------------|-----|
| Intended Use: | Scaphoid fractures<br>Carpal fractures & non-unions<br>Capitellum fractures<br>Metacarpal fractures<br>Phalangeal fractures<br>Distal radial fractures<br>Radial head fractures<br>Ulnar styloid fractures<br>Small joint fusions<br>Humeral head fractures<br>Glenoid fractures<br>Intercarpal fusions<br>Interphalangeal fractures<br>Metatarsal osteotomies<br>Tarsal fusions<br>Malleolar fractures<br>Patellar fractures<br>Osteo-chrondral fractures<br>Odontoid fractures<br>Mandibular fractures | Same | Navicular | Yes |
| Materials | Titanium | Same | Same | Yes |
| Lengths | 10-30 mm in 1mm increments | 10-30mm, 2mm | 10-30mm | Yes |
| Diameters | Proximal thread 4.7mm (S)<br>Proximal thread 3.2mm (M)<br>Distal thread 4.0mm (S)<br>Distal thread 2.5mm (M)<br>Shaft Dia. 3.0mm (S)<br>Shaft Dia. 1.5mm (M) | Prox.4.5mm<br>Distal 4.0mm | Unknown | Yes |
| Design | Headless, Threaded both ends | Same | Same | Yes |
| Sterilization | Steam | Steam | Steam | Yes |
| K-Number | Pending | 792022 | 931681 | Yes |
{5}------------------------------------------------
Image /page/5/Picture/1 description: The image contains the words "Public Health Service". The text is written in a simple, sans-serif font. The words are arranged on a single line, with "Public Health" appearing before "Service".
Image /page/5/Picture/2 description: The image is a black and white seal for the Department of Health & Human Services - USA. The seal is circular with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized image of a caduceus, a symbol of medicine and healing, with three intertwined snakes wrapped around a staff.
6 2002 אחר
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
Millennium Medical Technologies, Inc c/o Mr. David W. Schlerf Buckman Company, Inc. 200 Gregory Lane Suite C-100 Pleasant Hill, CA 94523-3389
Re: K020791
Trade/Device Name: HBS™ Headless Bone Screw Regulatory Number: 888.3040 Regulatory Name: Smooth or threaded metallic bone fixation fastener Regulatory Class: II Product Code: HWC Dated: March 7, 2002 Received: March 11, 2002
Dear Mr. Schlerf:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{6}------------------------------------------------
Page 2 - Mr. David W. Schlerf
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 21 CFR Part 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4659. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html
Sincerely yours.
Mach N Millhinn
Celia Witten, Ph.D., M.D. Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{7}------------------------------------------------
510(k) Number (if known):_____________________________________________________________________________________________________________________________________________________
#### HBS™ Headless Bone Screw Device Name:
# Indications For Use:
Headless bone screw system for the fixation of osseous fragments or fractures including:
| HBS - MINI | HBS - STANDARD |
|-------------------------------------|-------------------------------|
| Scaphoid fractures | Scaphoid fractures |
| Lunate fractures | Carpal fractures & non-unions |
| Capitate | Capitellum fractures |
| Trapezial fractures | Metacarpal fractures |
| Metacarpal and metatarsal fractures | Phalangeal fractures |
| Phalangeal fractures | Distal radial fractures |
| Radial head fractures | Radial head fractures |
| Ulnar styloid fractures | Ulnar styloid fractures |
| Osteo-chrondral | Small joint fusions |
| Small joint fusions | Humeral head fractures |
| | Glenoid fractures |
| | Intercarpal fusions |
| | Interphalangeal fractures |
| | Metatarsal osteotomies |
| | Tarsal fusions |
| | Malleolar fractures |
| | Patellar fractures |
| | Osteo-chrondral fractures |
| | Odontoid fractures |
| | Mandibular fractures |
PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NECESSARY
Concurrence of CDRH, Office of Device Evaluation (ODE)
for Mark A. Mellone
(Division Sign-Off)
Division of General, Restorative
and Neurological Devices
510(k) Number K020791
Prescription Use OR Over-The-Counter Use
(Per 21 CFR 801.109)
OR
Over-The-Counter Use _ (Optional format 1-2-96)
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.